ChapterPDF Available

Signals in bioimpedance measurement: Different waveforms for different tasks

Authors:

Abstract and Figures

Alternatives to the traditional sine wave excitation are studied in the paper. Impedance measurements can be performed much faster by using a broad bandwidth signal for excitation. Using of square wave pulses, Gaussian function and its derivative, also modifications of sinc and chirp signals, is analysed. Carefully designed pulse wave excitation can become to an alternative to established excitation waveforms, especially, when fast measurements with exact timing are required, and when the energy consumption is important.
Content may be subject to copyright.
Hermann Scharfetter, Robert Merva (Eds.): ICEBI 2007, IFMBE Proceedings 17, pp. 181–184, 2007
www.springerlink.com © Springer-Verlag Berlin Heidelberg 2007
Signals in bioimpedance measurement: different waveforms for different tasks
M. Min1, U. Pliquett2, T. Nacke2, A. Barthel2, P. Annus1 and R. Land1
1 Tallinn University of Technology, Tallinn, Estonia
2 Institute of Bioprocessing and Analytical Measurement Techniques, Heilbad Heiligenstadt, Germany
Abstract— Alternatives to the traditional sine wave excita-
tion are studied in the paper. Impedance measurements can be
performed much faster by using a broad bandwidth signal for
excitation. Using of square wave pulses, Gaussian function and
its derivative, also modifications of sinc and chirp signals, is
analysed. Carefully designed pulse wave excitation can become
to an alternative to established excitation waveforms, espe-
cially, when fast measurements with exact timing are required,
and when the energy consumption is important.
Keywords— Broadband excitation, pulse wave signals, fast
measurement, time-frequency analysis, low power devices.
I. INTRODUCTION TO THE PROBLEM
Electrical bioimpedance [1] is a well recognised parame-
ter for tissue characterisation in implantable medical tech-
nology, e.g. in cardiac pacemakers and defibrillators [2].
Measurement systems are downsized to the nanoliter size
biotechnology in the lab-on-the-chip type devices [3].
The frequency domain analysis is commonly used in bioim-
pedance measurement practice.
Impedance measurements can be performed much faster
by using a broad bandwidth signal for excitation and calcu-
lating the spectrum by means of Fourier transform. If the
signal is partially transformed, windowing functions are
used to extract the spectrum at a defined time point. This
approach is advantageous in cardiac applications [2] or in
high throughput bioprocessing [3].
Here, the joint time-frequency measurement and analysis
is one of the aims for development of new impedance meas-
urement methods. A generalised approach for bioimpedance
measurement is the current excitation and monitoring of the
voltage developing across the object (Fig.1). An excitation
generator G generates an AC excitation current ie(t), which
is injected into the complex impedance Ż to be measured
and analysed. The response to the excitation – a voltage vr(t)
carries information about the impedance Ż. This informa-
tion will be analysed in an impedance analyser – classically
in frequency domain, but also in time domain, e.g. for ob-
taining time varying spectrum. As a result, both, the spectral
and time based parameters of the impedance behaviour will
be obtained – Ż (ω, t). Typically the sine wave excitation
with adjustable frequency is used. To accelerate both, the
measurement and analysis processes, simultaneous multi-
sine measurements have been introduced [4, 5].
v
r
(t)
i
e
(t)
Ż(ω,t)
G Ż
excitation
generator
impedance
Impedance
Analyzer
response
voltage
Fig.1 Typical impedance measurement system
Compared to other biosensors, the bioimpedance meas-
urement requires generation of the excitation signal, which
means significant energy consumption. This is rather impor-
tant for wearable and other battery powered instruments,
and can be crucial in implantable device. Here we pay spe-
cial attention to the generation of minimal-energy excitation
signals where a sufficiently broad spectrum can be meas-
ured. It is expected that optimal excitation waveforms en-
able us to achieve both of these aims – to provide the short-
term but accurate spectral analysis, and to focus the excita-
tion energy onto the desired frequency range at the prede-
termined time instant.
II. WAVEFORMS OF EXCITATION SIGNALS
Sine wave excitations are commonly used for impedance
measurements. For fast measurements, the multi-sine simul-
taneous measurement [5] or time domain based approaches
(application of square wave pulses) are used [6]. Such a
simple broadband signal, as a short rectangular pulse
(Fig.2a), is typically used for excitation while the analysis
employs the Fourier-transformed signal [7]. This yields a
spectrum of sinc(ωT)=sin(ωT)/ωT type spectrum, which is
almost constant up to the frequency fmax=1/2T.
In Fig.2a, the duration of pulse is T=50μs. As a result,
the first zero value of the spectrum takes place at the
frequency f10= 1/T= (1/50)106 =20kHz, see Fig.2a and 2b,
and the fmax is there about 10kHz.
With such a quick method even fast changes of the object
impedance can be monitored. There are, however, some
drawbacks. For instance, the distribution of spectral energy
182 M. Min, U. Pliquett, T. Nacke, A. Barthel, P. Annus and R. Land
__________________________________________ IFMBE Proceedings Vol. 17 ___________________________________________
time
amplitude
log
f
1/2T
lin
f
lin magnlog magn
10kHz 20kHz 40kHz
T=50μs
20kHz 40kHz
10kHz
1/2T
(1)
(
2
)
a
)
b
)
c
)
Fig.2 Rectangular pulse excitations (a) with a DC component (1) and
without it (2), and their spectra in linear (b) and log (c) scales
is not adjustable, meaning that energy is wasted in fre-
quency regions without interest. The “wasted” energy can
reach about 40% of the useful one. As a result, the spectral
density of useful excitation remains low at the actual level
of energy consumption.
Our aim is to find waveforms for excitation pulses with
the energy concentrated in the frequency range of interest.
The measurement process must be fast at the predetermined
or precisely registered moment.
Therefore, there is an urgent need to find more effective
waveforms for the excitation pulses than the simplest rec-
tangular ones, having the two discrete values (+1 and 0, or
+1 and –1, see Fig.2a), see [6]. Min et al [8] proposed the
modified rectangular pulses with three discrete values
(+A, -A, and 0). However, these waveforms are no real
improvement to established waveforms in the present case
of application.
If only a limited bandwidth is required, excitation signals
like Gaussian function (see Fig.3a)
2
5.0
0
)(
=
σ
t
eAtG (1)
and preferably its derivatives, which do not contain any DC
component (Fig.4a), have an advantage due to the higher
energy density within the desired range of the spectrum
(Fig.3b and 3c). Anyway, the spectra of Gaussian pulse and
its first derivative (Figs 3 and 4) are not simply rectangle.
time
amplitude
log
f
lin
f
lin magnlog magn
10kHz 20kHz
20kHz
10kHz
(1)
(
2
)
a
)
b
)
c
2T=100μs
Fig.3 Gaussian (1) and windowed sinc (2) excitation pulses (a), and their
spectra in linear (b) and log (c) scales
Much better results gives the more complicated sinc(ωt)
type excitation pulse, especially when the windowing of the
pulse function is taken into use (Hanning window in Fig.3).
Some modification of sinc function, e.g. haversinc(ωt)=
=sin2(ωt)/ωt, is more preferred in biological systems,
because it does not contain the DC component, see Fig.4.
A difference of two sinc(ωt) type pulses as Asinc(ωt)-
-Bsinc(nωt) enables to shift the excitation bandwidth to the
desired higher frequency range from about ω to nω.
III. COMPARISON OF SINC AND GAUSSIAN PULSES
A shortened (duration 1ms) and windowed by Hanning
sinc function sin(ωt)/(ωt) in Fig.3a has a near to rectangular
spectrum with a bandwidth, which is proportional to the
frequency f = ω/2π of the sine function sin(ωt), see Fig.3b
and 3c. In principle, the sinc function has the perfectly rec-
tangular spectrum when its duration extends to infinity.
However, the practical case in Fig.3 is fully acceptable,
though not ideal. Less than 10% of the excitation energy
drops outside the bandwidth of interest. Thanks to certain
important properties of Gaussian function (simple wave-
form, smooth transient response, the spectrum is also the
Gaussian function without any side lobes), these pulses
(see Fig.3a) are of interest. Since the spectrum is not flat
(Fig.3b and 3c), the data processing requires as response as
well as the excitation signal.
Signals in bioimpedance measurement: different waveforms for different tasks 183
__________________________________________ IFMBE Proceedings Vol. 17 ___________________________________________
IV. COMPARISON OF HAVERSINC PULSE AND
THE DERIVATIVE OF GAUSSIAN FUNCTION
Any DC-application to biological structures should be
avoided. Therefore, the windowed haversinc pulse
sin2(ωt)/ωt and the first derivative of Gaussian function
dG(t)/dt are preferred for excitation (see Fig.4) instead of
original sinc and Gaussian pulses (Fig.3). Their spectra in
Fig.4a and 4b are similar to those of the original functions
in Fig.3b and 3c, only the DC component is absent and low
frequency parts are suppressed, the other properties are
almost the same.
The amplitude coefficient
(2)
of the n-th derivative (i = 1, 2, 3…, n) ensures equal ener-
gies for all the derivatives taking the values of amplitudes as
(3)
V. USING OF CHIRP PULSES
The chirp pulses or chirplets can be expressed mathe-
matically as
(4)
where 0< t T, B/T is a chirp rate, and T is duration of
the chirp pulse. These functions describe the radio impulse
with linearly increasing frequency and the bandwidth B
extending from ω/2π to (ω/2π +B) in Hz. In the simplest
case ω=0 (see Fig.5a), and the spectrum covers the range
from DC to B (see graphs 1 in Fig.5b and 5c, where
B=1MHz). The DC component is an average value over
time T, and it depends on the final value of the chirp signal
at the moment tfin=T. The DC component can be set to zero
choosing the “right” value for the final moment tfin.
The excitation spectrum approaches the ideal one by in-
creasing the time interval T.
We do not have any possibility to use the standard
windowing procedures for chirp function, but applying
some of their modifications can essentially improve the
shape of its spectrum [9].
A response spectrum of the β-region impedance
(frequency range from 10kHz to 1MHz) of a skeleton
muscle flap is given by graphs 2 in Fig. 5b and 5c.
time
amplitude
log
f
lin
f
lin magnlog magn
10kHz 20kHz
20kHz
10kHz
(1)
(
2
)
a
)
b
)
c
)
2T=100μs
Fig.4 Derivative of the Gaussian pulse (1) and haversinc (2) function (a),
and their spectra in linear (b) and log (c) scales
amplitude
log
f
lin
f
lin magnlog magn
1MHz
1MHz
a
)
b
)
c
)
time
100
μ
s
(1)
(
2
)
Fig.5 Initial 100μs part of the chirp pulse (a) and its spectrum (1) in linear
(b) and log (c) scale, and the response (2) of the impedance (β region) of
skeleton muscle flap to the chirp excitation
=
=
n
i
ni
C
15.0
σ
nn CAA ×= 0
)2/)/(2sin()( 2
tTBttch
πω
+=Ω
184 M. Min, U. Pliquett, T. Nacke, A. Barthel, P. Annus and R. Land
__________________________________________ IFMBE Proceedings Vol. 17 ___________________________________________
VI. DISCUSSIONS
Every excitation mode has its advantages and drawbacks
for practical applications. The Gaussian pulse and its
derivative are simple. The sinc function has a nice flat
spectrum, but the signal waveform is more complex.
The chirp excitation is the most favourable when the small
crest-factor (the peak value divided by the RMS value)
is required. The chirp pulse has a crest-factor about 1.4,
which is almost the same as of a single sine wave. Such the
signals as chirp and haversinc pulses and the derivatives of
Gaussian function might be preferred due to the absence of
the DC component. All the pulse wave excitations have a
wide frequency spectrum; therefore the noise level can be
relatively high in the response signal. The amplitude of
excitation pulses must be relatively high to obtain the inten-
sive broadband spectrum. Therefore, nonlinearity problems
can be more apparent than in the case of the periodic excita-
tion.
VII. CONCLUSIONS
Carefully designed pulse wave excitation can become to
an alternative to established excitation waveforms, espe-
cially, when fast measurements with exact timing are
required, and when the energy consumption is important
(implantable and wearable devices, laboratories on the
chip). The theoretical expectations were verified using an
electrical phantom of bioimpedance, an arbitrary waveform
generator AFG3252 as the excitation source, and a digital
oscilloscope DPO4104 with spectral analysis for measure-
ment and analysis (both from Tektronix).
The results were compared in different cases using both,
computer simulations and electrical experiments. For the
experiments, an arbitrary function generator, matched with
special amplifiers to the electrode system, and a digital high
speed oscilloscope for tracing the signals, were used.
The experiments were performed with RC-combination,
mimicking the biological object. All the above described
excitation waveforms can be used, but the most interesting
can give sinc and chirp pulses. The experiments show that
combining two pulses with different parameters, e.g. using
Asinc(x)-Bsinc(nx), we can get a smooth power spectrum
within the desired bandwidth, even when only a simple
windowing has been used. Simulation suggests a reasonable
outcome at even two or three full periods. Using of chirp
pulses can be the most informative, only the complicated
windowing problem needs to be solved before practical
implementations.
The results confirm that the described measurement
methods can become alternatives to the established ones,
e.g. to the multi-frequency simultaneous measurement.
ACKNOWLEDGMENT
The work was carried out in frames of the FP6 Marie
Curie Fellowship ToK project 29857 InFluEMP and
supported by grants no. 7212 and 7243 of Estonian Science
Foundation, and also by Enterprise Estonia through the
Competence Centre ELIKO.
REFERENCES
1. Grimnes S, Martinsen O.G (2000) Bioimpedance and Bioelectricity
Basics. Academic Press, London
2. Webster J.G (Ed.) (1995) Design of Cardiac Pacemakers. IEEE Press,
New York
3. Cheung K, Gawad S, Renaud P (2005) Impedance Spectroscopy Flow
Cytometry: On-chip Label-Free Cell Differentiation. Cytometry,
Part A, 65A:124-132
4. Gordon R, Land R, Min M, Parve T, Salo R W (2005) A virtual
system for simultaneous multi-frequency measurement of electrical
bioimpedance. Int J Bioelectromagnetism 7(2):243-246
5. Bragos R, Blanco-Enrich R, Casas O, Rosell J (2001) Characterisa-
tion of Dynamic Biologic Systems using Multisine based Impedance
Spectroscopy, Proc. IEEE Instrum. and Measurement Tech. Conf.
Budapest, Hungary, 2001, Vol.1, pp 44-47
6. Pliquett U, Gersing E, Pliquett F (2000) Evaluation of Time-domain
based Impedance Measurements on Biological Tissue. Biomediz-
inische Technik 45(1-2):6-13
7. Howie M B, Dzwonczyk R, McSweeney T D (2001) An Evaluation
of a new Two-Electrode Myocardial Electrical Impedance Monitor
for Detecting Myocardial Ischemia. Anesth Analg 92:12-18
8. Min M, Kink A, Land R, Parve T, Rätsep I (2004) Modification of
pulse wave signals in electrical bioimpedance analyzer for implant-
able medical devices, Proc. IEEE EMBS 26 International Conference,
San Francisco, California, 2004, pp 2263-2266
9. Xu F (2007) The Imaginary Part: A Source of Creativity in Instru-
mentation and Measurement. IEEE Instrumentation and Measurement
Magazine 10(1):29-33
Address of the corresponding author:
Author: Mart Min
Institute: Tallinn University of Technology
Street: Ehitajate tee 5
City: Tallinn
Country: Estonia
Email: min@elin.ttu.ee
... Broadband test signals with a spectrum covering a wide frequency range are used together with the discrete Fourier transform to obtain spectral information about the test object [26]. The excitation signals can have various shapes [24], e.g., rectangular [27], Gaussian functions [28], chirp signals [29], sinc signals [30], and multi-sine signals [31]. Although the idea of application of pulse excitation in impedance measurement is known, it has not been explored in ECT so far. ...
... Additive Gaussian noise was added to the generated digital samples. On the basis of the signal samples, the capacitance values for all electrode pairs were determined using Formulas (30) and (31). ...
... Additive Gaussian noise was added to the generated digital samples. On the basis of the signal samples, the capacitance values for all electrode pairs were determined using Formulas (30) and (31). (1)(2)(3)(4); (a,c) maps for capacitance measurement; (b,d) maps for conductance measurement. ...
Article
Full-text available
Electrical capacitance tomography (ECT) is a technique of imaging the distribution of permittivity inside an object under test. Capacitance is measured between the electrodes surrounding the object, and the image is reconstructed from these data by solving the inverse problem. Although both sinusoidal excitation and pulse excitation are used in the sensing circuit, only the AC method is used to measure both components of complex capacitance. In this article, a novel method of complex capacitance measurement using pulse excitation is proposed for ECT. The real and imaginary components are calculated from digital samples of the integrator response. A pulse shape in the front-end circuit was analyzed using the Laplace transform. The numerical simulations of the electric field inside the imaging volume as well as simulations of a pulse excitation in the front-end circuit were performed. The calculation of real and imaginary components using digital samples of the output signal was verified. The permittivity and conductivity images reconstructed for the test object were presented. The method enables imaging of permittivity and conductivity spatial distributions using capacitively coupled electrodes and may be an alternative measurement method for ECT as well as for electrical impedance tomography.
... Dielectric permittivity and electrical bioimpedance are well known parameters for material (biological/chemical, solid/liquid) or tissue characterizations in UWB spectroscopy systems [3], [4]. One of the fundamental considerations in the design of these systems is the UWB pulse shape. ...
... There are several types of pulse waveforms, such as sinusoidal, rectangular and Gaussian, which are typically used for excitation of different MUTs or impedance characterization systems whereas the biomedical applications (implantable/wearable biological systems) require a dc-free signal [4]. Most of the UWB systems use a Gaussian pulse (the spectrum is also Gaussian function without any side lobes) or its derivatives (the spectrum has a wide BW). ...
... A stepped-frequency continuous-wave (SFCW) or a frequency-modulated continuous-wave (FMCW) signal (commonly used in radars) which have a narrow bandwidth (∼500 MHz) can be used as an excitation pulse too, while a complicated circuit topology of pulse generation is necessity. Furthermore, a sinc function can be used as an excitation pulse, since its spectrum has a rectangular shape that preserves the most of the spectral energy inside the interested frequency range as shown in Fig to sinc function while it requires a careful adjustment of the pulse duration to eliminate the dc component [4]. For the fast measurements of MUTs using UWB systems, the time domain approaches can be utilized through a simple single UWB pulse covering the entire 3-10 GHz frequency range as similar to [5]. ...
... Excitation with a short duration pulse is often used to perform the fast spectroscopy -to cover the wide frequency range within short timeframe [1][2][3][4][5]. ...
... Variety of well known short excitation waveforms can produce wideband spectral distribution of power e.g. rectangular pulses, pseudo-random maximum length sequence (MLS) of rectangular pulses, Gaussian pulses and its derivatives, sinc and others [1][2][3][4][5][6]. Though a number of different waveforms are used, the chirp wave is expected to be the most suitable one in many cases [2][3][4][5]. ...
... rectangular pulses, pseudo-random maximum length sequence (MLS) of rectangular pulses, Gaussian pulses and its derivatives, sinc and others [1][2][3][4][5][6]. Though a number of different waveforms are used, the chirp wave is expected to be the most suitable one in many cases [2][3][4][5]. The power spectral density (PSD) of the chirp excitation pulse can cover tailored frequency range within given duration in time domain [4][5]. ...
Article
Full-text available
Wideband excitation is used to perform the fast spectroscopy - to cover the wide frequency range within short timeframe. Power spectrum of the chirp excitation pulse can cover tailored frequency range within given duration in the time domain. Linear chirp is a well-known wideband excitation waveform, where the instantaneous frequency varies linearly with time. Chirps with nonlinear relationship between time and frequency are also well known. This group of waveforms shape the power spectrum differently from its linear counterpart. The paper describes the results of the study of spectral properties of the nonlinearly modulated chirp signals, where the frequency of the signal varies according to multi-slope nonlinear relationship over the time. The advantage of this type of chirp signals is that the power spectrum of these signals has substantially steeper slope above the cut-off frequency, compared to its common linear and nonlinear counterparts.
... However, for the proposed system, a single Gaussian signal is not suitable to detect the effect of the MUT on the magnitude of the signal due to low amplitude at the lower and higher frequencies of the pulse spectrum. Although a wideband pulse can be generated utilizing the Gaussian derivatives, most of the energy still falls outside the interested frequency band as shown in Fig. 2(a,b) [5]. Furthermore, to utilize a homodyne transceiver, a 3-10 GHz image rejection circuit has to be implemented to avoid the effect of the dispersive material characteristics on the down-converted signal which is not easily implementable using CMOS process, otherwise; a heterodyne architecture has to be utilized which will be more complicated. ...
... Avoiding the DC component eliminates the flicker noise (1/f) and DC-offset effects in the RX unit resulting in SN R out improvement. The generated pulse has a similar characteristic properties with a haversinc function (sin 2 (ωt)/ωt) [5] as shown in Fig. 2(c,d). The proposed microwave frequency range (3-10 GHz) is then divided into 9 sub-bands with 100 MHz overlap from each side with an effective 800 MHz bandwidth (100-900 MHz centered at 500 MHz), they are up/down converted with LO frequencies of 2.8, 3.6, 4.4, 5.2, 6, 6.8, 7.6, 8.4, and 9.2 GHz as depicted in Fig. 1 Fig. 3. Circuit implementation of the proposed MBDS system including baseband pulse generator, transmitter, and receiver. ...
Article
This article presents an integrated time-domain (TD) microwave broadband dielectric spectroscopy (MBDS) system in the 3–10-GHz frequency span divided into nine subbands. The generated multitone input pulse with 1-GHz bandwidth is upconverted to RF frequencies in the transmitter using a broadband single-sideband (SSB) mixer to suppress the lower sideband of the excited signal and prevent the signal distortion at the configured dc-free direct conversion receiver with the center frequency of 500 MHz at the baseband. The upconverted pulse is passed through a contactless sensor in TD where the material under test (MUT) is placed in the middle of the two ultra-wideband (UWB) Vivaldi antennas coupled in their radiative near field. To identify both the real ( $\epsilon _{r}'$ ) and imaginary ( $\epsilon _{r}"$ ) parts of the MUT’s permittivity, the impact of the entire system on MUT characterization is alleviated by designing the system subblocks and the sensor with flat gain and constant group delay over the entire frequency range. The proposed MBDS system is fabricated in 65-nm CMOS and occupies an active area of 1.24 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> while consuming 64~69 mW from a 1-V supply. As a proof of concept, the complex dielectric permittivity of different pure organic chemical liquids has been detected within the desired frequency range with a customized quartz-glass cuvette with the MUT volume of ~ 5 mL by calibrating the sensor using the variations in the phase and the magnitude after FFT with respect to the reference material (air). The system results in an rms error of less than 0.2% and 0.4% for $\epsilon _{r}'$ and $\epsilon _{r}"$ , respectively, compared with theory. To the best of our knowledge, this is the first CMOS TD MBDS system with a homodyne RF transceiver architecture including an on-chip UWB excitation pulse generation and the contactless sensor.
Chapter
This chapter gives some background on impedance spectroscopy and tissue impedance but also about the instrumentation used for measuring transfer impedance of a biological material under study. The instrumentation concerned here is the current source for applying a current to tissue and the differential amplifier for measuring the resultant voltage. It shows also the main characteristics of the excitation source used for driving a constant current into the biological material under study and those of the receive circuit used to measure the resultant differential voltage. The major source of errors due to the nonideal characteristics of the instrumentation is also discussed, including the measuring circuit ones. The desired characteristics for a drive and measuring circuit on EIS are presented and some discussions are addressed.
Conference Paper
Electrical Impedance Spectroscopy (EIS) measurement of biological materials and systems allows to characterize the state of tissues, organs, etc. The study of a biological object whose electrical characteristics change quickly with time (lung, heart, etc) makes necessary to acquire information about the impedance modulation associated to the intrinsic cyclic movement of the tissue. Instead of using the classical frequency sweep, EIS should then be acquired performing a fast measurement using a broadband excitation signal in the desired frequency range. There are several signals proposed for biomedical applications such as pulses, filtered noise, pseudo-random binary sequences (PRBS), etc. The multisine burst signal satisfies optimization of Signal to Noise Ratio (SNR) at a limited set of spectral samples, limitation of energy to avoid biological stimulation and minimum measurement time. Equally spaced multisines have a lack of accuracy due to distortion. Pure logarithmic multisines are not optimized to detect impedance relaxation due to the fact that they group the excitation frequencies at the relaxation ends, which do not include much information. We propose a multisine signal based on a Bilateral Quasi-Logarithmic (BQL) frequency distribution which has more density around the central relaxation frequency. The initial bilateral logarithmically distributed frequency points are slightly moved to avoid harmonics and intermodulation effects at the measurement frequencies. Moreover, a genetic algorithm arranges the phase angles to minimize crest factor. This technique is implemented on a custom arbitrary signal waveform generator based on a Field Programmable Gate Array (FPGA) and a fast D/A converter. The first experiments have been carried out to validate the measurement method using a wideband front-end and an digital oscilloscope as acquisition system. Digital demodulation on a DSP for real time acquisition and processing is being developed. The application field is the real-time characterization of in-vitro and in-vivo engineered tissue for cardiac regeneration.
Conference Paper
Recent works indicate that through the non-invasive application of low current waveforms to the brain, systems implementing both cortical stimulation (Transcranial Direct Current Stimulation) and imaging (Electrical Impedance Tomography) with a high degree of accuracy and effectiveness can be realised. Safety and physical constraints along with the individually unique and fractal-like functional, structural and electrical complexity of the brain and surrounding cranial tissue hinders legacy systems from achieving enough precision and effectiveness for neurological treatment and investigation. This paper describes a novel system capable of achieving both stimulation and imaging while also ameliorating the shortcomings of legacy systems. Both these functions require multiple independently controlled electrodes distributed on the scalp and have complimentary functional requirements, therefore minimal additional resources are required to achieve both goals. Meeting these requirements also result in the ability to improve on legacy modalities. The system exceeds all appropriate safety requirements and is implemented with a modular architecture enabling cascading of the hardware such that the system capability and cost can be scaled according to the requirements of the application. While resource constraints precluded meeting critical functional requirements, tests and simulation of the subsystems proved the concept justifying further development.
Article
Electrochemical impedance spectroscopy (EIS) allows measuring the properties of the system as a function of the frequency as well as distinguishing between processes that could be involved: resistance, reactance, relaxation times, amplitudes, etc. Although it is possible to find related literature to in vitro and in vivo experiments to estimate glucose concentration, no clear information regarding the condition and precision of the measurements are easily available. This article first address the problem of the condition and precision of the measurements, as well as the effect of the glucose over the impedance spectra at some physiological (normal and pathological) levels. The significance of the measurements and the glucose effect over the impedance are assessment regarding the noise level of the system, the experimental error and the effect of using different sensors. Once the data measurements are analyzed the problem of the glucose estimation is addressed. A rational parametric model in the Laplace domain is proposed to track the glucose concentration. The electrochemical spectrum is measured employing odd random phase multisine excitation signals. This type of signals provides short acquisition time, broadband measurements and allows identifying the best linear approximation of the impedance as well as estimating the level of noise and non-linearities present in the system. All the experiments were repeated five times employing three different sensors from the same brand in order to estimate the significance of the experimental error, the effects of the sensors and the effect of the glucose over the impedance.
Article
Full-text available
Measuring the impedance frequency response of systems by means of frequency sweep electrical impedance spectroscopy (EIS) takes time. An alternative based on broadband signals enables the user to acquire simultaneous impedance response data collection. This is directly reflected in a short measuring time compared to the frequency sweep approach. As a result of this increase in the measuring speed, the accuracy of the impedance spectrum is compromised. The aim of this paper is to study how the choice of the broadband signal can contribute to mitigate this accuracy loss. A review of the major advantages and pitfalls of four different periodic broadband excitations suitable to be used in EIS applications is presented. Their influence on the instrumentation and impedance spectrum accuracy is analyzed. Additionally, the signal processing tools to objectively evaluate the quality of the impedance spectrum are described. In view of the experimental results reported, the impedance spectrum signal-to-noise ratio (SNRZ) obtained with multisine or discrete interval binary sequence signals is about 20–30 dB more accurate than maximum length binary sequence or chirp signals.
Article
Full-text available
The design and implementation of an instrument capable of simultaneously measuring bioimpedance on four channels at eight frequencies at a reasonable cost required a novel approach. Eight microampere-level sinusoidal excitation current generators operating at frequencies from 100 Hz to 5 MHz are multiplexed to external electrodes within a biological sample, while impedance signals from four pairs of electrodes are processed within the instrument. Four 16-bit A/D converters digitize the four analog signals, followed by detection using an original method of non-uniform synchronous undersampling. Instrumentation control, data acquisition and display software were developed in the LabView 7.1 VI environment. The software can also be used to run virtual experiments using a computer model (digital phantom) of an organ to be studied. This is useful for model development, planning of experiments and for training. The instrument will be used to characterize cardiac function as well as to assess vascular and respiratory systems.
Article
Bio-impedance measurements are widely used for characterization of biological objects. Although the measured impedance of such objects is independent of the measurement method used, slight differences between measurements in the frequency and time domain are found. For many practical applications time domain based measurements are advantageous, but they are often rejected as not accurate. In order to show their suitability for bio-impedance measurements we used a special arrangement of time domain and frequency domain based measurements at the same biological specimen (canine liver) with the same electrodes. A reasonable coincidence in the measurement results could be shown. Moreover we used only a fraction of the time domain measurement data in order to demonstrate a significant reduction in measurement time while maintaining a reasonable accuracy. An algorithm for fast processing of the time domain data without transformation into the frequency domain is provided.
Book
Key Features Second edition of this well regarded tet, with increased emphasis on bioelectricity, Clear explanation and development of key principles and mathematics to make the subject accessible to non-Physicists, New coverage of excitable tissue electrical properties, passive tissue electrical properties, geometrical analysis, dipole-dipole analytical solutions for transfer impedance, broader coverage of Cole theory, Geselowitz theory and multivariate statistical models, Clinical examples and problems to reinforce understanding, with solutions for adopting instructors from http://textbooks.elsevier.com. an excellent reference for students and researchers in those areas that make use of the concepts of bioimpedance and bioelectricity. (it provides), in a single work, an opportunity. to understand and compare the various methods of bioimpedance and bioelectricity measurement, as well as the several models used to describe and interpret the electrical behavior of biological tissues. Review of the First Edition, Annals of Biomedical Engineering.
Article
Bio-impedance measurements are widely used for characterization of biological objects. Although the measured impedance of such objects is independent of the measurement method used, slight differences between measurements in the frequency and time domain are found. For many practical applications time domain based measurements are advantageous, but they are often rejected as not accurate. In order to show their suitability for bio-impedance measurements we used a special arrangement of time domain and frequency domain based measurements at the same biological specimen (canine liver) with the same electrodes. A reasonable coincidence in the measurement results could be shown. Moreover we used only a fraction of the time domain measurement data in order to demonstrate a significant reduction in measurement time while maintaining a reasonable accuracy. An algorithm for fast processing of the time domain data without transformation into the frequency domain is provided.
Article
The objective of this study was to determine the efficacy of a two-electrode myocardial electrical impedance (MEI) monitor in reproducibly detecting induced myocardial ischemia by comparing MEI changes with hemodynamic changes, including sonomicrometric changes. With institutional approval, 80 dogs were anesthetized with sodium thiamylal, intubated, ventilated, and had venous, arterial, and pulmonary artery catheters placed. Medial sternotomy was performed to facilitate myocardial exposure and allow the left anterior descending coronary artery (LAD) to be isolated. Two pacing electrodes were attached to the myocardium to measure MEI with a monitor. Seventy dogs were randomly assigned to the 15, 30, 45, 60, or 120 min LAD occlusion group. Sonomicrometric transducers were attached to the myocardium of the ten remaining dogs and their LAD was occluded for 36 min. MEI increased immediately after LAD occlusion to a level significantly more (P < 0.05) than baseline and returned to the baseline level upon reperfusion. Twenty dogs developed ventricular fibrillation with no attempts at resuscitation. MEI changes paralleled the sonomicrometric changes expected with ischemia. No significant cardiovascular hemodynamic changes were found with less than 45 min of LAD occlusion. Sixty and 120 min LAD occlusion resulted in significant decreases in cardiac output. The results of these experiments demonstrate that the two-electrode MEI monitor reproducibly changes in response to myocardial ischemia.
Article
The microfabricated impedance spectroscopy flow cytometer used in this study permits rapid dielectric characterization of a cell population with a simple microfluidic channel. Impedance measurements over a wide frequency range provide information on cell size, membrane capacitance, and cytoplasm conductivity as a function of frequency. The amplitude, opacity, and phase information can be used for discrimination between different cell populations without the use of cell markers. Polystyrene beads, red blood cells (RBCs), ghosts, and RBCs fixed in glutaraldehyde were passed through a microfabricated flow cytometer and measured individually by using two simultaneously applied discrete frequencies. The cells were characterized at 1,000 per minute in the frequency range of 350 kHz to 20 MHz. Cell size was easily measured with submicron accuracy. Polystyrene beads and RBCs were differentiated using opacity. RBCs and ghosts were differentiated using phase information, whereas RBCs and fixed RBCs were differentiated using opacity. RBCs fixed using increasing concentrations of glutaraldehyde showed increasing opacity. This increased opacity was linked to decreased cytoplasm conductivity and decreased membrane capacitance, both resulting from protein cross-linking. This work presents label-free differentiation of cells in an on-chip flow cytometer based on impedance spectroscopy, which will be a powerful tool for cell characterization.
Article
The problems of application of pulse wave signals in electrical bioimpedance analyzers foreseen for using in implantable medical devices as diagnostical means are discussed in this paper. The main problem arises at measurement of phasor parameters by the aid of rectangular pulse wave signals. The specific measurement errors appear due to presence of higher harmonics in the spectra of pulse waveforms. These errors are discussed in two cases, in the case of full cycle rectangular waveform, and in the case of using the shortened pulses introduced specially for reduction of errors.
Conference Paper
The characterization of biological materials and systems using electrical impedance spectroscopy has traditionally been performed using the frequency sweep technique. When applied to in-vivo measurements, the movement induced modulation has often a period shorter than the sweep time. This drawback can be overcome using broadband signal bursts. Given that the energy amount to be injected to the biological material is limited for safety reasons, the best choice is the use of multisine signals, which concentrate all that energy in the measurement frequencies, then achieving an optimal signal-to-noise ratio. The uniform distribution of frequencies is not adequate due to the system nonlinearities and to the need of covering a three-decade frequency range. This work is concerned with the design of a quasilogarithmic multisine with a similar number of frequencies at each decade and with a safety band around each measurement frequency. This band will be free of harmonics and quadratic intermodulation products. The system has been implemented using a virtual instrument based on an arbitrary waveform generator, a digital oscilloscope and an analog frontend. The system has been validated using passive RC networks and has been applied to the in-vivo characterization of infarcted myocardium in pigs
Article
The discrete Fourier transform (DFT) is a widely used powerful signal analysis method. The method's discrete nature implies that the captured waveform being analyzed is a portion of a periodic signal. The procedure is called coherent sampling in which, no matter when the capture starts, the end point of the captured waveform in the same fashion as the next noncaptured sample. On the other hand, if a nonperiodic signal is captured, the captured waveform would be incoherently sampled and could not be connected to the first sample. In this situation, a special function needs to be applied to the captured waveform before applying the DFT. In window functions, the shapes of the windows are designed in such a way that the height reaches maximum at the center and gradually reduces to zero towards the edges. The windows are designed in such a way that its first derivative tends to be zero in the center and at both ends. However, since the procedure is two-dimensional, adding a third dimension can yield a solution that is not limited by the shape of windows. Real-valued window functions are limited to performing amplitude modulation, whereas complex-valued window functions have the advantage of being able to perform both amplitude and phase modulations. Such a window function has been found and has been applied to captured sine waves that include a fractional period of a sine wave and it produces no artifacts.