ArticlePDF Available

Chronic follow-up from the pilot study of a novel substernal extravascular implantable cardioverter-defibrillator

Authors:

Abstract

Background The investigational Extravascular Implantable Cardioverter Defibrillator (EV ICD) is an implantable defibrillator with defibrillation and pacing therapies, including the potential for antitachycardia pacing (ATP). Purpose To characterize chronic EV ICD system performance through 12 months of follow-up. Methods The first-in-human EV ICD Pilot study was prospectively conducted at four centers in Australia and New Zealand. Patients with Class I or IIa indications for ICD were enrolled between July and December of 2018. A defibrillation lead was inserted into the substernal space under radiologic guidance and a device was positioned close to the left midaxillary line. Figure 1 illustrates the implanted EV ICD. Defibrillation testing was conducted at implant and chronically in 5 patients. Follow-ups were conducted at pre-hospital discharge, 2 weeks, 4–6 weeks and 3 months post-implant, and every 6 months thereafter. Results A total of 26 patients were enrolled and 21 underwent the EV ICD implant procedure with no intraprocedural complications. Median energy for defibrillation at implant was 15 J; mean R-wave amplitude was 3.4±2.0 mV; mean pacing capture threshold was 5.1±2.0 volts. Seventeen of 21 implanted patients entered into chronic follow-up after successful system placement and implant testing, and 14 patients continue to be followed, representing >180 patient-months of total follow-up. Multiple episodes of spontaneous ventricular tachycardia (VT) were detected in one patient and either self-terminated or successfully cardioverted following unsuccessful ATP. One patient received an inappropriate shock secondary to lead tip displacement with subsequent P-wave oversensing. Elective chronic defibrillation testing was performed at physician discretion and converted 5 of 5 patients with ≤40 J (maximum energy of device). Three patients had the system electively removed due to 1) inappropriate shock after lead displacement, 2) recurrent VT without ATP termination, and 3) inadequate chronic safety margin for defibrillation. Conclusions The EV ICD is a novel platform for delivering high voltage therapy. Chronic system performance has generally remained stable during 12 months of follow-up and the device has demonstrated the ability to detect and treat VT. Fluoro images of implanted system Funding Acknowledgement Type of funding source: Private company. Main funding source(s): Medtronic plc
406 Arrhythmias and Device Therapy – Cardioversion and Defibrillation
Chronic follow-up from the pilot study of a novel substernal extravascular implantable
cardioverter-defibrillator
E. Kotschet1, I. Crozier2, H. Haqqani 3, J. Alison1, A. Thompson4,B.Portway
4, R. Sawchuk4, L. Sherfesee4, S. Liang4, L. Lentz4, P. Degroot4,
D. O’Donnell5
1Monash Medical Center, Clayton, Australia; 2Christchurch Hospital, Christchurch, New Zealand; 3The Prince Charles Hospital, Brisbane,
Australia; 4Medtronic, Inc., Minneapolis, United States of America; 5Austin Health Hospital, Melbourne, Australia
Funding Acknowledgement: Type of funding source: Private company. Main funding source(s): Medtronic plc
Background: The investigational Extravascular Implantable Cardioverter
Defibrillator (EV ICD) is an implantable defibrillator with defibrillation and
pacing therapies, including the potential for antitachycardia pacing (ATP).
Purpose: To characterize chronic EV ICD system performance through 12
months of follow-up.
Methods: The first-in-human EV ICD Pilot study was prospectively con-
ducted at four centers in Australia and New Zealand. Patients with Class
I or IIa indications for ICD were enrolled between July and December of
2018. A defibrillation lead was inserted into the substernal space under
radiologic guidance and a device was positioned close to the left midax-
illary line. Figure 1 illustrates the implanted EV ICD. Defibrillation testing
was conducted at implant and chronically in 5 patients. Follow-ups were
conducted at pre-hospital discharge, 2 weeks, 4–6 weeks and 3 months
post-implant, and every 6 months thereafter.
Results: A total of 26 patients were enrolled and 21 underwent the EV ICD
implant procedure with no intraprocedural complications. Median energy
for defibrillation at implant was 15 J; mean R-wave amplitude was 3.4±2.0
mV; mean pacing capture threshold was 5.1±2.0 volts. Seventeen of 21
implanted patients entered into chronic follow-up after successful system
placement and implant testing, and 14 patients continue to be followed,
representing >180 patient-months of total follow-up.
Multiple episodes of spontaneous ventricular tachycardia (VT) were de-
tected in one patient and either self-terminated or successfully cardioverted
following unsuccessful ATP. One patient received an inappropriate shock
secondary to lead tip displacement with subsequent P-wave oversensing.
Elective chronic defibrillation testing was performed at physician discretion
and converted 5 of 5 patients with 40 J (maximum energy of device).
Three patients had the system electively removed due to 1) inappropriate
shock after lead displacement, 2) recurrent VT without ATP termination,
and 3) inadequate chronic safety margin for defibrillation.
Conclusions: The EV ICD is a novel platform for delivering high voltage
therapy. Chronic system performance has generally remained stable dur-
ing 12 months of follow-up and the device has demonstrated the ability to
detect and treat VT.
Fluoro images of implanted system
ESC Congress 2020 – The Digital Experience
29 August – 1 September 2020
Downloaded from https://academic.oup.com/eurheartj/article/41/Supplement_2/ehaa946.0406/6002485 by guest on 17 January 2022
... Pacing output was subsequently increased and among four later episodes of VT, two self-terminated and two showed ATP reset without termination, requiring appropriate and successful shocks. 35,39 Mean R-wave amplitude in the ring-to-ring vector was 3.4 ± 2.0 mV at implant and remained stable over time (3.5 ± 2.0 mV at 6 months and 4.4 ± 2.2 mV at 12 months among patients evaluated in the supine posture). 35,39 Mean VF amplitude at implant was 2.8 ± 1.7 mV, and VF was detected at a sensitivity of ≥0.3 mV in all patients tested clinically (representing a two-fold safety margin over nominal). ...
... 35,39 Mean R-wave amplitude in the ring-to-ring vector was 3.4 ± 2.0 mV at implant and remained stable over time (3.5 ± 2.0 mV at 6 months and 4.4 ± 2.2 mV at 12 months among patients evaluated in the supine posture). 35,39 Mean VF amplitude at implant was 2.8 ± 1.7 mV, and VF was detected at a sensitivity of ≥0.3 mV in all patients tested clinically (representing a two-fold safety margin over nominal). 35 Postural effects on sensing performance were robustly evaluated. ...
Article
Full-text available
Introduction: The extravascular ICD (EV ICD) system with substernal lead placement is a novel non-transvenous alternative to current commercially available ICD systems. The EV ICD provides defibrillation and pacing therapies without the potential long-term complications of endovascular lead placement but requires a new procedure for implantation with a safety profile under evaluation. Methods: This paper summarizes the development of the EV ICD, including the pre-clinical and clinical evaluations that have contributed to system and procedural refinements to date. Results: Extensive pre-clinical research evaluations and 4 human clinical studies with >140 combined acute and chronic implants have enabled the development and refinement of the EV ICD system, currently in worldwide pivotal study. Conclusion: The EV ICD may represent a clinically valuable solution in protecting patients from sudden cardiac death while avoiding the long-term consequences of transvenous hardware. The EV ICD offers advantages over transvenous and subcutaneous systems by avoiding placement in the heart and vasculature; relative to subcutaneous systems, EV ICD requires less energy for defibrillation, enabling a smaller device, and provides pacing features such as anti-tachycardia and asystole pacing in a single system. This article is protected by copyright. All rights reserved.
ResearchGate has not been able to resolve any references for this publication.