ArticlePDF Available

Sensitive CE-ECL method by AuNPs enhanced signal for detection of β-blockers and study on interaction between drug and protein

Royal Society of Chemistry
RSC Advances
Authors:

Abstract and Figures

A sensitive capillary electrophoresis (CE) system coupled with electrochemiluminescence (ECL) of tris(2,2′-bipyridyl) ruthenium(II) is described for the detection of propranolol (Pro) and acebutolol (Ace). AuNPs were added to the end column luminescent reagents solution to enhance the sensitivity of the CE-ECL system. Experimental conditions of separation and detection as well as the amount of AuNPs were optimized. Under the optimum conditions, the linear range of Pro is 0.01 to 100 μmol L−1 with the LOD of 3.6 × 10−9 mol L−1 (S/N = 3) and with the LOQ of 1.1 × 10−7 mol L−1 in human urine samples (S/N = 10). For Ace, the linear range is 0.02 to 100 μmol L−1 with the LOD of 5.0 × 10−9 mol L−1 (S/N = 3) and with the LOQ of 9.5 × 10−8 mol L−1 in human urine samples (S/N = 10). RSDs (n = 3) of the migration time for Pro and Ace are from 1.9 to 2.4% intraday and from 2.9 to 3.8% interday. RSDs of the peak area for Pro and Ace are from 3.2 to 3.7% intraday and from 4.3 to 4.6% interday, respectively. The developed method was successfully used to determine two analytes in human urine samples. In addition, the interaction between Pro as a model analyte and human serum albumin (HSA) was investigated; the number of binding site and the binding constant of Pro with HSA are 1.0 and 2.3 × 104 L mol−1, respectively.
Content may be subject to copyright.
Sensitive CE-ECL method with AuNPs-enhanced
signal for the detection of b-blockers and the study
of drugprotein interactions
Hong-Bing Duan,
ab
Jun-Tao Cao,*
ab
Hui Wang
ab
and Yan-Ming Liu*
ab
A sensitive capillary electrophoresis (CE) system coupled with electrochemiluminescence (ECL) of tris(2,20-
bipyridyl) ruthenium(II) is described for the detection of propranolol (Pro) and acebutolol (Ace). AuNPs were
added to the end column luminescent reagents solution to enhance the sensitivity of the CE-ECL system.
Experimental conditions of separation and detection as well as the amount of AuNPs were optimized. Under
the optimum conditions, the linear range of Pro is 0.01 to 100 mmol L
1
with the LOD of 3.6 10
9
mol L
1
(S/N ¼3) and with the LOQ of 1.1 10
7
mol L
1
in human urine samples (S/N ¼10). For Ace, the linear
range is 0.02 to 100 mmol L
1
with the LOD of 5.0 10
9
mol L
1
(S/N ¼3) and with the LOQ of 9.5
10
8
mol L
1
in human urine samples (S/N ¼10). RSDs (n¼3) of the migration time for Pro and Ace are
from 1.9 to 2.4% intraday and from 2.9 to 3.8% interday. RSDs of the peak area for Pro and Ace are from
3.2 to 3.7% intraday and from 4.3 to 4.6% interday, respectively. The developed method was successfully
used to determine two analytes in human urine samples. In addition, the interaction between Pro as
a model analyte and human serum albumin (HSA) was investigated; the number of binding site and the
binding constant of Pro with HSA are 1.0 and 2.3 10
4
L mol
1
, respectively.
Introduction
Propranolol (Pro) and acebutolol (Ace) are nonselective b-
adrenergic blocking drugs (b-blockers) which are commonly
used to treat several diseases, such as cardiacarrhythmia,
hypertension, angina pectoris, sinus tachycardia, anxiety
disorders, thyrotoxicosis, hypertrophic subaortic stenosis,
myocardial infarction, and migraine in the clinic.
1,2
These drugs
are sometimes used with other drugs in poly-pharmacy. They
are also misused in some sports to reduce cardiac frequency,
contraction force and coronary ow.
3
The International Olympic
Committee has identied Pro as a doping substance and
included it in the list of forbidden substances.
4
In addition, an
overdose of b-blockers may lead to bradycardia, hypotension,
aggravation of cardiac failure, bronchospasm, hypoglycemia,
and fatigue.
5
The increasing demand for b-blocker detection
have driven tremendous eorts in the development of sensitive
and selective methods for the determination of b-blockers in
pharmaceutical formulations.
6,7
Therefore, sensitive and accu-
rate analytical methods for the quantication of these
compounds in biological uids are required. Several methods,
such as solid-phase extraction coupled to gas chromatography-
mass spectrometry (SPE-GC-MS),
8
GC-MS,
9
ow injection-
chemiluminescence (FI-CL),
10
high performance liquid
chromatography-photodiode-array UV (HPLC-UV)
11
and solid-
phase extraction-dierential pulse voltammetry (SPE-DPV)
12
have been reported for the determination of b-blockers.
However, these methods require the use of expensive instru-
mentation
9
and lack sucient sensitivity;
8,1012
also, some of
them involve long analysis times
8,9,11
and tedious preliminary
procedures such as pre-concentration in an organic solvent,
ltration, extraction and derivatization.
12
Therefore, the devel-
opment of sensitive and selective analytical methods to detect b-
blockers in biological uids is necessary.
Human serum albumin (HSA), the most abundant protein in
blood plasma, is a main carrier protein with high anity to
bind a wide range of endogenous and exogenous compounds,
such as fatty acids, metals, toxic metabolites, hormones, amino
acids, bile acids, and drugs.
13,14
HSA has long been a focus of the
pharmaceutical industry due to its ability to bind various drug
molecules. The interaction of HSA with drugs plays an essential
role in biological activity in the human body. Among its phar-
macokinetic functions, drug distribution is controlled by HSA
because most drugs transported in blood bind to serum
albumin and thereby reach the target tissues.
15
Studying the
interaction of drugs with HSA is of considerable importance to
understand the processes of drug transportation and predict
the concentration of free drug.
Capillary electrophoresis (CE) has attracted many
researchers' attention because of its advantageous properties of
a
College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang
464000, China. E-mail: liuym9518@sina.com; jtcao11@163.com; Fax: +86-376-
6392889; Tel: +86-376-6392889
b
Key Laboratory of Simulation and Control for Dabie Mountains Population Ecology,
Xinyang Normal University, Xinyang 464000, China
Cite this: RSC Adv.,2016,6, 45533
Received 16th March 2016
Accepted 25th April 2016
DOI: 10.1039/c6ra07003e
www.rsc.org/advances
This journal is © The Royal Society of Chemistry 2016 RSC Adv.,2016,6, 4553345539 | 45533
RSC Advances
PAPER
high separation eciency, short analysis time, low reagent
consumption, and ease of installation, operation and mainte-
nance.
1618
Electrochemiluminescence (ECL) occurs when elec-
trochemically generated reactants undergo a high-energy
electron transfer reaction to generate an excited state. ECL is
well known as a powerful analytical technique with the advan-
tages of high sensitivity, wide linear range, and simple instru-
mentation. ECL has been widely used in the areas of
immunoassay, food and water testing and biowarfare agent
detection.
19
Ru(bpy)
32+
-based ECL has received considerable
attention among a number of ECL systems, owing to its low
background signal, excellent chemical stability and high ECL
eciency in aqueous and non-aqueous solvents.
20
Therefore,
CE coupled with Ru(bpy)
32+
-ECL detection has been applied to
separate and detect compounds such as alkaloids,
21
hormones,
22
biogenic amines,
23
protein,
24
drugs
25
and small
organic molecules.
26
Over the past decade, gold nanoparticles (AuNPs) have
attracted much attention in biomolecular detection and clinical
diagnostic applications because of their facile synthesis, good
biocompatibility, relatively large surface and strong catalytic
properties. It has been found that the larger surface area of
AuNPs and the catalytic ability of AuNPs will improve the
number of transmission states, which is an important param-
eter for the intensity of ECL emission.
27,28
Inspired by the useful applications of AuNPs, a sensitive CE-
Ru(bpy)
32+
-ECL method based on the enhancement of AuNPs
was developed for the separation and detection of Pro and Ace.
The factors inuencing CE and ECL, including the pH and
concentration of the running buer, separation voltage and
detection potential as well as the amount of AuNPs were
investigated systematically. The proposed method was
successfully applied to the analysis of two analytes in human
urine samples and the study of the interaction between the
drugs and the protein. The number of binding site and the
binding constant of the drug and protein were obtained.
Experimental
Apparatus
The MPI-A CE-ECL system was produced by Remax Electronic
Science-Tech Co. Ltd. (Xi'an, China). The detection system
consists of four main parts: a high-voltage power, a potentiostat,
a multifunction chemiluminescence detector and a data
collection analyzer. All data collection was performed with MPI-
A analysis soware and recorded on a computer. CE separation
was performed in a 55 cm length 50 mm i.d. and 375 mm o.d.
uncoated fused-silica capillary obtained from Yongnian Optical
Conductive Fiber Plant (Hebei, China). ECL detection was
employed using a three-electrode system: a 500 mm diameter
platinum disk as the working electrode, a Pt wire as the auxiliary
electrode, and a Ag/AgCl electrode (in saturated KCl solution) as
the reference electrode. The transmission electron microscopy
(TEM) images were recorded using a Tecnai G
2
F20 TEM system
(FEI Co., America). UV-visible detection was carried out using an
UVmini-1240 UV-vis spectrophotometer (Shimadzu Corp.,
Kyoto, Japan).
Chemicals
All the reagents used were commercially available and of
analytical grade. Standard Ace was achieved from J&K Chemical
Ltd. (Shanghai, China). Standard Pro was purchased from the
National Institute for the Control of Pharmaceutical and Bio-
logical Products (Beijing, China). Hydrogen tetra-
chloroaurate(III) trihydrate (AuCl
3
$HCl$4H
2
O, Au > 47.8%) was
obtained from Alfa Aesar (A Johnson Matthey Company, Ward
Hill, MA). Tris(2,20-bipyridyl)-ruthenium(II) chloride hexahy-
drate (Ru(bpy)
3
Cl
2
) and HSA (>96%, M
w
¼66 kDa) were
purchased from Sigma-Aldrich (St Louis, MO, USA). Tween 20
was obtained from Seebio Biotech (Shanghai, China). Ultrapure
fresh water (18.2 MUcm) used was processed with an ultrapure
water system (Kangning Water Treatment Solution Provider,
China). All solutions were ltered through 0.22 mm cellulose
acetate membrane lters (Shanghai Xingya Purication Mate-
rial Factory, China) and stored under refrigeration at 4 C before
use.
Procedure
Before rst use, the new capillary was lled with 2.0 mol L
1
CH
3
OHNaOH, 1.0 mol L
1
NaOH, 1.0 mol L
1
HCl, H
2
O, and
nally with phosphate buer solution (PBS) for 30 min,
respectively. At the beginning use of each day, the capillary was
rinsed with 0.1 mol L
1
NaOH, water, and the corresponding
separation buer for 10 min successively until the baseline of
chemiluminescence was at to maintain an active and repro-
ducible capillary inner surface prior to a series of analyses. The
photomultiplier tube (PMT CR105, Beijing Binsong Photonics,
China) positioned under the detection cell for collecting the
ECL signal was set at 850 V in the process of detection. The
detection potential applied to the working electrode was xed at
1.25 V. Electrokinetic injections were performed at 10 kV for 10
s. The inlet end of the capillary was held at a positive potential
and the outlet end was maintained at ground. 5 mmol L
1
Ru(bpy)
32+
with 50 mmol L
1
PBS was added into the detection
cell as the ECL reagent. Prior to the experiments, the Pt disc
working electrode was polished with 0.3 and 0.05 mmAl
2
O
3
in
sequence, then cleaned with ultrapure water in an ultrasonic
cleaner. Elimination of the oxide layer from the Pt electrode was
performed in 50 mmol L
1
pH 7.0 PBS by cycling the potentials
of the working electrode from 0.5 to 0.0 V. At a scan rate of 100
mV s
1
, more than 10 cycles were required.
Preparation of AuNPs
All the glassware in the following experiments were cleaned in
freshly prepared HNO
3
: HCl (1 : 3) solution, rinsed with water
and dried in air. AuNPs were prepared according to a literature
report.
29
Briey, 5 mL of 40 mmol L
1
sodium citrate was rapidly
added to 50 mL of boiling 0.05% AuCl
3
$HCl$4H
2
O solution
under vigorous stirring; the colour of the solution changed from
pale-yellow to wine-red. Boiling was continued for an additional
15 min, and then the heating was stopped. The solution was
stirred for another 15 min until it cooled to room temperature.
Aer cooling, the synthesized AuNPs were ltered through
45534 |RSC Adv.,2016,6, 4553345539 This journal is © The Royal Society of Chemistry 2016
RSC Advances Paper
a 0.22 mm cellulose membrane and stored in a dark glass bottle
at 4 C before use.
Preparation of human urine sample
The urine samples were provided by a healthy male volunteer in
Xinyang Normal University. The male volunteer gave written
informed consent, and complete ethical approval was obtained.
The study was performed in compliance with the relevant laws
and institutional guidelines and was also approved by the
Institutional Review Board of Xinyang Normal University. Prior
to analysis, the fresh urine samples were immediately centri-
fuged at 2000 rpm for 15 min to remove as much precipitation
as possible. Then, the urine sample was diluted 100-fold with
water and then ltered through 0.22 mm cellulose acetate lters.
Results and discussion
AuNPs enhanced ECL of Ru(bpy)
32+
in the presence of Pro and
Ace
To investigate the eect of AuNPs (Fig. 1) on the ECL intensity in
the detection of Pro and Ace, cyclic voltammetry (CV) and ECL
measurements were conducted. Fig. 2 shows the CV and cor-
responding ECL curves before and aer adding analytes and
analytes with AuNPs, respectively. As is clearly shown in Fig. 2,
the ECL intensities of the two analytes increased markedly in
the presence of AuNPs. We can conclude that Pro and Ace can
react with the ruthenium species in the ECL process, and the
AuNPs can enhance the emitted light intensity. The enhance-
ment of the ECL emission in Ru(bpy)
32+
system can be inter-
preted as follows. In the ECL detection, the interaction of
negatively charged AuNPs with the positively charged
Ru(bpy)
32+
may decrease the ligand pp*energy gap, which
determines the rate and eciency of the charge transfer tran-
sitions between a d-orbital on the ruthenium and a p*anti-
bonding orbital on the ligand.
30
This interaction makes the
transition of the excited species Ru(bpy)
32+
*into the ground
state easier and more ecient, resulting in an enhancement of
the ECL emission. In addition, the large surface area and the
good catalytic ability of the AuNPs will also improve the number
of transmission states of Ru(bpy)
32+
, which is another impor-
tant parameter aecting the ECL intensity.
To achieve highly sensitive detection of Pro and Ace, the
amount of AuNPs in the detection cell was optimized. First, 300
mL of 5 mmol L
1
Ru(bpy)
32+
with 50 mmol L
1
PBS was injected
into the detection cell, and the ECL curve was recorded. Then,
30 mL solutions were extracted from the detection cell, followed
by the addition of 30 mL of AuNPs solution. The solutions were
immediately mixed with an injector and the corresponding ECL
spectrum was recorded. These actions were repeated eight
times. Fig. 3 shows the relationship of ECL intensity with the
volume ratio of AuNPs and Ru(bpy)
32+
solution for (a) Pro and
(b) Ace. As can be seen, with increasing ratio, the ECL intensity
for those two analytes exhibits approximately the same
tendency, an initial increase followed by a decrease, and the
maximum ECL intensity was obtained at a volume ratio of 0.7.
This can be explained as follows: at the rst stage, an increase in
AuNPs can greatly enhance the quantity of transmission states
of Ru(bpy)
32+
*; thus, the ECL intensity increases. However,
Fig. 1 TEM image of the as-prepared citric-acid stabilized AuNPs.
Fig. 2 Cyclic voltammograms (inset gure) and their corresponding
ECL curves. (a) 5 mmol L
1
Ru(bpy)
32+
+ 50 mmol L
1
PBS (pH 8.0); (b)
a + 0.1 mmol L
1
Pro; (c) a + 0.1 mmol L
1
Ace; (d) b + AuNPs; (e) c +
AuNPs; scan rate, 100 mV s
1
.
Fig. 3 Variation of the intensity of 5 mmol L
1
Ru(bpy)
32+
based ECL
emission in the presence of 0.1 mmol L
1
Pro (a) and Ace (b) with the
volume ratio of the suspension of AuNPs to Ru(bpy)
32+
solution.
This journal is © The Royal Society of Chemistry 2016 RSC Adv.,2016,6, 4553345539 | 45535
Paper RSC Advances
continuously increasing the amounts of AuNPs dilutes the
concentration of Ru(bpy)
32+
, causing the ECL intensity to
decrease; thus, a volume ratio of 0.7 was selected.
Eects of CE separation conditions
The concentration and pH of PBS play important roles in CE
separation. In order to acquire good CE separation of the two
analytes, the eects of the CE running buer on separation were
investigated. In this work, PBS was chosen as the running
buer. The buer concentration was tested from 10 to 30 mmol
L
1
, and 20 mmol L
1
was selected. It was found that baseline
separation results were not obtained when PBS alone was used
as the running buer. In order to further improve the separa-
tion and detection sensitivity, we choose Tween 20, Tween 80,
CTAB, PEG-20000, Brij-35 and PVP-K30 as additives respective
to the running buer. It was found that there were no obvious
improvements in resolution for any of these additives except for
Tween 20. Tween 20 concentrations ranging from 0 to 0.07% (v/
v) were tested. The results indicated that 0.03% Tween 20 in the
buer was good to increase the resolution of the two analytes.
The eect of the pH value of the running buer was also
investigated. The results showed that the ECL signal increased
with the running buer pH up to 8.0 and then decreased when
the pH was further increased from 8.0 to 8.5. Therefore, 8.0 was
chosen as the pH of the running buer for further experiments.
The separation voltage has a great eect on the eciency of
the separation and the migration time of analytes. The voltage
over the range of 10 to 16 kV was investigated. For two analytes,
the ECL intensity and the resolution reached a maximum value
up to a separation voltage at 13 kV. However, when the sepa-
ration voltage was greater than 13 kV, the baseline noise
increased. Furthermore, the increasing joule heat in the capil-
lary caused peak broadening and adversely aected the sepa-
ration eectiveness. Therefore, 13 kV was chosen as the
separation voltage.
Eect of detection potential
In ECL, the intensity depends on the oxidation rate of
Ru(bpy)
32+
and the potential applied to the working electrode.
Therefore, the relationship of ECL intensity with the applied
potential over the range of 1.10 to 1.35 V was studied. The
results indicated that when the applied potential increased
from 1.0 to 1.25 V, the ECL intensity increased; it reached
a maximum value at 1.25 V and then decreased slightly.
Therefore, the applied potential was set at 1.25 V.
Performance characteristics of the method
The optimal experimental conditions were used for the detec-
tion of Pro and Ace. The parameters were conrmed as follows:
20 mmol L
1
PBS containing 0.03% Tween 20 at pH 8.0 as
running buer; a volume ratio of AuNPs to 5 mmol L
1
Ru(bpy)
32+
with 50 mmol L
1
PBS of 0.7 at pH 8.0 as ECL reagent
solution in the detection cell; electrokinetic injection for 10 s at
10 kV; separation voltage at 13 kV; and detection potential at
1.25 V. A typical electropherogram of Pro and Ace is shown in
Fig. 4b; baseline separation of the two analytes could be ach-
ieved in 7 min.
The separation eciencies of the two analytes were evalu-
ated by the theoretical plate numbers, which were calculated
according to the following equation: N¼5.54 (t
R
/W
1/2
)
2
,
where Nis the theoretical plate number, t
R
is the migration time
and W
1/2
is the width at half the maximum peak height of the
analyte. The numbers of theoretical plates obtained for Pro and
Ace are 7.2 10
4
and 4.6 10
4
plates per m. The linear cali-
bration ranges, regression equations, limits of detection (LODs)
and limits of quantitation (LOQs) of the two analytes were also
examined, and the results are listed in Table 1. As can be seen
from Table 1, the linear range is 0.01 to 100 mmol L
1
for Pro
with the LOD of 3.6 10
9
mol L
1
(S/N ¼3) and with the LOQ
of 1.1 10
7
mol L
1
in urine samples (S/N ¼10). For Ace, the
linear range is 0.02 to 100 mmol L
1
with the LOD of 5.0 10
9
mol L
1
(S/N ¼3) and with the LOQ of 9.5 10
8
mol L
1
in
Fig. 4 Electropherograms of 1 10
7
mol L
1
propranolol (peak 1)
and 1 10
7
mol L
1
acebutolol (peak 2) without (a) and with (b)
AuNPs in the detection cell. Conditions: running buer, 20 mmol L
1
PBS containing 0.03% Tween 20 at pH 8.0; injection, 10 kV 10 s;
separation voltage, 13 kV; detection potential, 1.25 V. In the ECL
detection solution, the volume ratio of AuNPs to 5 mmol L
1
Ru(bpy)
32+
with 50 mmol L
1
PBS is 0.7 at pH 8.0.
Table 1 The performance characteristics of the proposed method
Analytes
Linear range
(mmol L
1
)
Calibration curves
LOD
(mol L
1
)
LOQ in urine
(mol L
1
)Slope Intercept R
Pro 0.01100 2958 16 714 0.998 3.6 10
9
1.1 10
7
Ace 0.02100 2607 32 363 0.995 5.0 10
9
9.5 10
8
45536 |RSC Adv.,2016,6, 4553345539 This journal is © The Royal Society of Chemistry 2016
RSC Advances Paper
urine samples (S/N ¼10). The relative standard derivations
(RSDs, n¼3) of the migration time for Pro and Ace (concen-
tration: 1 10
7
mol L
1
) are from 1.9 to 2.4% intraday and
from 2.9 to 3.8% interday. The RSDs of the peak area for Pro and
Ace are from 3.2 to 3.7% intraday and from 4.3 to 4.6% interday,
respectively.
Analytical applications
The proposed CE-ECL method was applied to the determination
of Pro and Ace in human urine samples. The analytes are not
found in urine samples of healthy person because the analytes
are exogenous substances. To evaluate the precision of the
proposed method, spiked urine samples were investigated. All
of the urine samples were spiked at three concentration levels to
determine the recoveries of the targeted analytes. For each
concentration level, ve replicates were conducted. The elec-
tropherograms of the blank urine sample from a healthy person
and a urine sample spiked with 5 10
7
mol L
1
Pro and Ace
are illustrated in Fig. 5a and b; no interferences from the
sample matrix overlapped with the two target analytes. The
recovery results are listed in Table 2. It can be seen that the
recoveries of Pro and Ace in urine samples are in the range of
98.0 to 106.7% and the RSD of the corresponding ECL peak
areas were less than 4.2%.
Comparison with other methods
A comparison of the proposed CE-ECL method with previous
reports
812
is listed in Table 3. Compared with other methods,
the proposed CE-ECL method exhibits high sensitivity and
a wide linear range, although the LOQ is lower in ref. 9. The
results demonstrated that the proposed CE-ECL method has
excellent application prospects in the eld of drug analysis.
Study of the binding of HSA with Pro
It is oen of great interest to investigate the interaction of HSA
with drugs. HSA-drug binding has become an important
research eld in chemistry, the life sciences and clinical medi-
cine; it is of great interest to evaluate this binding during the
drug development process. Therefore, the binding behaviour
between HSA and Pro as a model analyte was researched in this
work.
In order to obtain the equilibrium time, a mixture solution of
100 mL20mmol L
1
Pro and 100 mL 200 mmol L
1
HSA in
a dialysis bag was incubated at 37 C. The solution outside of
the dialysis bag was 4 mL PBS and the ECL intensity of Pro
outside of the bag was detected every half an hour until no
remarkable change in the ECL intensity was observed (a total of
about 8 h), which indicates that the binding has already reached
its equilibrium. The results show that the equilibrium time is
about 5 h. In order to explore whether Pro can react with HSA,
100 mL 20 mmol L
1
Pro, 100 mL 200 mmol L
1
HSA, and
a mixture of 100 mL 20 mmol L
1
Pro with 100 mL 200 mmol L
1
HSA balanced from the start to the end solution were subse-
quently examined by UV-vis spectrophotometry for comparison,
respectively. As shown in Fig. 6, there was a marked dierence
in absorbance between the balanced solution (curve d) and the
starting mixture of HSA and Pro (curve c). Both facts conrmed
the binding of HSA with Pro under the experimental conditions.
Fig. 5 Electropherograms of a blank urine sample (a) and a urine
sample spiked with Pro and Ace (b). Peak 1: Pro, peak 2: Ace, peak X:
unknown compounds. Other conditions are the same, as shown in
Fig. 4.
Table 2 Recoveries in human urine at dierent spiked levels
Analytes
Added
(mmol L
1
)
Found
(mmol L
1
)
Recovery
(%)
RSD (%)
(n¼5)
Pro 0.5 0.51 102.0 4.2
25 24.8 99.2 2.8
75 75.8 106.7 3.3
Ace 0.5 0.49 98.0 1.9
25 25.6 102.4 3.6
75 74.3 99.1 2.4
Table 3 Comparison of other methods for Pro and Ace determination with the proposed CE-ECL method
Methods
Linear range (mmol L
1
) LOD (mol L
1
) LOQ (mol L
1
)
Ref.Pro Ace Pro Ace Pro Ace
SPE-GC-MS 203400 203400 6.8 10
3
6.8 10
3
2.0 10
2
2.0 10
2
8
GC-MS 0.06817 2.1 10
8
6.3 10
8
9
FI-CL 06.8 10
4
1.3 10
7
———10
HPLC-UV 853400 853400 2.0 10
5
2.0 10
5
8.5 10
5
8.5 10
5
11
SPE-DPV 5100 20150 5.0 10
6
2.1 10
5
1.2 10
5
4.2 10
5
12
CE-ECL 0.01100 0.02100 3.6 10
9
5.0 10
9
1.1 10
7
9.5 10
8
This work
This journal is © The Royal Society of Chemistry 2016 RSC Adv.,2016,6, 4553345539 | 45537
Paper RSC Advances
Pro and HSA bind to form a complex structure; an improved
equation
31
was used to calculate the number of binding sites (n)
and the binding constant (K).
r¼½Cbound
½Ptotal¼X
m
i¼1
ni
KiCfree
1þKiCfree(1)
where [C
bound
], [P
total
], and [C
free
] represent the concentrations
of bound drug, total HSA and free drug, respectively; ris the
fraction of bound drug molecules per protein molecule, n
i
represents the number of sites of class iand K
i
is the binding
constant.
32
Drug protein data analysis oen assumes one type of
binding site on the protein; thus, eqn (1) can be simplied
to (2):
r¼n
KCfree
1þKCfree(2)
A series of dierent volumes (ranging from 40 to 160 mL, with
increments of 20 mL) of 20 mmol L
1
Pro was mixed with 100 mL
200 mmol L
1
HSA in the dialysis bag and incubated in 4 mL PBS
at 37 C. Aer equilibrium was reached, the ECL intensity
outside the dialysis bag was measured and the concentration of
the free Pro could be estimated. According to eqn (2), the
number of binding sites and binding constant can be obtained;
the binding curves of Pro with HSA were established, as shown
in Fig. 7. According to the calculated results, the number of
binding site and the binding constant of Pro with HSA are 1.0
and 2.3 10
4
L mol
1
, respectively.
Conclusions
In this study, a new and sensitive CE-ECL method for the
detection of Pro and Ace was described. The experimental
conditions were discussed in detail. The introduction of AuNPs
enhanced the detection sensitivity. Compared with other
methods, the proposed CE-ECL method has high sensitivity and
a wide linear range for Pro and Ace analysis. The applicability of
the proposed method was demonstrated in the determination
of Pro and Ace in human urine samples and investigation of the
interaction between drug and protein. The proposed method is
promising for biochemical analysis.
Acknowledgements
This study was supported by the National Natural Science
Foundation of China (Grant 21375114, 21405129) and Nan Hu
Young Scholar Supporting Program of XYNU.
References
1 M. A. Gotardo, J. O. Tognolli, H. R. Pezza and L. Pezza,
Spectrochim. Acta, Part A, 2008, 69, 11031109.
2 P. M. Vanhoutte and Y. S. Gao, Curr. Opin. Pharmacol., 2013,
13, 265273.
3 A. C. Anderson, Clin. Pediatr. Emerg. Med., 2008, 9,416.
4 P. Valderrama and R. J. Poppi, Anal. Chim. Acta, 2009, 651,
3135.
5 J. S. Huang, J. Y. Sun, X. G. Zhou and T. Y. You, Anal. Sci.,
2007, 23, 183188.
6 L. Zahalka, L. Matysova, Z. Sklubalova, S. Klovrzova and
P. Solich, Chromatographia, 2013, 76, 15531558.
7 O. Divya and M. Shinde, J. Appl. Spectrosc., 2013, 80, 326334.
8 M. R. Brunetto, S. Clavijo, Y. Delgado, W. Orozco,
M. Gallignani, C. Ayala and V. Cerda, Talanta, 2015, 132,
1522.
9 S. Magiera, C. Uhlschmied, M. Rainer, C. W. Huck,
I. Baranowska and G. K. Bonn, J. Pharm. Biomed. Anal.,
2011, 56,93102.
10 G. Z. Tsogas, D. V. Stergiou, A. G. Vlessidis and
N. P. Evmiridis, Anal. Chim. Acta, 2005, 541, 149155.
11 M. Delamoye, C. Duverneuil, F. Paraire, P. de Mazancourt
and J. C. Alvarez, Forensic Sci. Int., 2004, 141,2331.
12 I. G. Casella, R. Bonito and M. Contursi, Electroanalysis,
2015, 27,19.
13 X. Pan, P. Qin, R. Liu and J. Wang, J. Agric. Food Chem., 2011,
59, 66506656.
14 K. Yamasaki, V. T. Giam Chuang, T. Maruyama and
M. Otagiri, Biochim. Biophys. Acta, 2013, 1830, 54355443.
Fig. 6 UV-visible absorption spectroscopy of the solution inside the
dialysis bag. Curve (a) HSA; curve (b) Pro; curve (c) HSA + Pro; curve (d)
after Pro combination with HSA.
Fig. 7 The binding curve of HSA with Pro.
45538 |RSC Adv.,2016,6, 4553345539 This journal is © The Royal Society of Chemistry 2016
RSC Advances Paper
15 S. S. Kalanur, J. Seetharamappa and V. K. L. Kalalbandi, J.
Pharm. Biomed. Anal., 2010, 53, 660666.
16 Y. M. Liu, Y. Y. Liu, M. Zhou, K. J. Huang, J. T. Cao, H. Wang
and Y. H. Chen, J. Chromatogr. A, 2014, 1340, 128133.
17 Z. Q. Wu, W. B. Du, J. Y. Li, X. H. Xia and Q. Fang, Talanta,
2015, 140, 176182.
18 L. W. Shao, C. Q. Dong, X. Y. Huang and J. C. Ren, Chin.
Chem. Lett., 2008, 19, 707710.
19 G. F. Shi, J. T. Cao, J. J. Zhang, K. J. Huang, Y. M. Liu,
Y. H. Chen and S. W. Ren, Analyst, 2014, 139, 58275834.
20 X. B. Yin, B. B. Sha, X. H. Zhang, X. W. He and H. Xie,
Electroanalysis, 2008, 20, 10851091.
21 H. Guo, X. L. Wu, A. L. Wang, X. W. Luo, Y. J. Ma and
M. Zhou, New J. Chem., 2015, 39, 89228927.
22 G. M. Zhu, S. H. Long, H. Sun, W. Luo, X. Li and Z. B. Hao, J.
Chromatogr. B: Anal. Technol. Biomed. Life Sci., 2013, 941,62
68.
23 D. An, Z. Q. Chen, J. C. Zheng, S. Y. Chen, L. Wang,
Z. Y. Huang and L. Weng, Food Chem., 2015, 168,16.
24 Z. M. Zhou, Z. Feng, J. Zhou, B. Y. Fang, X. X. Qi, Z. Y. Ma,
B. Liu, Y. D. Zhao and X. B. Hu, Biosens. Bioelectron., 2015,
64, 493498.
25 S. J. Sun, C. J. Long, C. Y. Tao, S. Meng and B. Y. Deng, Anal.
Chim. Acta, 2014, 851,3742.
26 Y. Y. Xu, L. F. Xu, S. D. Qi, Y. L. Dong, Z. U. Rahman,
H. L. Chen and X. G. Chen, Anal. Chem., 2013, 85, 11369
11375.
27 L. H. Zhang, Z. A. Xu, X. P. Sun and S. J. Dong, Biosens.
Bioelectron., 2007, 22, 10971100.
28 Y. Xiao, H. X. Ju and H. Y. Chen, Anal. Biochem., 2000, 278,
2228.
29 Y. M. Liu, L. Mei, L. J. Liu, L. F. Peng, Y. H. Chen and
S. W. Ren, Anal. Chem., 2011, 83, 11371143.
30 J. W. Wang, Z. M. Yang, X. X. Wang and N. J. Yang, Talanta,
2008, 76,8590.
31 M. H. A. Busch, L. B. Carels, H. F. M. Boelens, J. C. Kraak and
H. Poppe, J. Chromatogr. A, 1997, 777, 311328.
32 A. V. Rudnev, S. S. Aleksenko, O. C. Semenova, G. Hartinger,
A. R. Timerbaev and B. K. Keppler, J. Sep. Sci., 2005, 28, 121
127.
This journal is © The Royal Society of Chemistry 2016 RSC Adv.,2016,6, 4553345539 | 45539
Paper RSC Advances
... However, CE suffers from matrix effects and lack of sensitivity, and therefore could not find widespread applications in b-blockers analysis. CE methods reported in the last decade for the determination of b-blockers are presented in Table 2. [122][123][124][125][126][127][128][129][130][131][132][133] Separations were carried out using fused silica capillaries of 45-65 cm length and 25-75 mm internal diameter. Analysis times were generally less than 20 min for three or fewer b-blockers. ...
... For these reasons, sample pretreatment is considered as a crucial step in CE to eliminate matrix effects and to remove interfering compounds. [135] Although some authors performed a simple dilution to deal with matrix effects, [126,127,129,130] the applications of more complex and effective sample pretreatment techniques including ultrafiltration, [122,123] tapered-capillary microextraction ( tCap-lEx), [125] vortex-assisted liquid-liquid extraction (VALLE), [128] salt assisted liquid-liquid extraction (SALLE), [131] and membrane supported liquid-liquid-liquid microextraction [132] have also been reported for determination of b-blockers. ...
Article
β-blockers are among the most prescribed drugs and frequently recommended in patients with hypertension, angina, myocardial infarction, arrhythmias, and heart failure. Additionally, since β-blockers cause low anxiety, they are abused by athletes in branches where aiming is important. β-blocker residues may be present in environmental samples due to high consumption. At this point, the development of sensitive, rapid, and reliable analytical methods is very important for determination of β-blockers in different matrices. Among all the analytical methods, chromatographic methods are used in a wide range for the determination of β-blockers as they enable simultaneous determination of many analytes and removal of interfering components. Electrochemical methods, which usually do not require any pretreatment process, have become important due to their reliability, high specificity, and fast response. Simple and rapid optical methods can be used in laboratories where modern and expensive instruments such as high-performance liquid chromatography or gas chromatography are not available. The objective of this review is to provide an overview of the analytical methods for determination of β-blockers in pharmaceuticals, biological fluids, and environmental samples. Some selected studies for the determination of β-blockers published in the last decade are discussed. Furthermore, current trends and future perspectives on β-blocker analyses are also suggested.
Article
Single-molecule fluorescence imaging is a promising strategy for biomolecule detection. However, the accuracy of single-molecule method is often compromised by the false-positive events at the ultralow sample levels that are caused by the non-specific adsorption of the fluorescent labeled probe and other fluorescent impurities on the imaging surface. Here, we demonstrate an ultrasensitive single molecule detection assay based on dual-color fluorescent colocalization of spilt aptamers that was implemented to the measurement of adenosine triphosphate (ATP). The ATP aptamer was split into two fragments and labeled with green and red dye molecules respectively. When the two probes of split aptamers were brought together by the target ATP molecule, the two colors of fluorescence of two probes were simultaneously detected through two channels and projected to the correlated locations in the two halves of image. The colocalizaiton imaging of two split apatamer probes greatly excluded the false detection of biomolecules that was usually caused by the fluorescent noise of single non-bound aptamer probes and impurities, and further improved the accuracy of measurement. The assay showed excellent selectivity and high sensitivity for ATP detection with linear range of 1 pM - 5 nM and a detection limit of 100 fM. This versatile protocol of single molecule colocalization of split apatamer can be widely applied to the ultrasensitive and highly accurate detection of many types of biomolecules in basic research and biomedical applications.
Article
The scope of this two-part review is to summarize publications dealing with capillary electrophoresis (CE) and nanomaterials together. This topic can be viewed from two broad perspectives, and this article is trying to highlight these two approaches: I) CE of nanomaterials, and II) nanomaterials in CE. The second part aims at summarization of publications dealing with application of nanomaterials for enhancement of capillary electrophoresis performance either in terms of increasing the separation resolution or for improvement of the detection. To increase the resolution, nanomaterials are employed as either surface modification of the capillary wall forming open tubular column or as additives to the separation electrolyte resulting in a pseudostationary phase. Moreover, nanomaterials have proven to be very beneficial for increasing also the sensitivity of detection employed in capillary electrophoresis or even they enable the detection (e.g., fluorescent tags of non-fluorescent molecules). This article is protected by copyright. All rights reserved.
Article
A Ru(bpy)32+-based electrochemiluminescence method coupled with capillary electrophoresis has been developed for the separation and detection of two tropane alkaloids, anisodamine and anisodine, within 7 min with a platinum microelectrode modified with a europium(III)-doped Prussian blue analog film as the working electrode. Under optimized conditions, the ECL intensity was in proportion to the log values of anisodine and anisodamine concentrations in the range 5.0 × 10−6–1.0 × 10−3 mol L−1 (r = 0.9992) for anisodine and 1.0 × 10−6–1.0 × 10−3 mol L−1 (r = 0.9991) for anisodamine, with detection limits of 1.2 × 10−7 mol L−1 for anisodine and 4.3 × 10−8 mol L−1 for anisodamine. The proposed method has been applied to identify and detect the two alkaloids in the different parts (seeds, roots and leaves) of Anisodus tanguticus. The results showed that anisodine existed in all samples, including the seeds, roots and leaves of the plant, and its content was highest in the seeds and lowest in the leaves. However, anisodamine was not found in the samples mentioned above due to the low levels in the plant and maybe the interreaction between anisodamine and other chemicals in the plant. Finally, average recoveries of 98.0–107.0% were obtained.
Article
A new strategy is proposed for determination of carcino-embryonic antigen (CEA) based on aptamer/graphene oxide (Apt/GO) by capillary electrophoresis-chemiluminescence (CE-CL) detection system. CEA aptamer conjugated with horseradish peroxidase (HRP) firstly mixes with GO, and the CL will be quenched because the stack of HRP-Apt on GO leads to chemiluminescence resonance energy transfer (CRET). When CEA exists, the specific combination of HRP-Apt and CEA can form HRP-Apt-CEA complex, which dissociates from GO. Then, the CL catalyzed by HRP-Apt-CEA complex can be detected without any CRET, and the content of CEA can be estimated by the CL intensity. It has been proved that the interference issue resulted from free HRP-Apt is solved well by mixing GO firstly with HRP-Apt, which blocks the free HRP-Apt's CL signal due to CL quenching effect of GO; and the interference resulted from GO to CL is also solved by CE, then the sensitivity and accuracy can be greatly improved. Results also showed that the CL intensity had a linear relationship with the concentration of CEA in the range from 0.0654 to 6.54 ng/mL, and the limit of detection was approximately 4.8 pg/mL (S/N=3). This proposed method with high specificity offers a new way for separation and determination of biomolecule, and has good potential in application of biochemistry and bioanalysis.
Article
Numerical simulation can provide valuable insights for complex microfluidic phenomena coupling mixing and diffusion processes. Herein, a novel finite element model (FEM) has been established to extract chemical reaction kinetics in a microfluidic flow injection analysis (micro-FIA) system using high throughput sample introduction. To reduce the computation burden, the finite element mesh generation is performed with different scales based on the different geometric sizes of micro-FIA. In order to study the contribution of chemical reaction kinetics under non-equilibrium condition, a pseudo-first-order chemical kinetics equation is adopted in the numerical simulations. The effect of reactants diffusion on reaction products is evaluated, and the results demonstrate that the Taylor dispersion plays a determining role in the micro-FIA system. In addition, the effects of flow velocity and injection volume on the reaction product are also simulated. The simulated results agree well with the ones from experiments. Although gravity driven flow is used to the numerical model in the present study, the FEM model also can be applied into the systems with other driving forces such as pressure. Therefore, the established FEM model will facilitate the understanding of reaction mechanism in micro-FIA systems and help us to optimize the manifold of micro-FIA systems. Copyright © 2015 Elsevier B.V. All rights reserved.
Article
Beta-adrenergic blocking agents or β-blockers are prescribed for the treatment of a broad array of common disorders. Their widespread use, coupled with a narrow therapeutic index, contributes to their being a significant cause of poisoning from overdose and the second most common cause of mortality from cardiovascular agents. This article provides an overview of beta-adrenergic system pathophysiology and the properties and pharmacokinetics of β-blockers. Also discussed are the various effects and management of β-blockers in overdose settings.
Article
An on-line solid-phase extraction (SPE) method coupled to gas chromatography - mass spectrometry (GC-MS) has been developed for the determination of atenolol (ATN) and propranolol (PRO) in human plasma. The hyphenation of SPE with multisyringe flow injection analysis (MSFIA) allows the simultaneous GC-MS determination of ATN and PRO with high selectivity and sensitivity. On-line preconcentration and derivatisation of the analytes were carried out by means of using restricted access materials (RAM) and microwave (MW) assisted derivatisation reactions with N-methyl-N-(trimethylsilyl) trifluoroacetamide (MSTFA) + 1% trimethylchlorosilane (TMCS). Multivariate optimization was applied to optimize experimental conditions. The whole procedure comprising sample pretreatment and analyte determination took about 15 min. However, the overlap of the automatic sample treatment with the GC separation increased the frequency to 7 samples h-1. The validated method was successfully applied to direct ATN and PRO determination in human plasma.
Article
A facile label-free electrochemiluminescence (ECL) aptasensor based on the ECL of cadmium sulfide-graphene (CdS-GR) nanocomposites with peroxydisulfate as coreactant was designed for the detection of carcinoembryonic antigen (CEA). Tripetalous CdS-GR nanocomposites were synthesized through a simple onepot solvothermal method and immobilized on the glassy carbon electrode surface. L-cystine (L-cys) could promote the electron transfer and enhance the ECL intensity largely. Gold nanoparticles (AuNPs) were assembled onto the l-cys film modified electrode to the aptamer immobilization and the ECL signal amplification. The aptamer modified with thiol was adsorbed onto the AuNPs surface through Au-S bond. Upon the hybridization of the aptamer with target protein, the sequence could conjugate CEA to form Y architecture. With the CEA as a model analyte, the decreased ECL intensity is proportional to CEA concentration in the range of 0.01 - 10.0 ng/mL with the detection limit of 3.8 pg/mL (S/N = 3). The prepared aptasensor has been applied to determination of CEA in human serum samples. The recoveries of CEA in human serum samples are between 85.0% and 109.5%, and the RSD values are no more than 3.4%.
Article
Changes in the concentrations of putrescine, histamine, tyramine, phenylethylamine and spermidine were studied by the method of capillary electrophoresis coupled with electrochemiluminescence (CE-ECL) during the storage of oysters at two different temperatures (0 °C and 4 °C). The results showed that, with an increase in the storage time, spermidine and putrescine became dominant. When the oysters were stored at 0 °C, the concentration of spermidine increased from 45.6 mg/kg to 68.5 mg/kg, and that of putrescine increased from 18.6 mg/kg to 28.3 mg/kg. When the storage temperature was controlled at 4 °C, the concentration of spermidine increased from 46.7 mg/kg to 119.7 mg/kg and that of putrescine increased from 19.4 mg/kg to 136.8 mg/kg, respectively. In contrast, the histamine, tyramine and phenylethylamine levels increased slightly throughout the storage period for all of the experimental conditions.
Article
The paper describes a homemade ultrasonic microdialysis device coupled with capillary electrophoresis electrochemiluminescence (CE-ECL) for studying the interaction between human serum albumin (HSA) and trimetazidine dihydrochloride (TMZ). The time required for equilibrium by ultrasonic microdialysis was 45min, which was far less than that by traditional dialysis (240min). It took 80min to achieve the required combination equilibrium by normal incubation and only 20min by ultrasonic. Compared with traditional dialysis, the use of ultrasonic microdialysis simplified experimental procedures, shortened experimental time and saved consumption of sample. A simple, sensitive and selective determination of TMZ was developed using CE-ECL and the parameters that affected ECL intensity were optimized. Under the optimized conditions, the linear range of TMZ was from 0.075 to 80μmol/L (r(2)=0.9974). The detection limit was 26nmol/L with RSD of 2.8%. The number of binding sites and binding constant were 1.54 and 15.17L/mol, respectively. Copyright © 2014 Elsevier B.V. All rights reserved.
Article
The Silences of the Archives, the Reknown of the Story. The Martin Guerre affair has been told many times since Jean de Coras and Guillaume Lesueur published their stories in 1561. It is in many ways a perfect intrigue with uncanny resemblance, persuasive deception and a surprizing end when the two Martin stood face to face, memory to memory, before captivated judges and a guilty feeling Bertrande de Rols. The historian wanted to go beyond the known story in order to discover the world of the heroes. This research led to disappointments and surprizes as documents were discovered concerning the environment of Artigat’s inhabitants and bearing directly on the main characters thanks to notarial contracts. Along the way, study of the works of Coras and Lesueur took a new direction. Coming back to the affair a quarter century later did not result in finding new documents (some are perhaps still buried in Spanish archives), but by going back over her tracks, the historian could only be struck by the silences of the archives that refuse to reveal their secrets and, at the same time, by the possible openings they suggest, by the intuition that almost invisible threads link here and there characters and events.
Article
Due to the unusual properties of the structure, the metal organic frameworks (MOFs) have received great interest in separation science. However, the most existing methods for the applications of MOFs in separation science require an off-line procedure to prepare the materials. Here, we report an in situ, layer-by-layer self-assembly approach to fabricate MIL-100(Fe) coated open tubular (OT) capillary columns for capillary electrochromatography. By a controllable manner, the OT capillary columns with a tailored MIL-100(Fe) coating have been successfully synthesized. The results of SEM, XRD, FT-IR, and ICP-AES indicated that MIL-100(Fe) was successfully grafted on the inner wall of the capillary. Some neutral, acidic and basic analytes were used to evaluate the performance of the MIL-100(Fe) coating OT capillary column. Because of the size selectivity of lattice aperture and hydrophobicity of the organic ligands, three types of analytes were well separated with this novel MIL-100(Fe) coating OT capillary column. For three consecutive runs, the intra-day relative standard deviations (RSDs) of migration time and peak areas were 0.4-4.6% and 1.2-6.6%, respectively. The inter-day RSDs of migration time and peak areas were 0.6-8.0% and 2.2-9.5%, respectively. The column-to-column reproducibility of retention time was in range of 0.6-9.2%. Additionally, the 10 cycles OT capillary column (10-LC) could be used for more than 150 runs with no observable changes on the separation efficiency.