Question
Asked 6th Jan, 2015

How can one perform time limited OCP (open circuit potential) measurement in Autolab Nova 1.10 ?

I want to measure OCP of a supercapacitor device for a time period (say for 10 hours). Is it possible with Autolab Nova 1.10?
I didn't find options for time limited OCP measurement, however, we have options for d(OCP)/dt limited OCP measurement.
Thanks

Most recent answer

Ana Baía
Universidade da Beira Interior
how can I clean my worked electrode (carbon rod/graphite) after cyclic voltammetry?

All Answers (6)

Ashwati Krishnan
Carnegie Mellon University
You will have to create your own procedure. If you go to "Commands", there should be a tree structure called "Plots - general", under which there is an option of adding "E vs t" under your OCP measurement in your procedure. Add "E vs t" , and link the E and t to WE(1).Potential and Time of the OCP measurement. Hope this helps.
Dominik Schäfer
Forschungszentrum Jülich
For the most basic measurement: just use a "Record signals" command to record the cell potential (potential between your working and reference electrode) for the desired amount of time without switching on the cell.
1 Recommendation
I totally agree with Dominik Schafer, you have to click and drag 'Record signals' into the window and specify the time duration
hello everyone
I would appreciate if you answer my question
can chronopotentiometery be considered for OCP measurement?
Albert Kazadi Mukenga Bantu
University of Kinshasa
Very good answer from Ashwati Krishnan.
Ana Baía
Universidade da Beira Interior
how can I clean my worked electrode (carbon rod/graphite) after cyclic voltammetry?

Similar questions and discussions

How can I get a good fit for my experimental EIS results, especially where the semi arch is not well defined?
Question
8 answers
  • Ibrahim MustafaIbrahim Mustafa
Dear All,
My Procedures:
I am performing 2 EIS experiments as follows: I used a 3 electrode cell (CE, RE, WE, S electrodes) with a 0.05M Vanadium (4) in 3M H2SO4 electrolyte. The frequency ranged from 100,000 to 0.01 Hz, and an amplitude of 10mV. I use Autolab NOVA software and instrumentation.
Experiment 1: WE is Carbon Paper
Experiment 2: WE is CNTs
After obtaining the EIS results, I performed fitting to quantify Rct and Cdl. My eq circuit was a randles circuit which included a resistor, and a (resistor, warburg term, and a constant phase element in parallel) as in FIGURE 5.
My Analysis of Results:
Experiment 1:
Fitting worked well for experiment 1 (carbon paper), which included a more apparent and defined semi-circle arch. The software automatically snapped to the experimental results as soon as I entered the equivalent circuit fit components. (SEE FIGURE 1)
Experiment 2:
However, fitting didn’t work well for experiment 2 (CNTs), which included a less defined semi circle arch. The software didn’t snap automatically to my experimental results, and showed me very wrong results. (SEE FIGURE 2)
So, I tried to enter some values manually (for Rs and Rct which I determined graphically from the EIS results), and the fit improved, however, I am not sure if this would be accurate, as the Warburg term is still well off. (SEE FIGURE 3) I couldn’t get a better fit of my EIS results than this. Please note I tried fitting with 3 different softwares (EISAnalyser, Nova Autolab, Zfit)
I used Randles equivilant circuit for all experiment (SEE FIG. 5).
My question is:
How can I get a more accurate fitting of my EIS results? What am I doing wrong please?
Your help and input is very much appreciated.
Thank you very much for your comments,
Ibrahim

Related Publications

Chapter
This paper presents an overview of sizing tasks in electronic circuit design and their corresponding formulations as optimization problems. We will start with the general multi-objective sizing problem. Then, the inclusion of statistically distributed parameters and of range-valued parameters into the scalar problems of yield optimization and desig...
Article
Thesis (M.S.)--University of New Mexico, 1988. Includes bibliographical references (leaf 116).
Article
En aquest projecte s’han realitzat tres pràctiques d’anàlisis de circuits on s’ha dissenyat i implementat els circuits a analitzar. El projecte consta d’una part d’amplificadors operacionals, on la finalitat es entendre com funciona un amplificador operacional i les propietats que te com a dispositiu electrònic, avantatges i utilitats.Una altra par...