Particle Distribution Dependent Inaccuracy of the Plantower PMS5003 low-
cost PM-sensor
Bernd Laquai, 22.10.2017
During a single measurement run, the Plantower PMS5003 and the Grimm were exposed to a strong
change in particle mass distribution versus size. In the first phase the particle spectrum only contained
particles sizes < 3um in the second phase the particle spectrum contained only particle sizes > 3um.
During the measurement, the Grimm 1.108 reference instrument reported the expected PM values:
For the first phase PM10 = PM2.5, for the second phase PM10 >> PM2.5.
The Plantower PMS5003 reported PM10>PM2.5 for the first phase and PM10=0 for the second phase.
For large PM2.5 mass concentrations (> 400ug/m^3) the PMS5003 reported slightly less (factor 0.5)
than the Grimm, for smaller PM2.5 mass concentration it reported slightly more (factor 1.5).
From these measurements, we conclude that the Plantower PMS5003 is able to measure PM2.5 more
or less correct. With a suitable calibration, it may be calibrated to a reference instrument such as the
Grimm. The PMS5003 however is not able at all to directly measure large particles > 3um. When it
outputs PM10 values, these are pure estimates based on particle sizes it can measure (d < 3um) and
under the assumption of a broadly distributed particle spectrum. When the particle spectrum is not
containing PM mass contributions at sizes < 3um, the PMS5003 erroneously outputs PM10 values close
to zero. It therefore behaves similar to the SDS011 low-cost PM-sensor from Nova Fitness.
These findings intensify the strong suspicion that low-cost PM-sensors, mainly produced by Chinese
manufacturers and designated as PM2.5 sensors but also reporting PM10 values, aren’t able to
measure particles larger than 3um and therefore solely extrapolate on PM10 values assuming a certain
particle distribution.
See also: Bernd Laquai, Impact of Particle Mass Distribution on the Measurement Accuracy of Low-
Cost PM-Sensors, 20.10.2017
2
Fig. 1a: The applied particle distribution over time as seen by the Grimm (lin. scale)
Fig. 1b: The applied particle distribution over time as seen by the Grimm (log. scale)
3
Fig 2a: A typical mass distribution plot of the Grimm SW during the first phase of the measurement
(dM_dlog(dp)-2017-10-22-07h23.png)
Fig 2b: A typical mass distribution plot of the Grimm SW during the second phase of the measurement
(dM_dlog(dp)-2017-10-22-08h32.png)
4
Fig. 3: Contour plot (log. scale) of the change between two different particle distributions during the
measurement versus size and time
Fig.4: PM10 and PM2.5 readings of the PMS5003, no indication for the large particles in phase 2
0
100
200
300
400
500
600
700
800
900
1000
00:00:00 00:28:48 00:57:36 01:26:24 01:55:12 02:24:00 02:52:48
ug/m^3
Zeit
PMS5003
PM10 PM2.5
5
Fig 5a: Comparison for PM10 values reported by the Grimm and the PMS5003
Fig 5b: Comparison for PM2.5 values reported by the Grimm and the PMS5003
0
100
200
300
400
500
600
700
800
900
1000
00:00:00 00:28:48 00:57:36 01:26:24 01:55:12 02:24:00 02:52:48
ug/m^3
Zeit
PM10 Vergleich
Grimm PMS5003
0
100
200
300
400
500
600
700
800
900
1000
00:00:00 00:28:48 00:57:36 01:26:24 01:55:12 02:24:00 02:52:48
ug/m^3
Zeit
PM2.5 Vergleich
Grimm PMS5003
6
Fig. 6: The ratio PMS5003/Grimm for PM2.5 during the measurement Fig 5a: Comparison for PM10
values reported by the Grimm and the PMS5003
Fig. 7: Plantower PMS5003 device connected to an Arduino based datalogger
0,00
0,50
1,00
1,50
2,00
2,50
3,00
00:00:00 00:14:24 00:28:48 00:43:12 00:57:36 01:12:00 01:26:24 01:40:48 01:55:12
PM2.5 PMS5003 vs. Grimm
PMS5003/Grimm
7
Fig. 8: The PMS5003 with attached product sticker designating it as PM2.5 sensor
Fig. 9: The measurement chamber of the particle generator with Grimm and PMS5003
... "Low-cost" refers to the purchase price of LCS [9] compared to the purchase and operating cost of reference analyzers [46] for the monitoring of regulated inorganic pollutants and PM, which can easily be an order of magnitude more costly. More recently, ultra-affordable OEMs have started to appear on the market for PM monitoring [47][48][49]. Many of them are designed to be integrated into Internet of Things (IoT) networks of interconnected devices. ...
... Those devices are manufactured in emerging economies, such as the Republic of China and the Republic of Korea [50]. Some of these LCS can achieve similar performance to more expensive OEMs [18,19,[29][30][31]37,48,49,[51][52][53][54][55][56]. ...
... A few studies also report the RMSE [10,20,22,36,[41][42][43]51,52,58,60,62,63,85] which clearly indicates that the magnitude of the error in LCS data is also sensitive to extreme values and outliers. Only a few studies report the measurement uncertainty [10,22,25,30,48,52,59,61]. Therefore, for the purpose of this work, we only focused on the analysis of the comparison of laboratory and field tests of LCS. ...
Article
Full-text available
A growing number of companies have started commercializing low-cost sensors (LCS) that are said to be able to monitor air pollution in outdoor air. The benefit of the use of LCS is the increased spatial coverage when monitoring air quality in cities and remote locations. Today, there are hundreds of LCS commercially available on the market with costs ranging from several hundred to several thousand euro. At the same time, the scientific literature currently reports independent evaluation of the performance of LCS against reference measurements for about 110 LCS. These studies report that LCS are unstable and often affected by atmospheric conditions—cross-sensitivities from interfering compounds that may change LCS performance depending on site location. In this work, quantitative data regarding the performance of LCS against reference measurement are presented. This information was gathered from published reports and relevant testing laboratories. Other information was drawn from peer-reviewed journals that tested different types of LCS in research studies. Relevant metrics about the comparison of LCS systems against reference systems highlighted the most cost-effective LCS that could be used to monitor air quality pollutants with a good level of agreement represented by a coefficient of determination R2 > 0.75 and slope close to 1.0. This review highlights the possibility to have versatile LCS able to operate with multiple pollutants and preferably with transparent LCS data treatment.
... Therefore, the GP2YX was selected for use in the low-cost multimodal device, but initial testing with the GP2YX produced erratic and highly inaccurate data. Alternatively, PlanTower sensors are often used in many commercial devices [48] and have been found to report data that correlates with reference equipment when measuring PM2.5 [49]- [51]. Therefore, a (PlanTower) PMSA003 (Table 1Table 1) sensor was selected here to measure PM2.5. ...
Article
Monitoring Indoor Environmental Quality (IEQ) is of growing interest for health and wellbeing. New building standards, climate targets and adoption of homeworking strategies are creating needs for scalable, monitoring solutions with onward Cloud connectivity. Low-cost Micro-Electromechanical Systems (MEMS) sensors have potential to address these needs, enabling development of bespoke multimodal devices. Here, we present insights into the development of a MEMS-based Internet of things (IoT) enabled multimodal device for IEQ monitoring. A study was conducted to establish the inter-device variability and validity to reference standard sensors/devices. For the multimodal, IEQ monitor, intraclass correlations and Bland-Altman analyses indicated good inter-sensor reliability and good-to-excellent agreement for most sensors. All low-cost sensors were found to respond to environmental changes. Many sensors reported low accuracy but high precision meaning they could be calibrated against reference sensors to increase accuracy. The multimodal device developed here was identified as being fit-for-purpose, providing general indicators of environmental changes for continuous IEQ monitoring.
Conference Paper
Air pollution has caused a vicious impact on the climate and human health. As a consequence, it has drawn the attention of researchers, policymakers, and citizens. Traditionally, air pollution is measured using heavy and expensive instruments at fixed monitoring stations. But the number of these monitoring stations is very few. Also, the scale of fire accidents due to flammable gas leakage is increasing rapidly. To control air pollution and to reduce fire accidents, a vastly installed device is a dire need. In the following research, we have proposed a low-cost sensor-based monitoring system with multiple network options, along with some preliminary data analysis results. Our system is capable of sharing quantitative data of air pollution globally using Global System for Mobile Communications (GSM), Wi-Fi, and Bluetooth. It also has a smart alarm system to alert authorized personals. The foreground of this research is implementing a feasible system that can benefit the 3rd world countries as well as the developed ones.KeywordsAir pollutionMobile sensingGas sensorsWireless communicationArduinoIoTSecurity System
Article
Full-text available
Low-cost particulate matter (PM) sensors have been under investigation as it has been hypothesized that the use of low-cost and easy-to-use sensors could allow cost-efficient extension of the currently sparse measurement coverage. While the majority of the existing literature highlights that low-cost sensors can indeed be a valuable addition to the list of commonly used measurement tools, it often reiterates that the risk of sensor misuse is still high and that the data obtained from the sensors are only representative of the specific site and its ambient conditions. This implies that there are underlying reasons for inaccuracies in sensor measurements that have yet to be characterized. The objective of this study is to investigate the particle-size selectivity of low-cost sensors. Evaluated sensors were Plantower PMS5003, Nova SDS011, Sensirion SPS30, Sharp GP2Y1010AU0F, Shinyei PPD42NS, and Omron B5W-LD0101. The investigation of size selectivity was carried out in the laboratory using a novel reference aerosol generation system capable of steadily producing monodisperse particles of different sizes (from ∼0.55 to 8.4 µm) on-line. The results of the study show that none of the low-cost sensors adhered to the detection ranges declared by the manufacturers; moreover, cursory comparison to a mid-cost aerosol size spectrometer (Grimm 1.108, 2020) indicates that the sensors can only achieve independent responses for one or two size bins, whereas the spectrometer can sufficiently characterize particles with 15 different size bins. These observations provide insight into and evidence of the notion that particle-size selectivity has an essential role in the analysis of the sources of errors in sensors.
Article
Full-text available
MoreAir is a low-cost and agile urban air pollution monitoring system. This paper describes the methodology used in the development of this system along with some preliminary data analysis results. A key feature of MoreAir is its innovative sensor deployment strategy which is based on mobile and nomadic sensors as well as on medical data collected at a children’s hospital, used to identify urban areas of high prevalence of respiratory diseases. Another key feature is the use of machine learning to perform prediction. In this paper, Moroccan cities are taken as case studies. Using the agile deployment strategy of MoreAir, it is shown that in many Moroccan neighborhoods, road traffic has a smaller impact on the concentrations of particulate matters (PM) than other sources, such as public baths, public ovens, open-air street food vendors and thrift shops. A geographical information system has been developed to provide real-time information to the citizens about the air quality in different neighborhoods and thus raise awareness about urban pollution.
Preprint
Full-text available
Abstract. Low-cost particulate matter sensors (PM) have been under investigation due to their prospective nature regarding spatial extension of measurement coverage. While majority of the existing literature highlights that low-cost sensors can be useful in achieving this goal, it is often reminded that the risk of sensor misuse is still high, and that the data obtained from the sensors is only representative of the specific site and its ambient conditions. This implies that there are underlying reasons yet to be characterized which are causing inaccuracies in sensor measurements. The objective of this study was to investigate the particle size-selectivity of low-cost sensors. Evaluated sensors were Plantower PMS5003, Nova SDS011, Sensirion SPS30, Sharp GP2Y1010AU0F, Shinyei PPD42NS, and Omron B5W-ld0101. The investigation of size-selectivity was carried out in laboratory using a novel reference aerosol generation system capable of steadily producing monodisperse particles of different sizes on-line. The results of the study showed that none of the low-cost sensors adhered exactly to the detection ranges declared by the manufacturers, and moreover, cursory comparison to a mid-cost aerosol spectrometer (GRIMM 1.108) indicated that the sensors could only achieve independent responses for 1–2 size bins whereas the spectrometer could sufficiently characterize particles with 15 different size bins. These observations provide insight and evidence to the notion that particle size-selectivity may have an essential role in the error source analysis of sensors.
ResearchGate has not been able to resolve any references for this publication.