With its unique structure, the Akiyama probe is a type of tuning fork atomic force microscope probe. The long, soft cantilever makes it possible to measure soft samples in tapping mode. In this article, some characteristics of the probe at its second eigenmode are revealed by use of finite element analysis (FEA) and experiments in a standard atmosphere. Although the signal-to-noise ratio in this environment is not good enough, the 2 nm resolution and 0.09 Hz/nm sensitivity prove that the Akiyama probe can be used at its second eigenmode under FM non-contact mode or low amplitude FM tapping mode, which means that it is easy to change the measuring method from normal tapping to small amplitude tapping or non-contact mode with the same probe and equipment.
Content may be subject to copyright.
Sensors 2015, 15, 28764-28771; doi:10.3390/s151128764
sensors
ISSN 1424-8220
www.mdpi.com/journal/sensors
Article
A Unique Self-Sensing, Self-Actuating AFM Probe at
Higher Eigenmodes
Zhichao Wu, Tong Guo *, Ran Tao, Leihua Liu, Jinping Chen, Xing Fu and Xiaotang Hu
State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University,
Tianjin 300072, China; E-Mails: bertwu@tju.edu.cn (Z.W.); sam1909@163.com (R.T.);
liuleihua@tju.edu.cn (L.L.); chenjinping@tju.edu.cn (J.C.); xingfu@tju.edu.cn (X.F.);
xthu@tju.edu.cn (X.H.)
* Author to whom correspondence should be addressed; E-Mail: guotong@tju.edu.cn;
Tel./Fax: +86-222-7892-691.
Academic Editor: Vittorio M. N. Passaro
Received: 3 October 2015 / Accepted: 9 November 2015 / Published: 13 November 2015
Abstract: With its unique structure, the Akiyama probe is a type of tuning fork atomic
force microscope probe. The long, soft cantilever makes it possible to measure soft
samples in tapping mode. In this article, some characteristics of the probe at its second
eigenmode are revealed by use of finite element analysis (FEA) and experiments in a
standard atmosphere. Although the signal-to-noise ratio in this environment is not good
enough, the 2 nm resolution and 0.09 Hz/nm sensitivity prove that the Akiyama probe can be
used at its second eigenmode under FM non-contact mode or low amplitude FM tapping
mode, which means that it is easy to change the measuring method from normal tapping to
small amplitude tapping or non-contact mode with the same probe and equipment.
Keywords: AFM; quartz tuning fork; higher eigenmode; non-contact; finite element analysis
1. Introduction
Quartz tuning forks are designed for high-precision frequency control and are widely used in clocks,
watches, and digital circuit frequency standards. By taking advantage of their extreme stability in
frequency, their high quality factor, their self-sensing and self-actuating capabilities, and the ease with
which the vibration signal may be obtained with fewer components than the conventional atomic force
OPEN ACCESS
Sensors 2015, 15 28765
microscopy (AFM) probes, and so on, they can be used as force sensors in AFM [1–4]. The tuning fork
AFM probes are typically realized in two forms (Figure 1). The tip of the probe could be a carbon
nanotube, a fiber, a conventional AFM cantilever, or another type of stylus.
Figure 1. Typical tuning fork probe structures.
These probes retain their high quality factor and high prong stiffness which makes them a stable
source for small vibration amplitudes. On the other hand, the high stiffness is a drawback when
measuring soft samples. To couple a soft cantilever to the quartz tuning fork, Bayat et al. designed a
novel probe [5] which has been commercialized by the Nanosensors Corporation (Neuchatel,
Switzerland). As shown in Figure 2, a U-shaped silicon nitride cantilever is combined in a symmetrical
arrangement with a quartz tuning fork. A slant probing tip is at the free end of the cantilever. The two
legs of the cantilever are fixed to the two prongs of the tuning fork respectively. The parameters of the
Akiyama probe are as follows: the resonant frequency is 45–55 kHz; the spring constant is about 5 N/m;
the cantilever’s length, width, thickness are respectively 310 μm, 90 μm, 3.7 μm; and the diameter of
the tip is less than 15 nm. The tip point is vertical, and lies perpendicular to the lateral plane defined by
the tuning fork and cantilever. Under the excitation of the tuning fork, the probe is self-sensing by
converting the deflection of the cantilever to a change in charge.
Figure 2. The appearance and motion of an Akiyama probe.
For the long, soft cantilever, this Akiyama probe cannot be used in non-contact AFM at its first
eigenmode like the conventional tuning fork probe can [6]. As is well known, a cantilever working at
higher eigenmode frequencies can increase the effective stiffness [7,8]. Thus, working at a higher
eigenmode can extend the range of application of the Akiyama probe.
Sensors 2015, 15 28766
2. Finite Element Analysis of the Probe
To account for realistic probe geometries and the electric field distribution, finite element
analysis (FEA) can be a powerful tool. ANSYS was used in this FEA and the parameters [2] are shown
in Table 1.
Table 1. FEA parameters.
Parameters Tuning Fork (Quartz) Cantilever (SiN)
Elastic constant
matrix/GPa
[]
11 12 13 14
11 13 14
33
11 12 14
44
44
00
00
000
()/20
0
ccc c
cc c
c
ANSYS c cc c
c
c




=





11 12 13
14 33 44
86.74 6.99 11.91
17.91 107.2 57.94
ccc
ccc
===
== =
,,
,,
None
Piezoelectric constant
matrix C/m2
[]
11
11
11
14
14
00
00
000
00
00
00
e
e
ANSYS e e
e
e
=
11 14
0.171 0.0406ee==
None
Permittivity F/m 4.43, 4.43, 4.63 (x-, y-, z-directions, respectively)
Young modulus, GPa None 180
Poisson’s ratio None 0.28
Density kg/m3 2290 2300
Length, μm 2690 310
Width, μm 220 90
Thickness, μm 100 3.7
Finite element SOLID226 SOLID95
Figures 3 and 4 are the results of the physical and electric analysis at the first and the second
eigenmodes respectively. As shown in Figure 5, the vibration amplitude of the second eigenmode is
much smaller than that of the first eigenmode, but the electric charge outputs are of the same order of
magnitude, which means that the Akiyama probe can obtain detectable low-level outputs.
1st Eigenmode
2nd Eigenmode
Figure 3. The vibration analysis of the Akiyama probe.
Sensors 2015, 15 28767
Figure 4. The electric field analysis of the Akiyama probe.
(a) (b)
Figure 5. The frequency response curve predicted by FEA. (a) The first eigenmode;
(b) The second eigenmode.
Conventional small-amplitude AFMs often use short, stiff cantilevers to keep the small vibrations
on a low level and ensure that the vibration can be detected. This Akiyama probe, with its long
cantilever, may manifest unique behavior while working as a small-amplitude AFM.
3. Results and Discussion
The Akiyama probe uses an amplification board based on a typical circuit [2] for I-V conversion
and capacitance compensation, as shown in Figure 6 and the influence of the parasitic capacitance is
shown in Figure 7.
Figure 6. Typical circuit diagram for an Akiyama probe amplification board.
Sensors 2015, 15 28768
(a) (b)
Figure 7. The electrical amplitude-frequency response of an Akiyama probe. (a) The first
eigenmode; (b) The second eigenmode.
3.1. Frequency Response of the Probe
The vibration amplitude of the cantilever is measured by laser Doppler vibrometer (LDV). As
shown in Figure 8, the amplitude of the first eigenmode is about 85 nm, the second eigenmode’s
amplitude is about 1.5 nm which is of the same order of magnitude as the conventional small
amplitude AFM. The LDV controller (OFV-3001, Polytec, Waldbronn, Germany) had a velocity decoder
resolution of 0.5 μm/s (RMS), a maximum frequency of 250 kHz, and a measuring range of 5 mm/s/V.
(a) (b)
Figure 8. The vibration amplitude-frequency response of an Akiyama probe. (a) The first
eigenmode; (b) The second eigenmode.
Differing from the FEA prediction, the electrical output from the second eigenmode was only 10%
of that of the first eigenmode, the sensitivity is 0.030 against 0.465 as shown in Figure 9; however, the
signal was still easily obtained with the same amplification as that for the first eigenmode, so the
amplification board can remain the same for different tests. This meant that the same system could
Sensors 2015, 15 28769
accomplish both normal tapping and small-amplitude tapping modes by changing the excitation signal
frequency alone.
(a) (b)
Figure 9. The scale (sensitivity) of the first and second eigenmodes. (a) The first
eigenmode; (b) The second eigenmode.
3.2. The Approach Curve
The tuning fork probes can work in FM non-contact mode [1]. In the test system, a phase-locked
loop (PLL) was used to trace the resonant frequency of the Akiyama probe during its approach to the
sample surface. The PLL (HF2PLL, Zurich Instruments, Zurich, Switzerland) had a frequency
resolution of 0.8 μHz. The approach curve shown in Figure 10 lay in the attraction region within which
non-contact AFM worked.
Figure 10. The approach curve for the second eigenmode in FM mode.
Compared with the 0.8 Hz/nm sensitivity of the first eigenmode in the repulsion region, the
sensitivity of the second eigenmode was about 0.09 Hz/nm; however, under standard atmospheric
conditions, the signal-to-noise ratio was not good enough to obtain a higher resolution.
Sensors 2015, 15 28770
3.3. The Resolution of the Probe
This test used a piezo-stage to raise the sample in steps of 2 nm after the probe touched the surface
with measurements taken at 20 points per step, the measured data and average data of each step are
shown in Figure 11.
Figure 11. The resolution of an Akiyama probe (2 nm increments).
Although the noise was apparent, the 2 nm steps in the average data plot are visible in Figure 11,
which meant that this mode can realize a resolution of 2 nm.
4. Conclusions
For a special structure, an Akiyama probe can be used as an FM-mode, non-contact AFM sensor in
its second eigenmode, although its cantilever is long and soft. According to the results, the second
eigenmode Akiyama probe was similar to the conventional tuning fork AFM probe working under
non-contact mode for which the signal-to-noise ratio is not good enough to obtain a better resolution
under atmospheric conditions. One advantage of the self-sensing, self-actuating AFM probe based on a
tuning fork was the simplicity with which it could be assembled for vacuum AFM to improve the
signal-to-noise ratio. Thus, it was appropriate for the measurement of different samples, such as soft or
movable samples, and for the use of the Akiyama probe in small-amplitude tapping mode or in non-contact
mode, under vacuum, without changing either the probe or the equipment.
Acknowledgments
The authors acknowledge the support of the Tianjin Natural Science Fund (No. 14JCZDJC39400),
the National Key Technology Research and Development MOST Programme (No. 2011BAK15B02),
and the 111 Project fund (B07014).
Sensors 2015, 15 28771
Author Contributions
Zhichao Wu, study design and software design; Tong Guo, literature research and study design;
Ran Tao, hardware design and data acquisition; Leihua Liu, test design and data acquisition;
Jinping Chen, data analysis and manuscript editing; Xing Fu, data analysis and manuscript revision;
Xiaotang Hu, manuscript revision and study concepts.
Conflicts of Interest
The authors declare no conflict of interest.
References
1. De Oteyza, D.G.; Gorman, P.; Chen, Y.-C.; Wickenburg, S.; Riss, A.; Mowbray, D.J.; Etkin, G.;
Pedramrazi, Z.; Tsai, H.-Z.; Rubio, A.; et al. Direct Imaging of Covalent Bond Structure in
Single-Molecule Chemical Reactions. Science 2013, 340, 1434–1437.
2. Zhao, J. Study on Development of Self-Sensing AFM Head and Hybrid Measurement System.
Ph.D. Thesis, Tianjin University, Tianjin, China, 2011.
3. Zhao, J.; Guo, T.; Ma, L.; Fu, X.; Hu, X. Metrological atomic force microscope with self-sensing
measuring head.Sens. Actuators A 2011, 167, 267272.
4. Guo, T.; Wang, S.; Dorantes-Gonzalez, D.J.; Chen, J.; Fu, X.; Hu, X. Development of a Hybrid
Atomic Force Microscopic Measurement System Combined with White Light Scanning
Interferometry.
Sensors 2012, 12, 175–188.
5. Bayat, D.; Akiyama, T.; de Rooij, N.F.; Staufer, U. Dynamic behavior of the tuning fork AFM
probe. Microelectron. Eng. 2008, 85, 10181021.
6. Akiyama, T.; de Rooij, N.F.; Staufer, U.; Detterbeck, M.; Braendlin, D.; Waldmeier, S.;
Scheidiger, M. Implementation and characterization of a quartz tuning fork based probe consisted
of discrete resonators for dynamic mode atomic force microscopy. Rev. Sci. Instrum. 2010, 81,
doi:10.1063/1.3455219.
7. Lozano, J.R.; Kiracofe, D.; Melcher, J.; Garcia, R.; Raman, A. Calibration of higher
eigenmode spring constants of atomic force microscope cantilevers. Nanotechnology 2010, 21,
doi:10.1088/0957–4484/21/46/465502.
8. Frentrup, H.; Allen, M.S. Error in dynamic spring constant calibration of atomic force microscope
probes due to nonuniform cantilevers. Nanotechnology 2011, 22, doi:10.1088/0957–4484/22/29/295703.
© 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article
distributed under the terms and conditions of the Creative Commons Attribution license
(http://creativecommons.org/licenses/by/4.0/).
... Akiyama probes are now commercially available and in active use as a characterization tool, yet relatively few studies have addressed their capability for multifrequency sensing [17,18]. To this end, measuring the spring constant, quality factor, and responsivity of the mechanical eigenmodes of Akiyama probes is essential to assess the viability of their utilization as force sensors with respect to more conventional cantilever and tuning fork probes. ...
... Finally, we also note the presence of a small, narrow, vibrational peak located at 197 kHz with a high Q factor of ∼7650. This peak has been identified as corresponding to the second flexural eigenmode of the tuning fork prongs [8,17,18,26], and is only weakly transduced into vertical motion of the cantilever. While small in amplitude, for multifrequency measurement applications it is noteworthy that vibrational motion associated with this eigenmode can be strongly transduced into electronic signals. ...
... FEA calculations were performed using COMSOL Multiphysics software and the solid mechanics module. All geometrical parameters of the tuning fork and Akiyama cantilever are shown in figure 3(a), and material properties are those reported in [18]. The tuning fork (Si cantilever) material has a density of 2290 kg m −3 (2300 kg m −3 ), Young's modulus of 76.5 GPa (180 GPa), and Poison ratio of 0.228 (0.28). ...
Article
Full-text available
Akiyama atomic force microscopy probes represent a unique means of combining several of the desirable properties of tuning fork and cantilever scanning probe designs. As a hybridized mechanical resonator, the vibrational characteristics of Akiyama probes result from a complex coupling between the intrinsic vibrational eigenmodes of its constituent tuning fork and bridging cantilever components. Through a combination of finite element analysis modeling and experimental measurements of the thermal vibrations of Akiyama probes we identify a complex series of vibrational eigenmodes and measure their frequency, quality factor, and spring constant. We then demonstrate the viability of Akiyama probes to perform bimodal multi-frequency force sensing by performing a multimodal measurement of a surface’s nanoscale photothermal response using photoinduced force microscopy (PiFM) imaging techniques. Further parametrically searching over alternative Akiyama probe geometries, we propose two modified probe designs to enhance the capability of Akiyama probes to perform sensitive bimodal multifrequency force sensing measurements.
... Certain researchers have enhanced the functions of AFM by modifying the structure of a QTF probe [7][8][9][10][11]. Bayat [7] coupled a cantilever probe called the Akiyama probe to a QTF. ...
... Taking a binomial expansion of those terms ( ) 8 1 wz , ; F w z k z w x t z F z =+ (6) where 9 3 ...
Article
Full-text available
A self-sensing and self-actuating quartz tuning fork (QTF) can be used to obtain its frequency shift as function of the tip-sample distance. Once the function of the frequency shift versus force gradient is acquired, the combination of these two functions results in the relationship between the force gradient and the tip-sample distance. Integrating the force gradient once and twice elucidates the values of the interaction force and the interatomic potential, respectively. However, getting the frequency shift as a function of the force gradient requires a physical model which can describe the equations of motion properly. Most papers have adopted the single harmonic oscillator model, but encountered the problem of determining the spring constant. Their methods of finding the spring constant are very controversial in the research community and full of discrepancies. By circumventing the determination of the spring constant, we propose a method which models the prongs and proof mass as elastic bodies. Through the use of Hamilton’s principle, we can obtain the equations of motion of the QTF, which is subject to Lennard-Jones potential force. Solving these equations of motion analytically, we get the relationship between the frequency shift and force gradient.
... The two probes are operated at similar setting that the amplitude is about 300mV and the maximum frequency shift is set to 50Hz (typical 100~400Hz). The 300mV amplitude is relatively small for the probe [46] and optimal for imaging based on our previous investigation [41], which is beneficial to the tip protection in the unfriendly sidewall measurement. The frequency-distance sensitivities are finally determined as 0.453 Hz/nm and 0.458 Hz/nm by fitting linearly the raw experimental data. ...
... Due to the aforementioned benefits, it would be advantageous if the PEGDA tip can be attached and regenerated onto QTFs. While attaching a PEGDA tip to QTFs seems much easier than attaching Open Access *Correspondence: jungchullee@kaist.ac.kr 3 Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology, 291 Daehak-ro, Yuseong-gu, Daejeon 34141, Republic of Korea Full list of author information is available at the end of the article a microfabricated silicon cantilever [4,5] or an etched tungsten wire [6], the negative tip mold previously used for tipless silicon cantilevers [2] is too small for QTFs: it may prevent easy use of the probe for imaging after tip attachment. A simple and straightforward solution for this issue is to make a larger negative tip mold. ...
Article
Full-text available
This paper reports the first demonstration of hydrogel conical tip attachment onto quartz tuning fork (QTF) by using an elastomeric tip mold that is soft-lithographically replicated from an electrochemically etched tungsten wire. The tungsten tip of 10–100 nm radius obtained by time-controlled electrochemical etching is replicated with h-polydimethylsiloxane (h-PDMS) to make negative conical tip molds large enough to be used for QTFs. By approaching a QTF to the negative h-PDMS tip mold filled with polyethylene glycol-diacrylate (PEGDA), a PEGDA tip is attached to the QTF without using an adhesive. Then, the PEGDA tip attached QTF is employed for shear force microscopy for calibration grating and atomic layers of hexagonal silicon carbide and also compared with a silicon tip attached QTF. Exclusively for the PEGDA tip attached QTF, we demonstrate that the imaging tip could be regenerated multiple times to address issues associated with tip wear. In a stark contrast with conventional QTF probes in attachment of electrochemically etched metallic wires or microfabricated AFM cantilevers, photocuring of liquid phase prepolymer within a tip mold demonstrated herein allows adhesive-free and exclusive attachment of the imaging tip onto a QTF. The relatively large PEGDA tip enables facile operation during approach and engagement. Moreover, the organic and inorganic combination of imaging tip and resonating body offers regeneration of the imaging tip upon its degradation.
... Due to the aforementioned reasons, it would be advantageous if the PEG-DA tip can be attached and regenerated onto QTFs. While the attachment of a PEG-DA tip to QTFs seems much easier than the attachment of a microfabricated silicon cantilever [2] or an etched tungsten tip [3], the negative tip mold previously used for tipless silicon cantilevers [1] is too small to be used for QTFs. The simple solution for this issue is to make a larger negative tip mold. ...
... In recent years, quartz tuning forks have been used as nanoscopy sensors due to their advantages under certain conditions over standard microcantilever sensors [17,18]. Some of the disadvantages of these QTF sensors are being overcome, such as the lack of a confident value for the spring constant [19] and the difficulty in making the tip radius comparable in size with the ones in commercial probes [20]; therefore, in the present work, QTF sensors were used to control the tip-sample distance. ...
Article
Full-text available
Here we devise a multiprobe electrical measurement system based on quartz tuning forks (QTFs) and metallic tips capable of having full 3D control over the position of the probes. The system is based on the use of bent tungsten tips that are placed in mechanical contact (glue-free solution) with a QTF sensor. Shear forces acting in the probe are measured to control the tip-sample distance in the Z direction. Moreover, the tilting of the tip allows the visualization of the experiment under the optical microscope, allowing the coordination of the probes in X and Y directions. Meanwhile, the metallic tips are connected to a current-voltage amplifier circuit to measure the currents and thus the impedance of the studied samples. We discuss here the different aspects that must be addressed when conducting these multiprobe experiments, such as the amplitude of oscillation, shear force distance control, and wire tilting. Different results obtained in the measurement of calibration samples and microparticles are presented. They demonstrate the feasibility of the system to measure the impedance of the samples with a full 3D control on the position of the nanotips.
Article
Full-text available
Quartz tuning fork (QTF) is often utilised for the design and manufacture of high-performance sensors, such as the force sensor for atomic force microscopy (AFM). In this study, the mechanical resonance and sensing characteristics of QTF with an additional mass on its prongs are studied by theoretical analysis and experiments. The results show that the position and size of additional mass on the QTF prong have an obvious influence on its resonance frequency and quality factor (Q-factor). The greater the mass of the additional component or the closer to the free end of the QTF prong is, the lower its resonance frequency and quality factor will become. Reducing the additional effective mass can significantly improve its quality factor, thus improving the sensing performance of the QTF sensor. This study can provide a quantitative reference for the design and optimisation of the QTF force sensor for AFM.
Article
Precision mechatronics and nanotechnology communities can both benefit from a course centered around an Atomic Force Microscope (AFM). Developing an AFM can provide precision mechatronics engineers with a valuable multidisciplinary hands-on training experience. In return, such expertise can be applied to the design and implementation of new precision instruments, which helps nanotechnology researchers make new scientific discoveries. However, existing AFMs are not suitable for mechatronics education due to their different original design intentions. Therefore, we address this challenge by developing an AFM intended for precision mechatronics education. This paper presents the design and implementation of an educational AFM and its corresponding precision mechatronics class. The modular educational AFM is low-cost (≤$4,000) and easy to operate. The cost reduction is enabled by new subsystem development of a buzzer-actuated scanner and demodulation electronics designed to interface with a myRIO data acquisition system. Moreover, the use of an active cantilever probe with piezoresistive sensing and thermomechanical actuation significantly reduced experiment setup overhead with improved operational safety. In the end, the developed AFM capabilities are demonstrated with imaging results. The paper also showcases the course design centered around selected subsystems. The new AFM design allows scientific-method-based learning, maximizes utilization of existing resources, and offers potential subsystem upgrades for high-end research applications. The presented instrument and course can help connect members of both the AFM and the mechatronics communities to further develop advanced techniques for new applications.
Article
Full-text available
AFM-in-SEM as a Tool for Comprehensive Sample Surface Analysis - Volume 28 Issue 3 - Veronika Novotna, Josef Horak, Martin Konecny, Veronika Hegrova, Ondrej Novotny, Zdenek Novacek, Jan Neuman
Article
Full-text available
Observing the intricate chemical transformation of an individual molecule as it undergoes a complex reaction is a long-standing challenge in molecular imaging. Advances in scanning probe microscopy now provide the tools to visualize not only the frontier orbitals of chemical reaction partners and products, but their internal covalent bond configurations as well. We used noncontact atomic force microscopy to investigate reaction-induced changes in the detailed internal bond structure of individual oligo-(phenylene-1,2-ethynylenes) on a (100) oriented silver surface as they underwent a series of cyclization processes. Our images reveal the complex surface reaction mechanisms underlying thermally induced cyclization cascades of enediynes. Calculations using ab initio density functional theory provide additional support for the proposed reaction pathways.
Article
Full-text available
A hybrid atomic force microscopic (AFM) measurement system combined with white light scanning interferometry for micro/nanometer dimensional measurement is developed. The system is based on a high precision large-range positioning platform with nanometer accuracy on which a white light scanning interferometric module and an AFM head are built. A compact AFM head is developed using a self-sensing tuning fork probe. The head need no external optical sensors to detect the deflection of the cantilever, which saves room on the head, and it can be directly fixed under an optical microscopic interferometric system. To enhance the system's dynamic response, the frequency modulation (FM) mode is adopted for the AFM head. The measuring data can be traceable through three laser interferometers in the system. The lateral scanning range can reach 25 mm × 25 mm by using a large-range positioning platform. A hybrid method combining AFM and white light scanning interferometry is proposed to improve the AFM measurement efficiency. In this method, the sample is measured firstly by white light scanning interferometry to get an overall coarse morphology, and then, further measured with higher resolution by AFM. Several measuring experiments on standard samples demonstrate the system's good measurement performance and feasibility of the hybrid measurement method.
Article
Full-text available
The quartz tuning fork based probe {e.g., Akiyama et al. [Appl. Surf. Sci. 210, 18 (2003)]}, termed "A-Probe," is a self-sensing and self-actuating (exciting) probe for dynamic mode atomic force microscope (AFM) operation. It is an oscillatory force sensor consisting of the two discrete resonators. This paper presents the investigations on an improved A-Probe: its batch fabrication and assembly, mounting on an AFM head, electrical setup, characterization, and AFM imaging. The fundamental features of the A-Probe are electrically and optically characterized in "approach-withdraw" experiments. Further investigations include the frequency response of an A-Probe to small mechanical vibrations externally applied to the tip and the effective loading force yielding between the tip and the sample during the periodic contact. Imaging of an electronic chip, a compact disk stamper, carbon nanotubes, and Si beads is demonstrated with this probe at ambient conditions in the so-called frequency modulation mode. A special probe substrate, which can snap on a receptacle fixed on an AFM head, and a special holder including a preamplifier electronic are introduced. We hope that the implementation and characterization of the A-Probe described in this paper will provide hints for new scanning probe techniques.
Article
A metrological atomic force microscope (AFM) adapting compact measuring head for the micro-nano dimensional measurement is introduced in this paper. Based on a novel self-sensing probe the head does not need any optical detector to measure the bending of the cantilever, significantly saving the working space to integrate with other measuring methods. This AFM operates in tapping mode and adapts frequency modulation (FM) mode to enhance measuring speed. A z-axis piezoelectric positioning stage (compact z-stage), with high resonant frequency, is responsible for the rapid motion of the sample in z-direction. A high-end digital signal processing (DSP) servo control system further guarantees high measurement speed. A nano-measuring machine (NMM) is equipped as the lateral moving stage to realize three-dimensional measurement. Three interferometers in the NMM enable the measurement to be traced to the meter definition. After calibration, the measurement of one-dimensional grating has been made. According to experimental results, the resolution of the AFM can reach nanometer level.
Article
Many common atomic force microscope (AFM) spring constant calibration methods regard the AFM probe as a uniform cantilever, neglecting the tip mass and any nonuniformity in the thickness of the probe along its length. This work quantifies the error in the spring constant estimated by the Sader and thermal calibration methods due to nonuniformity in the thickness of the cantilever and the influence of the mass loading effect of the probe tip. Formulae are presented that can be used to compute the uncertainty in cantilever calibration for an arbitrary thickness nonuniformity, or to correct the calibration methods if the thickness nonuniformity is known. The results show that both methods are quite sensitive to nonuniformity. When the first dynamic mode is used in the calibration, the error in the spring constant estimated by either method is between - 4% and 9% for a cantilever whose thickness increases or decreases linearly by 30% along its length. The errors are several times larger if the second or higher dynamic modes are used. To illustrate the proposed methods, a commercial AFM probe that has significant nonuniformity is considered and the error in calibrating this probe is quantified and discussed. For this particular probe, variations in the thickness of the probe over the last 15% of its length are found to significantly reduce the accuracy of the calibration when the thermal method is used, since that method is sensitive to changes in the shape of the eigenmode of the probe near its free end.
Article
Standard spring constant calibration methods are compared when applied to higher eigenmodes of cantilevers used in dynamic atomic force microscopy (dAFM). Analysis shows that Sader's original method (Sader et al 1999 Rev. Sci. Instrum. 70 3967-9), which relies on a priori knowledge of the eigenmode shape, is poorly suited for the calibration of higher eigenmodes. On the other hand, the thermal noise method (Hutter and Bechhoefer 1993 Rev. Sci. Instrum. 64 1868-73) does not require knowledge of the eigenmode and remains valid for higher eigenmodes of the dAFM probe. Experimental measurements of thermal vibrations in air for three representative cantilevers are provided to support the theoretical results.
Article
We recently introduced a new self-actuating and self-sensing atomic force microscope (AFM) probe based on a quartz tuning fork and a micro-fabricated cantilever. This system has two degrees of freedom, associated with its two components. We developed a model for describing how the sample-tip interaction is transduced to the tuning fork. It is based on two coupled spring-mass systems. In a first step, the coupling between the tuning fork and the cantilever was investigated to reveal the influential factors. The analysis of these factors enabled us to deduce their effect on the whole system and to optimize the sensitivity of this novel probe. The theoretical analysis was compared with experimental results and it was found that the model appropriately describes the probe in a qualitative manner while further refinement will be needed for achieving a correct quantitative description.
  • D G De Oteyza
  • P Gorman
  • Y.-C Chen
  • S Wickenburg
  • A Riss
  • D J Mowbray
  • G Etkin
  • Z Pedramrazi
  • H.-Z Tsai
  • A Rubio
De Oteyza, D.G.; Gorman, P.; Chen, Y.-C.; Wickenburg, S.; Riss, A.; Mowbray, D.J.; Etkin, G.; Pedramrazi, Z.; Tsai, H.-Z.; Rubio, A.; et al. Direct Imaging of Covalent Bond Structure in Single-Molecule Chemical Reactions. Science 2013, 340, 1434-1437.
Study on Development of Self-Sensing AFM Head and Hybrid Measurement System
  • J Zhao
Zhao, J. Study on Development of Self-Sensing AFM Head and Hybrid Measurement System. Ph.D. Thesis, Tianjin University, Tianjin, China, 2011.