Conference PaperPDF Available

Converting One Set of Mammograms to Simulate a Range of Detector Imaging Characteristics for Observer Studies

Authors:
  • Royal Surrey NHS Foundation Trust

Abstract and Figures

A methodology for adjusting mammographic images taken on a given imaging system to simulate their appearance if taken on a different system for use in observer studies is presented. The process involves adjusting the image sharpness and noise, which takes into account the detector, breast thickness, and beam quality. The method has been tested by converting images acquired using an a-Se detector of a CDMAM test object and ‘Rachel’ anthropomorphic breast phantom. They were degraded to appear as if acquired using a computed radiography (CR) detector. Good agreement was achieved in the resulting threshold gold thickness for the simulated CR images with measured real values for CDMAM images. Power spectra comparisons of real and simulated images of the ‘Rachel’ phantom agree with an average difference of 4%. This tool in conjunction with observer studies can be used to understand the effects of the detector characteristics on cancer detection in mammography.
Content may be subject to copyright.
A.D.A. Maidment, P.R. Bakic, and S. Gavenonis (Eds.): IWDM 2012, LNCS 7361, pp. 394–401, 2012.
© Springer-Verlag Berlin Heidelberg 2012
Converting One Set of Mammograms to Simulate
a Range of Detector Imaging Characteristics
for Observer Studies
Alistair Mackenzie1, David R. Dance1,
Oliver Diaz2, Annabel Barnard1, and Kenneth C. Young1
1 National Coordinating Centre for the Physics of Mammography, Royal Surrey County
Hospital, Guildford, GU2 7XX, UK and Department of Physics, University of Surrey,
Guildford, GU2 7XH, UK
2 Centre for Vision, Speech and Signal Processing, Faculty of Engineering and Physical
Sciences, University of Surrey, Guildford, GU2 7XH, UK
alistairmackenzie@nhs.net
Abstract. A methodology for adjusting mammographic images taken on a
given imaging system to simulate their appearance if taken on a different
system for use in observer studies is presented. The process involves adjusting
the image sharpness and noise, which takes into account the detector, breast
thickness, and beam quality. The method has been tested by converting images
acquired using an a-Se detector of a CDMAM test object and ‘Rachel’
anthropomorphic breast phantom. They were degraded to appear as if acquired
using a computed radiography (CR) detector. Good agreement was achieved in
the resulting threshold gold thickness for the simulated CR images with
measured real values for CDMAM images. Power spectra comparisons of real
and simulated images of the ‘Rachel’ phantom agree with an average difference
of 4%. This tool in conjunction with observer studies can be used to understand
the effects of the detector characteristics on cancer detection in mammography.
Keywords: simulation, noise power spectra, modulation transfer function.
1 Introduction
Clinical evaluation of image quality is expensive and time-consuming. Clinical trials
to compare the effectiveness of different systems are rarely conducted as they would
require large numbers of patients to achieve both sufficient numbers of detected
cancers and statistical significance. In particular it would be desirable to repeat
exposures on the same breast with the same positioning and compression to minimise
confounding differences in the projection of the breast tissues, but this raises ethical
issues. Alternative methods of evaluation involving some degree of image simulation
have the potential to enable comparisons at reduced cost and time and without
additional radiation exposure. For this purpose it is desirable to be able to acquire
images on a given system and to simulate their appearance on a second system, so the
Adjusting Image Quality of Mammograms 395
performance of the two systems can be compared. This may be possible when the
performance of the second system is inferior to that of the first system. Such a method
would enable the background tissue and compression to be matched in different arms
of a study, either using real cancers or the insertion of simulated cancers.
The aim of this work is therefore to develop and test a methodology for adjusting
mammographic images taken on a given imaging system to simulate their appearance
if taken on a different system. The methodology presented extends previous work [1]
with improved modelling of the noise power spectra, which takes into account the
breast thickness and beam quality.
2 Method
2.1 Summary of Methodology for Changing Image Quality
The conversion methodology blurs the original image to match the blurring of a target
system using measurements of the modulation transfer function. The difference in
noise between the original system and the target system is then calculated and added
in real space to the blurred image, ensuring that the magnitude and correlation of the
noise matches the total noise in the target system. The method also accounts for the
magnitude of the signal in the image. It has been validated using images of a contrast
detail test object for situations where the noise characteristics were measured at the
same beam quality as the image to be converted.
2.2 Linearisation of Images
The analysis below assumes that images have been linearised so that the pixel value is
a measure of the energy absorbed per unit area of the detector. This can be achieved
by a combination of measurements of the incident air kerma at the front face of the
detector and Monte Carlo simulations. The Geant4 Monte Carlo code
(http://geant4.cern.ch/) was used to calculate the absorbed energy per unit area for a
reference beam quality, and to relate this to the incident air kerma at the front face of
the image receptor. X-ray spectra from the work of Boone et al [2] filtered by the X-
ray tube window, filter, compression paddle, object being imaged (test phantom or
breast) and breast support were used for this purpose. The attenuation coefficients
were obtained from Berger et al [3]. The signal transfer properties (STP) thus
calculated for a given detector were assumed to apply to any beam quality for that
detector so that the pixel value was a measure of the energy absorbed per unit area for
all beam qualities.
2.3 Characterisation of Noise Power Spectra (NPS)
The noise was characterised by the NPS (W) which was split into components for
electronic, quantum and structure noise at a reference beam quality. For this purpose,
a series of collimated flat field images were acquired over a wide dose range using a
396 A. Mackenzie et al.
28 kV, Mo/Mo anode/filter combination and 4.5 cm polymethyl methacrylate
(PMMA) at the tube head and the NPS was calculated for each dose. The three noise
components for the reference condition were estimated by fitting a second order
polynomial of NPS against absorbed energy for each spatial frequency.
2.4 Correction of Noise and Signal for Beam Quality
Clinical mammograms are acquired over a range of compressed breast thicknesses
and radiographic factors and so the model needs to be able to take account of a range
of beam qualities. When originally developed the methodology used measurements of
the NPS appropriate to the beam quality used. In the present work we have improved
the modelling of the NPS to take account of beam quality and breast
thickness/composition. For this purpose, it was assumed that the three noise sources
are affected by beam quality as follows:
Electronic noise is independent of beam quality. No correction is required.
The quantum noise is comprised of a number of different sources (primary
quantum noise, excess noise, and secondary quantum noise) and is dependent
on the number of photons detected and energy absorbed.
Structure noise is proportional to the signal from the detector. No correction
for beam quality is required.
The beam quality affects both the proportion of energy absorbed and the quantum
noise. To estimate the effect of beam quality on the quantum noise, a series of flat
field images were taken as described in section 2.3, but using a range of PMMA
thicknesses, tube voltages and anode/filter combinations. These images were then
linearised using the reference STP, so that the linearised pixel value equalled the
absorbed energy per unit area irrespective of the beam quality.
Using the above flat field images, the NPS was calculated for each beam quality,
and the results used to determine the parameters in a model of the NPS applicable to
any beam quality (Eq. 1).
2
),(),()(),(),(
++=
o
s
o
qe E
E
vu
E
E
vuBvuvuW
ωωλω
(1)
In this equation ωe, ωq and ωs (with units of mm2) are ‘noise coefficients’ for
electronic, quantum and structure noise respectively at absorbed energy per unit area
at a reference beam quality Eo, E is the absorbed energy per unit area at the beam
quality under consideration and u and v are spatial frequencies. In accordance with
the assumptions in the bullet points above, the electronic and structure noise
coefficients are independent of beam quality. The quantum noise coefficient, ωq, is
also independent of beam quality; the variation of the quantum noise with beam
quality additional to the factor E in Eq. 1 is accounted for by the beam quality
correction factor (B). The beam quality parameter λ is average photon energy of the
beam incident on the detector and was calculated using the X-ray spectra model
Adjusting Image Quality of Mammograms 397
described in section 2.2. For breast images, λ can be calculated if an assumption about
the breast composition is made [4].
2.5 Validation of the Conversion Using Images of CDMAM Test Object
and Anthropomorphic Phantoms
The validation of the conversion from one detector to another was undertaken using
the following two systems:
ASE: Hologic Selenia X-ray system, amorphous selenium (a-Se) detector. Pixel
pitch 70 µm.
CR: Carestream CR900 reader with EHR-M2 CR plates. Pixel pitch 50 µm.
Sixteen images of the CDMAM contrast detail test object were acquired on both
systems at two beam qualities. The CDMAM test object was imaged on the breast
support on a base of 2 cm PMMA (‘thin’) using 26 kV, Mo/Mo and then with an
additional 4 cm PMMA (‘thick’ – total of 6 cm PMMA) on top of the test object and
imaged using 34 kV, Mo/Rh. The image conversion methodology was applied to the
ASE CDMAM images to convert them to appear at the same image quality and dose
as the CR images. The sets of target and simulated CDMAM images for CR were
automatically read using CDCOM software version 1.5.2 (www.euref.org). Contrast
detail curves were produced for both the simulated and target images, which were
then compared [5].
The Rachel anthropomorphic breast phantom (Gammex RMI, WI, USA) was
designed to mimic a 5 cm compressed breast. It was imaged five times using both
detectors at 31 kV, Mo/Rh. The phantom was shifted slightly between images. The
ASE images were converted to appear with the imaging characteristics of the CR
detector and X-ray system used. The largest rectangular region of interest (ROI) away
from the skin edge was extracted from the same location for the real CR and
simulated CR images. Smaller overlapping sub-ROIs of size 256 × 256 were extracted
from this ROI and the measured power spectra from all of the sub-ROIs were
averaged for each system.
2.6 Clinical Images: Subjective Evaluation and Demonstration of Image
Conversion
Mammography images have been collected for an image database. For the systems
included in this study, images from 234 women (ASE) and 233 (CR) were collected.
The ASE had 31 abnormal cases, while all of the CR images were as normal.
Firstly, a preliminary test of realism of the converted images was undertaken. A set
of normal ASE and CR images were selected. A reasonable match between ASE and
CR images in terms of compressed breast thickness, appearance of the breast and
radiographic factors was made. The ASE images were converted to appear with the
imaging characteristics of the CR system and its associated X-ray system. Both sets of
398 A. Mackenzie et al.
images were processed using Agfa Musica 2. A simulated CR and a real CR image
were shown to a set of 6 observers and they were asked to identify the real CR image.
The observers were shown 10 pairs of images.
Secondly, a sub-set (six images) of the ASE images containing a confirmed cancer
classified as being ‘subtle’ were selected. To demonstrate the conversion process,
these images were changed to appear with the imaging characteristics of the CR
detector using the same X-ray system and grid as the ASE system. Therefore no
correction was made for differences in scatter and grid attenuation. The simulated
images were visually examined for the effects of the conversion.
3 Results and Discussion
3.1 Conversion of CDMAM Test Object Images
The threshold gold thicknesses of simulated CR CDMAM images and the
corresponding real CR images for two beam qualities and PMMA thicknesses show a
close match (Fig. 1). The average differences between the results were 2.5% and 0.3%
for the thin and thick phantoms respectively. This is an encouraging result because the
CDMAM test object provides a good overall measure of the image quality in terms of
noise, sharpness and contrast.
32 mAs (ta rget CR)
32 mAs (AS E)
32 mAs (simu lated CR)
Diameter (mm)
Threshold gold thickness (μm)
0.1 0.13 0.16 0.2 0.250.31 0.4 0.5 0.63 0.8 1.0
0.03
0.05
0.10
0.50
1.00
2.00
CDMAM
+ 2 cm PMM A
125 mAs (ta rget CR)
130 mAs (AS E)
0.1 0.130.16 0.2 0.25 0.31 0.4 0.5 0.63 0.8 1.0
0.03
0.05
0.10
0.50
1.00
2.00
125 mAs (s imula ted CR)
Diameter (mm)
Threshold gold thickness (μm)
CDMA M
+ 6 cm PMMA
Fig. 1. Threshold gold thickness curves for CDMAM images acquired for the thin (2 cm
PMMA) phantom (left), the thick (6 cm PMMA) phantom (right). Results are shown for
simulated and target (real) images obtained with the CR detector and for the original ASE
images from which the CR images were simulated.
3.2 Power Spectra of Conversion of Images of Rachel Anthropomorphic
Phantom
No artefacts were seen in the simulated image of the Rachel phantom. The power
spectra of real and simulated Rachel phantom images obtained with the CR system
are shown in Fig. 2. The results show a good match between the images, with a
Adjusting Image Quality of Mammograms 399
maximum difference of 17% between the simulated and real images, the average
difference was 2% over all spatial frequencies.
Frequency (mm
-1
)
Power spectra (mm
2
)
0.1 110 100
0.001
0.01
0.1
1
10
100
Real CR imag e
Simulated CR image
Fig. 2. Power spectra of ‘Rachel’ phantom of target CR and simulated CR
3.3 Subjective Evaluation of Realism of Converted Images
Using 6 observers and 10 image pairs of a real and simulated CR image, the real
image was correctly identified 32 times out of 60 and the simulated CR image was
incorrectly selected as the real CR image 28 times out of 60. There was a slight
majority of the real image being correctly identified. While these number of results
are very small, it does give an indication that the images produced do look realistic.
3.4 Examples of Conversion of Clinical Images Suitable for an Observer Study
Figs. 3 & 4 show on the left a high quality image of a lesion acquired on a ASE
system. The image on the right shows the image after it has been degraded to have the
image quality of a CR system. There are noticeable differences between the ASE and
simulated CR images in terms of sharpness and noise. The cancers are still visible in
the CR images but the interest of this work is whether the detector used affects the
detection and diagnosis of cancer in breast imaging. An observer study has been
undertaken using this image modification process. The study showed a difference in
the detection of subtle calcifications between ASE detector and a generic CR detector
[6]. The advantage of this method is that the only difference between the two sets of
images is the detector, and that differences due to breasts, compression, anti-scatter
grid, X-ray tube have been removed.
This methodology can be applied to simulate images from different detectors or
even theoretical detectors. A range of observer studies can be undertaken on the effect
of the receptor performance and dose on the detection of different signs of breast
cancer (e.g. masses, calcifications).
400 A. Mackenzie et al.
Fig. 3. Example images of micro-calcification cluster for the original ASE system (left) and
simulated CR (right)
Fig. 4. Example images of mass for the original ASE system (left) and simulated CR (right)
4 Conclusions
We have developed a conversion methodology to change an image to appear as if
acquired on a different imaging system which accounts for the change in detector,
X-ray system and beam quality. The methodology has been successfully applied to
contrast detail measurements. Images with the appearance of a realistic clinical CR
mammogram can be produced without artefacts. The use of this tool in conjunction
with observer studies can be used to understand the effects of detector characteristics
on cancer detection.
Acknowledgements. This work is part of the OPTIMAM project and is supported by
Cancer Research-UK & EPSRC Cancer Imaging Programme in Surrey, in association
with the MRC and Department of Health (England).
The authors are grateful for the help and support of staff from St George’s Hospital,
London, and Jarvis Breast Screening Unit, Guildford. The authors acknowledge Hologic
Inc., MIS Healthcare, and Carestream Healthcare for their help in accessing images and
Agfa Healthcare for the use of their image processing package. We thank our NCCPM
colleagues Lucy Warren and Faith Green who have helped with the collection of
images, and all of our colleagues who took part in the observer study. We thank our
colleagues at Katholieke Universiteit Leuven for helpful discussion of this work.
Adjusting Image Quality of Mammograms 401
References
1. Mackenzie, A., Dance, D.R., Workman, A., Yip, M., Wells, K., Young, K.C.: Development
and validation of a method for converting images to appear with noise and sharpness
characteristics of a different detector and X-ray system. Med. Phys. 39, 2721–2734 (2012)
2. Boone, J.M., Fewell, T.R., Jennings, R.J.: Molybdenum, rhodium, and tungsten anode
spectral models using interpolating polynomials with application to mammography. Med.
Phys. 24, 1863–1874 (1997)
3. Berger, M.J., Hubbell, J.H., Seltzer, S.M., Chang, J., Coursey, J.S., et al.: XCOM: Photon
cross sections database. NIST Standard Reference Database 8, 87-3597 (1998)
4. Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., Kotre, C.J.: Additional factors for
the estimation of mean glandular breast dose using the UK mammography dosimetry
protocol. Phys. Med. Biol. 45, 3225–3240 (2000)
5. Young, K.C., Alsager, A., Oduko, J.M., Bosmans, H., Verbrugge, B., et al.: Evaluation of
software for reading images of the CDMAM test object to assess digital mammography
systems. In: Proc. SPIE, vol. 6913, pp. 69131C-1–69131C-11 (2008)
6. Warren, L.M., Mackenzie, A., Cooke, J., Given-Wilson, R., Wallis, M.G., et al.:
Mammographic calcification cluster detection and threshold gold thickness measurements.
In: Proc. SPIE, vol. 8313, p. 83130J (2012)
... The measurements included pre-sampled MTF (modulation transfer function), NNPS (normalized noise power spectrum), STP (signal transfer properties), flat fielding correction map, focal spot size and focus motion. For details on the measurement procedure, see Mackenzie et al (2012aMackenzie et al ( , 2012b. ...
... The following process, which is fully described in Mackenzie et al (2012aMackenzie et al ( , 2012b, was used to add noise to each image. First, the noise power spectra (NPS) were measured at a series of exposure levels and used to calculate the individual contributions of electronic, structure and quantum noise to the NPS as a function of absorbed energy (keV mm -2 ) and spatial frequency. ...
... To accurately simulate dose distribution across the detector, the non-uniformity of the incident x-ray flux due to the anode-heel effect was considered. The inverse of the calculated flat fielding correction map (Mackenzie et al 2012a(Mackenzie et al , 2012b was multiplied with the image pixels to apply the anode-heel effect prior to noise addition. After noise addition, the flat fielding correction map was reapplied to the images, along with a pixel offset correction. ...
Article
Full-text available
Planar 2D x-ray mammography is generally accepted as the preferred screening technique used for breast cancer detection. Recently, digital breast tomosynthesis (DBT) has been introduced to overcome some of the inherent limitations of conventional planar imaging, and future technological enhancements are expected to result in the introduction of further innovative modalities. However, it is crucial to understand the impact of any new imaging technology or methodology on cancer detection rates and patient recall. Any such assessment conventionally requires large scale clinical trials demanding significant investment in time and resources. The concept of virtual clinical trials and virtual performance assessment may offer a viable alternative to this approach. However, virtual approaches require a collection of specialized modelling tools which can be used to emulate the image acquisition process and simulate images of a quality indistinguishable from their real clinical counterparts. In this paper, we present two image simulation chains constructed using modelling tools that can be used for the evaluation of 2D-mammography and DBT systems. We validate both approaches by comparing simulated images with real images acquired using the system being simulated. A comparison of the contrast-to-noise ratios and image blurring for real and simulated images of test objects shows good agreement ( < 9% error). This suggests that our simulation approach is a promising alternative to conventional physical performance assessment followed by large scale clinical trials.
... We have developed methods to adapt an image acquired on one detector to appear with the characteristics of another detector. 6,7 The aim of this work was to use this methodology to create and conduct an observer study to compare cancer detection rates for different types of detector at the same radiation dose by using mammograms with normal, benign and malignant features. ...
... In order to allow for changes in beam quality, the effect of beam quality on the detector sensitivity and quantum noise was measured for one system. 6 In addition four other detectors were characterised: GE Essential (GE Medical Systems, Milwaukee, USA), Agfa Healthcare DX-M with NIP (Agfa Gevaert N.V., Mortsel, Belgium), Carestream NIP (Carestream, Rochester, USA) and Carestream CR900 with EHR2 image plates. ...
... The clinical images were acquired at a different beam quality from the beam quality at which the noise was measured. Therefore, a further correction was applied to the quantum noise to account for this difference following the method described in Mackenzie et al. 6 Finally, the flat field correction was re-applied to the images for Arms 1 and 2, as they would normally be flat fielded. This method has been validated using test objects and clinical images. ...
Article
Full-text available
Introduction: The effect that the image quality associated with different image receptors has on cancer detection in mammography was measured using a novel method for changing the appearance of images. Method: A set of 270 mammography cases (one view, both breasts) was acquired using five Hologic Selenia and two Hologic Dimensions X-ray sets: 160 normal cases, 80 cases with subtle real non-calcification malignant lesions and 30 cases with biopsy proven benign lesions. Simulated calcification clusters were inserted into half of the normal cases. The 270 cases (Arm 1) were converted to appear as if they had been acquired on three other imaging systems: caesium iodide detector (Arm 2), needle image plate computed radiography (CR) (Arm 3) and powder phosphor CR (Arm 4). Five experienced mammography readers marked the location of suspected cancers in the images and classified the degree of visibility of the lesions. Statistical analysis was performed using JAFROC. Results: The differences in the visibility of calcification clusters between all pairs of arms were statistically significant (p<0.05), except between Arms 1 and 2. The difference in the visibility of non-calcification lesions was smaller than for calcification clusters, but the differences were still significant except between Arms 1 and 2 and between Arms 3 and 4. Conclusion: Detector type had a significant impact on the visibility of all types of subtle cancers, with the largest impact being on the visibility of calcification clusters.
... We have previously shown that the energy of the X-rays incident on the detector can affect the quantum noise coefficients, and have demonstrated how a correction for this dependence on beam quality can be applied. 10 The model needs to be developed to apply lag and ghosting to the images. This is not straightforward as the lag in subsequent images and the effect of ghosting on the detector sensitivity are dependent on the history of the whole tomosynthesis run. ...
Conference Paper
Full-text available
The aim of this work is to characterise the image quality of a mammography system in both planar and tomosynthesis imaging modes for the purpose of realistic image simulation. The simulation technique will be applied to projected images from voxelised breast phantoms to investigate the imaging properties of different configurations of tomosynthesis systems. Methods: A Hologic Dimensions mammography system was characterised in terms of noise, sharpness, and lag in both planar and tomosythesis modes. The noise power spectra (NPS) were measured for both imaging modes. The contributions (noise coefficients) to the NPS from electronic, quantum and structure noise were calculated. Results: The MTF was shown to be affected by the focal spot size and the movement of the X-ray tube. These effects can be modelled in the tomosynthesis projection images using an extended source. The quantum noise coefficients for the two imaging modes showed a close match up to the Nyquist frequency of the tomosynthesis mode, while the electronic and structure noise showed differences. The effect of lag and ghosting caused the signal in the last projection of a tomosynthesis run to be between 2 and 4% higher than the first. Conclusions: We have characterised the noise and resolution properties of images from a Hologic Dimensions tomosynthesis system. This forms the basis of further work required to create a model for applying imaging characteristics to mathematically produced images. More work is required on detector lag and ghosting and the influence of oblique X-ray incidence.
Article
Full-text available
European Guidelines for quality control in digital mammography specify minimum and achievable standards of image quality in terms of threshold contrast, based on readings of images of the CDMAM test object by human observers. However this is time-consuming and has large inter-and intra-observer error. To overcome these problems a software program (CDCOM) is available to automatically read CDMAM images. After some further analysis the automated measurements can be used to predict the threshold contrast for a typical observer. The results of threshold contrast determination by human observers at three different centres were compared against automated readings. These data provide a means of predicting average human performance using the automated reading software. The coefficient of variation in automatically determined threshold gold thickness was about 4% for detail sizes from 0.2 to 1.0mm when 8 images were analysed. The coefficient of variation was about 10% at a detail size of 0.1mm. Using larger numbers of images improved reproducibility for all detail sizes. A change in phantom design could greatly improve reproducibility for the smallest detail sizes. Greater consistency of phantom construction would also be desirable as one of the four phantoms tested was significantly different from the other three. Despite some limitations automated reading of CDMAM images can provide a reproducible means of assessing digital mammography systems against European Guidelines.
Article
European Guidelines for quality control in digital mammography specify acceptable and achievable standards of image quality (IQ) in terms of threshold gold thickness using the CDMAM test object. However, there is little evidence relating such measurements to cancer detection. This work investigated the relationship between calcification detection and threshold gold thickness. An observer study was performed using a set of 162 amorphous selenium direct digital (DR) detector images (81 no cancer and 81 with 1-3 inserted calcification clusters). From these images four additional IQs were simulated: different digital detectors (computed radiography (CR) and DR) and dose levels. Seven observers marked and rated the locations of suspicious regions. DBM analysis of variances was performed on the JAFROC figure of merit (FoM) yielding 95% confidence intervals for IQ pairs. Automated threshold gold thickness (Tg) analysis was performed for the 0.25mm gold disc diameter on CDMAM images at the same IQs (16 images per IQ). Tg was plotted against FoM and a power law fitted to the data. There was a significant reduction in FoM for calcification detection for CR images compared with DR; FoM decreased from 0.83 to 0.63 (p<=0.0001). Detection was also sensitive to dose. There was a good correlation between FoM and Tg (R2=0.80, p<0.05), consequently threshold gold thickness was a good predictor of calcification detection at the same IQ. Since the majority of threshold gold thicknesses for the various IQs were above the acceptable standard despite large variations in calcification detection by radiologists, current EU guidelines may need revising.
Article
Undertaking observer studies to compare imaging technology using clinical radiological images is challenging due to patient variability. To achieve a significant result, a large number of patients would be required to compare cancer detection rates for different image detectors and systems. The aim of this work was to create a methodology where only one set of images is collected on one particular imaging system. These images are then converted to appear as if they had been acquired on a different detector and x-ray system. Therefore, the effect of a wide range of digital detectors on cancer detection or diagnosis can be examined without the need for multiple patient exposures. Three detectors and x-ray systems [Hologic Selenia (ASE), GE Essential (CSI), Carestream CR (CR)] were characterized in terms of signal transfer properties, noise power spectra (NPS), modulation transfer function, and grid properties. The contributions of the three noise sources (electronic, quantum, and structure noise) to the NPS were calculated by fitting a quadratic polynomial at each spatial frequency of the NPS against air kerma. A methodology was developed to degrade the images to have the characteristics of a different (target) imaging system. The simulated images were created by first linearizing the original images such that the pixel values were equivalent to the air kerma incident at the detector. The linearized image was then blurred to match the sharpness characteristics of the target detector. Noise was then added to the blurred image to correct for differences between the detectors and any required change in dose. The electronic, quantum, and structure noise were added appropriate to the air kerma selected for the simulated image and thus ensuring that the noise in the simulated image had the same magnitude and correlation as the target image. A correction was also made for differences in primary grid transmission, scatter, and veiling glare. The method was validated by acquiring images of a CDMAM contrast detail test object (Artinis, The Netherlands) at five different doses for the three systems. The ASE CDMAM images were then converted to appear with the imaging characteristics of target CR and CSI detectors. The measured threshold gold thicknesses of the simulated and target CDMAM images were closely matched at normal dose level and the average differences across the range of detail diameters were -4% and 0% for the CR and CSI systems, respectively. The conversion was successful for images acquired over a wide dose range. The average difference between simulated and target images for a given dose was a maximum of 11%. The validation shows that the image quality of a digital mammography image obtained with a particular system can be degraded, in terms of noise magnitude and color, sharpness, and contrast to account for differences in the detector and antiscatter grid. Potentially, this is a powerful tool for observer studies, as a range of image qualities can be examined by modifying an image set obtained at a single (better) image quality thus removing the patient variability when comparing systems.
Article
Computer simulation is a convenient and frequently used tool in the study of x-raymammography, for the design of novel detector systems, the evaluation of dose deposition, x-ray technique optimization, and other applications. An important component in the simulation process is the accurate computer-generation of x-ray spectra. A computermodel for the generation of x-ray spectra in the mammographic energy range from 18 kV to 40 kV has been developed. The proposed model requires no assumptions concerning the physics of x-ray production in an x-ray tube, but rather makes use of x-ray spectra recently measured experimentally in the laboratories of the Center for Devices and Radiological Health. Using x-ray spectra measured for molybdenum,rhodium, and tungstenanodex-ray tubes at 13 different kV’s (18, 20, 22,…,42 kV), a spectralmodel using interpolating polynomials was developed. At each energy in the spectrum, the x-ray photon fluence was fit using 2, 3, or 4 term (depending on the energy) polynomials as a function of the applied tube voltage (kV). Using the polynomial fit coefficients determined at each 0.5 keV interval in the x-rayspectrum, accurate x-ray spectra can be generated for any arbitrary kV between 18 and 40 kV. Each anode material (Mo, Rh, W) uses a different set of polynomial coefficients. The molybdenumanodespectralmodel using interpolating polynomials is given the acronym MASMIP, and the rhodium and tungstenspectralmodels are called RASMIP and TASMIP, respectively. It is shown that the mean differences in photon fluence calculated over the energy channels and over the kV range from 20 to 40 kV were −0.073% (σ=1.58%) for MASMIP, −0.145% (σ=1.263%) for RASMIP, and 0.611% (σ=2.07%) for TASMIP. The polynomial coefficients for all three models are given in an Appendix. A short C subroutine which uses the polynomial coefficients and generates x-ray spectra based on the proposed model is available on the World Wide Web at http://www.aip.org/epaps/epaps.html.
Article
The UK and European protocols for mammographic dosimetry use conversion factors that relate incident air kerma to the mean glandular dose (MGD) within the breast. The conversion factors currently used were obtained by computer simulation of a model breast with a composition of 50% adipose and 50% glandular tissues by weight (50% glandularity). Relative conversion factors have been calculated which allow the extension of the protocols to breasts of varying glandularity and for a wider range of mammographic x-ray spectra. The data have also been extended to breasts of a compressed thickness of 11 cm. To facilitate the calculation of MGD in patient surveys, typical breast glandularities are tabulated for women in the age ranges 40-49 and 50-64 years, and for breasts in the thickness range 2-11 cm. In addition, tables of equivalent thickness of polymethyl methacrylate have been provided to allow the simulation for dosimetric purposes of typical breasts of various thicknesses.