ArticlePDF Available

A FDR Sensor for Measuring Complex Soil Dielectric Permittivity in the 10-500 MHz Frequency Range

Authors:
  • Institute of Agrophysics of Polish Academy of Sciences

Abstract and Figures

Mechanical details as well as electrical models of FDR (frequency domain reflectometry) sensors for the measurement of the complex dielectric permittivity of porous materials are presented. The sensors are formed from two stainless steel parallel waveguides of various lengths. Using the data from VNA (vector network analyzer) with the connected FDR sensor and selected models of the applied sensor it was possible obtain the frequency spectrum of dielectric permittivity from 10 to 500 MHz of reference liquids and soil samples of various moisture and salinity. The performance of the analyzed sensors were compared with TDR (time domain reflectometry) ones of similar mechanical construction.
Content may be subject to copyright.
Sensors 2010, 10, 3314-3329; doi:10.3390/s100403314
sensors
ISSN 1424-8220
www.mdpi.com/journal/sensors
Article
A FDR Sensor for Measuring Complex Soil Dielectric
Permittivity in the 10–500 MHz Frequency Range
Wojciech Skierucha * and Andrzej Wilczek
Institute of Agrophysics, Polish Academy of Sciences, Doświadczalna 4, 20-290 Lublin, Poland;
E-Mail: a.wilczek@ipan.lublin.pl
* Author to whom correspondence should be addressed; E-Mail: w.skierucha@ipan.lublin.pl.
Received: 3 February 2010; in revised form: 5 March 2010 / Accepted: 19 March 2010 /
Published: 5 April 2010
Abstract: Mechanical details as well as electrical models of FDR (frequency domain
reflectometry) sensors for the measurement of the complex dielectric permittivity of porous
materials are presented. The sensors are formed from two stainless steel parallel waveguides
of various lengths. Using the data from VNA (vector network analyzer) with the connected
FDR sensor and selected models of the applied sensor it was possible obtain the frequency
spectrum of dielectric permittivity from 10 to 500 MHz of reference liquids and soil samples
of various moisture and salinity. The performance of the analyzed sensors were compared
with TDR (time domain reflectometry) ones of similar mechanical construction.
Keywords: FDR; TDR; dielectric permittivity; dielectric spectroscopy; soil moisture;
soil salinity
1. Introduction
The requirements for fast, reliable and automated spatially distributed soil water content monitoring
are not fully satisfied by the current commercially available technology measurements. The reasons for
that are high price of commercially available TDR meters working in the frequency range up to 1 GHz,
the need for site specific calibrations of FDR meters, especially for heavy and saline soils, or problems
with disposal of radioactive elements when the neutron scattering method is used. Scientists and
engineers are conducting research and building prototypes to develop new soil water and salinity
sensors and their physical models to increase measurement selectivity, accuracy, ease of operation and
availability, as well as to make them economically accessible for mass applications. Nondestructive
OPEN ACCESS
Sensors 2010, 10
3315
and in situ measurements of soil water content and salinity are fundamental in many agricultural
applications including hydrology, precision agriculture and irrigation scheduling [1–3]. The idea of
applying FDR sensors working in the frequency range above 150 MHz, which is the upper frequency
limit of currently available sensors, and reaching 500 MHz, seems to be promising, especially to
achieve more universal calibration than current FDR sensors and lower cost than the current TDR
meters [4].
At present, several methods are considered suitable for continuous soil water monitoring at the point
scale with minimum soil disturbance. Such methods are based on the dielectric properties of soil and
the propagation characteristics of electromagnetic (EM) waves in the soil, which change with water
content and salinity. A review of soil moisture and electrical conductivity measurement methods,
including the dielectric methods at different spatial and temporal scales is presented in [4]. The
relationship between the real part of the permittivity and soil volumetric water content is strong for the
quartz-dominated soils [5]. Although quartz-dominated materials show negligible frequency
dependence of dielectric properties, some clay minerals present dielectric dispersion [6–8]. As a result,
the relationship between soil moisture and electrical conductivity depends on the operating frequency
of the applied sensor. The least frequency dependence occurs above 500 MHz [7]. Also, accurate
measurement of the real part of the soil complex dielectric permittivity is required to fully characterize
its relationship with soil moisture in the frequency domain.
The aim of the study is to analyze the influence of soil water content and electrical
conductivity/salinity on the on the complex dielectric permittivity of soil measured by a developed
FDR (Frequency Domain Reflectometry) sensor and the measurement setup. The sensor and its
physical models are used to measure the complex dielectric permittivity in the 10–500 MHz frequency
range. The measurements are made in laboratory conditions using specialized equipment not suitable
for field use. The presented study is an initial step in works on field sensors and a prototype of a field
meter for the measurement of dielectric permittivity of soil in this frequency spectrum and hopefully
up to 1 GHz.
2. Sensor Description and Measurement Tools
The applied dielectric sensors were built with two stainless steel rods of 2 mm in diameter, 13 mm
separation distance and lengths of 1, 2 and 3 cm (Figure 1—left). The interface between the coax cable
and stainless steel rods was constructed to minimize the impedance discontinuity (Figure 1—right).
Figure 1. Two wire stainless steel rod FDR probes (left), the scheme of internal details
(middle) and the interface between coaxial and parallel waveguides (right).
Sensors 2010, 10
3316
The mechanical construction of these sensors is similar to the FP/m type manufactured in IA PAS
(Lublin, Poland) working with the TDR technique [9]. TDR probes with such short length are not used
because of timing restrictions imposed by this measurement method. The presented FDR probes with
parallel 1–3 cm length rods may fill the gap for dielectric sensors between open-ended coax [10] and
TDR sensors. The TDR measurements give averaged information about the apparent dielectric
permittivity and electrical conductivity of soil for a broad range of frequencies dependent n the rise
time of the analyzing pulse. In order to receive detailed frequency relations of the complex dielectric
permittivity it is necessary to apply FDR technique with appropriate models of the measurement
sensors. Several dielectric models of the constructed FDR probe were tested to select the one, which
performance on the reference liquids with known dielectric properties was superior over the others.
2.1. The Sensor Rods as a Loss Capacitor—the C0 Capacitance Model
This model is based on the open-ended coax sensor [11]. This type of sensor is very popular
because of the availability of an Agilent 85070E open-ended coaxial probe [12], which connected to an
Agilent Vector Network Analyzers equipped with specialized firmware, can produce a frequency
spectrum of the real and imaginary parts of the dielectric permittivity of the medium that is in contact
with this probe. Open-ended coax probes can be inserted in it, or used for non-invasive surface
measurements.
The termination of the open-ended sensor is described by capacitances Cf and C0
, where the former
is the fringing capacitance inside the coaxial waveguide of characteristic impedance ZC = 50 and the
latter is the fringing capacitance due to the test sample (Figure 2). The value of C0 represents the case
when the surroundings of the C0
capacitor is air.
Figure 2. Capacitance model CfC0 of an open-ended coax sensor.
The values of capacitance and loss of the C0
capacitor change with the real Re(
) and imaginary
Im(
) parts of the complex dielectric permittivity of the tested sample. The values of Cf and C0
are
determined for each angular frequency
= 2
f during the calibration process using dielectric media of
tabularized dielectric properties. The sensor impedance Zp is:
)(
1
0
CCj
Z
f
p
(1)
The Vector Network Analyzer (VNA) measures the reflection coefficient S11 defined as:
Sensors 2010, 10
3317
Cp
Cp
jZZ
ZZ
eSS
1111
(2)
where |S11| and are the module and phase of the complex reflection coefficient S11 measured by
the VNA.
Real and imaginary parts of
can be found after inserting (1) into (2) as shown below [13]:
)||cos||21(
||1
)Im(;
)||cos||21(
sin||2
)Re( 2
1111C0
2
11
0
f
2
1111C0
11 SSZC S
C
C
SSZC S
(3)
The capacitance C0 in a two wire sensor is much bigger as compared to Cf, which is not true for an
open-ended coax probe and Cf can be neglected [11] as presented in Figure 3.
Figure 3. Capacitance model C0 of a two-wire sensor in lossy dielectric medium.
The impedance of the C0 capacitor filled with lossy dielectric of the complex dielectric permittivity
equal to
is:
air
Z
Cj
Z
0
0
1
(4)
where Zair is the impedance of an air capacitor. The impedance Z0 of a two-wire sensor at the end of a
coaxial cable of characteristic impedance ZC is given as:
11
11
01
1
S
S
ZZ C
(5)
The formula for calculating the complex dielectric permittivity
of the measured medium can be
found after inserting (4) into (5):

mair
mair SS SS
1111
1111
11
11
(6)
where S11air and S11m are complex reflection coefficients measured by VNA in air and the measured
medium.
The capacitance model C0 does not include the effect of electromagnetic radiation from the sensor
open ending, which increases with the applied frequency f of the measurement signal. The radiation
increases the loss of the capacitor C0 and in consequence increases the imaginary part Im(
) of the
tested dielectric permittivity.
Sensors 2010, 10
3318
2.2. Four-Pole T Model of the Open-Ended Coax Sensor
A four-pole T model of the open ended coax sensor was described in [14] and it is a modified
capacitance model C0, where the capacitance Cf is replaced by a four pole type T in the form of three
impedances Z1, Z2, Z3 (Figure 4).
Figure 4. Four-pole T model of the open-ended coax sensor.
The impedance of the whole set can be substituted by ZT and rewritten in the linear form with
complex coefficients a = Z1Z2 + Z2Z3 + Z3Z1, b = Z1 + Z2 and c = Z2 + Z3 and then rewritten again to
incorporate the impedance of the sensor in air Zair:

T
air
T
air
TTT Z
Z
cZ
b
Za
ZZZcZba
ZZZ
ZZZ
ZZ
00
032
023
1
(7)
ZT can be calculated from the values of S11 measured by the VNA using Equation (5), where ZT will
replace Z0.
The parameters a, b and c can be determined by the measurement of ZT in three defined conditions:
short circuit of the sensor by means of mercury when Re(
) = and Im(
) = ,
measurement in air when Re(
) = 1, Im(
) = 0,
measurement in a liquid that has well defined dielectric properties by Deby’e or Cole-Cole
models, ex. distilled water, acetone or methanol (Table 2).
Having the parameters a, b, c from above described calibrations it is possible to calculate the
complex value of
after rearranging the Equation (7).
2.3. Impedance Model Zf – Z0
In this model the capacitances Cf and C0 (Figure 2) are replaced with the impedances Zf and Z0
because they include the conductances Gf and G0 describing the sensor attenuation and the loss of the
measured medium, respectively (Figure 5).
Figure 5. Localization of the sensor impedances in the model ZfZ0.
Sensors 2010, 10
3319
The impedance values of Zf and Z0 are determined for each measurement frequency during the
calibration process, similarly to the Cf and C0 values of an open-ended coax probe. The substitute
impedance Z of the impedances Zf and Z0 connected in parallel is:
0
1
1
ZZ
Z
f
(8)
Using VNA determined S11 and the formula (8) for two media, for example air and acetone, the
values of Z0 and Zf can be calculated from the following two equations:
00
1
1
;
1
1
ZZ
Z
ZZ
Zace
f
ace
air
f
air
(9)
where Zair is the sensor impedance in air, Zace is the sensor impedance in acetone,
air = 1+j0 is the
dielectric permittivity of air,
ace is the dielectric impedance of acetone determined from Debye model
using parameters from Table 1. Finally, the searched dielectric permittivity of the tested medium is
calculated from Equation (8).
2.4. Model of a Lossy Impedance Transformer—TT Model
For high frequency signals the two-wire sensor should be analyzed as a distributed parameter
system, opposite to a lumped parameter system, discussed in previous models. In this case a two-wire
sensor is a transmission line composed with four-pole lumped elements (Figure 6) of unit values of
resistance R, induction L, conductivity G and capacity C for the line unit length x [15].
Figure 6. Sensor rods as a transmission line.
The parameters describing the voltage and current in the transmission line are phase
(rad m–1),
attenuation
(dB m–1) and propagation
constants as well as characteristic impedance ZC of a
transmission line:
CjG LjR
ZCjGLjRj C
;))((
(10)
Assuming that soil is a paramagnetic material with relative magnetic permittivity
r = 1 and the
value of L does not depend on soil properties, the unit resistance R of the rods becomes zero and the
unit value of C can be substituted by C0
andEquation(10) can be substituted by:
Sensors 2010, 10
3320
air
CZ
Z
j;
c (11)
where c is light velocity in free space.
The termination of the open line ZK is modeled as a resistance of the 10 M value. It is transformed
to the impedance of the sensor rods beginning as follows [17]:
xZZ xZZ
ZZ
KC
CK
CL
tgh
tgh
(12)
Inserting
and ZC from Equation (11) to the Equation (12) gives Equation (13) as the impedance ZL
of the sensor rods of the x length inserted into the tested medium of known complex dielectric
permittivity
. It is much more difficult to determine in the analytical way the transposed function
= f(ZL). However the frequency spectrum of
can be found numerically:
c
tgh1
c
tgh
xj
Z
Z
xj
Z
Z
Z
air
K
air
K
L
(13)
Treating the sensor rods as a lossy impedance transformer enables to find the dielectric permittivity
values independent on the length of the sensor rods, but a new problem appears with the ambiguity of
the solution. The tangent hyperbolic function is periodic for the imaginary part of the argument. The
condition for the multiple solution of the Equation (13) is:
f
xc
for2
(14)
where: c is the velocity of light in free space, f – frequency of the applied measurement signal,
wavelength of the signal in a material of the refractive index
, x length of the sensor rods.
If condition (14) is met for the whole range of ZL values, multiple solutions can be obtained for
.
Inserting ZL as Zp into Equation (2) it is possible to visualize, with the help of the Matlab R2008b
program [16], the relation of the complex function
= f(S11), which is presented in Figure 7. The
assumed values of x = 10 cm, f = 500 MHz. For a line perpendicular to the S11 plane the values of
are
ambiguous. Therefore the numerical solution is enhanced using a tracing procedure that first calculated
at low frequencies, for which Equation (14) assures a single solution. Increasing measurement
frequency results in meeting conditions for two or more solutions. To distinguish these solutions the
procedure utilizes the former one to apply new limits of the function variability. Such approach
enables to interpret the measurements in high frequencies where the wavelength of the measurement
signal is shorter than the double length of the sensor rods. The detailed description of applied
algorithm and calculation software is presented in [18].
The calibration of the TT sensor model needs the application of at least three calibration points, for
example: (i) impedance of the shorted sensor at the place where the rods are connected with the epoxy
resin enclosure (Z = 0 + j0), (ii) impedance of the sensor in air (Z = 1 + j0), (iii) impedance of the
sensor fully inserted in acetone (or other calibration liquid from the Table 2) calculated from the
Debye model. The calibration details are described in [18].
Sensors 2010, 10
3321
Figure 7. Four-dimension S11/
chart (complex values) based on the model of lossy
impedance transformer (left) and its view in the Re(
)/Re(S11) plane (right).
3. Materials and Methods
The porous media used for testing the sensors consisted four mineral soils taken from 10–20 cm
layer below the ground surface that were characterized by parameters presented in Table 1. The soil
data were taken from the data bank of Polish soils [19].
Table 1. Localization and selected physical parameters of tested soils.
No. Soil type Soil localization Dry bulk density
(gcm–3)
Soil texture by FAO (%) Specific
surface (m2g–1)
Water holding
capacity
(kg kg–1)
Sand Silt Clay
569 Brown Majdan
Skierbieszowski 1.33 71 25 4 21 0.39
610 Brown Kol. Olempin 1.59 94 5 1 9 0.13
622 Chernozem Terebin 1.4 60 34 6 37 0.36
529 Black soil Annopol 1.76 100 0 0 8 0.16
The soils were especially selected to minimize all the factors other than water content and salinity
that influence the value of dielectric permittivity, i.e., non swelling or shrinking soils or those with no
organic content. There were 10 soil samples with different water content, evenly spaced covering the
range from air dry to saturation, for each soil 569, 610 and 622 (Table 1). The water holding capacity
was calculated by dividing the weight of the water held in saturated sample by the sample dry weight.
Homogeneity of water content in the samples was achieved by sealing the samples and maintaining
them at 45 °C during three days.
Because of non-homogeneity of porous materials, especially soil, the liquids of known dielectric
parameters (distilled water, acetone, methanol, isopropanol) described by Debye model [20] given by
Equation (15) were used to assess the accuracy of the applied FDR probe models (Table 2).
Sensors 2010, 10
3322

j
s
1 (15)
where
and
s are the values of dielectric permittivity above relaxation frequency and in static
electric field, respectively,
is the relaxation time of the dielectric medium.
Table 2. Parameters of Debye model of the applied reference liquids [20].
Name Temperature (C)
s
(ps)
Distilled water 20 80.4 5.2 9.45
Acetone 20 21.2 1.9 3.34
Methanol 20 33.64 5.7 53
Isopropanol 20 29 3.2 292
The influence of salinity on the complex dielectric permittivity was tested on 5 series (each with 10
samples) of the black soil 529. The samples were moistened with KCl solutions with variable electrical
conductivity values 0, 5, 10, 15 and 20 dSm–1. Similarly to the soils 569, 610, 622, the water content in
the samples were evenly distributed in steps of 0.1 of the maximum soil water capacity value.
Measurement signal generation as well as detection of the reflected one from the sensor is done by a
vector network analyzer (VNA) type ZVCE from Rohde and Schwarz [21] working in the reflection
mode in the frequency range 20 kHz – 8 GHz. It measured a complex reflection coefficient S11 when
the sensor rods were fully inserted into the tested material. The ambient temperature in was
20.6 ± 1 C. According to the instructions, before each series of measurement the VNA was warmed
for at least one hour.
Figure 8. Program window for processing the raw S11data measured by VNA and applied
to the models: C0 (Figure 3), CfC0 (Figure 2), four-pole T (Figure 4), ZfZ0 (Figure 5)
and TT (Figure 6), which are selected by the user.
Sensors 2010, 10
3323
The raw data S11 from calibration measurements, as well as final measurements in the tested
materials, were processed by a custom software application written in C++ Builder, where the final
frequency spectrum of the complex dielectric permittivity was calculated and presented in graphic
form. for each of the discussed model. Figure 8 presents the case with selected TT model. Real and
imaginary parts of the complex dielectric permittivity are shown at the upper and the lower part of the
diagram, respectively. More details of the application program are available in [18].
Evaluation of the sensor models used to determine the complex dielectric permittivity of tested
media were performed by calibration of the sensors in reference liquids (Table 2), measurements by
VNA in methanol and verification the calculated real and imaginary parts of the complex dielectric
permittivity spectrum with the one calculated with the use of Debye model. The influence of the
measurement frequency on the measured refractive index were analyzed for data collected on soil 529
samples using the sensor of 3 cm rods length.
3. Results and Discussion
3.1. Model Selection
The comparison of the complex dielectric permittivity values calculated on the base of calibration
measurements in reference liquids for various models described in earlier sections is presented in
Figure 9.
Figure 9. Performance comparison of the analyzed FDR sensor models for methanol as the
verification medium.
Sensors 2010, 10
3324
The artifacts on each graph (Figure 9), as well as in the respective graph of the complex reflection
coefficient S11 measured by VNA observed for the same frequencies (Figure 10), are discussed in [22]
for open-ended coax sensors. They are caused by resonance effects in an open transmission line. This
happens when the length of a cable with a sensor equals to the odd multiple wavelength of the
measurement signal. This was confirmed in separate tests not presented in the current study. These
effects resulted in limiting the frequency range of the measurement signal to 10–500 MHz.
Figure 10. Resonance effects (in circles) on the S11 parameter measured by VNA.
It is evident from Figure 9 that the model C0 can be successfully used only for the correct
determination of the real value of the complex dielectric permittivity Re(
) for the frequencies up to
200 MHz. The imaginary part Im(
) has the measurement error above 10% for all applied frequencies.
Similar results were found for the model CfC0 (Figure 2), which was not presented in Figure 9. This
can be explained by a minimal influence of the Cf capacitance in the two-wire sensor.
Better results were found for the impedance model ZfZ0, especially for the imaginary part Im(
).
However the useful frequency did not increase above 200 MHz. The graphs in Figure 9 concerning the
models C0 and ZfZ0 only refer to the sensor length equal to 3 cm, because similar results were
observed for other sensor length 1 cm and 2 cm.
The four-pole T model performs better for both components of the dielectric permittivity, but for a 3
cm length probe the upper measurement frequency limit is still around 200 MHz and 350 MHz for the
2 cm length sensor. This model correctly follows the values of Debye model calculated for methanol as
the reference liquid for the sensor of 1 cm rod length.
The best performance was observed for the model of lossy impedance transformer (TT model). The
real part of the dielectric permittivity is not dependent on the sensor length for the whole frequency
range of the measurement signal. Only a slight decrease of results are observed, especially for the real
part. This effect can be explained by the probable increase of dielectric permittivity of acetone used for
the sensor calibration. The sensor length affects the performance of the imaginary part of the model
and the length is shorter, the measurement error of Im(
) is bigger.
Sensors 2010, 10
3325
The combination of calibration points of the TT model also influenced the measurement error of

for both components of the dielectric permittivityThe best results were obtained for the combination
of water, air and acetone, where for Re(
) 2 the measurement error of the real part of dielectric
permittivity is less than 4% and it is below 2.2% for Re(
) 2.
The TT model was applied to the sensor of 10 cm length of parallel metal rods and its performance
is presented in Figure 11. In the frequency range 125–375 MHz the real part of dielectric permittivity
of methanol has a big maximal measurement error reaching 20% of the measured value, but its average
value is very close to the model value. The imaginary part of
behaves similarly. The erroneous effect
can be explained with the resonance of the open line described earlier in Figure 10 and they increase
with the increase of the sensor rods’ length. It seems that the TT model could be applied for 10 cm
length sensors which have bigger measurement volume as compared with 1–3 cm long ones. However
before that it is necessary to eliminate the resonance effect and develop a new calibration methodology
of these sensors.
Figure 11. Performance of TT model for the FDR sensor with 10 cm length of rods.
3.2. Influence of Measurement Frequency on the Measured Refractive Index
Analysis of the measurement frequency influence on the measurement result was done by the
comparison of the mean values of the refractive index for six selected frequency bands. The refractive
index defined as the square root of the real part of the dielectric permittivity was calculated from the TT
model of a lossy impedance transformer. The selected frequency bands were: 1030, 3550, 80120,
125190, 245300, 390485 MHz. These bands were not burdened with errors caused by the open
transmission line resonance described in the previous section.
Figure 12 presents the relation between the soil refractive index Re(
) and volumetric water
content for the soil 529 for various frequency ranges.
Sensors 2010, 10
3326
Figure 12. The influence of the measured frequency range on the value of the soil
refractive index for the samples of soil 529 wetted with distilled water (left-hand graph)
and 20 dSm–1 KCL solution (right-hand graph).
For soil samples wetted with distilled water the refractive index increases with the frequency
decrease for the whole range of soil moisture and the trend lines representing various frequency ranges
are nearly in parallel (
Figure 12—left). Trend lines for all frequency ranges have the determination coefficient R2 not less
than 0.989. For figure clarity, only extreme trend lines are presented.
For soil samples wetted with KCl solution of the electrical conductivity equal to 20 dSm–1, the
lower the measurement frequency, the bigger was the refractive index of soil samples with the same
thermogravimetric water content. In the frequency range 390–485 MHz the relation can be described
by a linear relation with determination coefficient equal to 0.997. For the frequency range 10–30 MHz
this relation is extremely far from linear and must be fitted with the polynomial of the 3rd degree to
achieve R2 = 0.994. This means that for the tested salted sandy soil low frequency dielectric
measurements of soil moisture require soil specific calibrations. However, it should be noted that the
shape of this calibration still depends on the salinity level.
3.3. FDR Soil Moisture Calibration
The analysis included data for the FDR sensor with 3 cm rod length and 390485 MHz frequency
range of the measurement signal, for which the real part of dielectric permittivity does not show any
frequency dependence. Data for 20 measurement points of Re(
) from this range were averaged for
each soil sample to calculate the square root as the refractive index. The standard deviation for all
measurements did not exceed 0.16 of the absolute value. The correlation of final data with the
corresponding thermogravimetric data is presented in
Sensors 2010, 10
3327
Figure 13. FDR soil moisture calibration for the measurement frequency range 390–480 MHz.
The fitted linear functions are presented for comparison with literature data. The trend line fitted to
the all data points is:
10
Re aa
, where a0 = 8.86, a1 = 1.48.
Table 3. Comparison of FDR calibrations of the analysed sensors with Malicki TDR calibrations [23].
1
0
Re aa
FDR measurements
(Figure 13) Malicki model Malicki model
(correction on soil density)
a0 8.86 7.16 8.96
a1 1.48 1.44 1.44
The values of the a0 and a1 coefficients determined for the Malicki model from the TDR calibration
measurements are very close to the ones from the presented FDR calibration. This is especially true for
density corrected calibration. This means that in the measurement frequency range of 390–480 MHz
the soil moisture values determined for presented FDR sensor and the applied sensor model are
comparable with the TDR ones.
4. Conclusions
The presented two-wire FDR sensors and the measurement methodology of the frequency spectrum
of the complex dielectric permittivity enable simultaneous and selective determination of soil
moisture for various frequency ranges of the measurement signal. The developed lossy impedance
transformer—TT model of the FDR two-wire sensor can be used to measure the frequency spectrum of
soil complex dielectric permittivity in the range 10–500 MHz, that up till now is not used in
commercial FDR meters of soil moisture and salinity. It was found that for the FDR measured soil
samples the real values of the complex dielectric permittivity determined for low frequencies
(10–50 MHz) are higher in relation to the ones determined for high frequencies (300–500 MHz),
Sensors 2010, 10
3328
especially for saline soils. The soil moisture values determined for the chosen mineral soil samples by
the applied FDR method and sensors are comparable to the ones determined by the TDR method.
References
1. Robinson, D.A.; Jones, S.B.; Wraith, J.M.; Or, D.; Friedman, S.P. A review of advances in
dielectric and electrical conductivity measurement in soils using time domain reflectometry.
Vadose Zone J. 2003, 2, 444-475.
2. Walker, J.P.; Willgoose, G.R.; Kalma, J.D. In situ measurement of soil moisture: a comparison of
techniques. J. Hydrol. 2004, 293, 85-99.
3. Jones, S.B.; Blonquist, J.M.; Robinson, D.A.; Rasmussen, V.P.; Or, D. Standardizing
characterization of electromagnetic water content. Vadose Zone J. 2005, 4, 1048-1058.
4. Robinson, D.A.; Campbell, C.S.; Hopmans, J.W.; Hombuckle, B.K.; Jones, S.B.; Knight, R.;
Ogden, F.; Selker, J.; Wendroth, O. Soil moisture measurement for ecological and hydrological
watershed-scale observatories: a review. Vadose Zone J. 2008, 7, 358-389.
5. Topp, G.C.; Davis, J.L.; Annan, A.P. Electromagnetic determination of soil water content:
measurements in coaxial transmission lines. Water Resour. Res. 1980, 16, 574-582.
6. Logsdon, S.; Laird, D. Cation and water content effects on dipole rotation. Soil Sci. Soc. Am. J.
2004, 68, 1586–1591.
7. Kelleners, T.J.; Robinson, D.A.; Shouse, P.J.; Ayars, J.E.; Skaggs, T.H. Frequency dependence of
the complex permittivity and its impact on dielectric sensor calibration in soils. Soil Sci. Soc. Am.
J. 2005, 69, 67-76.
8. Chen, Y.P.; Or, D. Geometrical factors and interfacial processes affecting complex dielectric
permittivity of partially saturated porous media. Water Resour. Res. 2006, 42, W06423.
9. Skierucha, W.; Malicki, M.A. TDR Method for the Measurement of Water Content and Salinity in
Porous Media; Institute of Agrophysics PAS: Lublin, Poland, 2004; pp. 1-152.
10. Skierucha, W.; Walczak, R.T.; Wilczek, A. Comparison of open-ended coax and TDR sensors for
the measurement of soil dielectric permittivity in microwave frequencies. Int. Agrophys. 2004, 18,
355-362.
11. Stuchly, M.A.; Stuchly, M.M. Coaxial line reflection methods for measuring dielectric properties
of biological substances at radio and microwave frequencies—a review. IEEE Trans. Instr.
Measur. 1980, IM-29, 176-182.
12. Agilent 85070E Dielectric Probe Kit 200 MHz to 50 GHz. Technical Overview; Agilent
Technologies: Santa Clara, CA, USA, 2008.
13. Athey, T.W.; Stuchly, M.M.; Stuchly, S.S. Measurement of radio frequency permittivity of
biological tissues with an open-ended coaxial line: Part I. IEEE Trans. Microwave Theory Techn.
1982, 30, 82-86.
14. Wei, Y.Z.; Sridhar, S. Technique for measuring the frequency dependent complex dielectric
constants of liquids up to 20 GHz. Rev. Sci. Instrum. 1989, 60, 3041-3046.
15. Magnusson, P.C.; Alexander, G.C.; Tripathi, V.K.; Weisshaar, A. Transmission Lines and Wave
Propagation; CRC Press LLC: Boca Raton, FL, USA, 2001.
Sensors 2010, 10
3329
16. Knight, A. Basics of Matlab and Beyond; Chapman & Hall/CRC Press, Inc.: Boca Raton, FL,
USA, 2000.
17. Chen, L.F.; Ong, C.K.; Neo, C.P.; Varadan, V.V.; Varadan, V.K. Microwave Electronics
Measurement and Materials Characterization; John Wiley & Sons Ltd.: Chichester, UK, 2004;
pp. 1-552.
18. Wilczek, A. Influence of moisture and salinity of soil on its dielectric permittivity determined by
frequency domain reflectometry. Ph.D. thesis (in Polish), Institute of Agrophysics PAS in Lublin:
Lublin, Poland, 2009.
19. Glinski, J.; Ostrowski, J.; Stepniewska, Z.; Stepniewski, W. Bank of soil samples representing
mineral soils in Poland (in Polish). Probl. Agrofiz. 1991, 66, 1-57.
20. Buckley, F.; Maryott, A.A. Table of Dielectric Dispersion Data for Pure Liquids and Dilute
Solutions; Bureau of Standards Circular: Washington, DC, USA, 1958; pp. 1-101.
21. Operating Manual, Vector Network Analyzer ZVR/ZVRE/ZVRL; Rohde & Schwarz: Munich,
Germany, 2001.
22. Sheen, N.I.; Woodhead, I.M. An open-ended coaxial probe for broad-band permittivity
measurement of agricultural products. J. Agric. Engineer. Res. 1999, 74, 193-202.
23. Malicki, M. A.; Plagge, R.; Roth, C.H. Reduction of soil matrix effect on TDR di-electric moisture
determination by accounting for bulk density or porosity. Eur. J. Soil Sci. 1996, 47, 357-366.
© 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.
This article is an open-access article distributed under the terms and conditions of the Creative
Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
... Using sensors in precision agriculture has been widely investigated in the literature, mostly using sensors to measure soil moisture [3][4][5] or using Frequency Domain Reflectometry (FDR) or Time Domain Reflectometry (TDR) to detect variations in the moisture content of soil [6,7]. There are approaches in the literature in which capacitive fringing fields use PCBs to measure the dielectric materials' variations in the soil [8,9]. ...
... The best fit that reflects the data trend of the measured data is the exponential function that shows the relationship between capacitance and moisture. The general form of the exponential function is shown in Equation (6) [21]: (6) where α is the value of y when x is set to Zero and β is the growth factor during each unit period of time. To find the values of α and β, the above equation is linearised, as shown in Equation (7). ...
Article
This paper focuses on building a non-invasive, low-cost sensor that can be fitted over tree trunks growing in a semiarid land environment. It also proposes a new definition that characterizes tree trunks' water retention capabilities mathematically. The designed sensor measures the variations in capacitance across its probes. It uses amplification and filter stages to smooth the readings, requires little power, and is operational over a 100 kHz frequency. The sensor sends data via a Long Range (LoRa) transceiver through a gateway to a processing unit. Field experiments showed that the system provides accurate readings of the moisture content. As the sensors are non-invasive, they can be fitted to branches and trunks of various sizes without altering the structure of the wood tissue. Results show that the moisture content in tree trunks increases exponentially with respect to the measured capacitance and reflects the distinct differences between different tree types. Data of known healthy trees and unhealthy trees and defective sensor readings have been collected and analysed statistically to show how anomalies in sensor reading based on eigenvectors and eigenvalues of the fitted curve coefficient matrix can be detected.
... For assessing crop growth and detecting soil moisture, soil moisture sensors are incredibly helpful tools [16]. Soil moisture sensors are a useful tool for a variety of applications, including domestic gardens, landscapes, rainfall monitoring, environmental testing, hydrology, and precision agriculture [17]. They also detect the quantity of water content in the root zone. ...
Article
The soil whose main role is to provide nutrients in the process of plant growth is the foundation and an important part of agriculture. Soil testing is an important step for increasing agricultural production and raising farm income. Traditional soil testing methods rely on chemical processes carried out in a laboratory. Sensors play a significant role in agriculture by direct measurement of soil chemistry through tests such as pH, moisture, nutrient content, humidity and temperature. The results of soil tests are important to get high yield with good quality produce. ISE (Ion Selective Electrode) and ISEFT (ion-sensitive field effect transistor) sensors have also been used to detect the uptake of ions by plants.
... The gravimetric method was recognised as a reference technique [12]. The method for point in situ measurements is currently mainly based on TDR [13] or FDR [14] probes. The main disadvantage of point measurements is related to the high variety of soil moisture in the environment caused by the high relationship of SM values with soil physical parameters [15,16]. ...
Article
Full-text available
The soil moisture at the medium spatial scale is strongly desired in the context of satellite remote sensing data validation. The use of a ground-installed passive L-band radiometer ELBARA at the Bubnów-Sęków test site in the east of Poland gave a possibility to provide reference soil moisture data from the area with a radius of 100 m. In addition, the test site comprised three different land cover types that could be investigated continuously with one day resolution. The studies were focused on the evaluation of the ω-τ model coefficients for three types of land cover, including meadow, wetland, and cropland, to allow for the assessment of the soil moisture retrievals at a medium scale. Consequently, a set of reference time-dependent coefficients of effective scattering albedo, optical depth, and constant-in-time roughness parameters were estimated. The mean annual values of the effective scattering albedo including two polarisations were 0.45, 0.26, 0.14, and 0.54 for the meadow with lower organic matter, the meadow with higher organic matter, the wetland, and the cropland, respectively. The values of optical depth were in the range from 0.30 to 0.80 for the cropland, from 0.40 to 0.52 for the meadows (including the two investigated meadows), and from 0.60 to 0.70 for the wetland. Time-constant values of roughness parameters at the level of 0.45 were obtained.
... This is an artefact of the limited accuracy and sensitivity of the setup within the range of values. In Masrakin et al. [9] and Skierucha [10], the probe used had direct contact with the material under test and had defined characteristics that were used to create and decouple the resonance circuit. However, in the case of a PCB with a cell, there is a hindrance in ensuring direct contact with the battery without its outer casing. ...
Article
Full-text available
Monitoring battery aging is crucial for maintaining reliability and performance. This study investigates Frequency Domain Reflectometry (FDR) as a tool for monitoring lithium-ion battery State-of-Health (SoH). While FDR has been applied in battery research, the existing literature fails to address SoH assessment and lacks studies on larger battery samples to provide more meaningful results. In this work, nineteen cells initially underwent Electrochemical Impedance Spectroscopy (EIS) to assess their degradation levels during cyclic aging. This work evaluates FDR’s effectiveness in monitoring battery health indicators, such as capacity and equivalent series resistance (ESR), by correlating these with FDR-measured impedance between 300 kHz and 1 GHz. Analytical comparison between impedance measured before and after de-embedding processes were presented. The results show FDR reactance within 300 kHz–40 MHz correlates with EIS-measured ESR, suggesting its potential as a SoH indicator. However, reduced sensitivity and accuracy, particularly after de-embedding, may limit practical applicability. Additionally, resonance-based analysis was conducted to explore the relationship between changes in circuit resonance and cell dielectric permittivity. Despite having the lowest sensitivity, the method showed that the resonance frequencies of cells remain relatively constant, mirroring behaviours associated with changes in resistive properties. Overall, this study provides insights into FDR’s potential for battery diagnostics while highlighting avenues for future research to enhance effectiveness in real-world scenarios.
... Free water refers to the water flowing freely in the soil pores, while hygroscopic water is water adsorbed by molecular suction in the soil particles. Among these, free water serves as the primary source of water for plant growth [3,4]. The nuclear magnetic resonance (NMR) method proves instrumental in measuring soil moisture content, the composition of various water types, soil porosity, compaction degree, and water movement characteristics in the soil. ...
Article
Full-text available
The application of inside-out magnetic resonance sensors has emerged as a focal point in soil moisture detection, offering a non-invasive approach crucial for evaluating soil physical structure. While existing inside-out magnetic resonance sensors with a unilateral structure can effectively measure the shallow soil layer, they fall short in assessing deeper soil layers (1–10 cm). However, the lack of optimal design and horizontal comparison among relevant sensors has impeded the widespread adoption of magnetic resonance sensors in soil moisture measurement. This paper addresses this gap by optimizing a magnetic resonance sensor with a dumbbell structure. The magnet structure is fine-tuned using a non-derivative optimization method, specifically the coordinate search, under practical engineering constraints to establish a uniform field in the region of interest. Employing the coordinate search method enhances the sensor’s relative signal-to-noise ratio (SNR) to its maximum. Artificial soil samples are then tested, and the T2 spectrum distribution reveals insights into water distribution characteristics across different pores.
... However, the method is affected by soil voids, salinity or electrical conductivity, and temperature (Xu et al., 2012); the temperature mainly affects the permittivity of soil unfrozen water by phase transition (Guo et al., 2018). The soil density and water near the electrode affect the measurement results (Skierucha and Wilczek, 2010). Temperature calibration is required for the data obtained from long-term fixed point real-time monitoring of soil water content changes (Dirksen and Hilhorst, 1994). ...
Chapter
IOUT has the ability to transform sewer control system. It has the ability to monitor parameters like flow speed, flow rate, etc. The monitoring capabilities of IOUT, when combined with the existing Gray infrastructure (pipes and sewer systems), and communication system can advance the ongoing research efforts in attaining an autonomous sewer system. In this chapter, sewer IOUTs are developed which use underground sensor nodes to gather and communicate the information to the cloud or central server for real-time decision-making.
Article
The global population is constantly increasing, reached 8 billion in November 2022 and is expected to reach 9 billion by 2037. This increased population is expected to increase the demand for food, clothing and shelter, which in turn are heavily dependent on limited water resources. The available freshwater resources for agricultural use are further declining due to overexploitation and changing climate in the major food baskets of the world. This increasing water scarcity is exacerbated by expanding cities due to increasing urbanization. This calls for a new look at the allocation of water to agriculture. Therefore, the development of new strategies to improve agricultural water use may serve as an important adaptation strategy. This review attempts to include a comprehensive review of the literature on (i) the status and definition of water productivity and (ii) factors responsible for low water productivity (WP) in Asian agriculture. Furthermore, it contains practical approaches to enhance water use efficiency at the farm level covering all field crops and a range of soil types, which include (i) agronomic interventions; (ii) genetic interventions, such as the identification and cultivation of crop cultivars with high WP; and (iii) genotype, environment and crop management interactions for higher WP.
Article
Full-text available
An open-ended coaxial line is analyzed as a sensor for in vivo measurement of the complex permittivity of biological substances. The measurement system, based on two computerized network analyzers operating at frequencies from 10 MHz to 1 GHz, is described. Experimental results are presented in a companion paper.
Article
Full-text available
At the watershed scale, soil moisture is the major control for rainfall-runoff response, especially where saturation excess runoff processes dominate. From the ecological point of view, the pools of soil moisture are fundamental ecosystem resources providing the transpirable water for plants. In drylands particularly, soil moisture is one of the major controls on the structure, function, and diversity in ecosystems. In terms of the global hydrological cycle, the overall quantity of soil moisture is small, ∼0.05%; however, its importance to the global energy balance and the distribution of precipitation far outweighs its physical amount. In soils it governs microbial activity that affects important biogeochemical processes such as nitrification and CO2 production via respiration. During the past 20 years, technology has advanced considerably, with the development of different electrical sensors for determining soil moisture at a point. However, modeling of watersheds requires areal averages. As a result, point measurements and modeling grid cell data requirements are generally incommensurate. We review advances in sensor technology, particularly emerging geophysical methods and distributed sensors, aimed at bridging this gap. We consider some of the data analysis methods for upscaling from a point to give an areal average. Finally, we conclude by offering a vision for future research, listing many of the current scientific and technical challenges.
Article
The influence of soiľs solid phase on the dielectric constant of the soil over a range of moisture contents has been studied. Samples of soil, soil-like, and also other porous materials were analysed using Time Domain Reflectometry (TDR) to determine the contribution of bulk density and porosity to the function that relates dielectric constant to water content. The study showed that bulk density, and thus also porosity, substantially affects the relation between dielectric constant and water content. Two equivalent, empirical, normalized conversion functions were found, one accounting for bulk density and the other for porosity. Each of them reduced the root mean square error of the dielectric TDR determinations of moisture to 0.03, regardless of the materialľs bulk density and porosity.
Article
The dependence of the dielectric constant, at frequencies between 1 MHz and 1 GHz, on the volumetric water content is determined empirically in the laboratory. The effect of varying the texture, bulk density, temperature, and soluble salt content on this relationship was also determined. Time‐domain reflectometry (TDR) was used to measure the dielectric constant of a wide range of granular specimens placed in a coaxial transmission line. The water or salt solution was cycled continuously to or from the specimen, with minimal disturbance, through porous disks placed along the sides of the coaxial tube. Four mineral soils with a range of texture from sandy loam to clay were tested. An empirical relationship between the apparent dielectric constant K a and the volumetric water content θ v , which is independent of soil type, soil density, soil temperature, and soluble salt content, can be used to determine θ v , from air dry to water saturated, with an error of estimate of 0.013. Precision of θ v to within ±0.01 from K a can be obtained with a calibration for the particular granular material of interest. An organic soil, vermiculite, and two sizes of glass beads were also tested successfully. The empirical relationship determined here agrees very well with other experimenters' results, which use a wide range of electrical techniques over the frequency range of 20 MHz and 1 GHz and widely varying soil types. The results of applying the TDR technique on parallel transmission lines in the field to measure θ v versus depth are encouraging.
Article
A technique for the measurement of the frequency‐dependent complex dielectric constant ϵ=ϵ’-jϵ‘ of liquids applicable to frequencies up to at least 20 GHz is described. The technique utilizes a coaxial probe dipped into the liquid. The reflection coefficient of the probe was measured using a network analyzer between 45 MHz and 20 GHz. A new de‐imbedding procedure for eliminating the connector and coax mismatches is described, during which the impedance Z(ϵ) of the probe end was determined. The dielectric constant ϵ was determined from Z(ϵ) by modeling the coax–liquid interface as a capacitance. The de‐imbedding procedure, which utilizes three calibrations, directly eliminates the (unknown) fringe‐field impedance. Radiation effects were minimized by using narrow (0.047‐in.) semirigid coax. The technique yields accurate results for ϵ’ and ϵ‘ of liquids such as methanol and water over the entire frequency range up to 20 GHz, and can be used to determine the relaxation spectra of liquid and liquidlike samples.
Article
The moisture content of most agricultural materials is closely related to their dielectric properties, and to characterize their response, measurement over a broad frequency range is desirable. A coaxial probe and calibration system has been designed for measurement of the permittivity of semi-solid or liquid samples over the frequency range 1 MHz–6 GHz. The reflection coefficient of the probe in the sample was measured using a vector network analyser. Calibration was achieved by using three references: an open circuit, short circuit (indium foil) and a reference liquid which was chosen so its permittivity was similar to that of the sample. The sample permittivity was calculated by an accurate numerical model which used as inputs, the true reflection coefficient obtained from the VNA, and physical measurements of the geometry of the system.The measurement system was tested using reference liquids, and the experimental configuration was optimized for accuracy over the whole frequency range. Numerical instability generated errors at low frequencies, and resonances limited accuracy at the top of the measurement frequency range. Over the frequency range 20 MHz–6 GHz and using an ethanol reference, the error in the measured relative permittivity of ethanol was less than 0·3. (or 1·3%).
Article
A number of automated techniques for point measurement of soil moisture content have been developed to an operational level over the past few decades. While each of those techniques have been individually evaluated by the thermogravimetric (oven drying and weighing) method, typically under laboratory conditions, there have been few studies which have made a direct comparison between the various techniques, particularly under field conditions. This paper makes an inter-comparison of the Virrib®, Campbell Scientific CS615 reflectometer, Soil Moisture Equipment Corporation TRASE® buriable- and connector-type time domain reflectometry (TDR) soil moisture sensors, and a comparison of the connector-type TDR sensor with thermogravimetric measurements for data collected during a 2-year field study. Both qualitative and quantitative comparisons between the techniques are made, and comparisons made with results from a simple water balance ‘bucket’ model and a Richards equation based model. It was found that the connector-type TDR sensors produced soil moisture measurements within the ±2.5% v/v accuracy specification of the manufacturer as compared to thermogravimetric data when using the manufacturer's calibration relationship. However, comparisons with the water balance model showed that Virrib and buriable-type TDR sensors yielded soil moisture changes that exceeded rainfall amounts during infiltration events. It was also found that the CS615 reflectometer yielded physically impossible soil moisture measurements (greater than the soil porosity) during periods of saturation. Moreover, the buriable-type TDR measurements of soil moisture content were systematically less than the Virrib measurements by approximately 10% v/v. In addition to the good agreement with thermogravimetric measurements, the connector-type TDR soil moisture measurements yielded the best agreement with Richards equation based model predictions of soil moisture content, with Virrib sensors yielding a poor agreement in the deeper layers. This study suggests that connector-type TDR sensors give the most accurate measurements of soil moisture content out of the sensor types tested.