ArticlePDF Available

Design of a dexterous robotic surgical instrument with a novel bending mechanism

Authors:

Abstract

Background The robot-assisted minimally invasive surgery (RMIS) has developed rapidly in recent years, requiring highly articulated instruments to enable surgeons to perform complicated and precise procedures. Methods A novel wrist-type surgical instrument was proposed for RMIS. The wrist consists of superelastic-wire-driven snake-like joints and universal joints, which could perform two deflections and one distal rotation. The bending mechanism and the kinematics of universal joints were analyzed. The forward and inverse kinematics of the wrist were derived. Results The performances of the instrument were evaluated using a prototype by experiments. The average motion deviation of the wrist’s deflection was 0.15 ± 0.08 mm, and the maximum deviation was 0.52 mm. The maximum payload capability was 10 N. The suture task and ex vivo procedure verified the effectiveness of the instrument. Conclusions The proposed instrument has high dexterity and payload capability, which contributes to improving the quality of the RMIS procedures. This article is protected by copyright. All rights reserved.
Received: 15 April 2021
-
Revised: 12 September 2021
-
Accepted: 16 September 2021
DOI: 10.1002/rcs.2334
ORIGINAL ARTICLE
Design of a dexterous robotic surgical instrument with a
novel bending mechanism
Yingkan Yang
|Jianmin Li |Kang Kong |Shuxin Wang
Key Laboratory of Mechanism Theory and
Equipment Design of Ministry of Education,
Tianjin University, Tianjin, China
Correspondence
Kang Kong, Key Laboratory of Mechanism
Theory and Equipment Design of Ministry of
Education, Tianjin University, Tianjin 300354,
China.
Email: kongkang103@tju.edu.cn
Funding information
National Key R&D Program of China, Grant/
Award Numbers: 2019YFC0118003,
2019YFB1311502; National Natural Science
Foundation of China, Grant/Award Numbers:
51805362, 51875390
Abstract
Background: The robotassisted minimally invasive surgery (RMIS) has developed
rapidly in recent years, requiring highly articulated instruments to enable surgeons
to perform complicated and precise procedures.
Methods: A novel wristtype surgical instrument was proposed for RMIS. The wrist
consists of superelasticwiredriven snakelike joints and universal joints, which
could perform two deflections and one distal rotation. The bending mechanism and
the kinematics of universal joints were analysed. The forward and inverse kine-
matics of the wrist were derived.
Results: The performances of the instrument were evaluated using a prototype by
experiments. The average motion deviation of the wrist's deflection was 0.15 0.08
mm, and the maximum deviation was 0.52 mm. The maximum payload capability
was 10 N. The suture task and ex vivo procedure verified the effectiveness of the
instrument.
Conclusions: The proposed instrument has high dexterity and payload capability,
which contributes to improving the quality of the RMIS procedures.
KEYWORDS
bending mechanism, RMIS, robotic surgical instrument, universal joint
1
|
INTRODUCTION
Minimally invasive surgery (MIS) is usually performed by inserting
several rigid, slim instruments into a patient's body through three or
four small incisions with image guidance by an endoscopy. Compared
with open surgery, MIS has numerous benefits for patients, such as
less blood loss, less postoperative pain and shorter hospital stays.
1
However, there exist some drawbacks due to the small incisions,
including limited degrees of freedom (DOFs) of the instrument, hand
eye incoordination due to fulcrum effect, a long learning curve and
easy fatigue. To solve these problems, surgical robotic systems have
been developed to improve the quality of the MIS.
2
Dexterous sur-
gical instrument is the core component of the surgical robot, and its
performance determines the quality of the surgery.
To improve the dexterity of the instrument, a wrist mechanism
with two or more DOFs is usually added to its distal end.
3
Various
kinds of dexterous instruments have been designed for surgical ro-
botic system, as shown in Figure 1. To achieve compact structure and
remote transmission, the cabledriven mechanism is widely used in
the design of robotic surgical instruments. The multiDOF instrument
of the da Vinci system is a typical cablepulley mechanism, as shown
in Figure 1A.
4
It has a dexterous twoDOF (pitchyaw, PY) wrist and a
rotational shaft. A cablepulley drive instrument with a yawroll (YR)
wrist was described by Wang et al., and the rotation at the distal end
of the instrument is demonstrated to assist the performance of su-
turing and knotting, as shown in Figure 1B.
5
Cable guide channels,
which eliminate the need for pulleys and reduce the overall length to
facilitate miniaturization, were utilized to develop a dexterous and
compact surgical instrument, as shown in Figure 1C.
6
However, the
slip, friction and stretching of cables reduce the motion accuracy and
stability of the instrument.
16
To gain precise transmission and
enhanced stiffness, Wang et al. designed a surgical manipulator with
Int J Med Robot. 2021;e2334. wileyonlinelibrary.com/journal/rcs © 2021 John Wiley & Sons Ltd.
-
1 of 12
https://doi.org/10.1002/rcs.2334
a multigear array mechanism, as shown in Figure 1D,
7
and a motor
driven laparoscopic needle holder with the bevel gear mechanism
was designed by Dexterite Surgical, as shown in Figure 1E.
8
However,
the size of such instruments is limited by the size of gears, and their
transmission structure is complicated and bulky. Adding the link drive
mechanism is also an effective way to design surgical instruments
due to its characteristics such as high accuracy and high load capa-
bility. Hong et al. designed an instrument with a threeDOF parallel
link mechanism, as shown in Figure 1F.
9
It can perform deflection in
two directions and axial rotational motion of the forceps. A robotic
visible forceps with a seriallinkage mechanism was developed by
Zhang et al., as shown in Figure 1G.
10
The forceps can perform one
bending motion and 3 N lifting force at the tip. Though such in-
struments have high payload capability, their structures are complex
and bulky. Moreover, shape memory alloys (SMAs), pneumatic actu-
ators, hydraulic actuators and magnetic actuators have been used to
design dexterous surgical tools.
17
Although these methods allow the
tools to achieve high strength and dexterity, they are difficult to
implement precise control due to the nonlinearity.
In recent years, flexible joints are increasingly used in the design
of surgical instruments to enhance dexterity and compliance. A
diameter of 5 mm instrument with a snakelike wrist was developed
for the surgery in confined space, as shown in Figure 1H.
11
The wrist
can perform bending motions in two orthogonal planes. The instru-
ment of the da Vinci SP system has a doublejointed distal end with
elbow and snakelike wrist joints, as shown in Figure 1I.
12
The wrist
and elbow joints have twoDOF, respectively. A highly flexible and
compliant instrument was designed for the SPORT singleport ro-
botic surgical system, as shown in Figure 1J.
13
The instrument con-
sists of threesection continuum joints, and each section has two
bending movements. Francis et al. designed two miniaturized flexible
instruments for operations in small and confined space, such as
neurosurgery, head and neck surgery, and paediatrics, as shown in
Figure 1K.
14
The instrument has a diameter of 2 mm with a three
DOF wrist. It is highly dexterous and compliant, but less rigid. The
instrument with flexible joints has high dexterity and compliance, but
its rigidity is generally low, which hinders its precise operation and
payload capability. To address this problem, researchers have
explored many methods. A flexible instrument with enhanced stiff-
ness has been developed, and its wrist is composed of a precision
spring and four superelastic wires, as shown in Figure 1L.
15
Due to
the rigidity of the spring and wires, the stiffness of the instrument is
improved. However, the machined spring is easy to fatigue. Wang
et al. designed a controllable stiffness continuum with phase change
materials to enhance the stiffness of the instrument, but the
switching time of the stiffness is long.
18
Methods to increase cable
tension
19
or optimize cable path
20
have been proposed to increase
the payload capability. Nevertheless, these designs are complex and
occupy a substantial space, making it difficult to create a compact and
lightweight instrument.
To develop a dexterous and compact surgical instrument with
reliable payload capability, a fourDOF wristtype robotic surgical
instrument is proposed in this paper by using a novel cabledriven
bending mechanism with serial universal joints. First, the overview
FIGURE 1 Various instruments. (A) Cabledriven Endowrist
4
; (B) Microhand instrument
5
; (C) instrument with cable guide channels
6
;
(D) geardriven instrument
7
; (E) needle holder with bevel gear mechanism
8
; (F) parallellinkageactuated instrument
9
; (G) seriallinkage
actuated instrument
10
; (H) snakelike wrist Endowrist
11
; (I) da Vinci SP instrument
12
; (J) SPORT instrument
13
; (K) miniaturized flexible
instruments
14
; (L) flexible instrument with enhanced stiffness
15
2 of 12
-
YANG
ET AL.
and detail design of the proposed instrument are given. Then, the
forward and inverse kinematics of the wrist are discussed. Further-
more, the performances of the prototype are evaluated by experi-
ments. Finally, discussions and conclusions are provided.
2
|
MATERIALS AND METHODS
2.1
|
Mechanical design
A fourDOF robotic surgical instrument with a snakelike wrist was
developed, as shown in Figure 2. The instrument is composed of four
parts, including the endeffector, wrist, shaft and transmission
interface. The endeffector performs opening and closing motions to
grip and dissect tissues. The wrist can bend in two directions and
rotate around the distal end of the instrument. The bending and
distal rotation movements can adjust the position and orientation of
the instrument tip, which contributes to performing dexterous
operation. The direction of jaws of the endeffector can be adjusted
by the distal rotation with ease. A compact cabledriven mechanism
is adopted to carry out remote drive from the transmission interface
to the wrist and the endeffector. The instrument can be attached to
the minimally invasive surgical robot system by the transmission
interface to implement operations.
The wrist with a compact structure is designed by using snake
like joints integrated with serial universal joints, as shown in
Figure 2. The snakelike joints consist of four hinge joints with five
links and are driven by cables to realize bidirectional bending. The
axes of the two adjacent hinge joints are perpendicular to each other.
The detailed structure of the wrist is shown in Figure 3. Two yokes
and one spider make up one universal joint, which is equipped in the
centre of the hinge joints, as shown in Figure 3A. The rotation axis of
the universal joint is designed to coincide with the rotation axis of the
hinge joint all the time, which ensures that the motion of the uni-
versal joint is in sync with the motion of the hinge joint, as shown in
Figure 3B. Two pairs of antagonistic drive cables slide in holes evenly
distributed on the links, which control the wrist to bend in two di-
rections, as shown in Figure 3C. At the same time, the universal joint
can rotate around its axis to realize the distal rotation of the wrist, as
shown in Figure 3D. The distal end of the universal joint and the base
of the endeffector are fixedly connected by the set pin, as shown in
Figure 2. Thus, the rotation of universal joints can realize the rotation
of the endeffector.
The endeffector is used to perform clamping, cutting, dissec-
tion and coagulation in surgery. Most of the endeffectors have two
jaws doing pivotal motions relative to each other. The pivotal mo-
tion is usually generated by linear displacement. Concerning the
proposed endeffector, the required linear movement is produced
by the rotary actuator coupled with a cable that has a reset spring,
as shown in Figure 2. The pin slides along the slot to change the
angle of the two jaws around the pivot pin. The closing of the
endeffector is implemented by the tension of the cable, and
the opening is performed by the restoring force of the spring. The
relationship between the angular position of the jaws and the lon-
gitudinal variation of the cable can be derived by geometry, which
can be expressed as follows:
FIGURE 2 Structure of the robotic surgical instrument
YANG ET AL.
-
3 of 12
Δl¼sgsin φg1
sinφgþθg1=21
sinφgþθg=2;ð1Þ
where Δlimplies the longitudinal variation of the cable, θ
g
and θ
g1
imply two different arbitrary angular position of the jaws
(θ
g
θ
g1
), and s
g
and φ
g
are the design parameters of the sliding
slot.
The compact cabledriven structure is used in the instrument to
perform long distance transmission, as shown in Figure 4. The
bending joint is driven by two pairs of antagonistic cables. The ends
of cables are attached to the distal end of the wrist and the other
ends are attached to drive shafts to form a closedloop cabledriven
structure. The distal rotation joint is connected to a rigid rod, which is
also driven by a closedloop cabledriven structure. An openloop
cabledriven structure with a reset spring is used to drive the end
effector. In this structure, the cable is connected to the drive shaft
by passing through the central axis of the serial universal joints and
the rigid rod.
2.2
|
Cabledriven mechanism
The hinge joints are fundamental elements of the wrist, and the total
bending angle of the wrist depends on the number of hinge joints and
the bending angle of each hinge joint. Figure 5A illustrates the
bending of one hinge joint. The black lines represent links, the blue
and red lines represent cables, and the circle is the rotation axis.
When the joint bends θleftwards, the length of the left cable de-
creases to l
l
(θ), and the length of the right cable increases to l
r
(θ). The
lengths of cables after bending can be expressed as follows
8
>
>
<
>
>
:
llðθÞ ¼ 2hgcos θ
2dsin θ
2
lrðθÞ ¼ 2hgcos θ
2þdsin θ
2;ð2Þ
where h
g
is the distance from the rotation axis to the link and dis the
distance from the centre of the cable hole to the centre of the link.
Based on the Equation (2), the length variations of the two cables
can be expressed as follows
8
>
>
<
>
>
:
Δll¼llðθÞllð0Þ ¼ 2hgcos θ
21dsin θ
2
Δlr¼lrðθÞlrð0Þ ¼ 2hgcos θ
21þdsin θ
2:ð3Þ
According to the Equation (3), the sum of two length variations
(Δl
l
+Δl
r
) is less than zero. Consequently, the cable slack tends to
occur in the closedloop cabledriven structure, which reduces the
motion accuracy and stiffness of the wrist. In order to address this
problem, superelastic wires can be used as drive cables and back-
bones of the wrist. With the superelastic wires, the discrete hinge
joints can form a continuous joint, whose centreline length is
invariant.
19
Figure 5B shows the schematic of the hinge joint with
superelastic wires. Based on the geometrical analysis, the length
variations of the two superelastic wires can be expressed as follows
ΔLl¼ ðRdÞΘRΘ¼dΘ
ΔLr¼ ðRþdÞΘRΘ¼dΘ;ð4Þ
where Rrepresents the bending radius of the wrist and Θis the
bending angle of the distal end of the wrist. The sum of two length
variations (ΔL
l
+ΔL
r
) is zero, which contributes to the elimination of
cable slack and variable cable tensions.
21
2.3
|
Universal joint kinematics analysis
According to the analysis above, the wrist consists of four hinge
joints with the same number of universal joints. The universal joint is
not a constantvelocity joint, and it needs to satisfy certain conditions
to realize a constantvelocity output.
22
The kinematic schematic of
universal joints is shown in Figure 6. The relationship between the
input and output of the double universal joint can be deduced by
geometry, which can be expressed as follows
FIGURE 3 Structure of the wrist. (A) Front view; (B) side view;
(C) top view; (D) axonometric drawing
FIGURE 4 Transmission of the instrument
4 of 12
-
YANG
ET AL.
tan β3¼tanarctantan β1
cos θ1γ1þψ1cos θ2;ð5Þ
ω3
ω1¼cos θ1
1sin2θ1cos2β1
cos θ2
1sin2θ2sin2arctantan β1
cos θ1γ1þψ1;
ð6Þ
where β
1
is the input angle, β
3
is the output angle, ω
1
is the input
angular velocity, ω
3
is the output angular velocity, γ
1
is the angle
between the plane containing the input yoke and intermediate
yoke (S
1
O
1
S
2
) and the plane containing intermediate yoke and output
yoke (S
2
O
2
S
3
), ψ
1
is the phase angle of the two ends of the inter-
mediate yoke, θ
1
is the angle between the input yoke and the in-
termediate yoke, and θ
2
is the angle between the output yoke and
the intermediate yoke.
Because the rotation axis of the universal joint coincides with
that of the hinge joint, as shown in Figure 3, the angle between the
yokes of the universal joint is equal to the bending angle of the hinge
joint. Due to the perpendicular configuration of the two adjacent
hinge joints, as shown in Figure 2,γ
1
is 90°. In order to gain a stable
output, ψ
1
is set to 90°. Hence, Equations (5) and (6) can be rewritten
as follows
tan β3
tan β1¼cos θ2
cos θ1
;ð7Þ
ω3
ω1¼cos θ1cos θ2
cos 2θ1cos 2β1þcos 2θ2sin 2β1
:ð8Þ
Further, the relationship of input angle and output angle of three
universal joints can be calculated as
tanðβ490Þ ¼
tanarctantan β1cos θ2
cos θ1γ2þψ2
cos θ3
:ð9Þ
Based on the above analysis, the superelastic wires can make the
discrete hinge joints form a continuous joint. Hence, the hinge joints
bending in the same direction have the same angular movement.
According to the structure of the wrist, as shown in Figure 2, the
angular movements of the first joint (θ
1
) and the third joint (θ
3
) are
the same and the angular movements of the second joint (θ
2
) and the
fourth joint (θ
4
) are the same. Due to the structure of the wrist, γ
2
is
90°. Hence, ψ
2
is set to to gain a stable output. Equation (9) can be
rewritten as
tan β4¼tan β1cos θ2:ð10Þ
The relationship of input angle and output angle of four universal
joints can be deduced as
tan β5¼tanðarctanðtan β1cos θ2Þγ3þψ3Þ
cos θ4
:ð11Þ
According to the structure of the wrist, γ
3
is 90° and θ
2
is equal
to θ
4
. Hence, ψ
3
is set to 90° to gain a stable output. Equation (11)
can be rewritten as
tan β5¼tan β1:ð12Þ
Hence, the input angle (β
1
) is equal to the output angle (β
5
), and
the four universal joints could achieve a constantvelocity output
(ω
1
=ω
5
) with the configuration analysed above.
2.4
|
Wrist kinematics
The snakelike wrist consisting of multiple links has two bending
DOFs. The angles of the wrist are determined by the length varia-
tions of cables. The position and orientation of the distal end of the
wrist in the space can be confirmed with given angles of the wrist.
Consequently, the kinematics analysis can be divided into two parts:
the mapping between actuator space and configuration space and the
mapping between configuration space and task space.
FIGURE 5 Kinematic diagram of the hinge.
(A) Hinge with cables; (B) hinge with
superelastic wires
FIGURE 6 Kinematic schematic of universal joints
YANG ET AL.
-
5 of 12
2.4.1
|
Mapping between actuator space and
configuration space
As shown in Figure 7, there are two pairs of antagonistic drive cables
which control the wrist to bend in arbitrary direction in space. The
bending angle is Θand the bending direction angle is Φ, as shown in
Figure 8. The length variations of the four cables after bending can be
calculated as follows
8
>
>
>
>
>
>
>
>
>
>
>
>
<
>
>
>
>
>
>
>
>
>
>
>
>
:
ΔL1¼dcosπ
4ΦΘ
ΔL2¼dcos3π
4ΦΘ
ΔL3¼dcos5π
4ΦΘ
ΔL4¼dcos7π
4ΦΘ
:ð13Þ
According to the cable length variations, the bending direction
angle and the bending angle of the wrist can also be determined
8
>
>
>
>
<
>
>
>
>
:
Φ¼π
4þarctan 2ΔL2ΔL4
ΔL1ΔL3
Θ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
ðΔL1ΔL3Þ2þ ðΔL2ΔL4Þ2
q2d
:ð14Þ
The inverse mapping between the actuator space and the
configuration space is defined by Equation (13). The forward mapping
is given by Equation (14).
2.4.2
|
Mapping between configuration space and
task space
In consideration of the wrist with superelastic wires, the centreline
length of the wrist is invariant and the hinge joints bending in the
same direction have the same bending angle. Hence, the curvature of
the wrist can be approximately regarded as a constant. The piece
wise constant curvature model
23
can be used to calculate the kine-
matics of the wrist. The coordinate of the wrist is shown in Figure 8.
The position and orientation of the endeffector in the space can be
represented by the following matrix
T0
n¼2
6
6
4
nxsxoxdx
nysyoydy
nzszozdz
0 0 0 1
3
7
7
5;ð15Þ
where [n
x
n
y
n
z
]
T
, [s
x
s
y
s
z
]
T
and [o
x
o
y
o
z
]
T
represent orientation vec-
tors in the x
n
,y
n
and z
n
directions in the coordinate o
0
x
0
y
0
z
0
.
[d
x
d
y
d
z
]
T
represents the vector from o
0
to o
n
in the coordinate
o
0
x
0
y
0
z
0
.
The forward kinematics of the wrist can be derived from the
homogeneous matrix transformation, and the position and orienta-
tion in the space can be calculated as follows
8
>
>
>
>
>
>
>
<
>
>
>
>
>
>
>
:
dx¼Ltcos Φsin ΘLccos Φðcos Θ1Þ
Θ
dy¼Ltsin Φsin ΘLcsin Φðcos Θ1Þ
Θ
dz¼Ltcos ΘþLcsin Θ
Θ
;ð16Þ
8
<
:
nx¼cos θrcos Θ1cos2Φþ1þcos Φsin Φsin θrðcos Θ1Þ
ny¼sin θrcos Θcos2Φðcos Θ1Þþcos Φsin Φcos θrðcos Θ1Þ
nz¼cosðΦθrÞsin Θ
;
ð17Þ
8
<
:
sx¼cos Φsin Φcos θrðcos Θ1Þsin θrcos Θ1cos2Φþ1
sy¼cos θrcos Θcos2Φðcos Θ1Þcos Φsin Φsin θrðcos Θ1Þ
sz¼sinðΦθrÞsin Θ
;
ð18Þ
8
<
:
ox¼cos Φsin Θ
oy¼sin Φsin Θ
oz¼cos Θ
;ð19Þ
FIGURE 7 Two pairs of antagonistic drive cables on the wrist
FIGURE 8 Coordinate of the wrist
6 of 12
-
YANG
ET AL.
where Lcis the length of the wrist, Ltis the distance from the wrist to
the distal end of the endeffector and θ
r
is the angle of the distal
rotation joint. The inverse kinematics can be derived from the posi-
tion and orientation, which can be found as follows
8
>
>
>
>
>
>
>
>
>
<
>
>
>
>
>
>
>
>
>
:
Φ¼arctandyLtoy
dxLtox
Θ¼2 arctan0
@ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
ðdxLtoxÞ2þ ðdyLtoyÞ2
qdzLtoz1
A
θr¼arcsin szoxnzoy
o2
xþo2
y!
:ð20Þ
3
|
EXPERIMENTS AND RESULTS
3.1
|
Prototype design
Based on the above analysis, a prototype of the robotic surgical in-
strument was developed, as shown in Figure 9. The instrument is
compact and lightweight with a total length of 550 mm, a diameter of
9.5 mm and a weight of 180 g. The ranges of two bending motions are
both 50°, and the angle of the distal rotation joint can reach 360°.
The maximum opening angle of the endeffector is 60°. The param-
eters of the jaw's sliding slot s
g
and φ
g
are 4.4 mm and 33°.
Superelastic NiTi wires of 0.55 mm diameter were used to drive
bending motions and a 7 19 SUS304 cable with a diameter of
0.5 mm was used as the drive cable of the endeffector. Various in-
struments, including needle holders, forceps, monopolar instruments
and bipolar instruments, as shown in Figure 9C, have been developed
for different surgical procedures. The instrument can be connected to
the “MicroHand” surgical robot
5
to perform operations. “MicroHand”
is a masterslave robot consisting of a surgeon console and a slave
cart, as shown in Figure 9D.
3.2
|
Motion accuracy
To measure the accuracy of the designed instrument, a motion ac-
curacy test was performed, as shown in Figure 10. The instrument
was actuated by a control device. The programmable multiaxis
controller (PMAC) processed the input commands from the panel
and output commands to motors to drive the corresponding joints of
the instrument. The instrument was driven repeatedly three times
within 50° in one direction and the position of the distal end was
measured with an electromagnetic tracking system (NDI Aurora,
Waterloo, Canada). The tracking sensor was attached to the end of
the instrument, as shown in Figure 10.
The deviation between the actual trajectory and desired tra-
jectory and the backlash hysteresis of the instrument were
measured. The motion trajectory is shown in Figure 11. The average
deviation was 0.15 0.08 mm, and the maximum deviation was 0.52
mm. The backlash hysteresis is shown in Figure 12. The average and
maximum hysteresis of the prototype were 4.46 2.34° and 7.78°
respectively.
3.3
|
Payload capability
To evaluate the payload capability of the instrument, the payloads at
various angles in one direction were tested by applying vertical
forces at the distal end of the instrument and measuring the
FIGURE 9 Prototype of the robotic surgical instrument. (A) Prototype of the instrument; (B) closeup view of the wrist; (C) various
instruments; (D) proposed instrument mounted on “MicroHand” surgical robot
YANG ET AL.
-
7 of 12
displacements. The distal end of the instrument was connected to a
dynamometer through a stainless cable and the instrument was fixed
with a rigid structure, as shown in Figure 13. According to related
research, the mean of average forces for retraction with grasping is
around 5 N and the force to lift onethird of an average human liver is
more than 4 N.
24,25
Hence, the maximum load in the test was set to 5
N which was sufficient for surgical operations. The results of the
payload capability test are shown in Figure 14. The largest defor-
mation of the wrist occurred at the horizontal position with a stiff-
ness of about 1.19 N/mm and the smallest deformation occurred at
the angular position of 50° with a stiffness of about 2.88 N/mm. The
extreme payload capability of the wrist at the horizontal position was
about 10 N, as shown in Figure 13.
3.4
|
Suture test
A suture task was carried out to evaluate the dexterity of the in-
strument for performing complicated operations. The “MicroHand”
surgical robot was used as the operating system for the proposed
FIGURE 11 Motion trajectory
FIGURE 12 Backlash hysteresis
FIGURE 10 Motion accuracy test
FIGURE 13 Payload capability test
FIGURE 14 Results of the payload capability test
8 of 12
-
YANG
ET AL.
instrument in this study. The task is to perform a running suture in
horizontal and vertical directions respectively. Four pairs of dots are
marked symmetrically on both sides of the line, and the diameter of
the dots is approximately 2 mm and the distance between the two
dots is 10 mm, as shown in Figure 15. In the test, the performance of
the proposed robotic instrument was compared with that of the
conventional manual laparoscopic instrument. The task performance
metrics are completion time and operation quality. The operation
quality is scored by the numbers of successful sutures within 5 min. A
successful suture is scored as one point, and nonsuccessful suture is
not scored. It is a successful suture that starts inside the entry dot
and ends in the exit dot, without digression of the needle from the
dot. Ten participants from Tianjin University were invited to partic-
ipate in the test. None of these participants was a lefthand user and
had experience in MIS operation. The participants were trained to
acquire basic laparoscopic skills and got familiar with the proposed
instrument and the conventional instrument before the test. The
experiment setup and operation condition were completely the same
for all participants. The data results were compared using Wilcoxon
signedranktest, a value of p<0.05 was considered significant.
Statistical analysis was run with SPSS version 19 software.
The average time of the horizontal suturing with the robotic
instrument was 221.0 68.29 s, and the average time for the lapa-
roscopic instrument was 282.3 32.89 s, which represents a sig-
nificant difference (p=0.007). The completion time of the vertical
suturing of the robotic instrument was significantly shorter than that
of the laparoscopic instrument (268.6 60.19 s vs. 371.4 77.74,
p=0.012). Compared with the laparoscopic instrument, the robotic
instrument performed a higher quality of the horizontal suturing
(2.3 1.25 vs. 1.7 0.67, p=0.196), and the vertical suturing
(2.0 0.67 vs. 1.6 0.84, p=0.102).
3.5
|
Ex vivo test
Furthermore, an ex vivo singlelayer continuous anastomosis test
with a porcine intestine was performed using the designed instru-
ment with the “MicroHand” system. The test aimed to evaluate the
usability of the instrument in the clinical procedures. A surgeon with
the experience of laparoscopic surgery was invited to perform the
test. Before the test, the surgeon had 30 min to get familiar with the
instrument and robotic system. During the test, as shown in
Figure 16, the ends of the two pieces of intestines were pulled
together using a needle holder on the right and a grasper on the left.
Then, a running suturing was performed to close the wound and the
thread was pulled tight, and a surgical knot was placed to secure the
closure. In the procedure, the suturing and knottying were
performed fluently. The needle could be held stably, and the tissue
could be clamped gently. The anastomosis procedure lasted for
approximately 40 min. The related research shows that the time of
the robotassisted laparoscopic intestinal anastomosis is about 25–
55 min and the time of the standard laparoscopic suturing is
approximately 90 min.
26
Hence, the anastomosis time of the ex vivo
test is acceptable.
4
|
DISCUSSION
The purpose of this study is to provide a dexterous and compact
surgical instrument with high payload capability for robotassisted
minimally invasive surgery (RMIS). The main contributions and fea-
tures of this study include: (1) a novel threeDOF wrist structure with
hinge joints and serial universal joints is proposed; (2) the distal
rotation motion with a large range enhances the ability to perform
complicated operations, especially sutures; (3) the wrist has a high
payload capability which improves the safety of the surgery; (4)
various instruments have been developed for different surgeries.
The comparison between this paper and previous studies has
been shown in Table 1. For many instruments,
4–8,1013,15
the wrist
generally has two DOFs. The proposed wrist has three DOFs, and its
movement is more dexterous. The rotation range (360°) is larger
than that of other instruments. The dexterity of the instrument has
been demonstrated by the suture and ex vivo test. Compared with
the cabledriven
46
and geardriven mechanisms,
7,8
the proposed
wrist has advantages in terms of the rotation range and payload
capability; Compared with the linkagedriven mechanism,
9,10
the
proposed wrist has a large distal rotation range; Compared with the
flexiblejointed mechanism,
1115
the proposed wrist has a larger
rotation range and a higher payload capability. Related studies have
shown that the snakelike Endowrist and SP instruments challenge
the surgeons due to the reduced strength.
27,28
Hence, their payload
capability is not high. Compared with the superelasticwiredriven
mechanism,
13,15
the proposed wrist mechanism has a higher rigidity
due to the composite structure with hinge joints and universal joints.
Overall, the proposed instrument has advantages over previous
studies in terms of rotation range and payload capability.
The structure of the wrist adopts a modular design method,
which makes it have good scalability. According to Figure 3, the
combination of a hinge joint and a universal joint can be considered
as a submodule, and several submodules make up a module. When
the axes of two adjacent hinge joints are perpendicular to each other,
the module can perform three DOFs (pitchyawroll, PYR)
FIGURE 15 Scene of the suture task
YANG ET AL.
-
9 of 12
simultaneously; when the axes of two adjacent hinge joints are par-
allel to each other, the module performs two DOFs (yawroll, YP)
simultaneously. Furthermore, two or more modules can be connected
as a multiDOF continuum arm, which is well suited to confined
spaces, especially for working in MIS, single port access surgery and
natural orifice transluminal endoscopic surgery.
The result of the motion accuracy test shows that the average
motion deviation of the instrument is 0.15 0.08 mm. In the delicate
suture of laparoscopic surgery, the diameter of the vessel is generally
greater than 1 mm, therefore, the motion accuracy of the instrument
can meet the needs of delicate surgical operations. Hence, the motion
of the instrument is precise. The deviation and hysteresis of the in-
strument are mainly caused by the following reasons. Frist, the
diameters of the holes and cables are 0.6 and 0.55 mm, respectively.
There are clearances between the holes and cables, which affects the
length variation that the cables should have when the wrist moves.
Second, the friction between the joints decreases the motion accu-
racy. Hence, reducing the clearances between the holes and cables
and decreasing the friction between the joints will improve the mo-
tion accuracy of the instrument. The control method is also an
important way affecting the motion accuracy. Hence, optimized
kinematic algorithm, dynamical model, feedback control, motion and
friction compensation will be considered to improve motion accuracy
in the future work. Furthermore, backlash hysteresis causes the
nonlinear motion, which makes precise control of the instrument
difficult. Various mathematical models for hysteresis such as Pre-
isach, Krasnosel'skii–Pokrovskii, Prandtl–Ishlinskii, MaxwellSlip,
Bouc–Wen and Duhem, have been reviewed and analysed by the
related reference.
29
Several backlash hysteresis models of cable
conduit actuators or cablepulley mechanisms have been proposed
and used to compensate for the hysteresis behaviour.
3032
Hence, to
avoid nonlinearities, the backlash hysteresis model and compensation
control of the instrument need to be analysed and implemented in
future work. Based on the analysis in Section 3.3, a 5 N payload
capability on the tip of instrument is required during a typical surgical
procedure. According to the payload test, the instrument has a high
stiffness and can take on a load of 10 N, which meets the clinical
requirements and ensures the stability of the surgical operation. The
results of the suture test show that the proposed robotic instrument
can perform the suturing in an awkward position which is difficult for
the conventional instrument. The distal rotation joint of the proposed
instrument makes it easy to adjust the posture of the needle, which
contributes to improving the performance of the suturing. The users
can complete the suturing with a high quality in a short time, which
FIGURE 16 Scene of the intestinal
anastomosis
TABLE 1Comparison between this paper and previous studies
Driving mechanism Diameter (mm) WristDOF Bending range Rotation range Payload capability
Endowrist
4
Cablepulley 8 PY 90° 270° High
Wang et al.
7
Geardriven 9.5 PY 80° Medium (2.5 N)
Hong et al.
9
Linkagedriven 8 PYR 64° 60° High (10 N)
Snakelike Endowrist
11
Cabledriven 5 PY 90° 270° Medium
27
SP instrument
12
Cabledriven 6 PY 90° 270° Medium
28
SPORT instrument
13
Superelasticwiredriven 8 PY 90° Medium (3.25 N)
Francis et al.
14
Concentric tube 2 RPR 70° 250° Low (0.26 N)
Haraguchi et al.
15
Superelasticwiredriven 10 PY 90° High (5 N)
This study Superelasticwiredriven 9.5 PYR 50° 360° High (10 N)
10 of 12
-
YANG
ET AL.
indicates that the proposed instrument can reduce the difficulty of
suturing and the learning time of complicated operations. The
feasibility of the instrument has been demonstrated by the ex vivo
anastomosis procedures. The instrument is dexterous enough to
perform the suturing and knotting. The force of the instrument is
sufficient to lift the tissue and tighten the suture stably. In consid-
eration of the requirements of clinical sterilization, the materials of
the instrument, including stainless steel, NiTi wires, carbon fibre and
polyetherimide plastic, are all biocompatible and autoclavable.
Hence, the proposed instrument can be sterilized by autoclave (134°
C, 201.7229.3 kPa).
However, some limitations of the proposed instrument need to
be further optimized. The diameter of the instrument is 9.5 mm
which satisfies the demands of the MIS, but it is still larger than that
of 5 and 8 mm instrument. Due to the constraint of the universal
joint, the range of the bending motion is 50°, which is not very
large compared with some robotic instruments.
4,5
Though the
payload test has investigated the stiffness of the instrument at
various angles in one deflecting direction, the payload capability with
different postures needs to be tested. In future works, the operating
angle of the universal joint will be increased to expand the motion
range of the bending joint. The instrument will be miniaturized,
tested by more comprehensive experiments and evaluated by the
animal experiment.
5
|
CONCLUSION
A novel wristtype fourDOF robotic surgical instrument for MIS is
presented. The wrist can perform three DOFs and provides high
payload capability with a compact structure. The results of the per-
formance evaluation experiments demonstrate that the dexterity,
payload capability, accuracy and feasibility of the designed instru-
ment satisfy the requirements of the MIS. The instrument can pro-
vide effective dexterity and operation force for surgical procedures,
and reduces the difficulty of suturing, which conduces to improving
the efficiency and safety of the MIS. The limitations of the instru-
ment, especially miniaturization and bending range, will be modified
and more comprehensive analysis and evaluation will be imple-
mented in the future.
ACKNOWLEDGEMENTS
This work was supported by the National Key R&D Program of China
(grant no. 2019YFC0118003 and grant no. 2019YFB1311502), the
National Natural Science Foundation of China (grant no. 51805362
and grant no. 51875390).
CONFLICT OF INTEREST
The authors have no conflicts of interest to declare.
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the
corresponding author upon reasonable request.
ORCID
Yingkan Yang
https://orcid.org/0000-0003-2525-3269
Kang Kong https://orcid.org/0000-0001-5646-8807
REFERENCES
1. Tonutti M, Elson DS, Yang G, Darzi AW, Sodergren MH. The role of
technology in minimally invasive surgery: state of the art, recent
developments and future directions. Postgrad Med J. 2017;93:
159167.
2. Chen Y, Zhang s, Wu Z, Yang B, Luo Q, Xu K. Review of surgical
robotic systems for keyhole and endoscopic procedures: state of the
art and perspectives. Front Med. 2020;14:382403.
3. Jelínek F, Arkenbout EA, Henselmans PWJ, et al. Classification of
joints used in steerable instruments for minimally invasive surgery: a
review of the state of the art. J Med Device. 2015;9(1):01080101.
4. Freschi C, Ferrari V, Melfi F, Ferrari M, Mosca F, Cuschieri A.
Technical review of the da Vinci surgical telemanipulator. Int J Med
Robot Comp. 2013;9(4):396406.
5. Wang W, Li J, Wang S, Su H, Jiang X. System design and animal
experiment study of a novel minimally invasive surgical robot. Int J
Med Robot Comp. 2016;12(1):7384.
6. Podolsky DJ, Diller E, Fisher DM, et al. Utilization of cable guide
channels for compact articulation within a dexterous three
degreesoffreedom surgical wrist design. J Med Device. 2018;13(1):
01100301.
7. Wang HB, Wang SX, Zuo SY. Development of visible manipulator
with multigear array mechanism for laparoscopic surgery. IEEE
Robot Autom Lett. 2020;5(2):30903097.
8. SánchezMargallo JA, SánchezMargallo FM. Initial experience using
a roboticdriven laparoscopic needle holder with ergonomic handle:
assessment of surgeons’ task performance and ergonomics. Int J
CARS. 2017;12:20692077.
9. Hong MB, Jo YH. Design of a novel 4DOF wristtype surgical in-
strument with enhanced rigidity and dexterity. IEEEASME T Mech.
2014;19(2):500511.
10. Zhang B, Liao Z, Yang P, et al. Robotic visible forceps manipulator
with a novel linkage bending mechanism. J Mech Robot. 2019;11(1):
011012.
11. Gobbi D, Midrio P, Gamba P. Instrumentation for minimally invasive
surgery in pediatric urology. Transl Pediatr. 2016;5(4):186204.
12. Maurice MJ, Ramirez D, Kaouk JH. Robotic laparoendoscopic single
site retroperitioneal renal surgery: initial investigation of a purpose
built singleport surgical system. Eur Urol. 2017;71(4):643647.
13. Robert R, Zitnick DA, Cameron PJK, et al. inventor; Titan Medical
Inc., assignee. Articulated tool positioner and system employing
same. United States patent, US 17242371. 2021.
14. Francis P, Eastwood KW, Bodani V, et al. Miniaturized instruments
for the da Vinci research kit: design and implementation of custom
continuum tools. IEEE Robot Autom Mag. 2017;24(2):2433.
15. Haraguchi D, Kanno T, Tadano K, Kawashima K. A pneumatically
driven surgical manipulator with a flexible distal joint capable of
force sensing. IEEEASME T Mech. 2015;20(6):29502961.
16. Miyasaka M, Haghighipanah M, Li Y, Matheson J, Lewis A, Hanna-
ford B. Modeling cabledriven robot with hysteresis and cablepulley
network friction. IEEEASME T Mech. 2020;25(2):10951104.
17. Le HM, Do TN, Phee SJ. A survey on actuatorsdriven surgical ro-
bots. Sensor Actuat A. 2016;247:323354.
18. Wang J, Wang S, Li J, Ren X, Briggs RM. Development of a novel
robotic platform with controllable stiffness manipulation arms for
laparoendoscopic singlesite surgery (LESS). Int J Med Robot Comp.
2018;14(1):e1838.
19. Kim YJ, Cheng S, Kim S, Iagnemma K. A stiffnessadjustable hyper-
redundant manipulator using a variable neutralline mechanism for
minimally invasive surgery. IEEE T Robot. 2014;30(2):382395.
YANG ET AL.
-
11 of 12
20. You JM, Kim H, Kim J, Kwon DS. Design and analysis of high
stiffness hyperredundant manipulator with sigmashaped wire path
and rolling joints. IEEE Robot Autom Lett. 2021;6(4):73577364.
21. Suh JW, Kim KY. Harmonious cable actuation mechanism for soft
robot joints using a pair of noncircular pulleys. J Mech Robot. 2018;
10(6):061002.
22. Zhang GQ, Du JJ, To S. Study of the workspace of a class of universal
joints. Mech Mach Theory. 2014;73:244258.
23. Webster RJ, Jones BA. Design and kinematic modeling of constant
curvature continuum robots: a review. Int J Robot Res. 2010;29(13):
16611683.
24. Golahmadi AK, Khan DZ, Mylonas GP, Marcus HJ. Tooltissue forces
in surgery: a systematic review. Ann Surg. 2021;65:102268.
25. Ishii C, Kobayashi K, Kamei Y, Nishitani Y. Robotic forceps manipulator
with a novel bending mechanism. IEEEASME T Mech. 2010;15(5):671684.
26. Ruurda JP, Broeders IA. Robotassisted laparoscopic intestinal
anastomosis. Surg Endosc. 2003;17(2):236241.
27. White MA, Autorino R, Spana G, et al. Robotic laparoendoscopic
singlesite radical nephrectomy: surgical technique and comparative
outcomes. Eur Urol. 2011;59:815822.
28. Moschovas MC, Bhat S, Rogers T, et al. Technical modifications
necessary to implement the da Vinci singleport robotic system. Eur
Urol. 2020;78:415423.
29. Hassani V, Tjahjowidodo T, Do TN. A survey on hysteresis
modeling, identification and control. Mech Syst Signal Pr. 2014;49(1
2):209233.
30. Do TN, Tjahjowidodo T, Lau MWS, Phee SJ. Nonlinear friction
modelling and compensation control of hysteresis phenomena for a
pair of tendonsheath actuated surgical robots. Mech Syst Signal Pr.
2015;60:770784.
31. Liang YL, Du ZJ, Wang WD, Sun L. A novel position compensation
scheme for cablepulley mechanisms used in laparoscopic surgical
robots. Sensors. 2017;17(10):2257.
32. Baek D, Seo JH, Kim J, Kwon DS. Hysteresis compensator with
learningbased hybrid joint angle estimation for flexible surgery
robots. IEEE Robot Autom Lett. 2020;5(4):68376844.
How to cite this article: Yang Y, Li J, Kong K, Wang S. Design
of a dexterous robotic surgical instrument with a novel
bending mechanism. Int J Med Robot. 2021;e2334. doi:10.
1002/rcs.2334
12 of 12
-
YANG
ET AL.
... To our knowledge, there was no previous work that developed an instrument that achieves a comparable bending angle. Several previous works reported instruments that could achieve bending angles from 50 to 90 • [27,[44][45][46][47]. In the work by Culmone et al., an instrument that facilitates the insertion of additional endoscopic instruments was developed [15]. ...
... Ad-ditionally, instead of bending force, several previous works focused exclusively on rigid laparoscopic instruments and reported on a similar metric called payload capability. In their cases, this metric ranged from 0.26 to 10 N [15,[47][48][49]. One noticeable difference is that payload capacity considers a range of tolerable deflection of the instrument under different loads that are applied to its tip. ...
Article
Full-text available
(1) Background: Interventional endoscopic procedures are growing more popular, requiring innovative instruments and novel techniques. Three-dimensional printing has demonstrated great potential for the rapid development of prototypes that can be used for the early assessment of various concepts. In this work, we present the development of a flexible endoscopic instrument and explore its potential benefits. (2) Methods: The properties of the instrument, such as its maneuverability, flexibility, and bending force, were evaluated in a series of bench tests. Additionally, the effectiveness of the instrument was evaluated in an ex vivo porcine model by medical experts, who graded its properties and performance. Furthermore, the time necessary to complete various interventional endoscopic tasks was recorded. (3) Results: The instrument achieved bending angles of ±216° while achieving a bending force of 7.85 (±0.53) Newtons. The time needed to reach the operating region was 120 s median, while it took 70 s median to insert an object in a cavity. Furthermore, it took 220 s median to insert the instrument and remove an object from the cavity. (4) Conclusions: This study presents the development of a flexible endoscopic instrument using three-dimensional printing technology and its evaluation. The instrument demonstrated high bending angles and forces, and superior properties compared to the current state of the art. Furthermore, it was able to complete various interventional endoscopic tasks in minimal time, thus potentially leading to the improved safety and effectiveness of interventional endoscopic procedures in the future.
... Arata et al. [16] developed a deformable wrist in which elastic strips are used instead of cables to increase rigidity, but the strips are susceptible to early fatigue. The use of superelastic wire-driven surgical instruments has been explored in [17], [18]. Haraguchi et al. proposed a flexible instrument composed of precision springs and four superelastic wires. ...
... Replacing Eq. (19) in Eq. (18) for b and c we get the following relation: ...
Article
Full-text available
Robot-assisted minimally invasive surgery (RMIS) has been shown to be effective in improving surgeon capabilities, providing magnified 3D vision, highly dexterous surgical tools, and intuitive human-robot interfaces for high-precision tool motion control. Robotic surgical tools (RST) are a critical component that defines the performance of an RMIS system. Current RSTs still represent a high cost, with few commercially available options, which limits general access and research on RMIS. We aim to take advantage of recent progress in biocompatible 3D printing and contribute to the development of RMIS technologies, presenting an open platform for low-cost, biocompatible, and customizable RSTs. The proposed design concept consists of a 3-DOF end-effector with a decoupled wrist mechanism, a tool interface module, and a tool drive unit. We validated our end-effector design using Finite Element Analysis (FEA) to confirm that stress generated by high grip forces is maintained below the material yield stress. Validation experiments showed that the proposed RST could provide up to 10N grip forces and up to 3N pulling forces. The proposed control framework exhibited a mean absolute positioning tracking error of approximately 0.1 rad. Finally, we also demonstrated the use of the proposed RST in two surgical training tasks: pick-and-place and stitching. The designs and software control framework are open-access and freely available for customization and fast development at https://github.com/jcolan/OpenRST.
... Flexible manipulators have become essential parts of robot-assisted gastrointestinal endoscopic surgery (GES) [1], such as flexible parallel robots [2] [3], single-backbone continuum robots [4] [5], and redundant robots [6]. The flexibility of the manipulators can adjust the long and winding working channel of GES well but also increase the difficulty of kinematic modeling. ...
Conference Paper
In the study of minimally invasive surgical robots, a mini parallel continuum robot has shown motion advantage after passing through a long and winding working channel. However, due to the interaction force between the elastic wires of the parallel robots during motion generation processes, the constant curvature assumption has shown modeling errors. This causes the current geometric kinematic model to become unreliable. Therefore, there is a need for a more accurate kinematic model in the absence of a complicated static model. This paper aims to solve this issue. The simulation in ANSYS is carried out, and the shape of one of the driving wires, when bending, is fitted by a two-segment polynomial curve. Then, the position of the distal wrist tip can be calculated based on the curve shape. To verify the accuracy of the proposed model, bending simulation and experiment are carried out. The accuracy of the proposed model is compared with that of the kinematic model based on constant curvature assumption. The result shows that the proposed model can get more accurate results, especially when the driving wire displacement increases. For a 10 mm parallel robot, when the displacements of the two pairs of wires are both 3.0 mm, the errors of the two models are 0.42 mm and 5.79 mm (4.2% and 57.9%), respectively.
... The dexterity of MIRIs is increased when adopting bending mechanisms, such as the one proposed in [21]. The cable-driven mechanism is used to actuate the MIRI and the bending mechanism, in the base of its design, and allows for the performance of predefined movements modifying the workspace. ...
Article
Full-text available
Cable-driven mechanism models are, usually, included in actuated systems; however, recently, their use for sensitive systems has been explored. In this paper, two cable-driven multi-body mechanism models are compared, underlining advantages and constraints in using sensitive cable-driven mechanisms for minimally invasive robotic instruments. The proposed approach could be useful in bypassing sterilization problems for surgical robotic instruments because our system allows for the separation of the robotic sterilizable part from the sensitive-actuated part of the surgical instrument. The real implementation of the proposed mechanism models, presented partially in other works, are validated in this paper, performing a simulation using a multi-body environment. Results confirm the feasibility of the proposed sensitive-actuated approach, defining new bases for the next challenges of the future.
... To offer a solution for these weaknesses, robotic systems for SILS were developed. The advantages of using robots in single incision laparoscopic surgery are presented in [13], and include excellent accuracy, repeatability, and the ability to work in a limited operating field enabling the use of highly dexterous instruments with multiple bends [14][15][16]. ...
Article
Full-text available
Augmented Reality (AR) became a very useful tool for various domains, with multiple applications in different medical fields, where this technology is integrated in medical training, image-based diagnosis, simulation of medical tasks, or in guiding systems for surgical procedures. The introduction of AR technology in robotic assisted surgery is relatively new and challenging considering the required precision for this type of medical task. The objective of this study is to develop an AR simulator for robotic assisted Single Incision Laparoscopic Surgery (SILS) where the robot performing the procedure is embedded in a virtual environment, providing both an interactive training and a simulation environment. The architecture and particularities of the proposed robot for SILS are analyzed for a better understanding of the structure, to obtain the validation of the kinematic model. The final result is represented in the form of an application which can be operated with Hololens 2 device.
... However, a serial linkage-based wrist has a finite offset between the rotation axes, resulting in inconsistent coordination compared with a human wrist. To solve this problem, different kinds of bending mechanism with 2 DoFs (pitch-yaw) were developed, such as cable-driven sphere type [4], snake-bone type [5], flexible tube (e.g. hypotube) [6]. ...
Preprint
Gastrointestinal endoscopic surgery (GES) has high requirements for instruments' size and distal dexterity, because of the narrow endoscopic channel and long, tortuous human gastrointestinal tract. This paper utilized Nickel-Titanium (NiTi) wires to develop a miniature 3-DoF (pitch-yaw-translation) flexible parallel robotic wrist (FPRW). Additionally, we assembled an electric knife on the wrist's connection interface and then teleoperated it to perform an endoscopic submucosal dissection (ESD) on porcine stomachs. The effective performance in each ESD workflow proves that the designed FPRW has sufficient workspace, high distal dexterity, and high positioning accuracy.
Article
Background Accurately estimating the 6D pose of snake‐like wrist‐type surgical instruments is challenging due to their complex kinematics and flexible design. Methods We propose ERegPose, a comprehensive strategy for precise 6D pose estimation. The strategy consists of two components: ERegPoseNet, an original deep neural network model designed for explicit regression of the instrument's 6D pose, and an annotated in‐house dataset of simulated surgical operations. To capture rotational features, we employ an Single Shot multibox Detector (SSD)‐like detector to generate bounding boxes of the instrument tip. Results ERegPoseNet achieves an error of 1.056 mm in 3D translation, 0.073 rad in 3D rotation, and an average distance (ADD) metric of 3.974 mm, indicating an overall spatial transformation error. The necessity of the SSD‐like detector and L1 loss is validated through experiments. Conclusions ERegPose outperforms existing approaches, providing accurate 6D pose estimation for snake‐like wrist‐type surgical instruments. Its practical applications in various surgical tasks hold great promise.
Article
In endoscopic submucosal dissection (ESD), the gastrointestinal (GI) tract warrants the surgical instruments to navigate through a long, narrow and tortuous endoscope. This poses a great challenge in developing ESD instruments with small dimensions, flexibility, and high distal dexterity. In this work, we propose the first Transendoscopic Flexible Parallel Continuum Robotic mechanism to develop a miniature dexterous flexible-stiff-balanced Wrist (FPCW). Besides, it can steer multifunctional instruments of diameters 2.5 mm to 3.5 mm, including the electrosurgical knife, injection needle, and forceps. Our FPCW instruments are adaptable to commercially available dual-channel endoscopes (diameter: <12 mm, channel width: 2.8 mm and around 3.8 mm). Furthermore, we develop a surgical telerobotic system, called DREAMS (Dual-arm Robotic Endoscopic Assistant for Minimally Invasive Surgery), by using our smallest FPCW instruments for bimanual ESD procedures. First, we conduct a series of experiments to determine the FPCW’s design and kinematics parameters and to verify the mechanical properties of the FPCW instruments’ prototypes, including workspace, stiffness, strength, and teleoperation accuracy. Second, we validate the functionality of the FPCW instruments through ex-vivo tests by performing ESD steps on porcine stomachs. Finally, we perform an invivo test on a live porcine model and showcase that our developed DREAMS can be teleoperated intuitively to perform bimanual ESD efficiently with an average dissection speed of 108.95 mm ² /min at the greater curvature in gastric body, which demonstrates that our DREAMS has satisfactory maneuverability as well as accuracy and is more competitive than counterpart robotic systems.
Article
Full-text available
High stiffness is desirable for wire-driven underactuated hyperredundant manipulators used in minimally invasive surgery. However, limited structural stiffness can cause lateral and angular deflections by the lateral force and moment applied to the distal end of the manipulator. This paper proposes a -shaped wire path, which is a combination of parallel and convergent wire paths, for a hyperredundant manipulator with rolling joints. The -shaped wire path applied to the manipulator has the ability to maintain the position and orientation of its distal end for an applied lateral force on the surgical instrument tip. The simple mechanics of deflection of the manipulator with the -shaped wire path are described, and straight-pose experiments were conducted to verify the deflection characteristics of the -shaped wire path. An arbitrary bending-pose experiment was also conducted to verify the stiffness performance. It was observed that the proposed method could enhance the stiffness of the manipulator in both straight and arbitrarily bent poses.
Article
Full-text available
Background Excessive tool-tissue interaction forces often result in tissue damage and intraoperative complications, while insufficient forces prevent the completion of the task. This review sought to explore the tool-tissue interaction forces exerted by instruments during surgery across different specialities, tissues, manoeuvres and experience levels. Materials & methods A PRISMA-guided systematic review was carried out using Embase, Medline and Web of Science databases. Results Of 462 articles screened, 45 studies discussing surgical tool-tissue forces were included. The studies were categorized into 9 different specialities with the mean of average forces lowest for ophthalmology (0.04N) and highest for orthopaedic surgery (210N). Nervous tissue required the least amount of force to manipulate (mean of average: 0.4N), followed by epithelial whilst connective tissue (including bone) required the most (mean of average: 45.8). For manoeuvres, drilling recorded the highest forces (mean of average: 14N), whilst sharp dissection recorded the lowest (mean of average: 0.03N). When comparing differences in the mean of average forces between groups, novices exerted 22.7% more force than experts, and presence of a feedback mechanism (e.g. audio) reduced exerted forces by 47.9%. Conclusions The measurement of tool-tissue forces is a novel but rapidly expanding field. The range of forces applied varies according to surgical speciality, tissue, manoeuvre, operator experience and feedback provided. Knowledge of the safe range of surgical forces will improve surgical safety whilst maintaining effectiveness. Measuring forces during surgery may provide an objective metric for training and assessment. Development of smart instruments, robotics and integrated feedback systems will facilitate this.
Article
Full-text available
In this paper, a method of modeling robotic systems with closed circuit cable-pulley transmissions using the independently developed hysteretic cable stretch model and cable-pulley network friction model is presented. The hysteretic cable stretch model captures the responses of the cable including elasticity, internal friction, material damping, and hysteresis. The cable-pulley network friction is a function of cable tension, average individual wrap angle of cable around pulley, and number of pulleys. For the verification of the method, these two models are integrated into the dynamic model of the first 3 joints of the RAVEN II robotic surgery platform. The parameters of the developed model are specifically tuned for the RAVEN II and the performance of the model is compared against the kinematic model and a previous, simplified dynamic model (a model with exponential cable stretch and linear damping model). The result showed that even though significant improvements were not observed for the first 2 joints, the average error and the maximum error of the 3rd joint which uses smaller diameter and longer cable and more guide pulleys could be reduced by 10-20% and 5-10% respectively over the previous, simplified dynamic model. Also, analysis with various input trajectories indicated that cable driven systems with very low speed, frequent change in the loading condition
Article
Full-text available
Hysteresis causes difficulties in precisely controlling motion of flexible surgery robots and degrades the surgical performance. In order to reduce hysteresis, model-based feed-forward and feedback-based methods using endoscopic cameras have been suggested. However, model-based methods show limited performance when the sheath configuration is deformed. Although feedback-based methods maintain their performance regardless of the changing sheath configuration, these methods are limited in practical situations where the surgical instruments are obscured by surgical debris, such as blood and tissues. In this paper, a hysteresis compensation method using learning-based hybrid joint angle estimation (LBHJAE) is proposed to address both of these situations. This hybrid method combines image-based joint angle estimation (IBJAE) and kinematic-based joint angle estimation (KBJAE) using a Kalman filter. The proposed method can estimate an actual joint angle of a surgical instrument as well as reduce its hysteresis both in the face of partial obscuration and in different sheath configurations. We use a flexible surgery robot, K-FLEX, to evaluate our approach. The results indicate that the proposed method has effective performance in reducing hysteresis.
Article
Full-text available
Various slim robots, such as surgical robots or humanoid robot fingers, are remotely actuated using transmission cables. If pull-pull drive is applied to actuate them using circular driving spools regardless of the shape of the joints, the tension of the driving cable becomes difficult to be maintained properly. Therefore, it is possible to solve such a cable slack problem by providing an appropriate cable actuation length to the joint structure of the robot from the cable driving unit. Therefore, we propose a harmonious nonlinear cable actuation mechanism suitable for driving non-circular shaped joints. The proposed cable driver can mechanically provide the required cable actuation length to suit the angle change of the target joint using a pair of non-circular pulleys without increasing the number of actuators. In this paper, a design methodology of a non-circular pulley that can be applied to pulleyless rolling joints as well as pulleyless hinge joints is shown. Moreover, a practical cable driver is designed for actuating a hyper-redundant discrete bending joint composed of pulleyless rolling joints, and its effectiveness is verified through experiments. This novel cable actuation mechanism using non-circular pulleys or gears is expected to be applicable to various miniature robots such as surgical robots and animal robots of continuum structure in the future.
Article
Full-text available
The tendon driven mechanism using a cable and pulley to transmit power is adopted by many surgical robots. However, backlash hysteresis objectively exists in cable-pulley mechanisms, and this nonlinear problem is a great challenge in precise position control during the surgical procedure. Previous studies mainly focused on the transmission characteristics of the cable-driven system and constructed transmission models under particular assumptions to solve nonlinear problems. However, these approaches are limited because the modeling process is complex and the transmission models lack general applicability. This paper presents a novel position compensation control scheme to reduce the impact of backlash hysteresis on the positioning accuracy of surgical robots’ end-effectors. In this paper, a position compensation scheme using a support vector machine based on feedforward control is presented to reduce the position tracking error. To validate the proposed approach, experimental validations are conducted on our cable-pulley system and comparative experiments are carried out. The results show remarkable improvements in the performance of reducing the positioning error for the use of the proposed scheme.
Article
Minimally invasive surgery, including laparoscopic and thoracoscopic procedures, benefits patients in terms of improved postoperative outcomes and short recovery time. The challenges in hand-eye coordination and manipulation dexterity during the aforementioned procedures have inspired an enormous wave of developments on surgical robotic systems to assist keyhole and endoscopic procedures in the past decades. This paper presents a systematic review of the state-of-the-art systems, picturing a detailed landscape of the system configurations, actuation schemes, and control approaches of the existing surgical robotic systems for keyhole and endoscopic procedures. The development challenges and future perspectives are discussed in depth to point out the need for new enabling technologies and inspire future researches.
Article
In recent years, robotic technology has been introduced to medical fields, and many surgical robots have been proposed for minimally invasive surgery (MIS). However, due to the limitations in dexterity imposed by surgical instruments and occlusion area, surgeons experience great difficulties during operations. In this paper, we propose a visible manipulator for laparoscopic surgery. Unlike other multiple degree-of-freedom (DOF) manipulators that utilize compliant parts or tendons and pulleys, our proposed manipulator adopts a multi-gear array mechanism to perform the yaw and pitch motions. The manipulator is integrated with a visualization unit to provide macroscopic images for observation in a constrained cavity. Moreover, flexible surgical tools with different functions can be inserted through the central channel of the manipulator to perform diagnostic or therapeutic procedures. A master-slave system is developed to control the bending motions. Bending characteristics experiments and load capacity experiments are performed. The experimental results demonstrate that the proposed manipulator can perform bending motions with a yaw angle range of -76.8°~76.2° and a pitch angle range of -75.2°~75.6°. The manipulator can lift a workload of 250 g during yaw motion and a workload of 150 g during pitch motion, demonstrating the potential clinical value of the visible manipulator for robot-assisted surgery.
Article
Background: Since the implementation of robotic surgery, the platforms have been updating constantly in terms of arm configuration, tool design, scope settings, and the number of trocars placed. The introduction of new robotic technology is challenging and requires studies followed by technique adaptions. Objective: This study aims to report a logical and technologically safe approach to the learning curve using the da Vinci single-port (SP) console and describes our robotic-assisted radical prostatectomy (RARP) technique step by step. Design, setting, and participants: A prospective study from 26 consecutive patients who underwent RARP with the da Vinci SP console from June to August 2019. Surgical procedure: All surgeries were performed with a transperitoneal technique; one robotic trocar was placed above the umbilicus and one additional 12 mm trocar was placed in the right lower quadrant. Measurements: We described the step-by-step technique and reported the perioperative and pathological data. In addition, we considered the hospital length of stay and pain scale following surgery. Continuous variables were reported as median and interquartile ranges. Categorical variables were reported as frequencies and proportions. Result and limitations: The total median operative time was 121 min, console time was 85 min, and blood loss was 50 ml. No complications were reported. In the final pathology, four patients had Gleason 6, 20 had Gleason 7, one had Gleason 8, and one had Gleason 9.Of the patients, 70% were ≤pT2 and 30% were ≥pT3a. Only 11% had positive surgical margins. This study is limited by the small number of patients and a short period of follow-up to evaluate functional and oncological outcomes of this new technology. Conclusions: RARP with the da Vinci SP is feasible and safe. Therefore, the step-by-step technique described in this study could be considered an option to perform radical prostatectomies. However, we still need better-designed studies to compare the outcomes with those of the multiport platform. Patient summary: We reported our step-by-step technique describing a safe approach to robotic-assisted radical prostatectomy during the transition from the Xi to the da Vinci single-port robot.
Article
In minimally invasive surgery (MIS), surgeons often suffer from occlusion region problems. It is difficult to solve these problems with traditional surgical instruments because of their size and rigid mechanical structure, such as endoscopes and corresponding operating tools. Thus, flexible manipulators and related robotic systems have been proposed for enhancing intraoperative inspection and surgical operation in MIS. Although a variety of flexible manipulators using different mechanisms have been developed, most of them are designed with a single function. In this paper, we present the concept of visible forceps that enriches the forceps function, which realizes the flexible bending capability and high output force, as well as the integrated endoscopic function. We developed a novel simplified linkage bending mechanism for forceps with a bendable tip and fabricated a robotic visible forceps manipulator system. According to this prototype, we performed experiments to evaluate the mechanical performance and the abdominal phantom test to evaluate the feasibility and usefulness. Preliminary results show that the forceps manipulator can realize both flexible bending capability and high output force, which implies promising applications in future MIS.