Content uploaded by Amy Thompson
Author content
All content in this area was uploaded by Amy Thompson on Jan 17, 2022
Content may be subject to copyright.
406 Arrhythmias and Device Therapy – Cardioversion and Defibrillation
Chronic follow-up from the pilot study of a novel substernal extravascular implantable
cardioverter-defibrillator
E. Kotschet1, I. Crozier2, H. Haqqani 3, J. Alison1, A. Thompson4,B.Portway
4, R. Sawchuk4, L. Sherfesee4, S. Liang4, L. Lentz4, P. Degroot4,
D. O’Donnell5
1Monash Medical Center, Clayton, Australia; 2Christchurch Hospital, Christchurch, New Zealand; 3The Prince Charles Hospital, Brisbane,
Australia; 4Medtronic, Inc., Minneapolis, United States of America; 5Austin Health Hospital, Melbourne, Australia
Funding Acknowledgement: Type of funding source: Private company. Main funding source(s): Medtronic plc
Background: The investigational Extravascular Implantable Cardioverter
Defibrillator (EV ICD) is an implantable defibrillator with defibrillation and
pacing therapies, including the potential for antitachycardia pacing (ATP).
Purpose: To characterize chronic EV ICD system performance through 12
months of follow-up.
Methods: The first-in-human EV ICD Pilot study was prospectively con-
ducted at four centers in Australia and New Zealand. Patients with Class
I or IIa indications for ICD were enrolled between July and December of
2018. A defibrillation lead was inserted into the substernal space under
radiologic guidance and a device was positioned close to the left midax-
illary line. Figure 1 illustrates the implanted EV ICD. Defibrillation testing
was conducted at implant and chronically in 5 patients. Follow-ups were
conducted at pre-hospital discharge, 2 weeks, 4–6 weeks and 3 months
post-implant, and every 6 months thereafter.
Results: A total of 26 patients were enrolled and 21 underwent the EV ICD
implant procedure with no intraprocedural complications. Median energy
for defibrillation at implant was 15 J; mean R-wave amplitude was 3.4±2.0
mV; mean pacing capture threshold was 5.1±2.0 volts. Seventeen of 21
implanted patients entered into chronic follow-up after successful system
placement and implant testing, and 14 patients continue to be followed,
representing >180 patient-months of total follow-up.
Multiple episodes of spontaneous ventricular tachycardia (VT) were de-
tected in one patient and either self-terminated or successfully cardioverted
following unsuccessful ATP. One patient received an inappropriate shock
secondary to lead tip displacement with subsequent P-wave oversensing.
Elective chronic defibrillation testing was performed at physician discretion
and converted 5 of 5 patients with ≤40 J (maximum energy of device).
Three patients had the system electively removed due to 1) inappropriate
shock after lead displacement, 2) recurrent VT without ATP termination,
and 3) inadequate chronic safety margin for defibrillation.
Conclusions: The EV ICD is a novel platform for delivering high voltage
therapy. Chronic system performance has generally remained stable dur-
ing 12 months of follow-up and the device has demonstrated the ability to
detect and treat VT.
Fluoro images of implanted system
ESC Congress 2020 – The Digital Experience
29 August – 1 September 2020
Downloaded from https://academic.oup.com/eurheartj/article/41/Supplement_2/ehaa946.0406/6002485 by guest on 17 January 2022