ArticlePDF Available

ECG Triggering in Ultra-High Field Cardiovascular MRI

Authors:

Abstract and Figures

Cardiac magnetic resonance imaging at ultra-high field (B0 ≥ 7 T) potentially provides improved resolution and new opportunities for tissue characterization. Although an accurate synchronization of the acquisition to the cardiac cycle is essential, electrocardiogram (ECG) triggering at ultra-high field can be significantly impacted by the magnetohydrodynamic (MHD) effect. Blood flow within a static magnetic field induces a voltage, which superimposes the ECG and often affects the recognition of the R-wave. The MHD effect scales with B0 and is particularly pronounced at ultra-high field creating triggering-related image artifacts. Here, we investigated the performance of a conventional 3-lead ECG trigger device and a state-of-the-art trigger algorithm for cardiac ECG synchronization at 7 T. We show that by appropriate subject preparation and by including a learning phase for the R-wave detection outside of the magnetic field, reliable ECG triggering is feasible in healthy subjects at 7 T without additional equipment. Ultra-high field cardiac imaging was performed with the ECG signal and the trigger events recorded in 8 healthy subjects. Despite severe ECG signal distortions, synchronized imaging was successfully performed. Recorded ECG signals, vectorcardiograms, and large consistency in trigger event spacing indicate high accuracy for R-wave detection.
Content may be subject to copyright.
ECG Triggering in Ultra-High Field
Cardiovascular MRI
Daniel Stäb
1,2
, Juergen Roessler
3
, Kieran O’Brien
4
, Christian Hamilton-Craig
5
, and Markus Barth
1
1
The Centre for Advanced Imaging, The University of Queensland, Brisbane, Queensland, Australia;
2
Department of Diagnostic and Interventional Radiology, University of
Würzburg, Würzburg, Germany;
3
Siemens Healthcare GmbH, Erlangen, Germany;
4
Siemens Healthcare Pty Ltd, Brisbane, Australia; and
5
Richard Slaughter Centre of
Excellence in CVMRI, The Prince Charles Hospital, Brisbane, Queensland, Australia
Corresponding Author:
Daniel Stäb
The Centre for Advanced Imaging,
The University of Queensland,
Brisbane St Lucia, QLD 4072, Australia;
E-mail: daniel.staeb@cai.uq.edu.au
Key Words: ECG, ultra-high field, magnetohydrodynamic effect, cardiac, MRI
Abbreviations: Magnetohydrodynamic (MHD), electrocardiogram (ECG), cardiovascular
magnetic resonance (CMR), vectorcardiography (VCG)
Cardiac magnetic resonance imaging at ultra-high field (B
0
7 T) potentially provides improved resolution
and new opportunities for tissue characterization. Although an accurate synchronization of the acquisition to
the cardiac cycle is essential, electrocardiogram (ECG) triggering at ultra-high field can be significantly im-
pacted by the magnetohydrodynamic (MHD) effect. Blood flow within a static magnetic field induces a volt-
age, which superimposes the ECG and often affects the recognition of the R-wave. The MHD effect scales
with B
0
and is particularly pronounced at ultra-high field creating triggering-related image artifacts. Here, we
investigated the performance of a conventional 3-lead ECG trigger device and a state-of-the-art trigger algo-
rithm for cardiac ECG synchronization at 7 T. We show that by appropriate subject preparation and by in-
cluding a learning phase for the R-wave detection outside of the magnetic field, reliable ECG triggering is
feasible in healthy subjects at 7 T without additional equipment. Ultra-high field cardiac imaging was per-
formed with the ECG signal and the trigger events recorded in 8 healthy subjects. Despite severe ECG sig-
nal distortions, synchronized imaging was successfully performed. Recorded ECG signals, vectorcardio-
grams, and large consistency in trigger event spacing indicate high accuracy for R-wave detection.
INTRODUCTION
Cardiovascular magnetic resonance (CMR) is an important and
well-established clinical tool for the diagnosis and management
of cardiovascular diseases, and it is the standard of reference for
the evaluation of cardiac morphology and function (1-3). CMR
must overcome the challenges introduced because of cardiac
and respiratory motion. In the clinic, CMR relies on accurate
cardiac gating alongside parallel imaging (4,5), simultaneous
multi-slice imaging (6-8), or other acceleration methods (9,10)
to address limitations due to motion. However, the fact remains
that CMR must always make a tradeoff between spatiotemporal
resolution and signal-to-noise ratio.
The signal-to-noise ratio gain inherent at higher field
strengths (11) has recently led to an increased use of high field
systems with B
0
3 T for clinical CMR (12), and moreover, it has
encouraged investigations into ultra-high field (B
0
7T)CMR
(13-15). Apart from enabling spatial resolutions that exceed
today’s limits (16), CMR at ultra-high field offers new opportu-
nities for magnetic resonance-based tissue characterization (17,
18) or metabolic imaging (19).
Cardiac gating is usually performed using electrocardio-
gram (ECG) triggering. In general, vectorcardiography (VCG)-
based QRS detection algorithms (20) are used, which aim to
detect the R-wave in their peak by recognizing the R-wave’s
rising edge. However, ECG signal distortions from several effects
have been challenging at ultra-high field. The interaction of the
conductive fluid blood with the static magnetic field B
0
, for
instance, induces a voltage perpendicular to B
0
and the direction
of flow that superimposes on the ECG signal (21). This so-called
magnetohydrodynamic (MHD) effect is particularly large during
the early systolic phase, when the blood is ejected from the left
ventricle into the aortic arch. Hence, it mainly affects the T-
wave of the ECG signal (22-24). The probability that a rising
edge of an MHD artifact is similar to the rising edge of the
R-wave is generally nonzero. Consequently, deteriorated cardiac
synchronization is likely in the presence of strong MHD artifacts
that are similar to the R-wave’s rising edge. Problems have been
observed at clinical field strengths such as3T(
25,26), and
because the MHD effect scales with B
0
, distortions have been
reported to be worse at ultra-high field (27-29). In addition, the
time-varying magnetic gradient fields, which induce voltage
perturbations in the ECG leads, also distort the ECG signal. To
avoid motion artifacts, the lengthening of scan times and scan
repetitions that result from poor ECG triggering, the establish-
ADVANCES IN BRIEF
ABSTRACT
© 2016 The Authors. Published by Grapho Publications, LLC This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
ISSN 2379-1381 http://dx.doi.org/10.18383/j.tom.2016.00193
TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016 167
ment of a stable cardiac synchronization technique is essential
to advance ultra-high field CMR.
Pulse triggering is often used in cases were the conventional
ECG approach fails (26). However, being derived from softly
shaped peaks in the pulse wave signal, the trigger events are
subject to immanent enhanced jittering (28,30), which com-
monly introduces trigger-related image artifacts. Because the
trigger events are also delayed with respect to the R-wave, pulse
triggering is unsuitable for modalities like myocardial tagging
that require an accurate detection of the R-wave.
Doppler ultrasound (30) and acoustic trigger devices (28,31)
as well as self-navigation (32-34) and pilot tone navigation (35)
have recently been explored as alternative tools to conventional
ECG triggering. In addition, advanced ECG detection algorithms
(36-40) have been proposed, and promising results have been
shown in initial studies.
Here, we explore the technical capabilities ofa7Tresearch
MRI system and state-of-the-art 3-lead ECG equipment for car-
diac synchronization at ultra-high field. Our initial study shows
that by including an appropriate ECG learning phase outside of
the magnetic field, existing ECG trigger technology in 7 T
research systems allows for generating stable and reliable ECG
trigger signals.
METHODOLOGY
All measurements were performed on a noncommercial 7 T
whole-body research MRI scanner (Siemens Healthcare GmbH,
Erlangen, Germany) under institutional review board permis-
sion. The gradient system provided a maximum gradient
strength of 70 mT/m and a slew rate of 200 T/m/s. A dedicated
7 T cardiac Tx/Rx array with 8 transmit and 32 receive channels
(MRI Tools GmbH, Berlin, Germany) was used for radiofre-
quency transmission and signal reception. The coil array was
operated in single-channel transmit mode. To improve the B
0
field homogeneity, third-order shimming was used. For all hu-
man in vivo experiments, written informed consent was ob-
tained before the examination as approved by the local ethics
committee.
Cardiac Synchronization
For cardiac synchronization, a 3-lead ECG trigger device
(Siemens Healthcare GmbH, Erlangen, Germany) using wire-
less signaling was used in conjunction with the standard trigger
algorithm provided by the device manufacturer. The basic prop-
erties of this algorithm are briefly described in the following
paragraph. For a detailed description, we refer the interested
reader to work by Frank et al. (41).
To ensure an accurate detection of the peak of the R-wave,
the trigger algorithm learns the shape of the rising edge of the
R-wave during an initial learning phase in both ECG channels.
Learning is performed while the subject is lying on the patient
table outside of the magnet bore, where the MHD effect is
typically negligible.
Once learning is completed, the trigger algorithm continu-
ously compares different derived entities (ie, derivatives, filtered
versions of derivatives) [for details refer Frank et al.’s study (41)]
of the incoming ECG signal with the corresponding entities of
the learned shape in real time. The comparisons are mainly
based on 2 filter functions. The first is a matched filter, which is
widely used in telecommunications (42) and mathematically
corresponds to forming the correlation of the 2 signals. The filter
function is given by the following equation:
mj(
)aj·i0
t
sj(
⫺⌬ttrj
*(t), (1)
where ais a normalization factor, tdepicts the period of
comparison, and jrefers to the signal entity for comparison. The
incoming and the reference signal entities are depicted by s(t)
and r(t), respectively. Both signal entities are complex with the
real and imaginary components derived from the 2 ECG chan-
nels. The second filter function sums up the squared differ-
ences between the incoming ECG signal and the learned
signal shape according to the following equation:
qj(
)bj·t0
t
(sj(
⫺⌬ttrj(t))2(2)
with baccounting for normalization. Trigger events are initiated
by thresholding the filtered signals, m
j
and q
j
. In addition to
using those filters, the angle of the VCG vector, which is spanned
by the signal in the 2 ECG channels at each time instant, is used
for R-wave detection. This angle criterion is implemented as a
necessary, but not sufficient, condition for a trigger generation.
The trigger algorithm showed an overall high performance at 1.5
T(
43).
In Vivo Measurements
To evaluate the performance of the underlying ECG trigger
algorithm, cardiac cine imaging at ultra-high field was per-
formed in 8 healthy volunteers. Before starting the imaging
procedure, ECG electrodes were placed onto the chest of the
subject, following the instructions of the trigger device manu-
facturer, and in conjunction with a senior electrophysiology
cardiac scientist. In 2 volunteers, the chest hair was removed in
the target area before electrode placement to ensure good cou-
pling at the body/electrode interface. After positioning the sub-
ject on the patient table, the ECG trigger device leads were
connected to the electrodes, which automatically started the
learning phase of the ECG algorithm. A pulse sensor (Siemens
Healthcare GmbH, Erlangen, Germany) was attached to the
subject’s index finger as a backup trigger device. After fin-
ishing all other preparatory steps, the learning phase of the
ECG algorithm was stopped to initiate the R-wave detection
mechanism. The signals detected in both available ECG chan-
nels and the generated trigger events were recorded through-
out all examinations.
Cardiac cine imaging was performed using a high-resolu-
tion breath-held ECG retro-gated segmented 2-dimensional
spoiled gradient echo (FLASH) sequence with the following
acquisition parameters: field of view 360 270 (360 338)
mm
2
, matrix 256 192 (256 240), slice thickness 4.0
mm, echo time 3.13 milliseconds, repetition time 6.1
milliseconds, receiver bandwidth 592 Hz/px, flip angle 60°,
phases 20, segments 7, and temporal resolution 43
milliseconds. Parallel imaging was used with an acceleration
factor of R 2. All images were reconstructed online using
GRAPPA with 34 reference lines for weight set calculation.
ECG Triggering in Ultra-High Field Cardiovascular MRI
168 TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016
Qualitative Evaluations
The performance of the underlying trigger algorithm and the
impact of the MHD effect onto the performance was qualita-
tively evaluated based on the recorded ECG signals. Signal time
curves for the individual ECG channels were visually assessed.
VCG plots were generated and examined to identify mistrigger-
ing. Finally, the time intervals between succeeding trigger
events were analyzed to obtain an estimate on the amount of
false positive and false negative trigger events.
RESULTS
A general overview of the MHD effect at ultra-high field can be
gained from Figure 1A. It shows the change of the detected
signal in one of the 2 ECG channels over several RR-intervals
during the transition of the examined subject into the isocenter
of the magnet. Outside of the magnet bore, the signal is
generally smooth and undistorted, and the R-wave is easy to
distinguish as the highest peak. With increasing magnetic flux
density, the MHD-related alterations of the ECG become increas-
ingly pronounced and result in a significant distortion of the
ECG signal at the magnet’s isocenter. In the depicted case, the
distorted T-wave clearly exceeds the R-wave.
Individual ECG channels can be affected in different ways,
as shown in Figure 1B, which compares the MHD effect on the
signals of the 2 different ECG channels. To reduce the influence
of inter-RR signal fluctuations, the ECG signals were averaged
over 40 consecutive RR-intervals. The signals detected in both
channels experience distortions at the isocenter of the magnet.
However, only in channel 2 does the overall shape of the signal
considerably change, and the R-wave is exceeded by
the distorted ECG segments. Despite the significant impact of the
MHD effect, the QRS complex is clearly identifiable in both
channels and—as can be seen from the accurate alignment of the
individual RR-intervals— has been accurately detected by the
trigger algorithm for each of the displayed cardiac cycles.
Exemplary VCG plots obtained in 3 volunteers outside and
at the isocenter of the magnet are given in Figure 2. In each
vectorcardiogram, the ECG signal recorded in channel 2 is plot-
ted against the signal measured in channel 1 over several RR-
intervals. Associated trigger events are superimposed (black
circles). The data were collected outside of the magnet bore (first
row, blue), at the isocenter in the absence of gradient activity
during free breathing (second row, red) and during a breath-held
cine scan (third row, yellow). The MHD effect-related increase of
the ECG signal variations within each RR-interval is apparent when
comparing the vectorcardiograms obtained outside and inside of
the magnet. In addition, the large magnetic field introduces con-
siderable changes in the shape of the vectorcardiograms. Altera-
tions can also be observed, when comparing data collected during
free breathing and breath-hold periods. The characteristic VCG
curves including the trigger events tend to be dispersed along the
vertical axis, when obtained during free breathing (second row). As
depicted by the enlarged section in Figure 2C (third row), gradient
activity seems to introduce only tiny additional deflections in the
VCG signal. For each subject, the location of the trigger events in
the VCG plots is preserved in the vast majority of cases when
exposing the subject to the ultra-high static magnetic field and
dynamic gradient fields.
The accuracy of the trigger algorithm is shown in more
detail in Figure 3. For one of the volunteers, ECG signal curves
obtained outside of the magnet bore (Figure 3A) and at the
magnet’s isocenter (Figure 3, B and C) are compared with each
other. The MHD effect-related distortions of the ECG signal are
clearly recognizable. Despite these distortions, trigger events are
typically placed accurately. Only on rare occasions, false nega-
Figure 1. Electrocardiogram (ECG) signal time curves obtained in a healthy subject. Evolution of the ECG signal in
channel 2 across consecutive RR-intervals during the transition from outside into the isocenter of the magnet (A). All
curves were aligned based on observed trigger events. RR-intervals recorded outside and at the isocenter of the magnet
are marked by blue and red lines, respectively. During the transition into the magnet (RR-intervals 30–100), the magne-
tohydrodynamic (MHD) effect increasingly impacts the ECG signal. Signal time curves averaged over 40 consecutive
RR-intervals (solid) and corresponding standard deviation (dotted) following alignment by the observed trigger events
observed outside (blue) and at the isocenter (red) of the magnet (B).
ECG Triggering in Ultra-High Field Cardiovascular MRI
TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016 169
tive and misplaced trigger events were observed. False positive
events were extremely rare. This is the case even in presence of
severe additional distortions (Figure 3C), which, in this case, can
be attributed to enhanced inhaling and exhaling in preparation
of a scan-related breath-hold. As seen in the enlarged section,
the gradient activity of the scan causes smaller variations of the
ECG signal. The start of the scan is marked by the dashed line.
An impression of the high trigger accuracy can also be
gained from Figure 4, which depicts histograms of the time
intervals between consecutive trigger events in 3 healthy sub-
jects. While the width of the histogram peak reflects the varia-
tion of the subject’s heart rate, outliers indicate false positive
trigger events and undetected RR-intervals. The spacing be-
tween almost all trigger events is in the range of a single
RR-interval. Only a few counts are spread out along the hori-
zontal axis of the histogram.
ECG-triggered cardiac cine imaging worked generally well
with the used synchronization setup. Representative exam-
ples of cardiac cine images obtained at 7 T are depicted in
Figure 5. Shown are diastolic and systolic time frames of a
short-axis (top row) and a 4-chamber long-axis view (bottom
row) of a healthy subject’s heart (see online Supplemental
Video 1 PLAY VIDEO and Video 2 PLAY VIDEO ). The myocardial walls
are well delineated, and the images are free of visible artifacts
that could be related to unsuccessful ECG triggering. The long-
axis views show slight signal inhomogeneities introduced by
Figure 2. Vectorcardiograms obtained in 3 healthy subjects outside of the magnet (first row, blue), at the isocenter of
the magnet during free breathing (second row, red), and during a breath-held cine acquisition (third row, yellow), each
within a time-frame of 20 seconds (A–C). The depicted scales reflect the actual relative signal amplitudes between the
channels. The generally small high-frequency signal variations induced by the imaging gradients can be seen in the en-
larged section in the third row for subject (C).
ECG Triggering in Ultra-High Field Cardiovascular MRI
170 TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016
destructive B1 interferences (arrows). Moreover, flow effects are
pronounced because of the high flip angle used.
DISCUSSION
For the medical application of CMR, the accurate synchroniza-
tion of the imaging protocol to the cardiac cycle is essential to
achieve high image quality and accurate results in functional
evaluations. In this initial study, we explored the applicability of
existing and state-of-the-art 3-lead ECG trigger technology for
cardiac synchronization at the ultra-high field strength of 7 T.
Without using additional hardware, the underlying trigger al-
gorithm generated reliable ECG trigger signals and provided the
Figure 3. ECG signal over time measured in one of the volunteers outside of the magnet (A) and at the isocenter of the
magnet (B–C) in the first (red) and second (blue) channel of the ECG. Trigger events (green vertical lines) are typically
placed accurately at the peak of the R-wave. The larger distortions in (C) can be attributed to deep breathing preceding
a breath-held cine acquisition. The enlarged section visualizes the start of the sequence (dashed line) and the effect of
the imaging gradients on the ECG signal (arrows).
Figure 4. Histograms analyzing the apparent time interval between succeeding trigger events in 3 healthy subjects
(A–C) outside (blue) and at the isocenter (red) of the magnet. The bins along the horizontal axis are separated by 100
milliseconds. For almost all events, the time elapsed with respect to the previous event is in the range of 1 RR-interval.
Counts at considerably larger or shorter intervals indicate false negative and false positive triggering.
ECG Triggering in Ultra-High Field Cardiovascular MRI
TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016 171
basis for high-fidelity cardiac imaging. Despite severe ECG sig-
nal distortions due to the MHD effect and breathing motion,
synchronized imaging was feasible without severe disruptions
in all healthy volunteers. The attached pulse sensor was not
required in any case as a substitute trigger device, and signifi-
cant synchronization-related image artifacts were not observed.
In general, ECG signal distortions can be introduced by
various motions and imaging gradients. In this initial study, the
distorting effect of imaging gradients turned out to be small,
although, as expected, considerable distortions were caused by
the MHD effect. Particularly irregular and, in some cases, severe
distortions could be attributed to breathing and subject motion.
Even in presence of these adverse effects, the utilized trigger
algorithm allows for an accurate R-wave detection, provided a
learning phase in the absence of the magnetic field has been
executed.
In the absence of the large magnetic field, ECG triggering
was found to be highly accurate with the used setup. In the
presence of the ultra-high field, the MHD effect led to significant
distortion of the ECG signal time curves in all volunteers. Nev-
ertheless, the overall synchronization accuracy remained high.
As indicated by the histogram analysis, the length of almost all
RR-intervals is within the range of a single RR-interval. The
relatively high accuracy can be explained by the fact that the
QRS complex of the ECG is typically only marginally impacted
by the MHD effect, and the used trigger algorithm relies on
real-time detection of the shape of the rising edge of the R-wave,
rather than signal thresholding of the R-wave. In this way, the
misplacement of trigger events and the generation of false
positive events is minimized. Consequently, even in cases where
the distorted T-wave of the ECG clearly exceeds the targeted
R-wave, accurate cardiac synchronization is feasible.
The generally high trigger accuracy is also recognizable in
the calculated vectorcardiograms. The location and the spread
of the trigger events were not significantly affected by the
exposure of the subject to the high magnetic field, despite the
considerable increase in signal fluctuations. Inside and outside
of the magnetic field, the VCG plots show different characteristic
patterns. Although the pattern change itself is governed by the
MHD effect, the additional dispersion of this pattern (Figure 2)
can be attributed to the breathing motion and the associated
movement of the chest with the electrodes through the static
magnetic field. In this study, imaging gradients did not have a
large influence on the VCG signal. As shown in Figures 2 and 3,
they only caused tiny and high-frequency signal deflections.
To achieve high trigger accuracy at ultra-high field, careful
subject preparation and electrode placement are essential to
ensure good connection of the electrodes with the subject’s skin
and to achieve high input signals in all ECG channels. As
depicted in Figures 2 and 3, gradient-induced ECG signal dis-
tortions can thus be kept small. A good preparation is particu-
larly important for applications such as retro-gated cine CMR,
where imaging gradients interfere with the QRS complex. Given
the additional effect of breathing motion corrupting the ECG
signal, properly instructing subjects on breathing techniques
might be helpful in achieving accurate triggering. This is par-
ticularly the case for free-breathing imaging applications.
In this work, accurate ECG triggering was achieved by fully
exploiting the technical capabilities of the ultra-high field scan-
ner and trigger equipment. Apart from conducting an appropri-
ate ECG learning phase, further hardware or software modifica-
tions were not required. Thus, the presented approach is widely
available and ready to use. Recently, published studies indicate
that using a larger number of ECG leads, the use of specifically
tailored trigger algorithms or combinations of both can be
advantageous in terms of trigger accuracy (36-40). Based on
this, we assume that the use of more leads or further refinements
of the algorithm could improve our results. However, the use of
a large number of leads also adds to patient discomfort and
preparation times.
Within the scope of this initial study, only a limited number
of healthy subjects could be examined. Thus, it is important to
note that the overall number of RR-intervals that could be
analyzed outside of the magnet bore is rather low. Moreover, it
is well-known that the success of ECG triggering can be highly
subject-dependent, and certainly, a larger number of subjects
need to be examined in future work to fully reveal the perfor-
mance of the trigger technology at hand. Apart from that, ECG
triggering can be particularly challenging in patient cohorts
with cardiac arrhythmia, where the MHD effect can be more
severe and variable.
In conclusion, we have shown that reliable cardiac
ECG triggering is feasible in healthy volunteers at ultra-high
Figure 5. Images obtained using fully ECG-trig-
gered cardiac cine imaging at ultra-high field.
Shown are diastolic (left) and systolic (right) time
frames of a short-axis (top) and horizontal long-
axis (bottom) view. Myocardial walls are accu-
rately delineated. Signal inhomogeneities induced
by destructive B1 interferences are depicted
(arrows).
ECG Triggering in Ultra-High Field Cardiovascular MRI
172 TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016
field by using a state-of-the-art 3-lead trigger device. The
used trigger algorithm provided sufficient accuracy for high-
fidelity cardiac cine imaging, despite severe ECG signal dis-
tortions due to the MHD effect. Future work will need to
further evaluate the algorithm in larger cohorts and pa-
tients with cardiac arrhythmia. Apart from CMR, other ultra-
high field imaging applications such as human brain func-
tional MRI with physiological noise correction may benefit
from the easy instrumentational setup and robust ECG
triggering.
Supplemental Materials
Video 1: http://dx.doi.org/10.18383/j.tom.2016.00193.vid.01
Video 2: http://dx.doi.org/10.18383/j.tom.2016.00193.vid.02
ACKNOWLEDGMENTS
CHC would like to acknowledge funding by the University of Queensland Academic Title
Holder Research Grant (ATHRF). The authors would like to thank Daniel Smith and Haris
Haqqani (Cardiac Electrophysiology Unit, The Prince Charles Hospital, Brisbane,
Queensland Australia) for their helpful advice on ECG devices, ECG triggering, and
ECG electrode placement. The authors acknowledge the facilities of the National
Imaging Facility (NIF) at the Centre for Advanced Imaging, University of Queensland,
and support by the NIF fellow Steffen Bollmann.
Disclosures: KO and JR disclose being employed by Siemens Healthcare. JR holds a
patent related to the work presented.
REFERENCES
1. Finn JP, Nael K, Deshpande V, Ratib O, Laub G. Cardiac MR imaging: state of
the technology 1. Radiology. 2006;241(2):338–354.
2. Earls JP, Ho VB, Foo TK, Castillo E, Flamm SD. Cardiac MRI: recent progress and
continued challenges. J Magn Reson Imaging. 2002;16(2):111–127.
3. Hundley WG, Bluemke DA, Finn JP, Flamm SD, Fogel MA, Friedrich MG, Ho VB,
Jerosch-Herold M, Kramer CM, Manning WJ, Patel M, Pohost GM, Stillman AE,
White RD, Woodard PK. ACCF/ACR/AHA/NASCI/SCMR 2010 expert consen-
sus document on cardiovascular magnetic resonance. J Am Coll Cardiol. 2010;
55(23):2614–2662.
4. Wintersperger BJ, Reeder SB, Nikolaou K, Dietrich O, Huber A, Greiser A,
Lanz T, Reiser MF, Schoenberg SO. Cardiac CINE MR imaging with a 32-
channel cardiac coil and parallel imaging: impact of acceleration factors on
image quality and volumetric accuracy. J Magn Reson Imaging. 2006;23(2):
222–227.
5. Schmitt M, Potthast A, Sosnovik DE, Polimeni JR, Wiggins GC, Triantafyllou C,
Wald LL. A 128-channel receive-only cardiac coil for highly accelerated cardiac
MRI at 3 Tesla. Magn Reson Med. 2008;59(6):1431–1439.
6. Stäb D, Ritter CO, Breuer FA, Weng AM, Hahn D, Köstler H. CAIPIRINHA accel-
erated SSFP imaging. Magn Reson Med. 2011;65(1):157–164.
7. Stäb D, Wech T, Breuer FA, Weng AM, Ritter CO, Hahn D, Köstler H. High reso-
lution myocardial first-pass perfusion imaging with extended anatomic coverage.
J Magn Reson Imaging. 2014;39(6):1575–1587.
8. Schmitter S, Moeller S, Wu X, Auerbach EJ, Metzger GJ, van de Moortele P-F,
Ug˘urbil K. Simultaneous multislice imaging in dynamic cardiac MRI at 7T using
parallel transmission. Magn Reson Med. 2016.
9. Manka R, Paetsch I, Schnackenburg B, Gebker R, Fleck E, Jahnke C. BOLD car-
diovascular magnetic resonance at 3.0 tesla in myocardial ischemia. J Cardio-
vasc Magn Reson. 2010;12:54.
10. Zhang S, Uecker M, Voit D, Merboldt K-D, Frahm J. Real-time cardiovascular
magnetic resonance at high temporal resolution: radial FLASH with nonlinear in-
verse reconstruction. J Cardiovasc Magn Reson. 2010;12:39.
11. Ohliger MA, Grant AK, Sodickson DK. Ultimate intrinsic signal-to-noise ratio for
parallel MRI: electromagnetic field considerations. Magn Reson Med. 2003;
50(5):1018–1030.
12. Gutberlet M, Noeske R, Schwinge K, Freyhardt P, Felix R, Niendorf T. Compre-
hensive cardiac magnetic resonance imaging at 3.0 Tesla: feasibility and impli-
cations for clinical applications. Invest Radiol. 2006;41(2):154–167.
13. Snyder CJ, DelaBarre L, Metzger GJ, van de Moortele P-F, Akgun C, Ugurbil K,
Vaughan JT. Initial results of cardiac imaging at 7 Tesla. Magn Reson Med.
2009;61(3):517–524.
14. van Elderen SGC, Versluis MJ, Westenberg JJ, Agarwal H, Smith NB, Stuber M,
de Roos A, Webb AG. Right coronary MR angiography at 7 T: a direct quantita-
tive and qualitative comparison with3Tinyoung healthy volunteers. Radiology.
2010;257(1):254–259.
15. von Knobelsdorff-Brenkenhoff F, Tkachenko V, Winter L, Rieger J, Thalhammer C,
Hezel F, Graessl A, Dieringer MA, Niendorf T, Schulz-Menger J. Assessment of
the right ventricle with cardiovascular magnetic resonance at 7 Tesla. J Cardio-
vasc Magn Reson. 2013;15:23.
16. Graessl A, Renz W, Hezel F, Dieringer MA, Winter L, Oezerdem C, Rieger J,
Kellman P, Santoro D, Lindel TD, Frauenrath T, Pfeiffer H, Niendorf T. Modular
32-channel transceiver coil array for cardiac MRI at 7.0T. Magn Reson Med.
2014;72(1):276–290.
17. Hezel F, Thalhammer C, Waiczies S, Schulz-Menger J, Niendorf T. High spatial
resolution and temporally resolved T2* mapping of normal human myocardium
at 7.0 tesla: an ultrahigh field magnetic resonance feasibility study. PLoS One.
2012;7(12).
18. Kober F, Jao T, Troalen T, Nayak KS. Myocardial arterial spin labeling. J Cardio-
vasc Magn Reson. 2016;18:22.
19. Clarke WT, Robson MD, Rodgers CT. Bloch-Siegert B1-mapping for human car-
diac (31) P-MRS at 7 Tesla. Magn Reson Med. 2015. doi: 10.1002/mrm.
26005 [Epub ahead of print].
20. Fischer SE, Wickline SA, Lorenz CH. Novel real-time R-wave detection algorithm
based on the vectorcardiogram for accurate gated magnetic resonance acquisi-
tions. Magn Reson Med. 1999;42(2):361–370.
21. Togawa T, Okai O, Oshima M. Observation of blood flow E.M.F. in externally applied
strong magnetic field by surface electrodes. Med Biol Eng. 1967;5(2):169 –170.
22. Krug JW, Rose G. Magnetohydrodynamic distortions of the ECG in different
MR scanner configurations. In: 2011 Computing in Cardiology; 2011: pp.
769–772.
23. Jekic M, Dzwonczyk R, Ding S, Raman V, Simonetti O. Quantitative evaluation of
magnetohydrodynamic effects on the electrocardiogram. In: 17th ISMRM Annual
Meeting and Exhibition; 2009: p. 3795.
24. Tenforde TS. Magnetically induced electric fields and currents in the circulatory
system. Prog Biophys Mol Biol. 2005;87(2–3):279–288.
25. Dietrich O, Reiser MF, Schoenberg SO. Artifacts in 3-T MRI: physical background
and reduction strategies. Eur J Radiol. 2008;65(1):29–35.
26. Sievers B, Wiesner M, Kiria N, Speiser U, Schoen S, Strasser RH. Influence of
the trigger technique on ventricular function measurements using 3-Tesla magnetic
resonance imaging: comparison of ECG versus pulse wave triggering. Acta
Radiol. 2011;52(4):385–392.
27. Suttie JJ, DelaBarre L, Pitcher A, van de Moortele PF, Dass S, Snyder CJ, Francis
JM, Metzger GJ, Weale P, Ugurbil K, Neubauer S, Robson M, Vaughan T.
7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and
SSFP to assess cardiac function: validation against 1.5 T and 3 T. NMR Biomed.
2012;25(1):27–34.
28. Frauenrath T, Hezel F, Renz W, d’Orth Tde G, Dieringer M, von Knobelsdorff-
Brenkenhoff F, Prothmann M, Menger JS, Niendorf T. Acoustic cardiac triggering:
a practical solution for synchronization and gating of cardiovascular magnetic
resonance at 7 Tesla. J Cardiovasc Magn Reson. 2010;12:67.
29. Krug J, Rose G, Stucht D, Clifford G, Oster J. Limitations of VCG based gating
methods in ultra high field cardiac MRI. J Cardiovasc Magn Reson. 2013;15.
30. Kording F, Schoennagel B, Lund G, Ueberle F, Jung C, Adam G, Yamamura J.
Doppler ultrasound compared with electrocardiogram and pulse oximetry cardiac
triggering: a pilot study. Magn Reson Med. 2015;74(5):1257–1265.
31. Maderwald S, Orzada S, Lin Z, Schäfer LC, Bitz AK, Kraff O, Brote I, Häring L,
Czylwik A, Zenge MO, Ladd SC, Ladd ME, Nassenstein K. 7 Tesla cardiac im-
aging with a phonocardiogram trigger device. In: 19th ISMRM Annual Meeting
and Exhibition; 2011: p. 1322.
32. Brau AC, Brittain JH. Generalized self-navigated motion detection technique:
preliminary investigation in abdominal imaging. Magn Reson Med. 2006;55(2):
263–270.
33. Buehrer M, Curcic J, Boesiger P, Kozerke S. Prospective self-gating for simultane-
ous compensation of cardiac and respiratory motion. Magn Reson Med. 2008;
60(3):683–690.
ECG Triggering in Ultra-High Field Cardiovascular MRI
TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016 173
34. Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Self-gated car-
diac cine MRI. Magn Reson Med. 2004;51(1):93–102.
35. Schroeder L, Wetzl J, Maier A, Rehner R, Fenchel M, Speier P. A novel method
for contact-free cardiac synchronization using the pilot tone navigator. In: 24th
ISMRM Annual Meeting and Exhibition; 2016: p. 3103.
36. Odille F, Pasquier C, Abacherli R, Vuissoz PA, Zientara GP, Felblinger J. Noise
cancellation signal processing method and computer system for improved real-
time electrocardiogram artifact correction during MRI data acquisition. IEEE Trans
Biomed Eng. 2007;54(4):630–640.
37. Krug JW, Rose GH, Stucht D, Clifford GD, Oster J. Filtering the magnetohydrody-
namic effect from 12-lead ECG signals using independent component analysis.
In: 2012 Computing in Cardiology; 2012:589–592.
38. Krug JW, Rose G, Clifford GD, Oster J. ECG-based gating in ultra high field car-
diovascular magnetic resonance using an independent component analysis ap-
proach. J Cardiovasc Magn Reson. 2013;15:104.
39. Gregory TS, Schmidt EJ, Zhang SH, Ho Tse ZT. 3DQRS: a method to obtain reli-
able QRS complex detection within high field MRI using 12-lead electrocardio-
gram traces. Magn Reson Med. 2014;71(4):1374–1380.
40. Zhang SH, Tse ZT, Dumoulin CL, Kwong RY, Stevenson WG, Watkins R, Ward J,
Wang W, Schmidt EJ. Gradient-induced voltages on 12-lead ECGs during high
duty-cycle MRI sequences and a method for their removal considering linear and
concomitant gradient terms. Magn Reson Med. 2016;75(5):2204–2216.
41. Frank M, Rößler J. Method for identifying an R-wave in an ECG signal, ECG
measuring device and magnetic resonance scanner. US patent application 2010
0191134 A1. 29 July 2010.
42. Proakis JG. Digital Communications. 3rd ed. New York, NY: McGraw-Hill; 1995.
43. Knesewitsch T, Meierhofer C, Rieger H, Rößler J, Frank M, Martinoff S, Hess J,
Stern H, Fratz S. Demonstration of value of optimizing ECG triggering for cardio-
vascular magnetic resonance in patients with congenital heart disease. J Cardio-
vasc Magn Reson. 2013;15(1):3.
ECG Triggering in Ultra-High Field Cardiovascular MRI
174 TOMOGRAPHY.ORG
|
VOLUME 2 NUMBER 3
|
SEPTEMBER 2016
... Vectorcardiogram (VCG) based triggering was performed using a three-lead wireless ECG trigger device (Siemens Healthineers GmbH, Erlangen, Germany), in conjunction with a matched filter based VCG trigger algorithm. To improve the synchronization performance, the VCG trigger algorithm was calibrated outside of the magnet bore where the MHD effect is negligible [5,7,8]. The learning phase of the algorithm was conducted over a period of at least 30 R-R intervals with the subjects lying on the patient table. ...
... Electrocardiographic signals are affected by the MHD effect due to the interaction between the blood, a ferromagnetic conductive fluid, and the surrounding magnetic field [4][5][6][7][8]. This interaction causes an electric field distortion, which is superimposed on the heart's intrinsic depolarization, thereby corrupting the signal transferred to cutaneous ECG electrodes. ...
... This can cause inaccuracies in detecting the QRS complex and may result in impaired cardiac synchronized imaging. Including a sufficiently long VCG learning phase outside of the magnet bore, where the MHD effect is negligible, enabled this problem to be overcome [8,10], resulted in effective R-wave recognition and successful cardiac synchronized cine imaging at 7T. ...
Article
Full-text available
Objective: Ultra-high-field B0 ≥ 7 tesla (7T) cardiovascular magnetic resonance (CMR) offers increased resolution. However, electrocardiogram (ECG) gating is impacted by the magneto-hydrodynamic effect distorting the ECG trace. We explored the technical feasibility of a 7T magnetic resonance scanner using an ECG trigger learning algorithm to quantitatively assess cardiac volumes and vascular flow. Methods: 7T scans were performed on 10 healthy volunteers on a whole-body research MRI MR scanner (Siemens Healthineers, Erlangen, Germany) with 8 channel Tx/32 channels Rx cardiac coils (MRI Tools GmbH, Berlin, Germany). Vectorcardiogram ECG was performed using a learning phase outside of the magnetic field, with a trigger algorithm overcoming severe ECG signal distortions. Vectorcardiograms were quantitatively analyzed for false negative and false positive events. Cine CMR was performed after 3rd-order B0 shimming using a high-resolution breath-held ECG-retro-gated segmented spoiled gradient echo, and 2D phase contrast flow imaging. Artefacts were assessed using a semi-quantitative scale. Results: 7T CMR scans were acquired in all patients (100%) using the vectorcardiogram learning method. 3,142 R-waves were quantitatively analyzed, yielding sensitivity of 97.6% and specificity of 98.7%. Mean image quality score was 0.9, sufficient to quantitate both cardiac volumes, ejection fraction, and aortic and pulmonary blood flow. Mean left ventricular ejection fraction was 56.4%, right ventricular ejection fraction was 51.4%. Conclusion: Reliable cardiac ECG triggering is feasible in healthy volunteers at 7T utilizing a state-of-the-art three-lead trigger device despite signal distortion from the magnetohydrodynamic effect. This provides sufficient image quality for quantitative analysis. Other ultra-high-field imaging applications such as human brain functional MRI with physiologic noise correction may benefit from this method of ECG triggering.
... The identified noise sources are IPs generated by the static magnetic field, radio frequency pulses, and gradient switching [10]. There are solutions to reduce the first two sources [11][12][13]. However, GIPs (Gradient Induced Potentials) are more challenging to eliminate. ...
... where μ is a scalar called the descent step and M k+1 is an interactive approximation of the inverse of the Hessian matrix of the cost function J calculated according to (12). ...
Article
Background: Magnetic resonance imaging (MRI) is the medical imaging technique that benefits most from recent technological innovations, particularly the constant proposal of new MRI sequences that refine clinical information from the obtained images. However, this generates new gradient-induced potential (GIP) morphologies. These induced potentials (IPs) pollute the electrophysiological signals possibly recorded simultaneously. Several algorithms developed to eliminate this noise rely on modelling the shape of the IP. As each new sequence has a different shape of IP, it might be interesting to find a mathematical approach to building sequence-specific models. In this article, we present a preliminary study that includes wavelet decomposition of contaminated electrocardiographic (ECG) to extract IP morphologies and whose time-frequency characterization allows the elaboration of a harmonic model, using sinusoidal decomposition. Method: The in vitro IPs are used to select analyzing wavelets. A broadband sensor (3.5Khz), placed inside a 3 T MRI scanner, is used to collect 3-lead ECGs while activating three sequences that generate very high noise levels. The in vivo IPs extracted from the polluted ECGs are characterized to verify their quasi-periodicity. Parameters of the sinusoidal model (amplitude, frequency, phase) are estimated using the Broyden-Fletcher-Goldfard-Shano optimization algorithm. Result: Four wavelets (sym7, coif3, bior2.2, bior3.3) showed efficient in vivo IP extraction results. Three evaluation criteria for the modelling algorithm, allowing the calculated models to be compared with the shapes of the extracted IPs, showed promising results. For example, for the chosen efficiency criterion Nash-Sutcliffe efficiency , the values obtained for the three leads are between 0.99980 and 1. Conclusion: Promising preliminary results have been obtained for the extraction on modelling of different IPs from noisy ECG signals. Continuing this preliminary study on more MRI sequences and subjects could help build a database of IP models to initiate deep learning filtering. Since these models are sequence-specific and integrate the distribution of induced voltages on the body surface, we hope to find a generic relationship that enables the prediction of IPs by new sequences and anticipate the development of purification algorithms in a near future.
... Third, the 5T system is capable of detecting almost distortion-free ECG signals for reliable gating during CMR scans, highly comparable to those from 3T ( Figure 1) and much better than those from 7T (26). ECG signals originate . ...
... However, electrode-based ECG detection in MRI is usually overlaid with the so-called magneto-hydrodynamic (MHD) effects (27) originating from the conducting fluid (i.e., cardiac blood flow) under the influence of the external magnetic field and will become stronger at higher fields. Practically, the MHD effect is negligible at 3T and lower fields, but it has been determined to severely affect ECG readings at 7T (26) to the extent of non-usability. Therefore, alternative ECG gating techniques have been generally needed at UHF, such as acoustic cardiac triggering (28) or the finger-clipping pulse oximetry. ...
Article
Full-text available
Background Cardiovascular magnetic resonance (CMR) imaging at ultra-high fields (UHF) such as 7T has encountered many challenges such as faster T 2 * relaxation, stronger B 0 and B 1+ field inhomogeneities and additional safety concerns due to increased specific absorption rate (SAR) and peripheral nervous stimulation (PNS). Recently, a new line of 5T whole body MRI system has become available, and this study aims at evaluating the performance and benefits of this new UHF system for CMR imaging. Methods Gradient echo (GRE) CINE imaging was performed on healthy volunteers at both 5 and 3T, and was compared to balanced steady-state-free-procession (bSSFP) CINE imaging at 3T as reference. Higher spatial resolution GRE CINE scans were additionally performed at 5T. All scans at both fields were performed with ECG-gating and breath-holding. Image quality was blindly evaluated by two radiologists, and the cardiac functional parameters (e.g., EDV/ESV/mass/EF) of the left and right ventricles were measured for statistical analyses using the Wilcoxon signed-rank test and Bland-Altman analysis. Results Compared to 3T GRE CINE imaging, 5T GRE CINE imaging achieved comparable or improved image quality with significantly superior SNR and CNR, and it has also demonstrated excellent capability for high resolution (1.0 × 1.0 × 6.0 mm ³ ) imaging. Functional assessments from 5T GRE CINE images were highly similar with the 3T bSSFP CINE reference. Conclusions This pilot study has presented the initial evaluation of CMR CINE imaging at 5T UHF, which yielded superior image quality and accurate functional quantification when compared to 3T counterparts. Along with reliable ECG gating, the new 5T UHF system has the potential to achieve well-balanced performance for CMR applications.
... 4) Similarly, POX cannot be used for in vitro experiments. Alternatives to ECG and POX triggers have been suggested, including optimizing ECG post-processing [17], developing self-gating [18], and using acoustic [16], magnetohydrodynamic [19] and optical [19] signals. However, these methods usually rely on specialized tools and techniques, and are not widely available or difficult to implement [16]. ...
Article
Objective: Cardiac gating, synchronizing medical scans with cardiac activity, is widely used to make quantitative measurements of physiological events and to obtain high-quality scans free of pulsatile artefacts. This can provide important information for disease diagnosis, targeted control of medical microrobots, etc. The current work proposes a low-cost, self-adaptive, MRI-compatible cardiac gating system. Method: The system and its processing algorithm, based on the monitoring and analysis of blood pressure waveforms, are proposed. The system is tested in an in vitro experiment and two living pigs using four-dimensional (4D) flow magnetic resonance imaging (MRI) and two-dimensional phase-contrast (2D-PC) sequences. Results: In vitro and in vivo experiments reveal that the proposed system can provide stable cardiac synchronicity, has good MRI compatibility, and can cope with the fringe magnetic field of the MRI scanner, radiofrequency signals during image acquisition, and heart rate changes. High-resolution 4D flow imaging is successfully acquired both in vivo and in vitro. The difference between the 2D and 4D measurements is ≤ 21%. The incidence of false triggers is 0% in all tests, which is unattainable for other known cardiac gating methods. Conclusion: The system has good MRI compatibility and can provide a stable and accurate trigger signal based on pressure waveform. It opens the door to applications where the previous gating methods were difficult to implement or not applicable.
... However, the most commonly applied adjustment methods are adversely affected by high electromagnetic interference, acoustic noise, and vibration. These effects become even more prominent with the increase in the MRI's magnetic field strength [4], [5]. ...
Article
Full-text available
During the past decades, fiber-optic technology has become a very popular tool for vital signs monitoring. Thanks to its advantageous properties, such as noninvasiveness, biocompatibility, and resistance to electromagnetic interferences, this methodology started to be explored under the conditions of a magnetic resonance (MR) environment. This review article presents the motivation and possibilities of using fiber-optic sensors (FOSs) in MR environment and summarizes the studies dealing with experimental validation of their compatibility with MR. Several aspects of the presented issue are highlighted and discussed, such as suitability of the fiber-optic approach for MR triggering, precision of vital sign detection, development of sensor designs, and its application to patient’s body. From the literature review, it can be concluded that FOSs have promising future in the field of cardiorespiratory monitoring in MR environment. This is mainly due to their advantages originating from sensing mechanical signals instead of electrical ones, which makes them resistant to MR interference and extrasystoles. Moreover, these sensors are easy to use, reusable, and suitable for combined monitoring. However, there are several shortcomings that should be solved in future research before introducing them to clinical practice, namely, signal’s delay or optimal placement of sensors.
... Third, despite careful placement, 37 acquisition of an acceptable ECG signal is error-prone and remains difficult compared with lower field strength. Despite other findings, 66 a reliable ECG signal was obtained in 10 of 11 volunteers in this study as in Ref. [67]. We speculate that the poor efficiency found in one volunteer may be related to the respiratory motion, but a separate analysis is necessary. ...
Article
Full-text available
Purpose Respiratory motion‐compensated (MC) 3D cardiac fat‐water imaging at 7T. Methods Free‐breathing bipolar 3D triple‐echo gradient‐recalled‐echo (GRE) data with radial phase‐encoding (RPE) trajectory were acquired in 11 healthy volunteers (7M\4F, 21–35 years, mean: 30 years) with a wide range of body mass index (BMI; 19.9–34.0 kg/m²) and volunteer tailored B1+ shimming. The bipolar‐corrected triple‐echo GRE‐RPE data were binned into different respiratory phases (self‐navigation) and were used for the estimation of non‐rigid motion vector fields (MF) and respiratory resolved (RR) maps of the main magnetic field deviations (ΔB0). RR ΔB0 maps and MC ΔB0 maps were compared to a reference respiratory phase to assess respiration‐induced changes. Subsequently, cardiac binned fat‐water images were obtained using a model‐based, respiratory motion‐corrected image reconstruction. Results The 3D cardiac fat‐water imaging at 7T was successfully demonstrated. Local respiration‐induced frequency shifts in MC ΔB0 maps are small compared to the chemical shifts used in the multi‐peak model. Compared to the reference exhale ΔB0 map these changes are in the order of 10 Hz on average. Cardiac binned MC fat‐water reconstruction reduced respiration induced blurring in the fat‐water images, and flow artifacts are reduced in the end‐diastolic fat‐water separated images. Conclusion This work demonstrates the feasibility of 3D fat‐water imaging at UHF for the entire human heart despite spatial and temporal B1+ and B0 variations, as well as respiratory and cardiac motion.
... On the other hand, WHOCARES allowed for the retrospective correction of such data (Figure 5,right). This has numerous applications; for example at ultra-high-field (≥ 7T ), where the effect of cardiac noise is more prominent and the ability to measure physiological traces is often hindered (Stäb et al. (2016)). ...
Preprint
Full-text available
Cardiac pulsation is a physiological confound of functional magnetic resonance imaging (fMRI) time-series that introduces spurious signal fluctuations in proximity to blood vessels. fMRI alone is not sufficiently fast to resolve cardiac pulsation. Depending on the ratio between the instantaneous heart-rate and the acquisition sampling frequency (1/TR, with TR being the repetition time), the cardiac signal may alias into the frequency band of neural activation. The introduction of simultaneous multi-slice (SMS) imaging has significantly reduced the chances of cardiac aliasing. However, the necessity of covering the entire brain at high spatial resolution restrain the shortest TR to just over 0.5 seconds, which is in turn not sufficiently short to resolve cardiac pulsation beyond 60 beats per minute. Recently, hyper-sampling of the fMRI time-series has been introduced to overcome this issue. While each anatomical location is sampled every TR seconds, the time between consecutive excitations is shorter and thus adequate to resolve cardiac pulsation. In this study, we show that it is feasible to temporally and spatially resolve cardiac waveforms at each voxel location by combining a dedicated hyper-sampling decomposition scheme with SMS. We developed the technique on 774 healthy subjects selected from the Human Connectome Project (HCP) and validated the technology against the RETROICOR method. The proposed approach, which we name Data-driven WHOle-brain CArdiac signal REgression from highly accelerated simultaneous multi-Slice fMRI acquisitions (WHOCARES), is fully data-driven, does not make specific assumptions on cardiac pulsatility, and is independent from external physiological recordings so that the retrospective correction of fMRI data becomes possible when such measurements are not available. WHOCARES is freely available at https://github.com/gferrazzi/WHOCARES.
... At 7T, as an alternative, acoustic triggering has been successfully used [32]. Prior studies demonstrated that ECG and acoustic triggering are interchangeable for volumetric analyses [16,32,48,49]. Adding to these findings, our data show that flow measurements are feasible with both triggering techniques and provide comparable results. ...
Article
Full-text available
Introduction Cardiac magnetic resonance (CMR) at ultrahigh field (UHF) offers the potential of high resolution and fast image acquisition. Both technical and physiological challenges associated with CMR at 7T require specific hardware and pulse sequences. This study aimed to assess the current status and existing, publicly available technology regarding the potential of a clinical application of 7T CMR. Methods Using a 7T MRI scanner and a commercially available radiofrequency coil, a total of 84 CMR examinations on 72 healthy volunteers (32 males, age 19–70 years, weight 50–103 kg) were obtained. Both electrocardiographic and acoustic triggering were employed. The data were analyzed regarding the diagnostic image quality and the influence of patient and hardware dependent factors. 50 complete short axis stacks and 35 four chamber CINE views were used for left ventricular (LV) and right ventricular (RV), mono-planar LV function, and RV fractional area change (FAC). Twenty-seven data sets included aortic flow measurements that were used to calculate stroke volumes. Subjective acceptance was obtained from all volunteers with a standardized questionnaire. Results Functional analysis showed good functions of LV (mean EF 56%), RV (mean EF 59%) and RV FAC (mean FAC 52%). Flow measurements showed congruent results with both ECG and ACT triggering. No significant influence of experimental parameters on the image quality of the LV was detected. Small fractions of 5.4% of LV and 2.5% of RV segments showed a non-diagnostic image quality. The nominal flip angle significantly influenced the RV image quality. Conclusion The results demonstrate that already now a commercially available 7T MRI system, without major methods developments, allows for a solid morphological and functional analysis similar to the clinically established CMR routine approach. This opens the door towards combing routine CMR in patients with development of advanced 7T technology.
Article
Objective: The aim of this study was to test the hypothesis that there are good agreements between cardiac functional and structural indices derived from magnetic resonance imaging (MRI) sequences triggered with pilot tone (PT) and electrocardiogram (ECG). Materials and methods: Sixteen healthy volunteers (11 male, age 21-76 years) underwent a cardiac MRI scan. Cine MRI, T1, and T2 mapping were acquired by using PT and ECG triggering. Quantitative measurements, including left and right ventricular end-diastolic volume, end-systolic volume, stroke volume, ejection fraction, longitudinal strain, left ventricular T1 and T2 values, left and right atrial longitudinal strain, and maximal/minimal volumes, were measured. The interclass correlation coefficient, coefficient of variation, and Bland-Altman plots were used to evaluate the agreements between measurements derived by MRI sequences triggered with 2 methods. Results: There were no significant differences among end-diastolic volume, end-systolic volume, stroke volume, ejection fraction, left ventricle mass, T1 and T2 values, or longitudinal strains acquired using PT and ECG. There were good agreements and low variations between the levels of these indices acquired with PT and ECG. Interclass correlation coefficients mainly ranged from 0.73 to 0.98. The coefficients of variation ranged from 1.4% to 22.6%. Conclusions: Pilot tone-triggered MRI provides comparable measurements of cardiac function, motion, and structure as ECG-triggered MRI. Pilot tone has the potential to become a backup of ECG gating in cardiovascular imaging.
Article
Full-text available
Arterial spin labeling (ASL) is a cardiovascular magnetic resonance (CMR) technique for mapping regional myocardial blood flow. It does not require any contrast agents, is compatible with stress testing, and can be performed repeatedly or even continuously. ASL-CMR has been performed with great success in small-animals, but sensitivity to date has been poor in large animals and humans and remains an active area of research. This review paper summarizes the development of ASL-CMR techniques, current state-of-the-art imaging methods, the latest findings from pre-clinical and clinical studies, and future directions. We also explain how successful developments in brain ASL and small-animal ASL-CMR have helped to inform developments in large animal and human ASL-CMR.
Article
Full-text available
Purpose: Phosphorus MR spectroscopy ((31) P-MRS) is a powerful tool for investigating tissue energetics in vivo. Cardiac (31) P-MRS is typically performed using surface coils that create an inhomogeneous excitation field across the myocardium. Accurate measurements of B1+ (and hence flip angle) are necessary for quantitative analysis of (31) P-MR spectra. We demonstrate a Bloch-Siegert B1+-mapping method for this purpose. Theory and methods: We compare acquisition strategies for Bloch-Siegert B1+-mapping when there are several spectral peaks. We optimize a Bloch-Siegert sensitizing (Fermi) pulse for cardiac (31) P-MRS at 7 Tesla (T) and apply it in a three-dimensional (3D) chemical shift imaging sequence. We validate this in phantoms and skeletal muscle (against a dual-TR method) and present the first cardiac (31) P B1+-maps at 7T. Results: The Bloch-Siegert method correlates strongly (Pearson's r = 0.90 and 0.84) and has bias <25 Hz compared with a multi-TR method in phantoms and dual-TR method in muscle. Cardiac 3D B1+-maps were measured in five normal volunteers. B1+ maps based on phosphocreatine and alpha-adenosine-triphosphate correlated strongly (r = 0.62), confirming that the method is T1 insensitive. Conclusion: The 3D (31) P Bloch-Siegert B1+-mapping is consistent with reference methods in phantoms and skeletal muscle. It is the first method appropriate for (31) P B1+-mapping in the human heart at 7T. Magn Reson Med, 2015. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Article
Full-text available
To restore 12-lead electrocardiographic (ECG) signal fidelity inside MRI by removing magnetic field gradient-induced voltages during high gradient duty cycle sequences. A theoretical equation was derived to provide first- and second-order electrical fields induced at individual ECG electrodes as a function of gradient fields. Experiments were performed at 3T on healthy volunteers using a customized acquisition system that captured the full amplitude and frequency response of ECGs, or a commercial recording system. The 19 equation coefficients were derived via linear regression of data from accelerated sequences and were used to compute induced voltages in real-time during full resolution sequences to remove ECG artifacts. Restored traces were evaluated relative to ones acquired without imaging. Measured induced voltages were 0.7 V peak-to-peak during balanced steady state free precession (bSSFP) with the heart at the isocenter. Applying the equation during gradient echo sequencing, three-dimensional fast spin echo, and multislice bSSFP imaging restored nonsaturated traces and second-order concomitant terms showed larger contributions in electrodes further from the magnet isocenter. Equation coefficients are evaluated with high repeatability (ρ = 0.996) and are dependent on subject, sequence, and slice orientation. Close agreement between theoretical and measured gradient-induced voltages allowed for real-time removal. Prospective estimation of sequence periods in which large induced voltages occur may allow hardware removal of these signals. Magn Reson Med, 2015. © 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.
Article
Full-text available
In Cardiovascular Magnetic Resonance (CMR), the synchronization of image acquisition with heart motion is performed in clinical practice by processing the electrocardiogram (ECG). The ECG-based synchronization is well established for MR scanners with magnetic fields up to 3 T. However, this technique is prone to errors in ultra high field environments, e.g. in 7 T MR scanners as used in research applications. The high magnetic fields cause severe magnetohydrodynamic (MHD) effects which disturb the ECG signal. Image synchronization is thus less reliable and yields artefacts in CMR images. A strategy based on Independent Component Analysis (ICA) was pursued in this work to enhance the ECG contribution and attenuate the MHD effect. ICA was applied to 12-lead ECG signals recorded inside a 7 T MR scanner. An automatic source identification procedure was proposed to identify an independent component (IC) dominated by the ECG signal. The identified IC was then used for detecting the R-peaks. The presented ICA-based method was compared to other R-peak detection methods using 1) the raw ECG signal, 2) the raw vectorcardiogram (VCG), 3) the state-of-the-art gating technique based on the VCG, 4) an updated version of the VCG-based approach and 5) the ICA of the VCG. ECG signals from eight volunteers were recorded inside the MR scanner. Recordings with an overall length of 87min accounting for 5457 QRS complexes were available for the analysis. The records were divided into a training and a test dataset. In terms of R-peak detection within the test dataset, the proposed ICA-based algorithm achieved a detection performance with an average sensitivity (Se) of 99.2%, a positive predictive value (+P) of 99.1%, with an average trigger delay and jitter of 5.8ms and 5.0ms, respectively. Long term stability of the demixing matrix was shown based on two measurements of the same subject, each being separated by one year, whereas an averaged detection performance of Se = 99.4% and +P = 99.7% was achieved. Compared to the state-of-the-art VCG-based gating technique at 7 T, the proposed method increased the sensitivity and positive predictive value within the test dataset by 27.1% and 42.7%, respectively. The presented ICA-based method allows the estimation and identification of an IC dominated by the ECG signal. R-peak detection based on this IC outperforms the state-of-the-art VCG-based technique in a 7 T MR scanner environment.
Article
Full-text available
Background The electrocardiogram (ECG) is important for gating purposes in cardiac magnetic resonance imaging (CMR). However, the magnetohydrodynamic (MHD) effect, which is caused by the flow of blood in the static magnetic field, makes it difficult to record clean ECG signals for gating. The vectorcardiogram (VCG), which can be derived from the ECG signal, is commonly used for gating purposes and has been well established for magnetic fields strength of up to 3T. However, for higher field strengths this method is prone to errors [1]. This work tries to explain the reasons for the VCG based methods not being suitable for cardiac gating in ultra high field MRI. Methods ECGs were recorded using a standard 12-lead Holter ECG (Getemed, Germany) inside a 7T MR scanner (Siemens Magnetom), a 3T MR scanner (Philips Achieva) and outside the MR scanner as references. Measurements were made on five healthy volunteers while MR imaging was switched off. The VCGs were derived from the 12-lead ECG using the inverse Dower matrix [2]. The VCG method [3] was implemented and applied to the acquired data sets. This method measures the distance in the VCG space between a static reference vector defined by the R-wave (R) peak of an ECG signal recorded outside the MR scanner (at 0T) and the VCG vector recorded inside the MR scanner. Those points where a minimal distance and angle between both vectors is reached are classified as R peaks.
Article
Purpose: Cardiac MRI at 7T suffers from contrast heterogeneity that can be mitigated with parallel transmission (pTX) and, when performed during breath-hold, from a limited number of slices that can be multiplied with multiband (MB) radiofrequency pulses by simultaneous excitation of multiple slices (SMS). The goal of this study was to apply both approaches simultaneously. Methods: Using a 16-channel transmit/receive body coil, pTX SMS was applied with/without CAIPIRINHA with a modified gradient echo cine sequence. Different calibration schemes were investigated for the slice-GRAPPA reconstruction kernels as a function of the cardiac cycle. Results: Excellent slice separation for MB = 2 was achieved with CAIPIRINHA, with slice leakage values below 3% for 99% of all voxels. A critical finding of this study was the variation of the MB leakage factor in the heart by as much as 30% throughout the cardiac cycle, which was reduced greatly when reconstruction kernels were calibrated on multiple cardiac phases. Acceptable results were still obtained when applying further acceleration with MB = 3 in combination with in-plane GRAPPA. In one case, two-spoke pulses were compared with one-spoke pulses, resulting as expected in improved homogeneity. Conclusion: pTX SMS imaging at 7T can address contrast heterogeneity while allowing larger slice coverage in cardiac MRI performed under breath-hold. Magn Reson Med, 2016. © 2016 Wiley Periodicals, Inc.
Article
PurposeAccurate triggering of the cardiac cycle is mandatory for optimal image acquisition and thus diagnostic quality in cardiac magnetic resonance imaging. The purpose of this work was to evaluate Doppler ultrasound as an alternative trigger method in cardiac MRI.Methods Steady-state free precession (SSFP) 2D cine CMR was performed in 11 healthy subjects at 1.5T. Doppler ultrasound (DUS), electrocardiogram (ECG) and pulse oximetry (POX) were used for cardiac triggering. DUS peak detection was verified in comparison to ECG. Quantitative analysis of image quality of each gating method was determined by calculating endocardial border sharpness (EBS) and left ventricular (LV) function parameters and compared with ECG.ResultsMean difference between DUS and ECG in detected RR intervals was 0.04 ± 63 ms (r = 0.96). Trigger jitter was not different between ECG and DUS (P = 0.15) but significant different between ECG and POX (P = 0.01). EBS was similar between each method (3.1 ± 0.2 / 2.6 ± 0.2 / 2.9 ± 0.2 pixel). Mean differences in stroke volume were not significantly different with −1 ± 7 mL (ECG/DUS, P = 0.9) and 2 ± 10 mL (ECG/POX, P = 0.8).Conclusion Cine cardiac MRI using DUS was successfully demonstrated. DUS triggering is an alternative method for cardiac MRI and may be applied in a clinical setting. Magn Reson Med, 2014. © 2014 Wiley Periodicals, Inc.
Article
To develop a technique that accurately detects the QRS complex in 1.5 Tesla (T), 3T, and 7T MRI scanners. During early systole, blood is rapidly ejected into the aortic arch, traveling perpendicular to the MRI's main field, which produces a strong voltage (VMHD ) that eclipses the QRS complex. Greater complexity arises in arrhythmia patients, since VMHD varies between sinus-rhythm and arrhythmic beats. The 3DQRS method uses a kernel consisting of 6 electrocardiogram (ECG) precordial leads (V1-V6), compiled from a 12-lead ECG performed outside the magnet. The kernel is cross-correlated with signals acquired inside the MRI to identify the QRS complex in real time. The 3DQRS method was evaluated against a vectorcardiogram (VCG)-based approach in two premature ventricular contraction (PVC) and two atrial fibrillation (AF) patients, a healthy exercising athlete, and eight healthy volunteers, within 1.5T and 3T MRIs, using a prototype MRI-conditional 12-lead ECG system. Two volunteers were recorded at 7T using a Holter recorder. For QRS complex detection, 3DQRS subject-averaged sensitivity levels, relative to VCG were: 1.5T (100% versus 96.7%), 3T (98.9% versus 92.2%), and 7T (96.2% versus 77.7%). The 3DQRS method was shown to be more effective in cardiac gating than a conventional VCG-based method. Magn Reson Med, 2014. © 2014 Wiley Periodicals, Inc.
Article
To evaluate and to compare Parallel Imaging and Compressed Sensing acquisition and reconstruction frameworks based on simultaneous multislice excitation for high resolution contrast-enhanced myocardial first-pass perfusion imaging with extended anatomic coverage. The simultaneous multislice imaging technique MS-CAIPIRINHA facilitates imaging with significantly extended anatomic coverage. For additional resolution improvement, equidistant or random undersampling schemes, associated with corresponding reconstruction frameworks, namely Parallel Imaging and Compressed Sensing can be used. By means of simulations and in vivo measurements, the two approaches were compared in terms of reconstruction accuracy. Comprehensive quality metrics were used, identifying statistical and systematic reconstruction errors. The quality measures applied allow for an objective comparison of the frameworks. Both approaches provide good reconstruction accuracy. While low to moderate noise enhancement is observed for the Parallel Imaging approach, the Compressed Sensing framework is subject to systematic errors and reconstruction induced spatiotemporal blurring. Both techniques allow for perfusion measurements with a resolution of 2.0 × 2.0 mm(2) and coverage of six slices every heartbeat. Being not affected by systematic deviations, the Parallel Imaging approach is considered to be superior for clinical studies.J. Magn. Reson. Imaging 2013;00:000-000. © 2013 Wiley Periodicals, Inc.
Article
To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1 (+) field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm(3) for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T. Magn Reson Med, 2013. © 2013 Wiley Periodicals, Inc.