To demonstrate the feasibility of a miniature handheld optical coherence tomography (OCT) imager for real time intraoperative vascular patency evaluation in the setting of super-microsurgical vessel anastomosis. A novel handheld imager Fourier domain Doppler optical coherence tomography based on a 1.3-µm central wavelength swept source for extravascular imaging was developed. The imager was minimized through the adoption of a 2.4-mm diameter microelectromechanical systems (MEMS) scanning mirror, additionally a 12.7-mm diameter lens system was designed and combined with the MEMS mirror to achieve a small form factor that optimize functionality as a handheld extravascular OCT imager. To evaluate in-vivo applicability, super-microsurgical vessel anastomosis was performed in a mouse femoral vessel cut and repair model employing conventional interrupted suture technique as well as a novel non-suture cuff technique. Vascular anastomosis patency after clinically successful repair was evaluated using the novel handheld OCT imager. With an adjustable lateral image field of view up to 1.5 mm by 1.5 mm, high-resolution simultaneous structural and flow imaging of the blood vessels were successfully acquired for BALB/C mouse after orthotopic hind limb transplantation using a non-suture cuff technique and BALB/C mouse after femoral artery anastomosis using a suture technique. We experimentally quantify the axial and lateral resolution of the OCT to be 12.6 µm in air and 17.5 µm respectively. The OCT has a sensitivity of 84 dB and sensitivity roll-off of 5.7 dB/mm over an imaging range of 5 mm. Imaging with a frame rate of 36 Hz for an image size of 1000(lateral)×512(axial) pixels using a 50,000 A-lines per second swept source was achieved. Quantitative vessel lumen patency, lumen narrowing and thrombosis analysis were performed based on acquired structure and Doppler images. A miniature handheld OCT imager that can be used for intraoperative evaluation of microvascular anastomosis was successfully demonstrated.
Content may be subject to copyright.
RESEARCH ARTICLE
MEMS-Based Handheld Fourier Domain
Doppler Optical Coherence Tomography
for Intraoperative Microvascular
Anastomosis Imaging
Yong Huang
1
*, Georg J. Furtmu¨ ller
2
, Dedi Tong
2,3
, Shan Zhu
2,4
,
W. P. Andrew Lee
2
, Gerald Brandacher
2
*, Jin U. Kang
1
1. Department of Electrical and Computer Engineering, Johns Hopkins University, 3400 N. Charles Street,
Baltimore, Maryland, 21218, United States of America, 2. Department of Plastic and Reconstructive Surgery,
Vascularized Composite Allotransplantation (VCA) Laboratory, Johns Hopkins University School of Medicine,
720 Rutland Avenue, Ross 749D, Baltimore, Maryland, 21205, United States of America, 3. Department of
Hand Surgery, Beijing Jishuitan Hospital, 31 Xinjiekou East Street, Xicheng District, Beijing, 10035, China, 4.
Peking Union Medical College and Chinese Academy of Medical Sciences, Department of Plastic and
Reconstructive Surgery, Plastic Surgery Hospital, 3 Ba-Da-Chu Road, Shijingshan District, Beijing, 10044,
China
*yhuang60@jhu.edu (YH);brandacher@jhmi.edu (GB)
Abstract
Purpose: To demonstrate the feasibility of a miniature handheld optical coherence
tomography (OCT) imager for real time intraoperative vascular patency evaluation
in the setting of super-microsurgical vessel anastomosis.
Methods: A novel handheld imager Fourier domain Doppler optical coherence
tomography based on a 1.3-mm central wavelength swept source for extravascular
imaging was developed. The imager was minimized through the adoption of a 2.4-
mm diameter microelectromechanical systems (MEMS) scanning mirror,
additionally a 12.7-mm diameter lens system was designed and combined with the
MEMS mirror to achieve a small form factor that optimize functionality as a
handheld extravascular OCT imager. To evaluate in-vivo applicability, super-
microsurgical vessel anastomosis was performed in a mouse femoral vessel cut
and repair model employing conventional interrupted suture technique as well as a
novel non-suture cuff technique. Vascular anastomosis patency after clinically
successful repair was evaluated using the novel handheld OCT imager.
Results: With an adjustable lateral image field of view up to 1.5 mm by 1.5 mm,
high-resolution simultaneous structural and flow imaging of the blood vessels were
successfully acquired for BALB/C mouse after orthotopic hind limb transplantation
using a non-suture cuff technique and BALB/C mouse after femoral artery
anastomosis using a suture technique. We experimentally quantify the axial and
OPEN ACCESS
Citation: Huang Y, Furtmu¨ller GJ, Tong D, Zhu S,
Lee WPA, et al. (2014) MEMS-Based Handheld
Fourier Domain Doppler Optical Coherence
Tomography for Intraoperative Microvascular
Anastomosis Imaging. PLoS ONE 9(12): e114215.
doi:10.1371/journal.pone.0114215
Editor: Tom Waigh, University of Manchester,
United Kingdom
Received: July 9, 2014
Accepted: November 4, 2014
Published: December 4, 2014
Copyright: ß2014 Huang et al. This is an open-
access article distributed under the terms of the
Creative Commons Attribution License, which
permits unrestricted use, distribution, and repro-
duction in any medium, provided the original author
and source are credited.
Data Availability: The authors confirm that all data
underlying the findings are fully available without
restriction. All relevant data are within the paper
and its Supporting Information files.
Funding: The research was supported by NIH/NEI
5R01EY021540-03. Yong Huang acknowledges
China Scholarship Council (CSC) for partial
support for his PhD study. The funders had no role
in study design, data collection and analysis,
decision to publish, or preparation of the manu-
script.
Competing Interests: The authors have declared
that no competing interests exist.
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 1/14
lateral resolution of the OCT to be 12.6 mm in air and 17.5 mm respectively. The
OCT has a sensitivity of 84 dB and sensitivity roll-off of 5.7 dB/mm over an imaging
range of 5 mm. Imaging with a frame rate of 36 Hz for an image size of
1000(lateral)6512(axial) pixels using a 50,000 A-lines per second swept source
was achieved. Quantitative vessel lumen patency, lumen narrowing and thrombosis
analysis were performed based on acquired structure and Doppler images.
Conclusions: A miniature handheld OCT imager that can be used for
intraoperative evaluation of microvascular anastomosis was successfully
demonstrated.
Introduction
Vascular anastomosis–the surgical connection of two blood vessels–is a common
procedure used in various surgical subspecialties. In particular, it is a critical
constituent of reconstructive microsurgery, vascular surgery, and transplant
surgery. The conventional suture-based anastomosis technique has seen many
technological advances since the introduction of the triangulation method by
Carrel [1]. The advances range from improved surgical microscopes to improved
suture material and vascular coupling devices [2], thermo-reversible poloxamers
[3], and non-suture cuff techniques [4]. However, quality assessment of surgical
outcomes depends merely on the experience of the surgeon and involves visual
inspection of the suture site as well as all tissues involved and perfused by the
vascular pedicle. Thus, despite the overall progress pushing the limits of
microsurgical possibilities, no significant clinically relevant advances have been
made allowing for the standardized postoperative evaluation of microvascular
anastomosis patency. In the field of imaging, technologies such as near-infrared
laser angiography may provide surrogate information about perfusion and
drainage by measuring the accumulation of fluorescent dyes, however an
application providing real time surgical guidance and the opportunity of
immediate postoperative microvascular anastomosis evaluation is deemed to be
the ultimate goal. In the era of super-microsurgery, there is a critical need for
devices that can evaluate the surgical outcome objectively and not solely based on
a surgeons’ subjective evaluation and accumulated experience over the years. Such
devices can also aid in training and evaluation of trainees.
Optical coherence tomography (OCT) has been successfully used in pre-clinical
and clinical applications due to its high resolution simultaneous structure and
Doppler flow imaging [5]. OCT has been demonstrated with the advantages of
fast temporal and spatial resolution compared to other medical imaging
modalities such as MRI, CT and ultrasound tomography. Our lab recently
demonstrated the capability of a real-time graphics processing units accelerated
Phase resolved Doppler OCT (PRDOCT) system for suturing guidance and
postoperative evaluation of microvascular anastomosis [6,7]. Compared to laser
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 2/14
speckle contrast blood flow imaging, PRDOCT has the advantage of simultaneous
high resolution depth-resolved structure and blood flow information [8]. While
Doppler optical microangiography imaging (DOMAG) can provide better
performance than PRDOCT, it requires more complicated system hardware
configuration [9]. Intraoperative application of OCT requires a task-oriented
OCT system optomechanical design to achieve bench-top to bed-side clinical
translation. Different approaches have been explored by researchers such as
integration of OCT to surgical microscope for vitreoretinal surgery [10]and
miniature handheld probes via fiber tip resonant vibration [5,11], scanning
mirrors combined with GRIN needles [12], endoscopic MEMS probe [13], and
MEMS mirrors-based handheld imaging instrument [14,15]. 2D MEMS mirrors
that scan in two axes as an alternative to the larger galvanometer scanner have
been used extensively to generate 3D volume images due to their low-cost,
compact size, and fast imaging speed [15]. MEMS mirror scanning is
conventional linear raster-scanning, which is uniform in spatial sampling and the
acquired image is more intuitive to understand compared to fiber tip resonant
vibration in 3D volume image generation mode. Due to the easy access of vessels
during the open anastomosis surgery, we chose to develop a handheld imager
using a 2D MEMS mirror-based scanner. The ergonomic design aim was that
surgeons can use the imager as a typical handheld surgical tool. By making the
OCT system portable and the probe miniature, the structure and Doppler OCT
imaging can be readily available for surgeons during the procedure.
In this work, we evaluated handheld OCT imager by performing the
intraoperative evaluation of microvascular anastomosis using a mouse model. The
proposed handheld MEMS scanning mirror-based imager is compact, light-
weight, and easy to manipulate. It can provide real-time simultaneous structure
and Doppler flow imaging at a speed of 36 frames per second while each frame
consists of 10006512 (lateral6axial) pixels. The Doppler flow imaging range was
characterized to be from 0.363 mm/s to 16.3 mm/s in both directions parallel to
the scanning beam.
Materials and Methods
Ethnic Statement
Johns Hopkins University Animal Care and Use Committee approved all
experiments with the approval number of MO12M348 and the animals were
handled in accordance with the Johns Hopkins University Animal Care and Use
Committee guidelines. Animals were housed under standard rodent housing
conditions (24C, 12:12 h dark/light cycle), food and water was provided ad
libitum. Analgesia with Buprenorphine (Dose: 0.05 mg/kg) was administered 1
hour prior to surgery and then at an interval of 60 min thereafter. Animals were
operated under sterile conditions and put on a heating pad to maintain body
temperature and all efforts were made to minimize suffering. Non-invasive
imaging procedures were performed immediately after the surgical procedure was
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 3/14
completed. At the end of the experiments animals were sedated with an overdose
of isoflurane inhalation anesthesia and euthanized by cervical dislocation.
Animals
Male 6–8 week old BALB/C mice with a BW of 33¡2.8 g (n52) were obtained
from Jackson Laboratories. Animals were anesthetized using oxygen and
isoflurane mixed gas and placed on a heating pad to maintain the body
temperature. In Mouse 1, femoral artery end-to-end anastomoses was performed
with a conventional suture technique. Mouse 2 received an orthotopic hind limb
transplantation surgery with a non-suture cuff-based technique for revascular-
ization. Both femoral vein and artery were anastomosed from the donor limb to
the recipient animal’s vessels.
Microvascular Anastomosis Procedures
Conventional interrupted suture technique: Supermicrovascular anastomosis was
performed in a mouse femoral artery cut and repair model using interrupted
suture technique (11-0 Nylon, S&T, Neuhausen, Switzerland). Femoral vessels
were dissected, cut and the proximal and distal stump mounted on an
approximating clamp (S&T, Neuhausen, Switzerland) to prevent blood flow and
provide optimal surgical exposure. Fig. 1(a) illustrates the procedures of suture
based vascular anastomosis. Sutures are placed in the collapsed vessel across the
full thickness of the vessel wall (adventitia, media and intima) on one end and
then from intima to adventitia in the other vessel end. Appropriate approximation
of proximal and distal vessel stumps was achieved with 6–8 sutures.
Non-suture cuff technique: Our group pioneered the non-suture-cuff
technique, which provides a rapid and reliable method for performing sub-
millimeter anastomosis [4,16]. Fig. 1(b) shows the procedures of the sutureless
cuff technique. First, a biocompatible cuff with proper diameter to fit the size of
vessel is chosen. The cuff is mounted on the distal vessel end, which is
subsequently everted and secured over the cuff using 10-0 circular ties. The distal
vessel end with the cuff is now secured in an approximating clamp to provide
optimal stability and surgical exposure. The proximal vessel is pulled over the
vessel-cuff complex providing intima-to-intima contact. Release of the approx-
imating clamp reestablished blood flow and patency could be evaluated
macroscopically.
Conventional rubber material as a background, is placed to both facilitate the
microsurgical procedure and provide optimal separation from surrounding tissue
for optimal and clear imaging conditions.
Handheld-Probe Design
Fig. 2(a) illustrated the optomechanical design of the probe. It consists of mainly
three parts: gray MEMS housing part with optical and electrical connectors
integrated to it, black lens tube part and a gray axial distance adjustment tip. To
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 4/14
Figure 1. Illustration of microvascular anastomosis procedures. (a) suture based vascular anastomosis;
(b) cuff based non-suture technique. Cuff is mounted over the vessel end and the end is everted over the cuff.
The other cut end of the vessel is then pulled over and tied to create intima-to-intima contact.
doi:10.1371/journal.pone.0114215.g001
Figure 2. Handheld probe illustration. (a) Probe compared to a quarter coin (H: height 37.24 mm, W: width
32.43 mm, L: length 136.7 mm). (b) Optics layout inside the probe. (c) Image of the probe held by hand.
doi:10.1371/journal.pone.0114215.g002
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 5/14
achieve light weight, the probe is made of aluminum and the weight of the probe
is measured to be 88 g. An FC/APC collimator with an output beam diameter of
1.5 mm was used as the optical input of the probe. MEMS scanning mirror
(AdvancedMEMS Inc.) with a diameter of 2.4 mm was chosen to accommodate
and stir the beam to perform scanning. MEMS scanning waveforms were supplied
through the electrical input connector located besides the optical input. A folding
mirror was used to fold the beam to a five half inch lens system, which is shown in
Fig. 2(b). Four half inches achromatic doublets (Thorlabs, AC127-050-C) form a
quasi-telecentric setup with two pairs of identical achromatic lenses to minimize
spherical aberration. One half inch achromatic lens with focal length of 30 mm
(Thorlabs, AC127-030-C) is used as the objective lens to focus the imaging beam.
The first pair of two lenses was placed within the gray MEMS housing part and the
remaining three lenses were placed in the lens tube. The length of the probe is
136.7 mm with 17.2 mm axial distance adjustment range. The axial distance
adjustment tip was designed for fine-tuning the sample position relative to the
zero-delay reference. It also gives us some flexibility when later longer focal length
lens is used. The height of the probe is 37 mm and width is 32 mm at the MEMS
housing and collimator region; after that the probe is a tube structure with an
outer diameter of 23 mm. A photograph showing a normal hand holding the
probe is shown in Fig. 2(c).
Swept Source-OCT System
Fig. 3(a) presents the system configuration. The swept source laser (Axsun
Technologies, Inc.) with a central wavelength of 1310 nm, tuning range of 100 nm
(from 1260 nm to 1360 nm), and A-scan rate of 50 kHz was used. Balanced
detector (Thorlabs, PDB480C AC) is utilized to detect the interference signal to
remove the common DC background. We used a quad-core@3.0 GHz Dell
Precision T7600 workstation to host a digitizer (AlazarTech, ATS9350), a DAQ
card (National Instrument, PCI-6221), one graphics processing unit (GPU)
(NVIDIA, Geforce GTX580). A-line trigger and k-clock signal from the laser
source are routed to the specific ports of the digitizer for equal k-space A-line data
sampling. The DAQ card generates MEMS control waveforms and synchronized
digital output as frame trigger to the digitizer for the acquisition start of a frame.
GPU is used to achieve real-time signal processing and image display. All the
scanning control and synchronization, data acquisition, image processing, and
rendering were performed on a multi-thread, CPU-GPU heterogeneous
computing system. The hybrid computing system and a customized user interface
was designed and programmed through Qt 4.8.5 and C++ (Microsoft Visual
Studio, 2008). We used computer unified device architecture (CUDA) version 5.0
from NVIDIA to program the GPU for general purpose computations.
Fig. 3(b) shows the carefully designed control waveforms generated by the DAQ
to the MEMS driver and digitizer for C-mode raster scanning imaging. An
internal clock with a sampling rate of 50 kHz was used as the clock source for the
fast-moving X and slow-moving Y waveforms. Each waveform consisted of
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 6/14
252 sections while the first and the last section were used to put and return the
MEMS mirror at a global starting point in the Y direction. Each section consists of
1400 sampling points. Fast-moving X waveform is quasi-sinusoidal with the first
and last 200 points of the section serving as the lead-in and lead-out points to put
the mirror from the zero point to acquisition start point and guide the mirror
back from the acquisition end point to the zero point in X direction. A
synchronized digital output from the DAQ with a rising edge as a frame trigger
was routed to the digitizer frame start trigger input, shown as green lines in
Figure 3. Schematic system setup and MEMS mirror control waveforms. (a) system configuration (C:
circulator, PC: polarization controller, DAQ: data acquisition card.) (b) filtered and scaled waveforms
generated by DAQ to MEMS driver to control the scanning mirror.
doi:10.1371/journal.pone.0114215.g003
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 7/14
Fig. 3(b). Once the trigger is detected, the system acquires 1000 A-lines in the
linear region, which is shown as the red dots for X imaged points and yellow dots
for Y imaged points. Each A-line consists of 1024 sampling points in the spectral
domain. The waveforms are updated periodically and iteratively through these
252 sections once a section is generated. For B-mode imaging, we simply set the Y
output to be zero all the time. Scan patterns can be switched at any time with a
delay of a few tens of milliseconds from B-mode to C-mode. Since the signal
generation source clock is 50 kHz and each section consists of 1400 points, this
gives us an effective imaging speed of 36 frames per second. It needs to be pointed
out that the imaging frame rate can be further increased by reducing the designed
imaged points or reducing the linearity of the scanning region by allowing the
MEMS mirror to perform more resonant scanning.
Imaging Protocol and Data Quantification Analysis
The OCT system was running at 36 fps with each frame size of 10006512 pixels,
corresponds to 1.5 mm (fast X)65 mm (axial) at 2D scanning mode. The 3D
volume scanning mode consists of 252 frames in the slow Y axis, which took
around 7 seconds to complete. Thus 252610006512 (1.561.565mm
3
) voxels
is acquired to cover the area of interest. Postoperative manual segmentation and
image analysis was through ImageJ 1.46r (NIH). Inner lumen area, lumen patency
rate, lumen narrowing rate and thrombosis percentage were extracted.
Results and Discussion
After assembling the probe and setting up the system, we carefully evaluated the
probe performance experimentally. Prior to the evaluation we carefully matched
the dispersion of the sample and reference through hardware first. Residual
second-order dispersion matching was achieved through our real-time numerical
dispersion compensation method [12]. We first analyzed axial resolution of the
system. Fig. 4(a) shows the measured results over the imaging range. An average
axial resolution of 12.6 mmis achieved. We evaluated the lateral resolution of our
system by imaging a USAF-1951 resolution target; the result is shown in Fig. 4(a)
inset. We can resolve up to the Group 5 Element 6, which implies a lateral
resolution of 17.5 mm. This is very close to the theoretically predicted value of
16 mmusing the formula 0.61fl/D assuming a focal length of objective lens (f)
30 mm and beam diameter (D) of 1.5 mm, where lis the central wavelength of
light source (1.31 mm). The system sensitivity was measured to be 84 dB by
putting a mirror as the sample at 0.4 mm from the zero-delay line and a neutral
density filter in front of the mirror. The system sensitivity roll-off is shown in
Fig. 4(b); the average decay is 5.7 dB/mm. We compared these values to a
standard galvanometer-based scanning system using an LSM04 3X OCT scan lens,
which has a sensitivity of 98 dB and a 4.7 dB/mm sensitivity roll-off. Reduced
sensitivity is probably due to the increased signal loss in the current MEMS probe
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 8/14
compared to a scan lens-based system. Larger sensitivity decay is due to the
reduced depth of focus of the current lens that has a focal length of 30 mm
compared to 54 mm of LSM04. Therefore further optical system optimization of
the probe through customizing lens design and coating will be achieved.
To evaluate the phase stability and thus the Doppler imaging sensitivity of the
imager, we imaged a stationary mirror and extracted the phase value of the mirror
surface position. A phase histogram is plotted in Fig. 4(c). By Gaussian fitting the
experimental data, the standard deviation of the fitting curve gave us a standard
deviation of 70 mrad, which is due to the combination of the system noise, MEMS
mirror-scanning noise, and the environmental noise. The Doppler flow imaging
speed range thus is calculated to be from 0.363 mm/s to 16.3 mm/s in both
directions parallel to the scanning beam using v(axical)5lDQ/4pDt), where Dt
equals 20 ms.
To evaluate the structure and Doppler imaging capability of our system before
animal experiments, we imaged an ex-vivo cow eye cornea shown in Fig. 4(d); all
layers are clearly visualized. We also imaged an in-vivo chorioallantoic membrane
(CAM) of a 16-day-aged chick embryo. Fig. 4(e) shows the structure image of one
vessel cross-section. Fig. 4(f) shows the corresponding Doppler image. We can see
the nearly parabolic distribution of the blood flow within the vessel lumen.
Figure 4. OCT System performance evaluation. (a) measured axial resolution over the imaging range,
inset: lateral resolution image of USAF target; (b) sensitivity roll-off over the imaging range; (c) measured
phase stability of a stationary mirror. (d) 5 frames averaged cornea imaging. (e) 5 frames averaged chicken
embryo imaging. (f) 5 frames averaged corresponding Doppler flow imaging of (e) (scale bar: 500 mm, Doppler
color bar range: 216.3 mm/s to 16.3 mm/s).
doi:10.1371/journal.pone.0114215.g004
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 9/14
Consecutively, the application of the handheld OCT imager for in-vivo mouse
microvascular anastomosis evaluation is tested. To describe how the probe was
used in animal experiment, Fig. 5(a) shows the probe held by the surgeon to
investigate the experimented mouse vessel. Fig. 5(b) demonstrates how the OCT
beam scans across the surgical site with a vessel groove adapted to the vessel. Note
that the vessel groove was not implemented in current probe version and will be
incorporated into our future version. The slow scanning direction indicated by
white arrow repeats every 252 frames to acquire volumetric images of the surgical
site. Fig. 6 shows selective frames of simultaneous structure and Doppler imaging
of the surgical sites from distal (left) to proximal (right) end of the blood vessel
under investigation: Fig. 6(a) (Mouse 1 Artery Cuff, Video S1), Fig. 6(b) (Mouse 1
Vein Cuff, Video S2) and Fig. 6(c) (Mouse 2 Artery Suture, Video S3). Successful
blood flow restoration, high vessel lumen patency, minimal lumen narrowing, and
thrombosis formation are important parameters for the surgeon to predict
successful surgical outcome. To analyze these parameters, all the selected frames
were manually segmented and highlighted using ImageJ 1.46r (NIH). The white
line encloses the inner lumen of the vessel; the yellow line outlines the blood
flowing area within the inner lumen if thrombosis was detected within the lumen.
The structure and Doppler images were cropped to a size of 1.5 mm (lateral) by
2 mm (axial). From Doppler images of Fig. 6, we can see clearly that all the blood
flow was restored successfully. Strong pulsatile artery flow was clearly visualized.
As vessel lumen diameter narrows, flow turbulences were also captured at suture-
rich sites. Blood flowing area was segmented using Doppler images. For Fig. 6(a),
the area of the artery inner lumen goes from 0.21 mm
2
(left) to 0.16 mm
2
(middle) and 0.17 mm
2
(right). Using the left figure as naive reference–as there is
no surgical operation such as cuff and suture–the inner lumen narrowing can by
characterized by 23% decrease at middle image and 16% decrease at right image.
One probable reason for this inner lumen area drop is that as shown in Fig. 1(b)
Figure 5. Application of handheld probe in the mouse experiment. (a) Handheld probe investigating the
surgical site of an experimented mouse; (b) Illustration of how the beam scans across the vessel region (red
lines indicate that fast B-mode scanning beam, white arrow indicates slow beam scanning direction, yellow
circles marks vessel groove to be implemented in our future probe optimization).
doi:10.1371/journal.pone.0114215.g005
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 10 / 14
the top vessel end needs to be expanded and then attached to the bottom vessel
end, which will cause the inner lumen area increase for the top vessel. The lumen
patency was 100% at reference (left) and drops to 56% at the middle image, and
to 80% at the right image, which corresponds to 44% and 20% thrombosis
formation crossectionally, respectively. The same analysis was applied to Fig. 6(b).
Cuff-based vein inner lumen area are 0.46 mm
2
(left), 0.09 mm
2
(middle) and
0.11 mm
2
(right). The inner lumen narrowing are 80% at the middle and 76% at
the right image. This is a relatively large shrinkage in the vein. This might be an
indication for changing the cuff with a more appropriate diameter. The patency of
the left is 55% while 100% for the middle and right images. For suture-based
artery anastomosis, the inner lumen area is 0.18 mm
2
(left), 0.11 mm
2
(middle)
and 0.13 mm
2
(right), respectively. The inner lumen narrowing values are 39% at
the middle and 28% at the right. The patency was all 100% over the images. No
thrombosis was detected. Currently, there is no detailed study of correlating inner
lumen area, restenosis, and thrombosis formation percentage with long-term
Figure 6. Selective frames of simultaneous structure and Doppler imaging of the surgical sites from distal (left) to proximal (right). White lines
represent manually segmented blood vessel inner lumen, red lines represent manually segmented thrombosis, blue lines outline the blood flowing area
within the vessel inner lumen. (a) Mouse 1 artery (Video S1); (b) Mouse 1 vein (Video S2) and (c) Mouse 2 artery (Video S3). Videos were played back at 24
frames per second. Images were cropped to best fit the area of interest (scale bar: 500 mm, Doppler color bar range: 216.3 mm/s to 16.3 mm/s).
doi:10.1371/journal.pone.0114215.g006
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 11 / 1 4
surgical outcome; however, we believe that all these parameters are not only of
vital importance for immediate surgical outcome evaluation but also for long-
term anastomosis patency prediction. With detailed quantitative image analysis,
handheld OCT intraoperative imaging might also help the surgeons to optimize
their technique.
Compared to our previous imaging results using 850 nm band OCT systems,
1.3 mm central wavelength shows larger image penetration depth, which enables
us to evaluate the inner lumen structure of vessels more accurately. At current
study, the vessel inner diameter is around 0.5 mm. For larger size vessels (0.5–
1 mm), when penetration depth becomes an issue for imaging, an angular
compounding method will be used and evaluated. Currently detailed quantitative
blood flow information analysis is limited by the phase wrapping of blood flow at
high speed combined with turbulence. Proper correction methods need to be
implemented in the future to resolve this issue. Noticeably from the video images
there is hand tremor imposed on vessel imaging. A special vessel adapter, such as
V-groove on the probe tip will be fabricated so that the probe can rest on the
mouse limb without pressing the vessel. Furthermore, three-dimensional motion-
compensation algorithms will be developed and applied to remove motion
artifacts. Thus, more comprehensive volumetric information of the vessel under
investigation can be extracted.
Conclusion
We have demonstrated a compact user-friendly handheld MEMS scanning
mirror-based imager for intraoperative evaluation of microvascular anastomosis
and the proof was tested using a mouse model. It can provide real-time
simultaneous structure and Doppler flow imaging at a speed of 36 frames per
second. The imager performance was carefully characterized experimentally and
later evaluated through in-vivo mouse femoral vessel anastomosis via both suture
and non-suture cuff techniques. Parameters such as vessel inner lumen area,
lumen narrowing, and partial thrombosis formation were extracted quantitatively
from the imaging results. In conclusion, we believe that the intraoperative
application of this handheld OCT imager will greatly improve the quality of
microvascular repair and vascular patency and thereby significantly increase long
term results after microsurgical vascular repair. Additionally this application
allows for real-time performance evaluation and thus provides an optimal
platform for in situ performance evaluations in microsurgical trainees.
Supporting Information
Checklist S1. ARRIVE checklist.
doi:10.1371/journal.pone.0114215.s001 (DOCX)
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 12 / 14
Video S1. Crossectional inspection of Mouse 1 Artery (using cuff technique)
over the surgical site.
doi:10.1371/journal.pone.0114215.s002 (MP4)
Video S2. Crossectional inspection of Mouse 1 Vein (using cuff technique) over
the surgical site.
doi:10.1371/journal.pone.0114215.s003 (MP4)
Video S3. Crossectional inspection of Mouse 2 Artery (using suture technique)
over the surgical site.
doi:10.1371/journal.pone.0114215.s004 (MP4)
Author Contributions
Conceived and designed the experiments: YH DT SZ WPAL GB JUK. Performed
the experiments: YH GJF. Analyzed the data: YH DT SZ GJF GB JUK.
Contributed reagents/materials/analysis tools: YH GJF. Contributed to the writing
of the manuscript: YH GJF DT SZ WPAL GB JUK.
References
1. Rothwell A (2011) Alexis Carrel: innovator extraordinaire. J Perioper Pract 21: 73–76.
2. Ahn CY, Shaw WW, Berns S, Markowitz BL (1994) Clinical experience with the 3M microvascular
coupling anastomotic device in 100 free-tissue transfers. Plast Reconstr Surg 93: 1481–1484.
3. Chang EI, Galvez MG, Glotzbach JP, Hamou CD, El-Ftesi S, et al. (2011) Vascular anastomosis using
controlled phase transitions in poloxamer gels. Nature Med 17: 1147–1152.
4. Sucher R, Lin CH, Zanoun R, Atsina KK, Weinstock M, et al. (2010) Mouse hind limb transplantation:
a new composite tissue allotransplantation model using nonsuture supermicrosurgery. Transplantation
90(12): 1374–1380.
5. Sun C, Lee KKC, Vuong B, Cusimano MD, Brukson A, et al. (2012) Intraoperative handheld optical
coherence tomography forward-viewing probe: physical performance and preliminary animal imaging.
Biomed Opt Express 3: 1404–1412.
6. Huang Y, Liu X, Kang JU (2012) Real-time 3D and 4D Fourier domain Doppler optical coherence
tomography based on dual graphics processing units. Biomed Opt Express 3: 2162–2174.
7. Huang Y, Ibrahim Z, Tong D, Zhu S, Mao Q, et al. (2013) Microvascular anastomosis guidance and
evaluation using real-time three-dimensional Fourier-domain Doppler optical coherence tomography.
J Biomed Opt 18: 111404.
8. Boas DA, Dunn AK (2010) Laser speckle contrast imaging in biomedical optics. J Biomed Opt 15:
011109.
9. Wang RK, Lin A (2009) Doppler optical micro-angiography for volumetric imaging of vascular perfusion
in vivo. Opt Express 17: 8926–8940.
10. Tao YK, Ehlers JP, Toth CA, Izatt JA (2010) Intraoperative spectral domain optical coherence
tomography for vitreoretinal surgery. Opt Lett 35: 3315–3317.
11. Huo L, Xi JF, Wu YC, Li XD (2010) Forward-viewing resonant fiber-optic scanning endoscope of
appropriate scanning speed for 3D OCT imaging. Opt Express 18: 14375–14384.
12. Liang CP, Wierwille J, Moreira T, Schwartzbauer G, Jafri MS, et al. (2011) A forward-imaging needle-
type OCT probe for image guided stereotactic procedures. Opt Express 19: 26283–26394.
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 13 / 14
13. Wang DL, Fu L, Wang X, Gong Z, Samuelson S, et al. (2013) Endoscopic swept-source optical
coherence tomography based on a two-axis microelectromechanical system mirror. J Biomed Opt 18:
086005.
14. Lu CD, Kraus MF, Potsaid B, Liu JJ, Choi W, et al. (2014) Handheld ultrahigh speed swept source
optical coherence tomography instrument using a MEMS scanning mirror. Biomed Opt Express 5: 294–
311.
15. Jung W, Kim J, Jeon M, Chaney EJ, Stewart CN, et al. (2011) Handheld optical coherence
tomography scanner for primary care diagnostics. IEEE Trans Biomed Eng 58: 741–744.
16. Martins PN, Theruvath TP (2007) Detachable cuffs for sutureless microvascular anastomosis.
Microsurgery 27: 678–679.
Handheld OCT Imager for Vascular Anastomosis
PLOS ONE | DOI:10.1371/journal.pone.0114215 December 4, 2014 14 / 14
... With spectroscopic information, photothermal effect, or pump-probe mechanism, molecular contrast can be achieved with OCT [85], providing the potential for molecular OCT imaging of mouse embryonic hearts. Innovative designs and implementations of OCT probes [86,87] can potentially inspire new strategies of integrating the OCT imaging head with the embryo culturing setup, which might lead to an improved experimental environment and condition for live mouse embryonic imaging. Image processing methods to extract The major feature of OCT angiography in mouse embryonic imaging is endogenous contrast. ...
... With spectroscopic information, photothermal effect, or pump-probe mechanism, molecular contrast can be achieved with OCT [85], providing the potential for molecular OCT imaging of mouse embryonic hearts. Innovative designs and implementations of OCT probes [86,87] can potentially inspire new strategies of integrating the OCT imaging head with the embryo culturing setup, which might lead to an improved experimental environment and condition for live mouse embryonic imaging. Image processing methods to extract specific features are increasingly demanded for the assessment of rich information provided by OCT, and to this end, deep-learning-enabled algorithms developed for OCT embryonic heart images [88,89] could have a profound impact on increasing the efficiency of analyzing the OCT cardiodynamic data. ...
Article
Full-text available
The embryonic heart is an active and developing organ. Genetic studies in mouse models have generated great insight into normal heart development and congenital heart defects, and suggest mechanical forces such as heart contraction and blood flow to be implicated in cardiogenesis and disease. To explore this relationship and investigate the interplay between biomechanical forces and cardiac development, live dynamic cardiac imaging is essential. Cardiodynamic imaging with optical coherence tomography (OCT) is proving to be a unique approach to functional analysis of the embryonic mouse heart. Its compatibility with live culture systems, reagent-free contrast, cellular level resolution, and millimeter scale imaging depth make it capable of imaging the heart volumetrically and providing spatially resolved information on heart wall dynamics and blood flow. Here, we review the progress made in mouse embryonic cardiodynamic imaging with OCT, highlighting leaps in technology to overcome limitations in resolution and acquisition speed. We describe state-of-the-art functional OCT methods such as Doppler OCT and OCT angiography for blood flow imaging and quantification in the beating heart. As OCT is a continuously developing technology, we provide insight into the future developments of this area, toward the investigation of normal cardiogenesis and congenital heart defects.
... The transverse and axial resolutions of the DOCT system were 25 and 12 µm (on the air), respectively. The A-scan rate was equal to 16 kHz, which allowed us to measure absolute velocities up to ∼5.5 mm/s [100,101]. A 30 min OCT recording of CBF (a.u.) was performed on the same healthy rats (n = 10) before, during and after one-time LT exposure 12.6 kJ/cm 2 and expressed as an average of the sum of the CBF values. ...
Article
Full-text available
Over sixty years, laser technologies have undergone a technological revolution and become one of the main tools in biomedicine, particularly in neuroscience, neurodegenerative diseases and brain tumors. Glioblastoma is the most lethal form of brain cancer, with very limited treatment options and a poor prognosis. In this study on rats, we demonstrate that glioblastoma (GBM) growth can be suppressed by photosensitizer-free laser treatment (PS-free-LT) using a quantum-dot-based 1267 nm laser diode. This wavelength, highly absorbed by oxygen, is capable of turning triplet oxygen to singlet form. Applying 1267 nm laser irradiation for a 4 week course with a total dose of 12.7 kJ/cm2 firmly suppresses GBM growth and increases survival rate from 34% to 64%, presumably via LT-activated apoptosis, inhibition of the proliferation of tumor cells, a reduction in intracranial pressure and stimulation of the lymphatic drainage and clearing functions. PS-free-LT is a promising breakthrough technology in non- or minimally invasive therapy for superficial GBMs in infants as well as in adult patients with high photosensitivity or an allergic reaction to PSs.
... Transverse and axial resolutions of the DOCT system were 25 and 12 µm (on the air), respectively. A-scan rate was equal to 16 kHz, which allowed us to measure absolute velocities up to ∼5.5 mm/s [97,98]. The 30-min OCT recording of CBF (a.u.) was performed on the same healthy rats (n=10) before, during and after LT-onetime exposure 12.6 kJ/cm 2 and expressed as an average of the sum of the CBF values. ...
Preprint
Full-text available
For sixty years, laser technologies have undergone a technological revolution and become one of the main tools in biomedicine, particularly in neuroscience, neurodegenerative diseases and brain tumours. Glioblastoma is the most lethal form of brain cancer with very limited treatment options and a poor prognosis. In this study on rats we demonstrate that the glioblastoma (GBM) growth can be suppressed by photosensitiser-free laser-treatment (PS-free-LT) using a quantum-dot based 1267 nm laser diode. This wavelength, highly absorbed by oxygen, is capable of turning triplet oxygen to singlet form. Applying the 1267nm laser irradiation for a 4 week course with a total dose of 12.7 kJ/cm ² firmly suppresses GBM growth and increases the survival rate from 34% up to 64%, presumably via the LT-activated apoptosis, inhibition of proliferation of the tumour cells, reduction of the intracranial pressure and stimulation of the lymphatic drainage and clearing functions. The PS-free-LT is promising breakthrough technology when developed and clinically approved becomes a successful non- or minimally-invasive therapy for superficial GBMs in infants as well as in adult patients with high photosensitive or an allergic reaction to PSs.
... This flexibility opens up the possibility to image the retina of immobile/bedridden patients or infants, skin areas that are not accessible easily, ear-nose-throat measurements, or even large animals [31] and plants [32] can be imaged. In the last decade, handheld probes have proven their value in imaging the retina [33][34][35][36][37][38][39][40][41][42][43][44][45][46], hair follicles [47] and the tympanic membrane in the ear [30,48,49], to detect and monitor skin cancer [50,51] or skin and mucosal lesions [52][53][54][55][56][57][58], and as an imaging tool to assist in surgery [59,60]. ...
Article
Full-text available
Optical coherence tomography (OCT) has revolutionized ophthalmic diagnosis as a non‐invasive, cross‐sectional imaging technique in the last 30 years and hence is one of the fastest adopted advanced imaging technologies in the history of medicine. A miniaturization of OCT devices would not only reduce size but ideally also reduce costs and therefore create potential new markets. OCT systems based on photonic integrated circuits (PIC) could enable a significant miniaturization of complex systems with high degree of integration as well as low costs of goods. These therefore have a potential to enable portable, cost‐effective, high performing, real handheld OCT devices. This review identifies three main categories towards miniaturized OCT devices: Handheld imaging probes interfaced to a (mobile) base station, compact home/self‐ OCT and PIC‐based OCT. Imaging performance parameters and technical readiness levels of the identified miniaturized OCT systems for (non‐) ophthalmic applications are presented. Special attention is paid to PIC‐based OCT applications and their progress. Picture: First in vivo human retinal tomogram using a PIC‐based subcomponent (arrayed waveguide grating, AWG) of a spectral domain OCT system. Adapted with permission from E. A. Rank et al., Light Sci Appl. 2021, 10, 1, CC By 4.0. Copyright © 2021, The Author(s).
... For future different OCT applications, it is important to mention that OCT probe can be customized on demand and the size of the FOV can be optimized depending on the application. For example, clinical intraoperative vascular imaging requires miniature probe as was used in this manuscript due to tight space limitation [35]. For large-scale nondestructive material evaluation such as artwork inspection, an OCT probe with larger FOV is preferred. ...
Article
Full-text available
Optical coherence tomography (OCT) is a three-dimensional non-invasive high-resolution imaging modality that has been widely used for applications ranging from medical diagnosis to industrial inspection. Common OCT systems are equipped with limited field-of-view (FOV) in both the axial depth direction (a few millimeters) and lateral direction (a few centimeters), prohibiting their applications for samples with large and irregular surface profiles. Image stitching techniques exist but are often limited to at most 3 degrees-of-freedom (DOF) scanning. In this work, we propose a robotic-arm-assisted OCT system with 7 DOF for flexible large FOV 3D imaging. The system consists of a depth camera, a robotic arm and a miniature OCT probe with an integrated RGB camera. The depth camera is used to get the spatial information of targeted sample at large scale while the RGB camera is used to obtain the exact position of target to align the image probe. Eventually, the real-time 3D OCT imaging is used to resolve the relative pose of the probe to the sample and as a feedback for imaging pose optimization when necessary. Flexible probe pose manipulation is enabled by the 7 DOF robotic arm. We demonstrate a prototype system and present experimental results with flexible tens of times enlarged FOV for plastic tube, phantom human finger, and letter stamps. It is expected that robotic-arm-assisted flexible large FOV OCT imaging will benefit a wide range of biomedical, industrial and other scientific applications.
... The phase stability of the OCT system is 70 mrad. The Doppler flow imaging speed range thus is calculated to be from 0.363 mm/s to 16.3 mm/s in both directions parallel to the scanning beam with adjacent A-scan time difference of 20 µs [36]. ...
Article
Full-text available
To solve the phase unwrapping problem for phase images in Fourier domain Doppler optical coherence tomography (DOCT), we propose a deep learning-based residual en-decoder network (REDN) method. In our approach, we reformulate the definition for obtaining the true phase as obtaining an integer multiple of 2π at each pixel by semantic segmentation. The proposed REDN architecture can provide recognition performance with pixel-level accuracy. To address the lack of phase images that are noise and wrapping free from DOCT systems for training, we used simulated images synthesized with DOCT phase image background noise features. An evaluation study on simulated images, DOCT phase images of phantom milk flowing in a plastic tube and a mouse artery, was performed. Meanwhile, a comparison study with recently proposed deep learning-based DeepLabV3+ and PhaseNet methods for signal phase unwrapping and traditional modified networking programming (MNP) method was also performed. Both visual inspection and quantitative metrical evaluation based on accuracy, specificity, sensitivity, root-mean-square-error, total-variation, and processing time demonstrate the robustness, effectiveness and superiority of our method. The proposed REDN method will benefit accurate and fast DOCT phase image-based diagnosis and evaluation when the detected phase is wrapped and will enrich the deep learning-based image processing platform for DOCT images.
Article
Coins are used for size comparison in photographs. This interdisciplinary study compiles a comprehensive dataset of over one thousand instances where coins are utilized as size references in scientific publications. Employing quantitative analysis, a numismatic measurement system is uncovered that’s characterized by widespread adoption and discernible conventions governing coin use. The findings are presented in this article. The analysis reveals a prototypical numismatic measurement system with widespread use and distinguishable rules how coins get used. The prevalence of coins as a size reference is within the single-digit percentage range in several technical journals. Distributions examine the number of coins per article and explore correlations of coins by number of pages, figures and photos. Text analysis reveals the authors’ deliberate choice to use coins for scale and size comparison. Coins lack many aspects of modern measurement tools. They are not used as a measurement unit unlike the meter or the inch and they don’t provide a lower and upper bound on size. The dataset provides evidence that coins are used as size reference for objects with similar dimensions and are occasionally used in conjunction with other measurement tools. Coin properties other than their size are not widely used for measurement applications. A numismatic analysis provides statistics of used currencies, denominations and mint years. Analysis reveals a trend towards larger currencies when using foreign coins, but generally coins are chosen from the first author’s country. Similar items like banknotes, credit cards or jewelery are not used. The investigation extends to the arrangement in scenes, whether they are placed next to objects or on them, the displayed sides and whether they are standing or lying horizontally. The study concludes by discussing notable outliers within the dataset.
Article
Full-text available
Significance: After three decades, more than 75,000 publications, tens of companies being involved in its commercialization, and a global market perspective of about USD 1.5 billion in 2023, optical coherence tomography (OCT) has become one of the fastest successfully translated imaging techniques with substantial clinical and economic impacts and acceptance. Aim: Our perspective focuses on disruptive forward-looking innovations and key technologies to further boost OCT performance and therefore enable significantly enhanced medical diagnosis. Approach: A comprehensive review of state-of-the-art accomplishments in OCT has been performed. Results: The most disruptive future OCT innovations include imaging resolution and speed (single-beam raster scanning versus parallelization) improvement, new implementations for dual modality or even multimodality systems, and using endogenous or exogenous contrast in these hybrid OCT systems targeting molecular and metabolic imaging. Aside from OCT angiography, no other functional or contrast enhancing OCT extension has accomplished comparable clinical and commercial impacts. Some more recently developed extensions, e.g., optical coherence elastography, dynamic contrast OCT, optoretinography, and artificial intelligence enhanced OCT are also considered with high potential for the future. In addition, OCT miniaturization for portable, compact, handheld, and/or cost-effective capsule-based OCT applications, home-OCT, and self-OCT systems based on micro-optic assemblies or photonic integrated circuits will revolutionize new applications and availability in the near future. Finally, clinical translation of OCT including medical device regulatory challenges will continue to be absolutely essential. Conclusions: With its exquisite non-invasive, micrometer resolution depth sectioning capability, OCT has especially revolutionized ophthalmic diagnosis and hence is the fastest adopted imaging technology in the history of ophthalmology. Nonetheless, OCT has not been completely exploited and has substantial growth potential-in academics as well as in industry. This applies not only to the ophthalmic application field, but also especially to the original motivation of OCT to enable optical biopsy, i.e., the in situ imaging of tissue microstructure with a resolution approaching that of histology but without the need for tissue excision.
Article
Full-text available
We developed an ultrahigh speed, handheld swept source optical coherence tomography (SS-OCT) ophthalmic instrument using a 2D MEMS mirror. A vertical cavity surface-emitting laser (VCSEL) operating at 1060 nm center wavelength yielded a 350 kHz axial scan rate and 10 µm axial resolution in tissue. The long coherence length of the VCSEL enabled a 3.08 mm imaging range with minimal sensitivity roll-off in tissue. Two different designs with identical optical components were tested to evaluate handheld OCT ergonomics. An iris camera aided in alignment of the OCT beam through the pupil and a manual fixation light selected the imaging region on the retina. Volumetric and high definition scans were obtained from 5 undilated normal subjects. Volumetric OCT data was acquired by scanning the 2.4 mm diameter 2D MEMS mirror sinusoidally in the fast direction and linearly in the orthogonal slow direction. A second volumetric sinusoidal scan was obtained in the orthogonal direction and the two volumes were processed with a software algorithm to generate a merged motion-corrected volume. Motion-corrected standard 6 x 6 mm² and wide field 10 x 10 mm² volumetric OCT data were generated using two volumetric scans, each obtained in 1.4 seconds. High definition 10 mm and 6 mm B-scans were obtained by averaging and registering 25 B-scans obtained over the same position in 0.57 seconds. One of the advantages of volumetric OCT data is the generation of en face OCT images with arbitrary cross sectional B-scans registered to fundus features. This technology should enable screening applications to identify early retinal disease, before irreversible vision impairment or loss occurs. Handheld OCT technology also promises to enable applications in a wide range of settings outside of the traditional ophthalmology or optometry clinics including pediatrics, intraoperative, primary care, developing countries, and military medicine.
Article
Full-text available
A microelectromechanical system (MEMS) mirror based endoscopic swept-source optical coherence tomography (SS-OCT) system that can perform three-dimensional (3-D) imaging at high speed is reported. The key component enabling 3-D endoscopic imaging is a two-axis MEMS scanning mirror which has a 0.8×0.8 mm2 mirror plate and a 1.6×1.4 mm2 device footprint. The diameter of the endoscopic probe is only 3.5 mm. The imaging rate of the SS-OCT system is 50 frames/s. OCT images of both human suspicious oral leukoplakia tissue and normal buccal mucosa were taken in vivo and compared. The OCT imaging result agrees well with the histopathological analysis.
Article
Full-text available
Vascular and microvascular anastomoses are critical components of reconstructive microsurgery, vascular surgery, and transplant surgery. Intraoperative surgical guidance using a surgical imaging modality that provides an in-depth view and three-dimensional (3-D) imaging can potentially improve outcome following both conventional and innovative anastomosis techniques. Objective postoperative imaging of the anastomosed vessel can potentially improve the salvage rate when combined with other clinical assessment tools, such as capillary refill, temperature, blanching, and skin turgor. Compared to other contemporary postoperative monitoring modalities-computed tomography angiograms, magnetic resonance (MR) angiograms, and ultrasound Doppler-optical coherence tomography (OCT) is a noninvasive high-resolution (micron-level), high-speed, 3-D imaging modality that has been adopted widely in biomedical and clinical applications. For the first time, to the best of our knowledge, the feasibility of real-time 3-D phase-resolved Doppler OCT (PRDOCT) as an assisted intra- and postoperative imaging modality for microvascular anastomosis of rodent femoral vessels is demonstrated, which will provide new insights and a potential breakthrough to microvascular and supermicrovascular surgery.
Article
Full-text available
We present real-time 3D (2D cross-sectional image plus time) and 4D (3D volume plus time) phase-resolved Doppler OCT (PRDOCT) imaging based on configuration of dual graphics processing units (GPU). A GPU-accelerated phase-resolving processing algorithm was developed and implemented. We combined a structural image intensity-based thresholding mask and average window method to improve the signal-to-noise ratio of the Doppler phase image. A 2D simultaneous display of the structure and Doppler flow images was presented at a frame rate of 70 fps with an image size of 1000 × 1024 (X × Z) pixels. A 3D volume rendering of tissue structure and flow images—each with a size of 512 × 512 pixels—was presented 64.9 milliseconds after every volume scanning cycle with a volume size of 500 × 256 × 512 (X × Y × Z) voxels, with an acquisition time window of only 3.7 seconds. To the best of our knowledge, this is the first time that an online, simultaneous structure and Doppler flow volume visualization has been achieved. Maximum system processing speed was measured to be 249,000 A-scans per second with each A-scan size of 2048 pixels.
Article
Full-text available
A prototype intraoperative hand-held optical coherence tomography (OCT) imaging probe was developed to provide micron resolution cross-sectional images of subsurface tissue during open surgery. This new ergonomic probe was designed based on electrostatically driven optical fibers, and packaged into a catheter probe in the form factor of clinically accepted Bayonet shaped neurosurgical probes. Optical properties of the probe were measured to have a ~20 μm spot size, 5 mm working distance and 4 mm field of view. Feasibility of this probe for structural and Doppler shift imaging was tested on porcine femoral blood vessel imaging.
Article
Full-text available
A forward-imaging needle-type optical coherence tomography (OCT) probe with Doppler OCT (DOCT) capability has the potential to solve critical challenges in interventional procedures. A case in point is stereotactic neurosurgery where probes are advanced into the brain based on predetermined coordinates. Laceration of blood vessels in front of the advancing probe is an unavoidable complication with current methods. Moreover, cerebrospinal fluid (CSF) leakage during surgery can shift the brain rendering the predetermined coordinates unreliable. In order to address these challenges, we developed a forward-imaging OCT probe (740 μm O.D.) using a gradient-index (GRIN) rod lens that can provide real-time imaging feedback for avoiding at-risk vessels (8 frames/s with 1024 A-scans per frame for OCT/DOCT dual imaging) and guiding the instrument to specific targets with 12 μm axial resolution (100 frames/s with 160 A-scans per frame for OCT imaging only). The high signal-to-background characteristic of DOCT provides exceptional sensitivity in detecting and quantifying the blood flow within the sheep brain parenchyma in real time. The OCT/DOCT dual imaging also demonstrated its capability to differentiate the vessel type (artery/vein) on rat's femoral vessels. We also demonstrated in ex vivo human brain that the location of the tip of the OCT probe can be inferred from micro-anatomical landmarks in OCT images. These findings demonstrate the suitability of OCT guidance during stereotactic procedures in the brain and its potential for reducing the risk of cerebral hemorrhage.
Article
Full-text available
Vascular anastomosis is the cornerstone of vascular, cardiovascular and transplant surgery. Most anastomoses are performed with sutures, which are technically challenging and can lead to failure from intimal hyperplasia and foreign body reaction. Numerous alternatives to sutures have been proposed, but none has proven superior, particularly in small or atherosclerotic vessels. We have developed a new method of sutureless and atraumatic vascular anastomosis that uses US Food and Drug Administration (FDA)-approved thermoreversible tri-block polymers to temporarily maintain an open lumen for precise approximation with commercially available glues. We performed end-to-end anastomoses five times more rapidly than we performed hand-sewn controls, and vessels that were too small (<1.0 mm) to sew were successfully reconstructed with this sutureless approach. Imaging of reconstructed rat aorta confirmed equivalent patency, flow and burst strength, and histological analysis demonstrated decreased inflammation and fibrosis at up to 2 years after the procedure. This new technology has potential for improving efficiency and outcomes in the surgical treatment of cardiovascular disease.
Article
Alexis Carrel was awarded the 1912 Nobel Prize for Physiology or Medicine for pioneering surgical techniques in vascular surgery. His triangulation technique enabled successful anastomosis of blood vessels and subsequently led to advances in blood vessel and organ transplantation. However, Alexis Carrel did not limit his research to the vascular speciality. During the First World War he developed the Carrel-Dakin method of wound irrigation and later his interests lay in aging cells.
Article
The development of microsurgical techniques has facilitated the establishment of vascularized composite tissue transplant models in small mammals. Because the mouse would be the ideal model to study various composite tissue allotransplantation (CTA)-related problems, we designed two new surgical techniques for orthotopic (ORT) and heterotopic (HET) hind limb transplantation. BALB/c hind limbs were transplanted to BALB/c or C57BL6 recipients using a nonsuture cuff technique. ORT: donor femoral vessels were anastomosed to recipient femoral vessels, the sciatic nerve approximated end-to-end and osteosynthesis was performed using an intramedullary rod. HET/cervical: Donor femoral vessels of a reduced size osteomyocutaneous hind limb CTA were anastomosed to recipient common carotid artery and external jugular vein without nerve approximation. Both procedures could be performed with a high success rate (ORT: 62%; HET: 90%). Donor operation lasted for 100±12 min and recipient operation 114±27 min (ORT) and 54±16 min (HET). Complication rates in terms of bleeding, and thrombosis at the cuff side was slightly higher in the ORT group. All syngeneic grafts survived long term (>100 days). FK506 (2 mg/kg) significantly prolonged graft survival (87±22 days) when compared with untreated controls (6±1 day). Functional evaluation of ORT grafts by means of video gait kinematics and CatWalk analysis revealed specific differences of gait parameters when compared with nontransplanted controls (P<0.05). The ORT hind limb transplant model seems to be best suited to study functional outcome and nerve regeneration in CTA. The technically less demanding HET/cervical model may be used to investigate basic immunology and clinically relevant questions related to acute and chronic rejection, and ischemia reperfusion injury in reconstructive transplantation.