ArticlePDF Available

Rapid volumetric imaging with Bessel-Beam three-photon microscopy

Optica Publishing Group
Biomedical Optics Express
Authors:

Abstract and Figures

Owing to its tissue-penetration ability, multi-photon fluorescence microscopy allows for the high-resolution, non-invasive imaging of deep tissue in vivo; the recently developed three-photon microscopy (3PM) has extended the depth of high-resolution, non-invasive functional imaging of mouse brains to beyond 1.0 mm. However, the low repetition rate of femtosecond lasers that are normally used in 3PM limits the temporal resolution of point-scanning three-photon microscopy. To increase the volumetric imaging speed of 3PM, we propose a combination of an axially elongated needle-like Bessel-beam with three-photon excitation (3PE) to image biological samples with an extended depth of focus. We demonstrate the higher signal-to-background ratio (SBR) of the Bessel-beam 3PM compared to the two-photon version both theoretically and experimentally. Finally, we perform simultaneous calcium imaging of brain regions at different axial locations in live fruit flies and rapid volumetric imaging of neuronal structures in live mouse brains. These results highlight the unique advantage of conducting rapid volumetric imaging with a high SBR in the deep brain in vivo using scanning Bessel-3PM.
This content is subject to copyright. Terms and conditions apply.
Rapid volumetric imaging with Bessel-Beam
three-photon microscopy
BINGYING CHEN,1,7 XIAOSHUAI HUANG,2,7 DONGZHOU GOU,2 JIANZHI
ZENG,3 GUOQING CHEN,2 MEIJUN PANG,2 YANHUI HU,4 ZHE ZHAO,5
YUNFENG ZHANG,4 ZHUAN ZHOU,2 HAITAO WU,5 HEPING CHENG,2 ZHIGANG
ZHANG,1 CHRIS XU,2,6 YULONG LI,3 LIANGYI CHEN,2,8 AND AIMIN WANG1,9
1State Key Laboratory of Advanced Optical Communication System and Networks, School of Electronics
Engineering and Computer Science, Peking University, Beijing 100871, China
2State Key Laboratory of Membrane Biology, Beijing Key Laboratory of Cardiometabolic Molecular
Medicine, Institute of Molecular Medicine, Peking University, Beijing 100871, China
3State Key Laboratory of Membrane Biology, School of Life Sciences, PKU-IDG/McGovern Institute for
Brain Research, Peking University, Beijing 100871, China
4School of Electronics Engineering and Computer Science, Peking University, Beijing 100871, China
5Department of Neurobiology, Institute of Basic Medical Sciences, Beijing, China
6School of Applied and Engineering Physics, Cornell University, Ithaca, NY 14853, USA
7These authors contributed equally to this work
8lychen@pku.edu.cn
9wangaimin@pku.edu.cn
Abstract: Owing to its tissue-penetration ability, multi-photon fluorescence microscopy
allows for the high-resolution, non-invasive imaging of deep tissue in vivo; the recently
developed three-photon microscopy (3PM) has extended the depth of high-resolution, non-
invasive functional imaging of mouse brains to beyond 1.0 mm. However, the low repetition
rate of femtosecond lasers that are normally used in 3PM limits the temporal resolution of
point-scanning three-photon microscopy. To increase the volumetric imaging speed of 3PM,
we propose a combination of an axially elongated needle-like Bessel-beam with three-photon
excitation (3PE) to image biological samples with an extended depth of focus. We
demonstrate the higher signal-to-background ratio (SBR) of the Bessel-beam 3PM compared
to the two-photon version both theoretically and experimentally. Finally, we perform
simultaneous calcium imaging of brain regions at different axial locations in live fruit flies
and rapid volumetric imaging of neuronal structures in live mouse brains. These results
highlight the unique advantage of conducting rapid volumetric imaging with a high SBR in
the deep brain in vivo using scanning Bessel-3PM.
© 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
OCIS codes: (110.0180) Microscopy; (180.2520) Fluorescence microscopy; (180.5810) Scanning microscopy;
(190.4180) Multiphoton processes.
References and links
1. W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science
248(4951), 73–76 (1990).
2. W. Denk and F. Helmchen, “Deep tissue two-photon microscopy,” Nat. Methods 2(12), 932–940 (2005).
3. W. R. Zipfel, R. M. Williams, and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,”
Nat. Biotechnol. 21(11), 1369–1377 (2003).
4. D. Kobat, M. E. Durst, N. Nishimura, A. W. Wong, C. B. Schaffer, and C. Xu, “Deep tissue multiphoton
microscopy using longer wavelength excitation,” Opt. Express 17(16), 13354–13364 (2009).
5. N. G. Horton, K. Wang, D. Kobat, C. G. Clark, F. W. Wise, C. B. Schaffer, and C. Xu, “In vivo three-photon
microscopy of subcortical structure within an intact mouse brain,” Nat. Photonics 7(3), 205–209 (2013).
6. D. G. Ouzounov, T. Wang, M. Wang, D. D. Feng, N. G. Horton, J. C. C. Hernandez, Y. Cheng, J. Reimer, A. S.
Tolias, N. Nishimura, and C. Xu, “In vivo three-photon imaging of activity of GCaMP6-labeled neurons deep in
intact mouse brain,” Nat. Methods 14(4), 388–390 (2017).
7. L. Cheng, N. G. Horton, K. Wang, S. Chen, and C. Xu, “Measurements of multiphoton action cross sections for
multiphoton microscopy,” Biomed. Opt. Express 5(10), 3427–3433 (2014).
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1992
#322760
https://doi.org/10.1364/BOE.9.001992
Journal © 2018
Received 9 Feb 2018; revised 22 Mar 2018; accepted 23 Mar 2018; published 29 Mar 2018
8. E. J. Botcherby, R. Juskaitis, M. J. Booth, and T. Wilson, “An optical technique for remote focusing in
microscopy,” Opt. Commun. 281(4), 880–887 (2008).
9. W. Yang, J. K. Miller, L. C. Reid, E. Pnevmatikakis, L. Paninski, R. Yuste, and D. S. Peterka, “Simultaneous
multi-plane imaging of neural circuits,” Neuron 89(2), 269–284 (2016).
10. G. D. Reddy, K. Kelleher, R. Fink, and P. Saggau, “Three-dimensional random access multiphoton microscopy
for fast functional imaging of neuronal activity,” Nat. Neurosci. 11(6), 713–720 (2008).
11. E. J. Botcherby, R. Juskaitis, and T. Wilson, “Scanning two photon fluorescence microscopy with extended
depth of field,” Opt. Commun. 268(2), 253–260 (2006).
12. P. Dufour, M. Piché, Y. D. Koninck, and N. McCarthy, “Two-photon excitation fluorescence microscopy with a
high depth of field using an axicon,” Appl. Opt. 45(36), 9245–9252 (2006).
13. G. Theriault, Y. D. Koninck, and N. McCarthy, “Extended depth of field microscopy for rapid volumetric two-
photon imaging,” Opt. Express 21(8), 10095–10104 (2013).
14. R. Lu, W. Sun, Y. Liang, A. Kerlin, J. Bierfeld, J. D. Seelig, D. E. Wilson, B. Scholl, B. Mohar, M. Tanimoto,
M. Koyama, D. Fitzpatrick, M. B. Orger, and N. Ji, “Video-rate volumetric functional imaging of the brain at
synaptic resolution,” Nat. Neurosci. 20(4), 620–628 (2017).
15. B. Chen, T. Jiang, W. Zong, L. Chen, Z. Zhang, and A. Wang, “910nm femtosecond Nd-doped fiber laser for in
vivo two-photon microscopic imaging,” Opt. Express 24(15), 16544–16549 (2016).
16. B. Chen, H. Rong, X. Huang, R. Wu, D. Wu, Y. Li, L. Feng, Z. Zhang, L. Chen, and A. Wang, “Robust hollow-
fiber-pigtailed 930 nm femtosecond Nd:fiber laser for volumetric two-photon imaging,” Opt. Express 25(19),
22704–22709 (2017).
17. L. Liang, Y. Li, C. J. Potter, O. Yizhar, K. Deisseroth, R. W. Tsien, and L. Luo, “GABAergic projection neurons
route selective olfactory inputs to specific higher-order neurons,” Neuron 79(5), 917–931 (2013).
18. M. Ng, R. D. Roorda, S. Q. Lima, B. V. Zemelman, P. Morcillo, and G. Miesenboeck, “Transmission of
olfactory information between three populations of neurons in the antennal lobe of the fly,” Neuron 36(3), 463–
474 (2002).
19. C. Xu and W. W. Webb, “Multiphoton Excitation of Molecular Fluorophores and Nonlinear Laser Microscopy,”
in Topics in Fluorescence Spectroscopy, J. R. Lakowiczv, ed. (Kluwer Academic Publishers, 1997).
20. O. Brzobohaty, T. Cizmar, and P. Zemanek, “High quality quasi-Bessel beam generated by round-tip axicon,”
Opt. Express 16(17), 12688–12700 (2008).
21. R. F. Stocker, G. Heimbeck, N. Gendre, and J. S. Belle, “Neuroblast Ablation in Drosophila P[GAL4] Lines
Reveals Origins of Olfactory Interneurons,” J. Neurobiol. 32(5), 443–456 (1997).
22. J. W. Wang, A. M. Wong, J. Flores, L. B. Vosshall, and R. Axel, “Two-Photon Calcium Imaging Reveals an
Odor-Evoked Map of Activity in the Fly Brain,” Cell 112(2), 271–282 (2003).
23. J. B. Connolly, I. J. H. Roberts, J. D. Armstrong, K. Kaiser, M. Forte, T. Tully, and C. J. O’Kane, “Associative
Learning Disrupted by Impaired Gs Signaling in Drosophila Mushroom Bodies,” Science 274(5295), 2104–2107
(1996).
24. M. Heisenberg, “Mushroom Body Memoir: From Maps to Models,” Nature 4(4), 266–275 (2003).
1. Introduction
Multi-photon microscopy (MPM) holds the promise of unique and impactful applications in
cell biology and neuroscience research, where it can enable the high-resolution imaging of the
neuronal structures buried deep inside the intact brain [1–3]. Benefiting from higher order
nonlinear excitation and longer excitation wavelengths, 3PM has shown better performance in
terms of the signal-to-background ratio (SBR) than has two-photon microscopy (2PM), and it
has extended the non-invasive functional imaging of the mouse brain to beyond 1.0 mm.
Using 3PM, the spontaneous activity recording in the hippocampal stratum pyramidale has
been realized [4–6].
MPM normally uses focused femtosecond laser pulses to illuminate the sample, which
limits the excitation volume to the focal point. Therefore, the sample must be imaged pixel by
pixel to obtain a 3D image stack. In particular, the mechanical movement of the objective to
perform the axial scanning decreases the imaging rate. This hampers its application in
recording the activity of the large neuronal circuits distributed in a 3D volume. Moreover, the
low repetition rate of femtosecond lasers (1-2 MHz) optimized for deep-tissue 3PE further
limits the temporal resolution of point-scanning 3PM [7].
Various alternative axial scanning methods have been proposed to address this problem,
such as fast z-scanning with remote focusing [8], wavefront shaping using spatial light
modulators (SLM) [9], and random access scanning with acousto-optic deflectors (AOD)
[10]. For sparsely labeled samples, using a Bessel beam to image a 3D volume is another
feasible approach. The Bessel beam has an axially elongated, needle-like optical intensity
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1993
distribution and maintains the same lateral resolution over an extended depth of field [11, 12].
Hence, a single scan can generate a 2D projection of the 3D volume image. Given that a
single neuron often has dendrites that are sparsely distributed in a 3D volume, Bessel beam
scanning is an excellent fast imaging method for obtaining volumetric information about the
neuronal structures and activities. Moreover, the image obtained with Bessel-3PM is a 2D
projection of a 3D object, which cannot offer the information in z-axis. Therefore, we can use
Bessel-3PM to obtain the temporal information at different depths and use point-scanning
3PM to obtain the imaging depth as assistance. Nevertheless, the sample must be sparsely
labeled; otherwise, this approach will cease to be effective if too much axial information is
stacked together.
The Bessel beam has been used to extend the depth of field in 2PM of neurons [13, 14].
However, the intensity distribution of the zeroth-order Bessel beam in the transverse plane
has strong side lobes under two-photon excitation (2PE) that generate an undesirable
background and decrease the SBR. Because 3PE and 2PE fluorescence obey the cubic and
quadratic dependence on the excitation power, respectively, a Bessel beam with 3PE has
much smaller side lobes than does that with 2PE. In this paper, we compared the performance
of 2PE and 3PE Bessel-beam using femtosecond lasers at 920 nm and 1300 nm, respectively,
and demonstrate the superior SBR of volumetric Bessel 3PM in imaging zebrafish larva, fruit
flies and mice in vivo.
2. Method and materials
2.1 Three- and two-photon imaging systems with Bessel-beam or Gaussian-beam
excitation
A schematic diagram of the optical setup is shown in Fig. 1.
Fig. 1. Schematic diagram of the apparatus. Flipper mirrors: F1 was used to switch between
920 nm and 1300 nm lasers; F2 and F3 were used to switch between the Gaussian and Bessel
beam excitation. Axicon, α = 2 degrees. Lenses: L1 and L2, 50 mm (or 75, 100 mm) focal
length, determines the effective excitation NAs and lengths of the Bessel beam. L3 (100 mm),
L4 (100 mm), SL (54 mm) and TL (200 mm) were used to create multiple conjugate planes of
the back focal plane (BFP) of the objective in the light path; L5, collection lens (30 mm).
Galvo mirrors: GMX/GMY, for the x/y direction of beam scanning. DM, 735-nm long-pass
dichroic mirror; a 770-nm short pass filter and a 525/40 nm bandpass filter were used here
(Filters) for emission collection.
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1994
The light source for 3PM was a non-collinear optical parametric amplifier (NOPA,
Spectra Physics) pumped by a 1040 nm high-power femtosecond laser (Spirit, Spectra
Physics). The NOPA produced 50-fs pulses at 1300 nm with an average power over 500 mW
and a repetition rate of 400 kHz. The femtosecond laser at 1300 nm could also be switched to
a home-built femtosecond Nd:fiber laser, which delivered up to 800 mW, 150-fs pulses at 930
nm with a repetition rate of 50 MHz [15, 16]. This configuration enabled a comparison of the
performance of 2PM and 3PM using the same specimen. After using a pair of scan mirrors
(6215H, Cambridge Technology), the laser pulses were directed into a 54-mm scan lens and a
200-mm tube lens. A water immersion objective lens (XLPLN25XWMP2, 25X, NA 1.05,
Olympus) was used to focus the light into the sample. The fluorescence emitted from the
sample was epi-collected and directed to a GaAsP PMT detector (H7422p-40, Hamamatsu)
by a long-pass dichroic mirror (FF735-Di02-25 × 36, Semrock). A collection lens, a short-
pass filter (ET770sp-2p 1500IR, Chroma) and a 525/40 nm bandpass filter (FF02-525/40-25,
Semrock) were used to collect the fluorescence signals. The microscope was compatible with
two scanning modes: the Bessel-beam scanning mode and the point-scanning of Gaussian
focal spot mode. For the Bessel-beam mode, a plano-convex axicon (1-APX-2-VIS-G254-P,
Altechna) was used to generate the Bessel beam to excite the sample. The axial FWHM of
Bessel beams depends on the different combinations of the angle α of the axicon (angle
between the sloped and plane faces) and the focal length of Lens 2. For the point-scanning
mode, lens L1 was used to collimate the beam before directing to the scan mirror. The
transmission efficiencies through the system under both modes were approximately 25%.
The length of the Bessel-beam depends on optical components; however, for a certain
biological experiment, the maximum extension is limited by the sample’s property and the
single pulse energy of the laser source. For tissues with high transparency, a long Bessel
beam could be used to illuminate the entire imaging region (e.g. zebrafish). However, the
length of the Bessel-beam should not be much longer than the thickness of the sample;
otherwise, the excessive background fluorescence would be produced. Moreover, because of
the limited excitation power, the beam length should be adjusted to even shorter to increase
the imaging depth (e.g. mouse brain in our case). Hence, the appropriate length of depends on
the sample’s thickness and transparency, the depth of the imaging region, the single pulse
energy of the laser source and other factors.
2.2 Sample preparation
In this work, three representative tissues (zebrafish, Drosophila and mouse brain) were used
to characterize the properties of the imaging system.
The zebrafish was used to compare the SBR between 2PE and 3PE. The vascular
epithelial cells were labeled with EGFP in the Tg (kdrl:EGFP) transgenic fish line. For
imaging, the fish was raised in an embryo medium containing 0.002% phenylthiourea (PTU,
Sigma) to suppress pigmentation synthesis. Prior to live imaging, the fish was anaesthetized
with 0.01% tricaine (Sigma) to minimize the movement artifacts during imaging. Then, the
anaesthetized zebrafish was embedded in a 1% ultra-pure agarose (Invitrogen) and immersed
in E3 medium containing 0.01% tricaine.
For functional imaging, flies expressing GCaMP were reared at 25 °C on standard
cornmeal for 8-12 days after eclosion before experiments. The procedures used for mounting
and dissecting the flies for imaging were the same as previously described [17]. Flies were
mounted to a small dish by tape, and the antennae were exposed to the air. The cuticle, fat
bodies and air sacs between the eyes were removed to expose the brain. The exposed brains
were perfused with adult-like hemolymph (ALH). Isoamyl acetate (IA, Sigma-Aldrich; Cat#
306967) was initially diluted by 100-fold in mineral oil (Sigma-Aldrich; Cat# 69794) (100 μl
IA in 900 μl mineral oil) and then placed in a glass bottle. During olfactory stimulation, the
airflow carrying IA (200 μl/min) was mixed with purified air (1000 ml/min) and delivered to
the antennae of the flies. The identification of glomeruli was based on the map of the antenna
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1995
lobe published previously [18]. The Fiji software was used to process the images of
functional imaging experiments, including subtracting the background, smoothing with a
Gaussian blur filter and measuring the fluorescence within the ROIs. The Origin software was
used to process the raw data with binning and plot the traces. The pseudo-color snapshots
(color-encoded by ΔF/F0) generated by Matlab with custom scripts were the averaged results
of 2-4 repeats.
We used male C57BL/6J mice (20 g, 8 postnatal weeks, Thy-1-YFP) for in vivo brain
imaging. Animals were prepared using the methods described in Ref [4]. The craniotomies
were performed centered at 2.2 mm posterior and 3 mm lateral to the Bregma point. All
procedures were approved by the Peking University Animal Use and Care Committee and
complied with the standards of the Association for Assessment and Accreditation of
Laboratory Animal Care, including the animal breeding and experimental manipulation.
3. Results and discussions
Following the analysis of Xu et al. [19], we can obtain the n-photon fluorescence:
()
n0
C() ()
nn
n
Ft I t dS=rr (1)
where n
C is a constant that depends on the dye concentration, the n-photon fluorescence
cross-section, and the collection efficiency of the system. 0()
n
I
t and ()
n
Sr are, respectively,
the temporal and spatial intensity distributions of the excitation beam. Because the Bessel
beam was transformed from a Gaussian beam by an axicon, ()Sr is given by [13]
()
12
,()()Srz S zS r=(2)
()
22 22
12
0
4πNA 2NA
()
zz
S z exp w
λ
=(3)
()
2
20
2NA
()
r
Sr J
π
λ
=(4)
where NA is the effective numerical aperture, λ is the wavelength of the laser, 0
w is the
initial Gaussian beam width, and z, r and θ are the cylindrical coordinates. Using the
equations above, the calculated point-spread functions (PSFs) in the x-z plane are shown in
Fig. 2(a) (3PM) and Fig. 2(b) (2PM). The x-y intensity distributions after integrating the z-
axis are shown in Fig. 2(e) (3PM) and Fig. 2(g) (2PM).
The total fluorescence of the m-th lobe for n-photon excitation can be calculated as
()
0
0
1
22NA
n0 1 2
02NA
C()() ()
m
m
x
nn n
nm x
Ft I t S dz d rzSrdr
λ
π
λ
θ
π
π
= (5)
where 0
m
x
is the m-th node (m > 0) of the zeroth order Bessel function, and 0
0
is 0. From
Eqs. (4) and (5), and with '2NA
r r
π
λ
=, we obtain the ratio of fluorescence in the m-th lobe
to fluorescence in the main lobe:
()
()
0
0
1
0
1
'2n
0
'2n
10
0
(') '
.
(') '
m
m
x
x
nm
x
n
rJ r dr
Ft
Ft rJ r dr
=
(6)
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1996
Equation (2) shows that the transverse distribution of the fluorescence generated by the
Bessel beam is invariant along the z-axis, and Eq. (6) shows the relative fluorescence in the
side lobes does not depend on the excitation NA, the length of the Bessel beam, or the pulse
width. We define the fluorescence in the main lobe as the signal, and we define the rest,
including the fluorescence in the side lobes and in regions far away from the focal area, as the
background.
As shown in Fig. 2(e) and Fig. 2(g), the three-photon fluorescence in the 2nd lobe (i.e.,
m = 2) is approximately 5.3% that in the main lobe, while the two-photon fluorescence in the
2nd, 3rd and 4th lobes is approximately 27.2%, 15.2% and 10.6% that in the main lobe,
respectively. The fluorescence generated by the side lobes of two-photon Bessel-beam
significantly degrades the SBR of the image.
Fig. 2. Theoretically simulated and experimentally determined PSFs. (a)(b), simulated PSFs of
3PM and 2PM along the x-z planes. (c)(d), the experimental results. Scale bars, 100 μm (z), 2
μm (x). 0.5-μm fluorescent beads were used to measure xyz profiles of Bessel-3PM and
Bessel-2PM. (e)(g), x-y cross section after integration of the z-axis of the Bessel beam to
numerically simulate the fluorescence distribution of the side lobes (e, 3PM; g, 2PM), (f)(h),
the experimental results. Scale bar, 2 µm. (i-l), comparison of blood vessels in live zebrafish
imaged with Bessel-3PM and 2PM. (i), a representative example of the fish under Bessel 2PM
and 3PM, in which the boxed region was magnified in j and k. (l), fluorescence intensities
along the line across the same blood vessels.
We measured the lateral intensity distribution of fluorescent beads throughout the length
of the Bessel beam for 2PE and 3PE. A stack of x-y images of 0.5-µm yellow/green
(505/515) beads (Invitrogen) were recorded with a 2-µm depth increment using a constant
excitation power. Figure 2(c), 2(d) show the axial PSFs of the Bessel beam of 3PM and 2PM.
For focal length of L2 of 50 mm, the axial FWHMs of the Bessel beam were ~250 µm for
3PM and ~300 µm for 2PM. These values are in good agreement with theoretical predictions.
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1997
The intensity modulation along the z-axis was caused by the round tip of the axicon [20].
Figures 2(f) and 2(h) show the x-y cross sections after integration of the z-axis of the Bessel
beam using 3PE and 2PE. The lateral resolution (defined as the intensity FWHM) for 3PE and
2PE are approximately 0.7 µm and 0.6 µm, respectively, with an effective NA of 0.4. The
lateral resolution can be improved with higher NA (e.g., 0.6 to 0.8). After the subtraction of
the background, defined as the average pixel value from regions without fluorescent beads,
the total fluorescence from all planes was integrated to estimate the side lobes, as shown in
Fig. 2(f) and 2(h). The three-photon fluorescence of the 2nd lobe is 5.1% of the main lobe,
and the two-photon fluorescence of the 2nd, 3th and 4th lobes are 27.8%, 14.6% and 10.1%
of the main lobe, respectively, in good agreement with the calculated values. By integrating
the fluorescence intensity from the first six side lobes, the combined fluorescence intensity of
Bessel-3PM (8.7%) is less than one-eighth that of the Bessel-2PM (73.1%), indicating a
much-improved SBR for Bessel-3PM.
We compared the performance of Bessel-3PM and Bessel-2PM in imaging blood vessels
from a large volume (500 μm × 500 μm × 300 μm) in the same zebrafish (3 days post-
fertilization) in vivo. We used the Tg (kdrl:EGFP) transgenic fish, in which EGFP was
selectively expressed in vascular epithelial cells. The average power after the objective was
90 mW at 1300 nm for 3PM and 127 mW at 920 nm for 2PM. In agreement with the
theoretical predictions and the experimental results with fluorescent beads, images obtained
with Bessel-3PM were sharper than those obtained with Bessel-2PM [Fig. 2(i-l)].
Fig. 3. Projection of volumetric imaging of GCaMP-labeled neuronal Ca2+ transients in
olfactory centers of Drosophila in vivo. We stimulated the fly with isoamyl acetate (IA) for 5 s
to trigger neuronal activities at the LH and the AL (DM2 glomerulus) of ePNs. (a), 3D
neuronal structures under the point-scanning 3PM with color-coded by depth. (b), projection of
3D volume of neuronal structure under the Bessel-3PM. (c)(d), pseudo-color snapshots of the
Ca2+ signals upon mineral oil and IA. (e), time-dependent Ca2+ responses from regions at the
LH and the DM2 were recorded at a scanning rate of 5 Hz in experiments (c) and (d). Blue bar,
stimulation duration. Individual traces are averaged results of 2-4 repeats and are processed
with 2 × binning. Error bars, ± s.e.m. (f-j), similar experiments as (a)-(e), except that Kenyon
cells at the MB were imaged. Axial FWHM of the Bessel beam, 110 μm. NA, 0.6.
Given the superior temporal resolution and comparative spatial resolution with point-
scanning 3PM, we used the Bessel-3PM to monitor the fast dynamics of different olfactory
centers of Drosophila along the axial direction in vivo. We genetically expressed GCaMP5
(GH146>GCaMP5 strain) [21] in the excitatory projection neurons (ePNs) at the antenna lobe
(AL), where they relay crucial olfactory information from peripheral olfactory receptor
neurons to higher brain centers, the mushroom body (MB) and the lateral horn (LH). The
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1998
application of odor isoamyl acetate (IA) could evoke calcium responses specifically in the
DM2 glomerulus of the AL, as reported by point-scanning 2PM [22]. Figure 3(a) shows the
3D neuronal structure with depth coded by color under point-scanning 3PM. We acquired a
120-μm-deep stack taken at 2-μm depth increments, and each stack was averaged from 3
frames. Figure 3(b) shows the projective neuronal structure under Bessel-3PM (averaged
from 10 frames). When imaging at the same frame rate (without averaging), point-scanning
3PM takes 60 times longer than Bessel-3PM. The small structures discernible in Fig. 3(a) and
3(b) (labeled by arrow) indicate the comparable spatial resolution for Bessel-3PM. Using
Bessel-3PM, we simultaneously detected IA-evoked responses from discrete regions of the
DM2 glomerulus and the LH at a frame rate of 5 Hz (256 × 256 pixels/frame). The
application of the control mineral oil evoked no response [Fig. 3(c)], indicating the specificity
of responses [Fig. 3(d)]. We also probed Ca2+ responses in Kenyon cells (KCs) at the MB (in
OK107>GCaMP6s strain) [23], which are the third-order olfactory neurons thought to be
critical for olfactory associative learning [24]. As shown in Fig. 3(f)-3(j), the amplitudes of
IA-evoked Ca2+ responses from the calyx and the vertical lobe of the MB, which are 40 μm
apart in axial distance, could be recorded simultaneously.
Fig. 4. In vivo three-photon images of neural cortex in a Thy1-YFP transgenic mouse. (a),
projection of 3D volume neurons and neurites in cortex of the awake mouse taken with Bessel-
beam at a frame rate of 1 Hz (image averaged from 10 frames afterwards). (b), mean intensity
projection of a 65-μm-thick image stack collected at 1-μm z steps in point-scanning mode,
from the same region as in (a). The stack covers from 620 μm to 685 μm below the dura, with
structures color-coded by depth. The image of each layer was averaged from 3 frames, with a
post objective power of 10 mW for Gaussian beam scanning and 110 mW for Bessel beam
scanning.
Furthermore, we benchmarked the performance of Bessel-3PM to point-scanning 3PM in
imaging 3D neuronal structures deeply buried in mice. We altered the focal length of L2 to
100 mm, which resulted in a 60 μm focal-depth Bessel beam for 3PM and an NA of 0.8. We
used Bessel-beam scanning 3PM to image the brain of an awake mouse from 620 μm to 685
μm below the dura. The imaging speed was 1 Hz (512 × 512 pixels/frame) for a volume of
300 μm × 300 μm × 65 μm. Imaging the same volume with point-scanning 3PM required 1
min, which might contribute to the skewed structures due to motion artifacts (Fig. 4). These
data highlights the importance and unique advantage of conducting rapid volumetric imaging
in deep brain in vivo with scanning Bessel-3PM.
4. Summary
In this work, we presented the results of using a rapid volumetric 3PM method based on the
axially elongated Bessel beam in imaging zebrafish larva, fruit flies and mice in vivo. The
proposed method enables volume scanning as large as 500 µm × 500 µm × 300 µm at 5 Hz
rate. Bessel-3PM has superior performance in terms of the penetration depth, SBR, and
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 1999
imaging speed and potentially provides a new multi-photon imaging tool to study the
information processing within three-dimensional neural circuits in vivo.
Funding
The authors gratefully acknowledge funding from the National Natural Science Foundation of
China (31327901, 61475008, 31428004, 31570839, and 31521062) and the National Science
and Technology Major Project Program (2016YFA0500400).
Acknowledgment
The authors thank Hao Rong and Wan Yang for their help with the mechanical design. The
Tg (kdrl:EGFP) fish was provided by Jingwei Xiong at Peking University.
Disclosures
The authors declare that there are no conflicts of interest related to this article.
Vol. 9, No. 4 | 1 Apr 2018 | BIOMEDICAL OPTICS EXPRESS 2000
... Bessel beam is a special diffraction-free beam satisfying the Helmholtz wave equation [1] and has a longer transmission distance than Gaussian-like beam [2]. Therefore, it is widely used in high-resolution imaging [3,4], laser processing [5][6][7], particle manipulation [8][9][10][11], optical communication [12][13][14] and other fields. Orthogonal electric field vector composed of linear and circular polarization can be utilized to construct arbitrary polarization distributions, such as radial and angular polarization, which satisfy the Helmholtz wave equation. ...
Article
According to the current research on polarization generation methods, it is difficult to create multi-mode vector Bessel beams in free space by extracting arbitrary polarization modes from a single beam. This is due to the fact that the polarization and phase distributions between multiple polarization modes can interfere with each other. In this paper, the mode extraction principle is combined with the optical pen technique to extract arbitrary polarization modes from a single Bessel beam, and the number, position and phase of Bessel beams can be arbitrarily regulated to achieve the multi-mode coexistence of vector Bessel beams in free space. The experimental results are consistent with the theoretical analysis. This work is not only important for the in-depth study of vector Bessel beams, but also will facilitate the development of optical manipulation, optical communication, microscopic imaging and other applications.
... 1960 年激光器的发明,使得以激光为基础的各个科学技术领域得到迅速发展,进而催生出许多 新型的光子技术,并广泛应用于光学操纵、精密测量、激光通信、激光加工、显微成像等领域。随 着激光技术的快速发展,传统高斯激光模式的传输规律单一,在各应用领域逐渐遇到技术瓶颈。由 此,光场调控技术应运而生。通过调制激光的振幅、相位和偏振函数,人们开发了许多新型的激光 模式,统称为空间结构光场。最著名的例子便是光学涡旋和矢量光束。为满足对激光模式抗扰动、 抗衍射特性的需求,人们发现了一类新的空间结构光场,即无衍射光束(non-diffracting beams)。无 衍射光束的概念由 Durnin 于 1987 年提出 [1] 。尽管"无衍射"这个概念在提出时引起了争议 [2,3] ,但 该光束优秀的抗衍射特性很快引起了研究人员们的着重关注, "无衍射"的概念也逐渐为人们接受。 理想条件下无衍射光束可以在传输过程中不改变横向强度分布,且受遮挡时可以快速恢复。与传统 的高斯型激光模式相比,无衍射光束可以进行超长距离传输,且能量损耗很小。随着研究的深入, 人们发现了一系列具有不同空间结构的无衍射光束,如马丢光束、韦伯光束、艾里光束、贝塞尔光 束等。这些新型光束均展现出了无衍射光束的共同特性。 作为亥姆霍兹方程传输不变模式的精确解 [1,4] ,贝塞尔光束是最典型的一类无衍射光束,其主 瓣在长距离传输中不发生明显展宽,表现出无衍射特性。贝塞尔光束的另一个特性是自恢复,也被 称为自重建 [5] 。当贝塞尔光束被部分遮挡时,其在传输一定距离后会恢复初始形貌。基于这些优异 的特性,贝塞尔波束被引入光学 [6,7] 和声学 [8][9][10] 的各个应用研究领域中。例如,在光学显微成像中 [11][12][13][14][15][16] ,贝塞尔光束的自恢复特性可以使透过强散射介质后的光场快速重建,从而能抑制光束散射,提 高照明深度 [17] ;其无衍射特性确保了光束在生物样品内长距离传输的稳定性 [18] ,有益于提高成像景 深和信噪比 [19,20] 。在微粒操控中,可利用贝塞尔光束的自重建特性来同时捕获多个样品池中的粒子 [21][22][23][24] 。在声学领域,贝塞尔波束可用于超声成像系统 [25,26] ,其长距离传输和细束宽可在成像系统中 获得更高分辨率和衬比度的图像,其无衍射特性使其能够深入样品内部,并提供比高斯波束更高的 轴向分辨率。此外,贝塞尔光束在光通信 [27][28][29][30][31][32] 和材料加工 [33][34][35][36][37][38][39] ...
... 10 Physical slits 11,12 or scientific complementary metal-oxide-semiconductor (CMOS) shutters 13 can be used to filter out background noise introduced by sidelobes. Higher-order nonlinear effects (such as three-photon excitation), 14,15 structured light illumination, 16 deconvolution algorithms, 17 and a "Bessel droplet" 18,19 with periodic axial light and dark intensity distributions formed by interference of multiple coaxial Bessel beams can also effectively reduce sidelobe effects. Although several methods have been reported to suppress the sidelobes and enhance image contrast, they often involve complex optical setups or image postprocessing. ...
Article
Full-text available
Significance Two-photon fluorescence microscopy (TPFM) excited by Gaussian beams requires axial tomographic scanning for three-dimensional (3D) volumetric imaging, which is a time-consuming process, and the slow imaging speed hinders its application for in vivo brain imaging. The Bessel focus, characterized by an extended depth of focus and constant resolution, facilitates the projection of a 3D volume onto a two-dimensional image, which significantly enhances the speed of volumetric imaging. Aim We aimed to demonstrate the ability of a TPFM with a sidelobe-free Bessel beam to provide a promising tool for research in live biological specimens. Approach Comparative in vivo imaging was conducted in live mouse brains and transgenic zebrafish to evaluate the performance of TPFM and Bessel-beam-based TPFM. Additionally, an image-difference method utilizing zeroth-order and third-order Bessel beams was introduced to effectively suppress background interference introduced by sidelobes. Results In comparison with traditional TPFM, the Bessel-beams-based TPFM demonstrated a 30-fold increase in imaging throughput and speed. Furthermore, the effectiveness of the image-difference method was validated in live biological specimens, resulting in a substantial enhancement of image contrast. Importantly, our TPFM with a sidelobe-free Bessel beam exhibited robustness against axial displacements, a feature of considerable value for in vivo experiments. Conclusions We achieved rapid, high-contrast, and robust volumetric imaging of the vasculature in live mouse brains and transgenic zebrafish using our TPFM with a sidelobe-free Bessel beam.
... In response to the challenges in near-infrared 2PM, three-photon microscopy (3PM) using high pulse energy infrared radiation has been developed, to allow access to deeper tissue layers in various organs [29][30][31][32] . Technological improvements such as the design of 3PM with Bessel beam illumination 33,34 , multicolor 3PM 35 , adaptive-optics enhanced 3PM 36,37 , or miniaturized 3PM in freely moving mice 38 helped to perform imaging at unprecedented tissue depths in brain cortex and hippocampus. Moreover, 3PM allowed for structural and functional murine brain imaging through the intact skull (up to 100 µm thick) 32 , which requires wave-front distortion correction using customized adaptive optics 36 . ...
Preprint
Intravital multi-photon imaging of the bone marrow is crucial to the study of cellular dynamics, communication with the microenvironment and functions, however, imaging of deep tissue areas is challenging and minimally invasive methods for deep-marrow imaging in intact long bones are needed. We developed a high pulse energy 1650 nm laser prototype, which permits to surpass >100 μm thick cortical bone and to perform three-photon microscopy (3PM) in more than 400 μm depth in the marrow cavity of intact mouse tibia in vivo. Its unique 3 and 4 MHz laser repetition rates allowed us to analyze motility patterns of rare cells over large fields of view deep within the unperturbed marrow. In this way, we found a bi-modal migratory behavior of marrow plasma cells. Besides, the analysis of third harmonics generation (THG) in the tibia identified this signal to be a label-free indicator of the abundance of cellular organelles, in particular the endoplasmic reticulum, reflecting protein biosynthesis capacity. We found that only one third of the plasma cells in the tibia marrow of adult mice have a strong THG signal and, thus, a high protein synthesis capacity, while the other two thirds of plasma cells display a low THG signal. Finally, we identified an inverse link between migratory behavior and THG signal strength in marrow plasma cells. As in these cells, the protein biosynthesis capacity indicated by a strong THG signal is mainly associated with antibody secretion, we could relate motility to functional states of plasma cells in vivo. Our 3PM method retains the ability to connect cellular dynamics to protein biosynthesis capacity in various marrow cell types beyond plasma cells, as THG is a ubiquitous signal, opening new perspectives on understanding how tissue microenvironment impacts on cellular functions in the bone marrow.
... (l) projection of mouse brain slices based on Gaussian beam, color corresponding to different depths; (m) projection of mouse brain slices after subtraction of 0 -order and 1 -order Bessel beams [45,[47][48] ; (n)(o) the PSF experimental images of the Gaussian and Airy for a 1 µm fluorescent particle in the twophoton microscope on the xy (left) and xz (right) plane; (p) strength curve corresponding to Fig. 4(n), (o); (q) (r) mouse brain slice imaging, in which the left image corresponds to Gaussian beam z scan imaging, and the depth is color coded, and the right image is a single frame measurement image of Airy beam [49] 封面文章·特邀综述 第 60 卷第 20 期/2023 年 10 月/激光与光电子学进展 ...
... Wavelengthversatile BBs with three-dimensional tunable shapes have been experimentally generated in a wide range of the electromagnetic spectrum and have seen tremendous applications in modern science and engineering technology owing to their unique propagation properties. For instance, a long, needle-like structure with low-intensity side rings of a zero-order BB is used for the rapid volumetric imaging of biological tissues in the presence of twophoton absorption and three-photon absorption [10]. The signal to background noise in bio-imaging due to BBs' side lobes can be successfully suppressed through multi-photon absorption. ...
Article
Full-text available
The generation of Bessel beams (BBs) and their characterization in a wide range of the electromagnetic spectrum are well established. The unique properties of BBs, including their non-diffracting and self-healing nature, make them efficient for use in material science and engineering technology. Here, I investigate the polarization components (s-polarization, p-polarization, transverse polarization, and longitudinal polarization) created in scalar BBs owing to their conical wave front. For emphasis, I provide a theoretical analysis to characterize potential experimental artifacts created in the four polarization components. Further, I provide a brief discussion on how to prevent these artifacts in scalar BBs. To my knowledge, for the first time, I can generate vector BBs in s-polarization and p-polarization via the superposition of two orthogonally polarized scalar BBs. This method of generation can provide the four well-known types of vector modes categorized in the V-point phase singularity vector modes. I suggest a suitable experimental configuration for realizing my theoretical results experimentally. The present analysis is very practical and beneficial for young researchers who seek to utilize BBs in light applications of modern science and technology.
Article
Full-text available
Over the past 36 years much research has been carried out on Bessel beams (BBs) owing to their peculiar properties, viz non-diffraction behavior, self-healing nature, possession of well-defined orbital angular momentum with helical wave-front, and realization of smallest central lobe. Here, we provide a detailed review on BBs from their inception to recent developments. We outline the fundamental concepts involved in the origin of the BB. The theoretical foundation of these beams was described and then their experimental realization through different techniques was explored. We provide an elaborate discussion on the different kinds of structured modes produced by the BB. The advantages and challenges that come with the generation and applications of the BB are discussed with examples. This review provides reference material for readers who wish to work with non-diffracting modes and promotes the application of such modes in interdisciplinary research areas.
Article
Full view observation throughout entire specimens over a prolonged period is crucial when exploring the physiological functions and system-level behaviors. Multi-photon microscopy (MPM) has been widely employed for such purposes owing to its deep penetration ability. However, the current MPM struggles with balancing the imaging depth and quality while avoiding photodamage for the exponential increasement of excitation power with the imaging depth. Here, we present a dual-objective two-photon microscope (Duo-2P), characterized by bidirectional two-photon excitation and fluorescence collection, for long-duration volumetric imaging of dense scattering samples. Duo-2P effectively doubles the imaging depth, reduces the total excitation energy by an order of magnitude for samples with a thickness five times the scattering length, and enhances the signal-to-noise ratio up to 1.4 times. Leveraging these advantages, we acquired volumetric images of a 380-μm suprachiasmatic nucleus slice for continuous 4-h recording at a rate of 1.67 s/volume, visualized the calcium activities over 4000 neurons, and uncovered their state-switching behavior. We conclude that Duo-2P provides an elegant and powerful means to overcome the fundamental depth limit while mitigating photodamages for deep tissue volumetric imaging.
Article
Full-text available
We demonstrate a robust high power 930 nm femtosecond Nd:fiber laser system with hollow-core photonic bandgap fiber (HC-PBGF) as the output delivery, which can be easily integrated into compact two-photon microscopy system for bio-imaging. The whole laser system can deliver up to 17.4 nJ, 220-fs pulses at 930 nm with repetition rate of 46 MHz. In this paper, this laser was demonstrated as the light source for volumetric imaging of zebrafish blood vessel.
Article
Full-text available
Neurons and neural networks often extend hundreds of micrometers in three dimensions. Capturing the calcium transients associated with their activity requires volume imaging methods with subsecond temporal resolution. Such speed is a challenge for conventional two-photon laser-scanning microscopy, because it depends on serial focal scanning in 3D and indicators with limited brightness. Here we present an optical module that is easily integrated into standard two-photon laser-scanning microscopes to generate an axially elongated Bessel focus, which when scanned in 2D turns frame rate into volume rate. We demonstrated the power of this approach in enabling discoveries for neurobiology by imaging the calcium dynamics of volumes of neurons and synapses in fruit flies, zebrafish larvae, mice and ferrets in vivo. Calcium signals in objects as small as dendritic spines could be resolved at video rates, provided that the samples were sparsely labeled to limit overlap in their axially projected images.
Article
Full-text available
Pre-chirp technique was used in an Nd-doped fiber amplifier to optimize high-quality 910 nm pulses with the pulses width of 114 fs and pulse energy of 4.4 nJ. The in vivo zebrafish imaging results from our totally home-made microscopy proves our femtosecond Nd fiber laser an ideal source in two-photon microscopic imaging.
Article
Full-text available
We report quantitative measurements of two-, three-, and four-photon excitation action cross sections of several commonly used fluorophores and fluorescent proteins at three different excitation wavelengths of 800 nm, 1300 nm, and 1680 nm. The measured cross section values are consistent with simple quantum mechanic estimations. These values indicate that the optimum repetition rate for deep tissue 3-photon microscopy is approximately 1 to 2 MHz. We further demonstrate that it is feasible to perform 4-photon fluorescence microscopy of GFP labeled microglia in mouse brain in vivo at 1700 nm. 4-photon excitation increases the accessibility of fluorophores at the long wavelength spectral window of 1700 nm.
Article
Full-text available
Two-photon fluorescence microscopy (2PM)(1) enables scientists in various fields including neuroscience(2,3), embryology(4), and oncology(5) to visualize in vivo and ex vivo tissue morphology and physiology at a cellular level deep within scattering tissue. However, tissue scattering limits the maximum imaging depth of 2PM within the mouse brain to the cortical layer, and imaging subcortical structures currently requires the removal of overlying brain tissue(3) or the insertion of optical probes(6,7). Here we demonstrate non-invasive, high resolution, in vivo imaging of subcortical structures within an intact mouse brain using three-photon fluorescence microscopy (3PM) at a spectral excitation window of 1,700 nm. Vascular structures as well as red fluorescent protein (RFP)-labeled neurons within the mouse hippocampus are imaged. The combination of the long excitation wavelength and the higher order nonlinear excitation overcomes the limitations of 2PM, enabling biological investigations to take place at greater depth within tissue.
Article
High-resolution optical imaging is critical to understanding brain function. We demonstrate that three-photon microscopy at 1,300-nm excitation enables functional imaging of GCaMP6s-labeled neurons beyond the depth limit of two-photon microscopy. We record spontaneous activity from up to 150 neurons in the hippocampal stratum pyramidale at ∼1-mm depth within an intact mouse brain. Our method creates opportunities for noninvasive recording of neuronal activity with high spatial and temporal resolution deep within scattering brain tissues.
Article
Molecular excitation by two or more infra red photons simultaneously absorbed from strongly focused 100 femtosecond mode locked laser pulses provides intrinsic 3-d resolution for fluorescence microscopy and photochemical micro pharmacology in living biological cells.1 Confinement of non-linear excitation to the focal volume illuminated by ~10 photons/nm s eliminates out of focus photobleaching and photodamage and provides intrinsic 3-d resolution. To facilitate this technology new methods have developed for accurate measurements of multi-photon excitation cross sections and spectra in the absence of adequate data to guide non-linear laser microscopy applications.2,3 Absolute cross section measurements have been facilitated by development of a method that takes advantage of the interference of excitation pulse trains that are time shifted by a conventional Michaelson interferometer to allow variable relative delay of two spatially superimposed beams. Two photon excitation with single mode CW lasers provides an independent absolute calibration of cross sections. Measurements of more than twenty useful two-photon excitation spectra (690–1050nm) have revealed some with large blue shifts of the excitation peak. Two photon excitation spectra of some asymmetric molecules superimpose on the one photon absorption spectra (with wavelength doubled) presumably due to relaxation of parity selection rules in these cases. Most of the tested fluorescent molecules show blue shifted excitation peaks with strongly enhanced cross sections exceeding those observed at twice the one photon absorption. No red shifts of the excitation spectra have been detected.
Article
Recording the activity of large populations of neurons is an important step toward understanding the emergent function of neural circuits. Here we present a simple holographic method to simultaneously perform two-photon calcium imaging of neuronal populations across multiple areas and layers of mouse cortex in vivo. We use prior knowledge of neuronal locations, activity sparsity, and a constrained nonnegative matrix factorization algorithm to extract signals from neurons imaged simultaneously and located in different focal planes or fields of view. Our laser multiplexing approach is simple and fast, and could be used as a general method to image the activity of neural circuits in three dimensions across multiple areas in the brain. Yang et al. demonstrate a novel approach for simultaneously imaging multiple layers of the mouse cortex. They combine holographic two-photon microscopy with advanced computational source extraction to create a flexible platform for studying mesoscale neural circuits at multiple depths of the brain with cellular resolution.
Article
We characterize an inhibitory circuit motif in the Drosophila olfactory system, parallel inhibition, which differs from feedforward or feedback inhibition. Excitatory and GABAergic inhibitory projection neurons (ePNs and iPNs) each receive input from antennal lobe glomeruli and send parallel output to the lateral horn, a higher center implicated in regulating innate olfactory behavior. Ca(2+) imaging of specific lateral horn neurons as an olfactory readout revealed that iPNs selectively suppressed food-related odor responses, but spared signal transmission from pheromone channels. Coapplying food odorant did not affect pheromone signal transmission, suggesting that the differential effects likely result from connection specificity of iPNs, rather than a generalized inhibitory tone. Ca(2+) responses in the ePN axon terminals show no detectable suppression by iPNs, arguing against presynaptic inhibition as a primary mechanism. The parallel inhibition motif may provide specificity in inhibition to funnel specific olfactory information, such as food and pheromone, into distinct downstream circuits.