ArticlePDF Available

Wireless Medical Device Coexistence

Authors:

Figures

Content may be subject to copyright.
Published on Medical Electronics Design (http://www.medicalelectronicsdesign.com)
Home > Wireless Medical Device Coexistence
Wireless Medical Device Coexistence
Nickolas J. LaSorte, Hazem H. Refai, Donald M. Witters Jr., Seth J. Seidman, Jeffrey L. Silberberg
Created 2011-08-10 18:10
[1]
Wireless Medical Device Coexistence [2]
August 10, 2011
By: Nickolas J. LaSorte, Hazem H. Refai, Donald M. Witters Jr., Seth J. Seidman, Jeffrey
L. Silberberg
[3]
Find more content on: Wireless Technology [4]
Tests to assess the risks associated with coexistence of wireless technologies are
necessary for safe and effective RF wireless medical devices.
At any given time, a typical home or hospital uses a number of wireless systems (e.g.,
IEEE 802.11a/b/g/n, or WiFi; Bluetooth; ZigBee; cordless phones) operating on the same
industrial, scientific, and medical (ISM) band.1,2 Given the increasing use of wireless, RF
wireless medical devices and other wireless systems operating nearby can interfere with
each other. If a collision between their respective transmissions occurs, data packets
transmitted by medical devices could be delayed or blocked, potentially interfering with
timely transmissions of critical data. Techniques such as retransmission and forward error
correction might no longer be sufficient to overcome interference and spectrum
congestion. Hence, methods to design and test wirelessly enabled medical devices for
risks associated with coexistence of wireless technologies are essential for innovative,
safe, and effective RF wireless medical devices.
Although there is some overlap between electromagnetic compatibility (EMC) and wireless
coexistence, differences exist. Wireless coexistence is the ability of one wireless system
to perform a task in an environment where other systems that may or may not be using
the same set of rules can also perform their tasks.3 EMC is the ability of a device to
function properly in its intended electromagnetic environment without introducing
excessive electromagnetic energy that could interfere with other devices. Manufacturers of
electrically powered medical devices routinely test their equipment to applicable national
Share
Pa
g
e 1 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
and international consensus safety standards. EMC test results are often used to support
safety claims to regulatory agencies such as FDA. Less well-known are the issues and
concerns associated with wirelessly enabled medical devices, although this is changing
thanks to FDA’s guidance document on wireless medical devices.4
To date, no consensus standards adequately address the risks associated with wireless
coexistence for medical devices and systems. Current methods of evaluating wireless
coexistence use ad hoc test methods that vary widely among device manufacturers and
test facilities. Moreover, current medical device EMC standards have no requirements or
test procedures to assess the performance of systems containing RF receivers in the
presence of in-band transmitters. This article examines the limitations of present medical
device EMC standards for coexistence evaluation, identifies factors to be examined when
testing for coexistence, and discusses the status of plans to develop a wireless
coexistence test method.
Limitations of Medical Device EMC Standards
International Electrotechnical Commission (IEC) 60601-1-2— a collateral to IEC 60601-1,
the general safety standard for medical electrical equipment—is the primary standard
used for EMC testing of nonimplanted and nonin-vitro diagnostic electrical medical
devices.5,6 Even if the wireless technology is considered part of the essential performance
and function of the device, meaning its absence would result in unacceptable risk, the
testing and requirements in these standards and the related documents do not fully
address the characteristics and performance associated with the wireless technology.
One major limitation of the IEC 60601-1-2 standard is that RF receivers are exempt from
immunity testing in the exclusion band (passband). This exemption is granted because a
test signal in the passband of a traditional RF receiver would be expected to cause
interference. The exclusion band is defined in 3.10 of IEC 60601-1-2 as the:
…frequency band for intentional receivers of RF electromagnetic energy that
extends from -5 % to +5 % of the frequency, or frequency band, of reception
for frequencies of reception greater than or equal to 80 MHz and from -10 % to
+10 % of the frequency, or frequency band, of reception for frequencies of
reception less than 80 MHz”
For a wireless medical device system operating at 2.45 GHz, the exclusion band covers
the entire 2.4 GHz ISM band. In other words, a medical device with RF wireless
technology is not required by IEC 60601-1-2 to be able to maintain wireless
communication when subjected to RF signals in its passband. Therefore, as published,
this standard does not provide means for assessing the performance of the medical
device wireless communication system. A new draft of Edition 4 of IEC 60601-1-2 would
require that medical equipment remain safe (i.e., provide basic safety and essential
performance) when exposed to an RF immunity test signal in the passband but still would
not assess nonsafety performance under these conditions.7
No Standard RF Wireless Coexistence Test
Wireless coexistence can also be defined as the ability of multiple wireless systems to
share the same or adjacent frequency spectrum without undue interaction or interference
Pa
g
e 2 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
affecting performance and transmission or reception of signals and data. IEEE 802.15.2
discusses computer modeling and design issues for improving wireless coexistence
between IEEE 802.15 and IEEE 802.11 communication systems.8 The document does
not, however, specify a method or pass-fail criteria for testing coexistence, leaving a gap
in the ability to assess the characteristics and performance of wireless systems.
Due to concern for patient safety and medical device effectiveness, FDA drafted a
guidance document to assist in the design and testing of RF wireless technology in
medical devices.4 The document calls attention to general risks and concerns for devices
operating in the crowded RF spectrum—particularly in the ISM frequency bands. Other
risks, including those that might be affected by quality of service, data integrity, wireless
security, and EMC of wireless technologies and medical device functions, are also
outlined. In addition, the document provides information to assist in preparing regulatory
submissions. FDA remains active and has an interest in the development of standards
and evaluation methods for determining and validating the performance of RF wireless
technology in medical devices in general and wireless coexistence in particular.9
Considerations for RF Wireless Medical Devices and Wireless
Coexistence
The following are major factors influencing the coexistence of wireless medical devices
operating in the presence of other heterogeneous wireless networks, including radio
channel characteristics, antenna and signal polarization, frequency bands, separation
distances, and cochannel and adjacent-channel interference. These factors should be
considered when developing a standardized test protocol for RF wireless coexistence.
Radio Channel Characteristics. The environment in which an RF wireless medical
device is evaluated for coexistence is critical and must be well-characterized to mimic the
expected environment and typical deployment. Medical device deployments include those
that can be connected to or implanted in stationary or mobile patients. Device deployment,
configuration, and environment affect the wave propagation through the wireless channel.
Two such wave propagation scenarios, line-of-sight (LOS) and nonline-of-sight (NLOS),
must be examined. LOS testing can be performed in an anechoic chamber, ensuring
reproducibility by isolating the medical device and its wireless technology from spurious
interference. Conversely, NLOS testing is performed outside an anechoic chamber to
account for the effects of reflection-caused multipath on the received RF signal.
Performance in an NLOS deployment depends upon radio channel path loss among the
transmitter and receiver, frequency band, and transmission power. In an NLOS test setup,
as the signal-to-interference ratio of the wireless medical device decreases, the likelihood
of interference from nearby wireless systems increases.10,11 The delay in the traffic flows
is also exponentially related to the signal-to-noise ratio of the wireless medical device
network under test, which would cause an increase in latency for the wireless medical
device data.12,13
This type of testing for the NLOS configuration outside of an anechoic chamber raises the
possibility of collateral phenomena, including reflections from nearby structures causing
multipath. To account for this, multiple test arrangements of the transmitter and receiver
terminals should be considered. For each test setup, the wireless channel can be
Pa
g
e 3 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
Figure 1. LOS Test Setup for Wireless Medical Device Telemetry
System
characterized by finding its path loss attenuation and power delay spread, improving test
result reproducibility from one testing environment to another.
The wireless medical device transceivers should be evaluated separately (see Figure 1)—
each separately exposed to one or multiple interfering wireless networks. Various
interference phenomena arise depending on whether the interfering wireless network is in
the proximity of the wireless medical device transmitter or receiver. When a medical
device receiver is surrounded by an interfering network or networks, packet collisions
increase at the receiver (i.e., the hidden terminal effect). In contrast, when a medical
device transmitter is surrounded by one or more interfering networks, channel utilization
decreases (i.e., the exposed terminal effect). Decreased channel utilization is the result of
busy channel sensing followed by increased backoff (contention) windows.
Polarization. Polarization and cross-polarization of both the wireless medical device and
interfering terminals should be considered when testing for coexistence. For example, if
the RF fields from the wireless medical device are intended to be horizontally polarized
when operated, the interfering network should be deployed in both horizontal and vertical
polarizations because an antenna is never 100% polarized in a single mode. Additionally,
radio channel multipath could cause cross polarization in an NLOS test environment.
Cochannel and
Adjacent-
Channel
Interference.
Cochannel and
adjacent-
channel
interference
must be
considered to
effectively
characterize the
probability of
packet
collision.14,15 The
effect of
adjacent-
channel
interference on
the bit-error rate
has been shown to strongly depend on the frequency offset between the channel under
test and the interfering carriers.16,17 The correlation between spatial distance and channel
spacing to control interference between concurrent transmissions in a wireless sensor
multichannel network has also been investigated.18 The impact of cochannel interference
was tested experimentally, and performance was analyzed for IEEE 802.11g networks.19
The impact of adjacent channel interference in 802.15.4 (ZigBee) networks has also been
studied.14 Coexistence testing was performed with a single interfering terminal and
multiple interfering terminals in different combinations, including two interferers in the
same adjacent channel, two interferers in two different adjacent channels, and three
interferers in the same adjacent channel. It has been suggested that the key to
Pa
g
e 4 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
Transmitter
Power Initial
Distance
m
Minimum
Distance
m
<600 mW 1 0.25
600 mW–2
W20.5
2 W–8 W 3 1
Table 1. Initial and minimum
test distances specified by
ANSI C63.18.
determining coexistence in heterogeneous networks is the study of the effects of multiple
interferers.20
While a number of studies were limited to a specific wireless technology, the
aforementioned practices are recommended across various wireless technology
platforms. Hence, to evaluate a medical device for RF wireless coexistence, testing should
consider a pair (server,client) of interfering terminals configured to operate on the
cochannel and subsequently on the two immediate adjacent channels of the wireless
medical device under test. Does one pair of interfering terminals operating on the
cochannel frequency of the medical device present the worst-case scenario? The answer
depends on the duty cycle (i.e., packet generation and transmission rate) of the interfering
terminals. If the duty cycle is high, the cochannel will be proportionally highly utilized. If
multiple pairs of interfering terminals are configured to operate on the cochannel, the
exposed terminal problem arises, resulting in cochannel underutilization.
Most wireless technologies (e.g, WiFi and ZigBee) employ carrier sensing multiple access
(i.e., CSMA/CA protocol) to avoid packet collision. This regulates channel accessibility by
permitting one terminal to transmit at any given time. Testing with multiple interfering
terminals—where one pair of terminals is set on the cochannel frequency and two others
are set on two different adjacent channels—is highly advantageous, as it evaluates
performance of medical devices designed to automatically select an alternate channel
when interference exists on its current channel. Testing will evaluate the accuracy of the
new selection and the time required for the medical device to switch to the new channel.
Distance. The distance between the interfering wireless
terminal or network and the RF wireless medical device
transmitter or receiver requires consideration because
the likelihood of interference with the medical device
increases as the signal-to-interference ratio
decreases.10,11 Adjusting the distance between the
medical device transmitter or receiver and the
interfering wireless terminal or network is one way to
control the signal-to-interference ratio. The initial and
minimum test separation distances recommended by
ANSI C63.18 should be used, as shown in Table 1.21
Initial and minimum distances are determined based on
the transmitting power of the interfering terminals. The
initial test distance is calculated to expose the device
under test to approximately 3 V/m. The minimum test distances were calculated so that
the device under test would be exposed to approximately 20 V/m. In the C63.18 test
method, if there is no interference at the initial distance, the distance between the wireless
medical device receiver and interfering networks is systematically reduced until
interference occurs or the minimum test distance is reached. Auto-power-leveling
algorithms should be disabled in interfering terminals. Allowed maximum power should be
configured for each interfering wireless device used. The C63.18 test procedure
recommends configuring transmitters to maximum output power. If interference is present
at the initial distance, then the distance between the medical device terminal and the
interfering network terminals is increased until interference ceases. Results are most
reproducible when the output power of the interfering terminals does not change during
the test. The results obtained at various separation distances are used to characterize the
Pa
g
e 5 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
wireless medical device and inform policies and procedures for EMC management in
healthcare facilities.
Medical Wireless Transmission Parameters
The transmission parameters of the wireless medical device (e.g., packet size, polling
window, clear channel assessment threshold, and duty cycle) can alter the outcome of
coexistence testing. Studies have shown that as the packet size increases, the probability
of packet loss increases.10,11,22 Studies have also shown that as the polling window
increases, the probability of packet loss decreases.11,22 Additionally, results indicate that
when the interference level is below a certain level specified by the device sensitivity, the
channel is sensed as idle, or clear, and the interference does not affect
communication.10,11,14,22 Duty cycle, or channel utilization, is mainly dependent on the
amount of traffic generated and transmitted by the interfering wireless networks. Studies
show that as the interfering device or network increases its duty cycle, the victim network
packet loss ratio increases, causing either temporary or permanent interference.10,11,22
Therefore, two transmission parameter settings should be used during coexistence
testing: typical or manufacturer-suggested default settings and worst-case-scenario
settings (e.g., as suggested by previous work published in the literature).
Case Study
Assimilating the aforementioned coexistence factors into a test protocol is dependent
upon the medical device under test. A case study is offered to serve as an example. Of
note is that a test setup is unique to a specific wireless medical device or system.
The wireless medical system under test was a telemetry system implementing a ZigBee
chipset to establish bi-directional communication between two parts of the system,
hereafter designated as device #1 and device #2. ZigBee is a low-power communication
protocol with a bit rate of 250 kilobits per second in the 2.4 GHz band. When using
ZigBee, selecting channel 13 or 21 for coexistence testing is suggested, the justification
for which is explained below. Channel 21, i.e., 2.455 GHz, was used by the medical
wireless system for this case study.
The RF wireless medical device system was first baselined under static conditions without
interference. The test layout is shown in Figure 1. Medical devices were separated by a
distance of 5 m and elevated on a wood table to a height of 1 m. The separation distance
and height were chosen based on the typical deployment of the wireless medical
telemetry system.
The wireless medical system transmission parameters were evaluated using two
configurations: typical settings (i.e., default manufacturer settings) and worst-case-
scenario settings, as explained previously. The medical device system communicated a
predetermined number of packets to establish a statistical baseline of the wireless link for
each transmission parameter setting. To characterize the wireless medical telemetry
system for this case study, the measured parameters consisted of packet loss ratio, delay,
and throughput. These parameters are used to assess the quality of service provided by
the wireless medical device. The desired quality of service is dependent upon the critical
functions specified by the medical device manufacturer.
Pa
g
e 6 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
After establishing a baseline, the wireless medical device system was evaluated for
coexistence with other wireless technologies. An IEEE 802.11g network, e.g. Wi-Fi, was
deployed around medical device #2. The 802.11g network consisted of up to three pairs,
each operating on a different channel. It should be noted that auto-power-leveling
algorithms are disabled in the 802.11g network, and the maximum power allowed by the
wireless network standard is used. The initial separation distance between the medical
device and the interfering network was determined based upon the transmission power of
the interfering network, which is less than 600 mW. The ANSI C63.18 table (Table 1)
suggested that the initial distance should be 1 m.
For the first set of coexistence tests, only a single 802.11g network was introduced to the
wireless medical device network. Communication was first established between the
wireless medical devices and then in the 802.11g network. In an attempt to cause
interference with the wireless medical device, the testing parameters were: channels 1-11,
data rate of 1-26 Mbit/s, horizontal and vertical polarization, variable separation distance
between the medical device and interfering network and a single 802.11g interfering
network. After an adequate number of packet transmission attempts between the medical
devices, the interfering 802.11g network was turned off. If interference occurred during
testing, the amount of time was determined that the wireless medical device telemetry
system needed to restore communication after the cause of interference was removed.
After all variables were tested, the positions of medical device #1 and medical device #2
were exchanged, and testing was repeated so that each medical device was tested
separately with the interfering network.
The RF wireless medical device system was then exposed to three 802.11g wireless
networks simultaneously. Channels 9, 10, and 11 were chosen. The channel selection
allows co-channel interference (10) and two different adjacent channel interferences (9,
11) to be tested for the wireless medical device telemetry system. The data rate,
separation distance, and polarization of the three 802.11g networks were variables that
were evaluated. Again, medical device #1 and medical device #2 were separately
exposed to the interfering network.
The NLOS test setup was then performed. The experimental methodology for LOS and
NLOS testing was the same, i.e., the RF wireless medical telemetry system was first
baselined without an interfering network and then interfering networks were introduced
into the environment.
A LOS and an NLOS test setup were then repeated with different interfering networks in
the same frequency band as the wireless medical telemetry system. Examples include
ZigBee, IEEE 802.11b, IEEE 802.11g, IEEE 802.11n, Bluetooth, and cordless phones,
which transmit at 2.4 GHz.
The wireless medical network did not experience a loss of network connectivity during the
LOS testing with one or three 802.11g wireless networks operating simultaneously in the
same unlicensed spectrum band at a separation distance of 0.25 m between the
interfering networks and the wireless medical device under test. All of the wireless
functions operated as intended.
The wireless medical network experienced a loss of network connectivity during the NLOS
testing when one or three 802.11g wireless networks were operating simultaneously in the
same unlicensed spectrum band with a separation of 3 m or less between the interfering
Pa
g
e 7 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
networks and the wireless medical device under test. When the separation distance was
greater than 3 m, network connectivity was restored to the wireless medical network.
When the throughput (duty cycle) of the 802.11g network was decreased, network
connectivity was restored to the wireless medical network at a separation distance of 1 m.
It was also observed that shorter wireless medical device packets had a higher probability
of successful transmission while longer data packets had a lower probability of successful
transmission.
General Considerations for Coexistence Testing
The RF wireless parameters discussed and the recommendations made in this article are
intended to serve as a starting point for medical equipment manufacturers and healthcare
organizations in assessing the coexistence of RF wireless communication in modern
dynamic electromagnetic environments. Coexistence testing is different from EMC RF
immunity testing, under which the RF immunity of a medical device is characterized over a
range of frequencies.
Pass-fail criteria for coexistence testing must be identified and need to be quantified in
terms of the critical functions of the device under test as determined by risk assessment.
The following parameters can be considered:
Packet error rate.Latency.Jitter.Network throughput.
In other words, coexistence testing should demonstrate that the effects of the wireless
medical device on nearby RF wireless equipment and networks is minimal and effects of
nearby RF wireless equipment and networks on the functions of the wireless medical
device would not result in unacceptable risk to the patient or user.
Status
Several groups interested in RF wireless coexistence testing of wireless medical devices
have begun work in this area. A working group of ANSI-accredited committee C63 held a
preliminary teleconference in January 2011, and FDA sponsored a teleconference in
February 2011. The University of Oklahoma Wireless and Electromagnetic Compliance
and Design Center, in Tulsa, has an evolving test protocol based on the parameters listed.
The University of Oklahoma is currently experimentally testing medical devices for
coexistence. Plans for future work include proposing that ANSI-accredited committee C63
establish an official working group to develop consensus coexistence standards or
recommended practices.
References
1. N. LaSorte, Y. Burette, H. Refai, “Experimental Characterization of Electromagnetic
Propagation of a Hospital from 55-1950MHz”, in Proceedings of IEEE Asia-Pacific
Symposium on Electromagnetic Compatibility, April, 2010, pp. 826 – 829.
2. N. LaSorte, W. Barnes, H. Refai, “Characterization of the Electromagnetic Environment
Pa
g
e 8 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
in a Hospital and Propagation Study”, in Proceedings of IEEE International Symposium on
Electromagnetic Compatibility, August 2009, pp. 135-140.
3. Institute of Electrical and Electronics Engineers. IEEE Recommended Practice for
Information technology—Telecommunications and information exchange between
systems — Local and metropolitan area networks — Specific requirements, Part 15.2:
Coexistence of wireless personal area networks with other wireless devices operating in
unlicensed frequency bands. IEEE 802.15.2:2003. New York, NY: IEEE, 2003.
4. http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDo... [5]
5. International Electrotechnical Commission. Medical electrical equipment—Part 1:
General requirements for basic safety and essential performance. IEC 60601-1:2005.
Geneva (Switzerland): IEC, 2005.
6. International Electrotechnical Commission. Medical electrical equipment—Part 1:
General requirements for basic safety and essential performance. 2. Collateral standard:
Electromagnetic compatibility—Requirements and tests. IEC 60601-1-2:2007. Geneva
(Switzerland): IEC, 2007.
7. International Electrotechnical Commission. Medical electrical equipment—Part 1:
General requirements for basic safety and essential performance. 2. Collateral standard:
Electromagnetic disturbances—Requirements and tests. IEC 62A/746/CD, draft Edition 4
of IEC 60601-1-2. Geneva (Switzerland): IEC, 2011.
8. Institute of Electrical and Electronics Engineers. IEEE Recommended Practice for
Information technology — Telecommunications and information exchange between
systems — Local and metropolitan area networks — Specific requirements, Part 15.2:
Coexistence of wireless personal area networks with other wireless devices operating in
unlicensed frequency bands. IEEE 802.15.2:2003. New York, NY: IEEE, 2003.
9. D. Witters, S. Seidman, H. Bassen, “EMC and Wireless Healthcare,” IEEE 2010
APEMC, 2010.
10. L. Angrisani, and M. Vadursi, “Cross-Layer Measurements for a Comprehensive
Characterization of Wireless Networks in the Presence of Interference,” IEEE Trans.
Instrum. Meas., vol. 56, no. 4, pp. 1148-1156, Aug. 2007.
11. L. Angrisani, M. Bertocco, D. Fortin, and A. Sona, “Experimental study of coexistence
issues between IEEE 802.11b and IEEE 802.15.4 wireless networks,” IEEE Trans.
Instrum. Meas., vol. 57, pp. 1514–1523, 2008.
12. Tianlin Wang, and Hazem H. Refai, “The Development of an Empirical Delay Model
for IEEE 802.11b/g based on SNR Measurements”, Proceedings of the 1st IEEE
International Symposium on Wireless Quality-of-Service, Maui, Hawaii, June 13-16, 2005.
13. Tianlin Wang, and Hazem H. Refai,” Empirical Network Performance Analysis on IEEE
802.11a/b/g with Different Protocols and Signal to Noise Ratio Values”, Proceedings of the
IEEE International Conference on Wireless and Optical Communications Networks
(WOCN 2005), Dubai, UAE, March 6-8, 2005.
14. L. Lo Bello, E. Toscano. “Coexistence Issue of Multiple Co-Located IEEE
802.15.4/ZigBee Networks Running on Adjacent Radio Channels in Industrial
Environments,” IEEE Trans. Industrial Infomatics, vol. 5, pp.157-167, 2009.
15. I. Howitt, “WLAN and WPAN coexistence in UL band,” Transactions on Vehicular
Technology, vol. 50, pp. 1114–1124, 2001.
16. J. C. Haartsen, “The bluetooth radio system,” IEEE Personal Communications, vol. 7,
pp. 28–36, 2000.
17. A. Kamerman, “Coexistence between Bluetooth and IEEE 802.11 CCK solutions to
avoid mutual interference,” IEEE 802.11-00/162, 2000.
18. O. Incel, S.O. Dulman, P.G. Jansen, “Multi-channel support for dense wireless sensor
networking,” in Proceedings 31st IEEE Conference on Local Computer Networks,
November 2006, pp. 694-701.
Pa
g
e 9 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
19. G. Cena, I. Cibrario Bertolotii, A. Valenzano, C. Zunino, “Reasoning About
Communication Latencies in Real WLANs,” in Proceedings 12th IEEE Conference on
Emerging Technology Factory Automation, September 2007, pp. 187-194.
20. P. Pinto, A. Giorgetti, M. Chiani, and M. Win, “A stochastic geometry approach to
coexistence in heterogeneous wireless networks,” IEEE Journal on Sel. Areas in
Communications, vol. 7, pp 1268-1282, 2009.
21. Institute of Electrical and Electronics Engineers, Inc. American national standard
recommended practice for on-site ad hoc test method for estimating radiated
electromagnetic immunity of medical devices to specific radio-frequency transmitters.
(standard C63.18). Piscataway, NJ: IEEE, 1997.
22. M. Bertocco, A. Sona, F. Tramarin, “Design of Experiments for the Assessment of
Coexistence Between Wireless Networks,” in Proceedings of the Instrumentation and
Measurement Technology Conference, May 3-6, 2010, pp. 928-932.
Nickolas LaSorte is pursuing a PhD in electrical engineering at the University of
Oklahoma. Hazem H. Refai is an associate professor in the university’s school of electrical
and computer engineering and founding director of the WECAD Center. Seth Seidman is
a research electrical engineer at FDA. Donald Witters is chairman of the CDRH EMC and
Wireless Group. Jeffrey L. Silberberg is senior electronics engineer for CDRH, Office of
Science and Engineering Laboratories, Division of Electrical and Software Engineering.
Author:
Nickolas J. LaSorte, Hazem H. Refai, Donald M. Witters Jr., Seth J. Seidman, Jeffrey L.
Silberberg
Featured Image:
Pa
g
e 10 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
Wireless Technology
Privacy Policy | Contact | Advertise | Subscribe | Sitemap
© 2011 UBM Canon
Related Sites from UBM Canon:
- Qmed - Qualified Medical
Suppliers
- Medical Device +
Diagnostic Industry
- European Medical Device
Technology
- Medical Product
Manufacturing News
- IVD Technology
- OrthoTec
- China Medical Device
Manufacturer
- medtechinsider
- medtechinsider auf Deutsch
- Pharmaceutical & Medical
Packaging News
- Pharmalive
Pa
g
e 11 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
Source URL: http://www.medicalelectronicsdesign.com/article/wireless-medical-device-coexistence
Links:
[1] http://www.medicalelectronicsdesign.com/
[2] http://www.medicalelectronicsdesign.com/article/wireless-medical-device-coexistence
[3] http://www.facebook.com/sharer.php
[4] http://www.medicalelectronicsdesign.com/categories/wireless-technology
[5]
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm077210.htm
Pa
g
e 12 of 12Wireless Medical Device Coexistence
8/23/2011htt
p
://www.medicalelectronicsdesi
g
n.com/
p
rint/1297
... Wireless communication is accomplished using either published IEEE standards or proprietary protocols, many of which operate at unlicensed frequencies in the Industrial, Scientific, and Medical (ISM) bands. Coexistence of wireless medical devices operating in the ISM band is a growing concern for the Food and Drug Adminis- tration [1], [2]. To date, no standard has been implemented to adequately assess these risks. ...
Article
We describe the development of free-field test methods for wireless electronic safety equipment that replicate field-test conditions in a laboratory environment. The test methods can be used to verify the performance of wireless devices, such as those used by emergency responders, in the presence of known attenuation and under RF interference conditions. The test methods presented here were developed to support the National Fire Protection Association (NFPA) in the revision of NFPA 1982: Standard on Personal Alert Safety Systems (PASS), but would be applicable to other types of RF-based equipment as well. In Part 1, we illustrated methods for extracting performance metrics from a series of field tests conducted by NIST researchers. In Part 2, we replicate the key field test conditions in the laboratory and verify device performance under those conditions.
Conference Paper
Full-text available
Currently, most wireless sensor network applications assume the presence of single-channel Medium Access Control (MAC) protocols. When sensor nodes are densely deployed, single-channel MAC protocols may be inadequate due to the higher demand for the limited bandwidth. To overcome this drawback, we propose multiple channel support for improving the performance. Our method allows the nodes to utilize new frequency channels which results in the significant increase on the number of nodes that are granted access to the wireless medium. The method requires only one half-duplex transceiver per node, which is capable of sending and receiving over distinguished frequency channels. Simulation results show that, method successfully utilizes multiple channels and increases the performance proportional to the number of available frequencies for an example single-channel MAC protocol, LMAC.
Conference Paper
Full-text available
This work presents the results of a survey of the electromagnetic environment from 30 MHz-7 GHz in St. Francis Hospital, in Tulsa, Oklahoma. Measurements were considered short term as the full spectrum was measured in twelve minutes. The results show the dynamic environment apparent in today's hospitals. There were no recorded cases in which the IEC immunity standard specifications, 3V/m, was exceeded. Another contribution by this paper was to show an example of cross-floor attenuation in a hospital.
Article
Full-text available
With the increasing proliferation of different communication devices sharing the same spectrum, it is critical to understand the impact of interference in heterogeneous wireless networks. In this paper, we put forth a mathematical model for coexistence in networks composed of both narrowband (NB) and ultrawideband (UWB) wireless nodes, based on fundamental tools from stochastic geometry. Our model considers that the interferers are spatially scattered according to a Poisson field, and are operating asynchronously in a wireless environment. We first determine the statistical distribution of the aggregate interference for both cases of NB and UWB emitters. We then provide error probability expressions for two dual configurations: 1) a NB victim link subject to the aggregate UWB interference, and 2) a UWB victim link subject to the aggregate NB interference. The results show that while the impact of a single interferer on a link is often negligible due to restrictions on the transmitted power, the aggregate effect of multiple interferers may cause significant degradation. Therefore, aggregate interference must be considered to ensure coexistence in heterogeneous networks. The proposed analytical framework shows good agreement with physical-level simulations of the system.
Conference Paper
Full-text available
An examination of the electromagnetic environment and characteristics from 30MHz-7GHz in a hospital is discussed in this paper. Empirical propagation constants, building attenuation, floor height gain, and the CDF of the coverage, were calculated from 50-1950MHz. This is carried out to further understand the propagation of radio waves through hospitals. The understanding of wave propagation through hospitals will help advance the knowledge to control the electromagnetic environment of hospitals to decrease electromagnetic interference to medical devices. The understanding of the RF hospital environment is even more pressing given that many hospitals are implementing electronic medical records that would require the co-location and co-existence of many different types of RF emitters. The measurement parameters calculated are useful for propagation modelling.
Article
Full-text available
A few years ago it was recognized that the vision of a truly low-cost, low-power radio-based cable replacement was feasible. Such a ubiquitous link would provide the basis for portable devices to communicate together in an ad hoc fashion by creating personal area networks which have similar advantages to their office environment counterpart, the local area network. Bluetooth/sup TM/ is an effort by a consortium of companies to design a royalty-free technology specification enabling this vision. This article describes the radio system behind the Bluetooth concept. Designing an ad hoc radio system for worldwide usage poses several challenges. The article describes the critical system characteristics and motivates the design choices that have been made.
Article
The characteristics of the IEEE 802.15.4 physical and medium access layers make such a protocol a suitable candidate to support communication between sensors and actuators in industrial environments. As industrial networks may comprise a large number of sensors and actuators and the delay increases with the increased number of nodes, a possible solution to keep the delay small is the use of multiple radio channels to implement different small low-latency communication cells. Although in IEEE 802.15.4 the radio channels do not overlap, recent literature showed that some interference may actually occur. This paper provides a better understanding of cross-channel interference in co-located IEEE 802.15.4 industrial networks and proposes a general methodology for the assessment of IEEE 802.15.4 performance under different cross-channel interference conditions. This methodology allows a network designer to perform on-site but accurate assessments and can be easily deployed in real industrial environments to perform measurements directly in the environment-under-test. A case study based on COTS IEEE 802.15.4 devices is presented to show how to apply our methodology to a real scenario and to discuss the results obtained with one or multiple interferers and varying some MAC level parameters.
Conference Paper
The paper investigates the reliability of simulation models for the assessment of wireless network performances under the effects of radio interference. In particular, some critical issues affecting the accuracy of modern simulation tools are addressed. The main purpose is to highlight the role of measurements in the optimization of theoretical models. Accordingly, some meaningful results from experiments are compared with the corresponding ones obtained from simulations, and simulation models have been consequently improved. Experiments and simulations have been carried out by using a suitable testbed, described in the paper, enlisting a set of IEEE 802.15.4 and 802.11b compliant devices operating in the 2.4 GHz industrial scientific and medical band.
Conference Paper
Large diffusion of wireless communications is a trend that will soon involve industrial environments, too, and in particular those applications where granting limited response times is of utmost importance. This paper analyzes the impact of a number of causes, besides the medium access technique, on the performance of real 802.11g networks. In particular, the extent to which beacon frames and traffic on nearby channels can decrease the communication performance is taken into account, as well as the effect of the adapter's and access point's overheads on frame transmission latencies.
Conference Paper
Wireless local area networks (WLAN) based on IEEE 802.11b/g standards' are widely deployed to provide wireless interconnectivity around university campuses, business offices, and homes. A WLAN is inexpensive, easy to install, and provides data rates of up to 54 Mbps. This study experimentally analyzed the network delay performance parameter and the wireless physical channel signal-to-noise ratio (SNR) of an IEEE 802.11b/g WLAN under different signal strength values. The measurement results indicated that the IEEE 802.11 b/g delay parameter could be predicted and modeled as a function of SNR. The experiments were conducted with and without source of interference; a cordless phone operating in 2.4 GHz was used to provide interference. Given SNR values. The achieved model is useful in determining optimal placements of the wireless network access points prior to deployment. The values of SNR of a specific site are calculated using the site path loss models.
Conference Paper
Wireless local area network (WLAN) based on IEEE 802.11 standards has experienced tremendous growth in recent years. The 802.11g is one of 802.11 standards that offer data transmission at up to 54 Mbps, compared with the 11 Mbps theoretical maximum with the earlier 802.11b standard. This study experimentally examined network performance of an IEEE 802.11g WLAN. It focused on the effect of signal-to-noise ratio (SNR), transport protocol, chipset design, and the impact of file size or the number of users on the network throughput and delay performances. The measurement results indicated that the network delay is exponentially related to SNR. Hence an empirical delay model was developed. The measurements also showed that for a given file size, the network throughput reached a saturation value at a critical SNR. This result confirmed with a throughput model developed for the 802.11b network by Rappaport. Both developed models can be used in conjunction with propagation models and site-specific information to optimally place network access points prior to actual 802.11g wireless network deployment. Throughput measurements were also obtained using transmission control protocol (TCP), multiple network users, and different packet sizes.