ArticlePDF Available

The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging

Authors:

Abstract and Figures

Efforts are made in radiographic examinations to obtain the best image quality with the lowest possible absorbed dose to the patient. In dental radiography, the absorbed dose to patients is very low, but exposures are relatively frequent. It has been suggested that frequent low-dose exposures can pose a risk for development of future cancer. It has previously been reported that there was no significant difference in the diagnostic accuracy of approximal carious lesions in radiographs obtained using tube voltages of 60 and 70 kV. The aim of this study was, therefore, to evaluate the patient dose resulting from exposures at these tube voltages to obtain intraoral bitewing radiographs. The absorbed dose distributions resulting from two bitewing exposures were measured at tube voltages of 60 and 70 kV using Gafchromic(®) film and an anatomical head phantom. The dose was measured in the occlusal plane, and ± 50 mm cranially and caudally to evaluate the amount of scattered radiation. The same entrance dose to the phantom was used. The absorbed dose was expressed as the ratio of the maximal doses, the mean doses and the integral doses at tube voltages of 70 and 60 kV. The patient receives approximately 40 - 50% higher (mean and integral) absorbed dose when a tube voltage of 70 kV is used. The results of this study clearly indicate that 60 kV should be used for dental intraoral radiographic examinations for approximal caries detection.
Content may be subject to copyright.
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
The Effects on Absorbed Dose Distribution in Intraoral X-ray
Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing
Imaging
Kristina Hellén-Halme1, Mats Nilsson1,2
1Department of Oral and Maxillofacial Radiology, Faculty of Odontology, Malmö University, Malmö, Sweden.
2Department of Radiation Physics, Skåne University Hospital, Malmö, Sweden.
Corresponding Author:
Kristina Hellén-Halme
Department of Oral and Maxillofacial Radiology, Malmö University
SE-205 06 Malmö
Sweden
Phone: +46 40 665 8414
E-mail: Kristina.Hellen-Halme@mah.se
ABSTRACT
Objectives: Efforts are made in radiographic examinations to obtain the best image quality with the lowest possible
absorbed dose to the patient. In dental radiography, the absorbed dose to patients is very low, but exposures are
relatively frequent. It has been suggested that frequent low-dose exposures can pose a risk for development of
future cancer. It has previously been reported that there was no signicant difference in the diagnostic accuracy of
approximal carious lesions in radiographs obtained using tube voltages of 60 and 70 kV. The aim of this study was,
therefore, to evaluate the patient dose resulting from exposures at these tube voltages to obtain intraoral bitewing
radiographs.
Material and Methods: The absorbed dose distributions resulting from two bitewing exposures were measured at
tube voltages of 60 and 70 kV using Gafchromic® lm and an anatomical head phantom. The dose was measured
in the occlusal plane, and ± 50 mm cranially and caudally to evaluate the amount of scattered radiation. The same
entrance dose to the phantom was used. The absorbed dose was expressed as the ratio of the maximal doses, the
mean doses and the integral doses at tube voltages of 70 and 60 kV.
Results: The patient receives approximately 40 - 50% higher (mean and integral) absorbed dose when a tube
voltage of 70 kV is used.
Conclusions: The results of this study clearly indicate that 60 kV should be used for dental intraoral radiographic
examinations for approximal caries detection.
Keywords: dental radiography; dental digital radiography; bitewing radiography; radiation dosage; radiographic
image enhancement.
Accepted for publication: 7 July 2013
To cite this article:
Hellén-Halme K, Nilsson M. The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube
Voltages of 60 and 70 kV for Bitewing Imaging.
URL: http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.pdf
doi: 10.5037/jomr.2013.4302
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.1
(page number not for citation purposes)
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.2
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
INTRODUCTION
The purpose of all radiographic examinations is to
provide reliable diagnostic information allowing rapid
and suitable treatment of the patient. These examinations
must be performed with great care to ensure sufcient
image quality while exposing the patient to the lowest
dose possible. In order to increase the sensitivity and
specicity of a particular diagnostic method, every
link in the diagnostic chain must be optimized and
evaluated for the specic task at hand. Many studies
have been performed in dental digital radiography to
evaluate digital detectors [1-5], monitors [6-8], viewing
conditions [7,9,10] and tube voltage [11-14].
The effect of tube voltage on radiographic image quality
and diagnostic accuracy for dental carious lesions has
been investigated by several authors. Svenson et al.
[12] concluded that an optimal balance was obtained
between the absorbed dose to the patient and diagnostic
accuracy with an analogue lm technique using a
tube voltage of 60 kV. In a later study using a digital
sensor technique [13] no signicant difference was
found in the diagnostic accuracy of approximal carious
lesions when using tube voltages of 60 kV and 70 kV.
In another previous study by Vandenberge and Jacobs
[15] it was concluded that 63 kV and 70 kV provided
a similar diagnostic accuracy and image quality for
periodontal disease. The main opinion among vendors
and many users is that digital sensors often perform
with a higher subjective image quality at a higher tube
voltage, although no studies could be found supporting
this belief.
Optimization of any radiological procedure is a matter of
obtaining adequate image quality at the lowest possible
absorbed dose to the patient. In general dental practice
radiographs are often taken every time the patient
attends the clinic. Carious lesions are small, faint objects
in the X-ray image, superimposed on a background of
anatomical structures, which may impede detection.
The dose administered by standard dental X-ray units
can be adjusted by changing the exposure time or the
tube voltage.
Self-developing Gafchromic® lm (XR-QA2,
International Specialty Products, Wayne, NJ, USA)
has been used previously to measure absorbed dose
and its distribution in phantoms simulating the clinical
situation [16-18]. This offers a simple and accurate way
of mapping the dose distributions from radiographic
examinations. In a recent study [19] some support was
found of the hypothesis that exposure to dental X-rays,
particularly multiple exposures, may be associated with
an increased risk of thyroid cancer. Since it has been
shown that reducing the voltage from 70 kV to 60 kV
does not reduce image quality, we have investigated the
effect of voltage reduction on the absorbed dose to the
patient at these two voltages.
MATERIAL AND METHODS
Self-developing Gafchromic® lm (XR-QA2,
International Specialty Products, Wayne, NJ, USA) was
used to measure the absorbed dose and its distribution
in phantoms simulating the clinical situation. This lm
has a sensitive layer containing a crystalline diacetylene
monomer which polymerises and, as a result, darkens
when irradiated. This provides a simple and accurate
way of mapping the dose distributions from radiographic
examinations. The response of Gafchromic® lm is not
linear to the absorbed dose [20]. The response curve of
the lm was obtained by irradiating the lm with X-rays
when it was placed adjacent to a calibrated ionisation
chamber (Radcal 10X6-6, Radcal Corporation,
Monrovia, CA, USA) which measured the absorbed
dose to the lm. The lm, in which different sections
were irradiated with different absorbed doses, was
digitalized using a high-quality at-bed scanner (Epson
Perfection 4990, Seiko Epson Corporation, Nagano,
Japan). The results were used to obtain a polynomial
calibration curve which was then used to calculate the
actual absorbed dose distributions in the lms irradiated
in the phantom. Response curves were also obtained
for 60 and 120 kV, respectively, and were found to be
identical to that for 70 kV. Therefore, the same response
curve could be used for the experiments with 60 and 70
kV, respectively.
As the output from a dental intraoral X-ray unit is
very low and the Gafchromic® lm has low sensitivity,
a standard X-ray tube for medical radiology (A-196,
Varian Medical Systems, Inc., Salt Lake City, UT, USA)
with a standard collimating device (Svendx SX100-
MF, Santax Medico A/S, Aarhus, Denmark) was - for
practical reasons - used to irradiate the lm in the head
phantom. The output from a medical radiology X-ray
tube is 50 - 100 times higher than that from an intraoral
X-ray tube. This means that the experiments could be
carried out using a few exposures with the medical unit
instead of having to make more than 500 exposures
with the intraoral unit. The radiation eld produced
by the medical X-ray unit was compared (uniformity,
penumbra regions axial and transverse) with that of a
standard intraoral dental unit (Planmeca Intra, Planmeca
Oy, Helsinki, Finland). For that purpose, two pieces of
Gafchromic® lm were irradiated with identical eld
size and focal distance. The ltration of the beam from
the standard X-ray tube was adjusted so that the half-
value layer was the same as for the intraoral unit.
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.3
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
The basis for comparison of the two X-ray tube voltages
was that the signal-to noise ratio in the images was the
same. The ratio was measured in images produced in
a geometry used in a previous study [10] simulating
the clinical case with extracted teeth mounted in
PRESIDENT putty (Coltène Whaledent AG, Cuyahoga
Falls, Ohio, USA). In order to obtain the same signal-
to-noise ratio, the exposure time for 70 kV had to be
reduced with 20%. This reduction also resulted in an
approximately equal entrance skin dose for 60 and
70 kV, respectively.
To simulate a dental patient, the head of an anatomical
phantom (Rando/RAN100, The Phantom Laboratory,
Salem, NY, USA) was used. The Rando head phantom
consists of natural bone, full dentition and a soft plastic
simulating tissue, and is well suited and frequently
used in dosimetry studies. The Gafchromic® lm was
cut with a pair of scissors to t between the slices of
the anatomical phantom (Figure 1). The anatomical
phantom was irradiated corresponding to two bitewing
exposures. This was done by using the same entrance
angle for the X-ray eld as for normal intraoral units.
The dose distributions were measured in the occlusal
plane, and ± 50 mm cranially and caudally to evaluate the
primary and scattered dose distributions, respectively.
Following irradiation the lms were digitized in the
scanner and read into an image processing program
(ImageJ, NIH, Bethesda, MD, USA). The measured
pixel values were converted to absorbed dose using the
polynomial calibration curve. The dose distributions
were recalculated in order to correspond to exposure of
the patient from two standard bitewing images.
The sensor used when obtaining a radiographic image
in a patient is in itself an efcient beam stopper.
When placed intraorally, the absorbed dose behind the
sensor is drastically reduced. However, an intraoral
digital sensor could not be placed inside the phantom.
Therefore, the attenuation of two types of sensors:
Planmeca DIXI2 (Planmeca Oy, Helsinki, Finland)
and a CDR wireless sensor (Schick Technologies, Inc.,
Long Island City, NY, USA) was measured at 60 and
70 kV with an ionization chamber (Radcal 10X6-6,
Radcal Corporation, Monrovia, CA, USA) with 4 cm
of plexiglass in front of the sensor in order to produce
a similar amount of scatter as in the clinical case. Both
sensors are scintillation detectors using CsI (Tl).The
dose distributions behind the position where the sensor
would have been placed in the mouth were corrected
for sensor attenuation by scaling the dose values in the
region affected by attenuation of the sensor.
RESULTS
The radiation eld of the standard X-ray tube used was
found to have properties very similar to those of the
dental X-ray unit, as can be seen in Figure 2, where dose
proles along the main axes of the radiation eld are
shown. The use of the standard X-ray unit was therefore
considered representative of the clinical situation.
The transmission of the DIXI2 intraoral sensor is
4.4% for a tube current of 60 kV and 4.6% for 70 kV.
The corresponding values for the Schick CDR sensor
were 2.4% and 2.7%. The lower values for the Schick
sensor are explained by the fact that this sensor is
wireless, and is powered by a small battery which
increases its attenuation. The dose distributions in the
Gafchromic® lm for the two different tube voltages
are shown in Figure 3, while Figure 4 shows the dose
distributions as isodose curves for the same entrance
dose at tube voltages of 60 and 70 kV. It should be
noted that the absorbed dose outside of the primary
radiation eld, i.e. in the cranial and caudal sections,
is only a few percent of that inside the primary eld.
The effect of using sensors in the clinical situation on
Figure 1. The anatomical head phantom and the Gafchromic® lm cut to t between the different layers of the phantom.
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.4
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
Figure 4. Dose distributions represented as isodose lines. The values given on the
right are absorbed doses in mGy for exposures of 0.12 s at 60 kV and 0.1 s at 70 kV:
A = cranial level, B = occlusal level, C = caudal level.
Figure 3. Gafchromic® lm after exposure in the
phantom at 60 kV and 70 kV: A = cranial level,
B = occlusal level, C = caudal level.
Figure 2. Signal proles along the minor and major axes for the standard X-ray unit used in this study and for a conventional dental X-ray
unit, showing the similarity between them.
Dental unit
Standard unit
Minor axis
Major
axis
Signal proles - major axis
Signal proles - minor axis
0 200 400 600
0 200 400 600
210
200
190
180
170
160
150
210
200
190
180
170
160
150
Relative lm blackening Relative lm blackening
Position
Position
60 kV 70 kV
A
B
C
60 kV 70 kV
5
4
3
2
1
Absorbed dose (mGy)Absorbed dose (mGy)Absorbed dose (mGy)
175
150
100
50
25
10
5
5
4
3
2
1
A
B
C
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.5
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
the dose distributions is illustrated in Figure
5. The lack of scatter caused by the presence
of the sensor can be estimated by integrating
the dose distribution that would have been
shadowed by the sensor in relation to the
total integrated absorbed dose within the
primary beam. Table 1 gives the absorbed
dose expressed as the ratio of the maximal,
mean and integral doses resulting from
exposure using tube voltages of 70 and
60 kV for the same entrance dose.
DISCUSSION
The purpose of this study was to evaluate
how tube voltage affected the absorbed
dose within the primary radiation eld,
and outside the primary radiation eld
Figure 5. The values of absorbed dose at the occlusal level without a sensor and
the simulated levels with a sensor in place.
Without sensor - 60 kV With sensor - 60 kV
Absorbed dose (mGy)
175
150
100
50
25
10
5
Table 1. Ratios of the absorbed doses (mGy) resulting from exposure using tube
voltages of 70 kV and 60 kV
Layer Maximal dose Mean dose Integral dose
50 mm cranially 2.46 1.4 1.4
Occlusal 0.99 1.5 1.49
50 mm caudally 2.49 1.43 1.4
due to scattered radiation. The major principles when
undertaking any radiological procedure are justication
and optimisation. Optimisation means that the absorbed
dose to the patient is kept as low as reasonably
achievable while the diagnostic value of the procedure
is maintained. It can be argued that the absorbed dose,
and hence the effective dose, are very low for a dental
intraoral exposure. On the other hand, the number of
intraoral X-ray examinations performed is relatively
high, and is the most common X-ray procedure in
the Western world. Despite the low individual dose,
the effects on the population as a whole cannot be
neglected. In a recent publication, the risk of thyroid
cancer as a result of dental X-ray examinations was
extensively discussed, and it was concluded that dose
optimization in dental radiography should be urgently
addressed [19].
In intraoral imaging, only a few parameters that affect
the absorbed dose to the patient can be altered. Given
proper ltration and collimation, only the tube voltage
and the exposure time can be adjusted to change the
absorbed dose. When using digital detectors, it is the
responsibility of the dentist to use a dose at which the
quantum noise will not impair the diagnostic accuracy.
Therefore, the parameter affecting the dose which
should be studied in detail is the tube voltage. Today,
the lowest tube voltage (kV) permitted and used in the
Western world is 50 kV [21,22]. In Europe, there is an
on-going discussion on increasing the lower limit to
60 kV [22]. In Sweden, the permitted tube voltage
interval is 60 - 75 kV for general dental practitioners
[23]. This study was based on a comparison of 60 and
70 kV, which are the two most common tube voltages
used in Sweden.
Previous studies have been carried out to evaluate
different tube voltages in intraoral imaging. Svenson
et al. [12] concluded that 60 kV was preferable when
using analogue lm. Kaeppler et al. [24] showed that
increasing the tube voltage from 60 to 90 kV did not
have any effect on either the local absorbed dose or the
effective dose. They did not investigate how the image
quality or the diagnostic accuracy was affected when
the tube voltage was increased. In a study on a charge-
coupled device (CCD) Kitagawa et al. [25] found that
the estimated signal-to-noise ratio improved at a lower
tube voltage. Results reported by Hayakawa et al. [26]
showed that the low-contrast resolution of a CCD
sensor decreased when the tube voltage was increased
from 60 to 70 kV. In a previous study [10], we found
no signicant difference in the diagnostic accuracy for
any approximal carious lesions when evaluating digital
radiographs using tube voltages of 60 and 70 kV.
Due to the higher photon energies using tube voltage of
70 kV, a larger fraction of the photons is scattered than
at 60 kV. Additionally, the mean energy of the scattered
photons generated at 70 kV is higher than those at
60 kV, and their range is thus longer. This should result
in a higher absorbed dose outside the primary radiation
eld at 70 kV than at 60 kV. This is conrmed by the
results of this study, and is illustrated in Figure 4. The
low dose levels outside of the primary eld should not be
neglected, since the scattered radiation causing the dose
will inevitably hit sensitive tissues as the brain, thyroid
and salivary glands. Here, it is obvious that 70 kV will
cause a signicantly higher dose outside of the primary
radiation eld. It should also be noted that the dose
distributions cranially and caudally of the occlusal plane
were measured without a sensor blocking the primary
photons (as it was not possible to insert a sensor inside
the phantom). Since most of the scattered radiation
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.6
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
is produced in front of the sensor, due to the high intensity
of photons in this area, the lack of a blocking sensor will
have a small effect on the amount of scattered radiation.
Since the sensor will block the beam to an extent that
approximately 10% less scatter will be generated in the
phantom, the measured dose outside of the primary eld
may be slightly overestimated. Furthermore, as most
scattered radiation is produced in front of the sensor
and has an angular distribution that is generally in the
forward direction with respect to the primary beam, this
overestimation is clearly below 10%.
Several studies [13-15] have shown that the accuracy and
reliability of Gafchromic® lm for dose measurements
are adequate. The results obtained in this study clearly
conrm these previous ndings. Furthermore, the lm
is extremely user friendly and makes it possible to
measure absorbed dose distributions with an almost
unsurpassed spatial resolution. Its only drawback is its
low sensitivity which requires repeated exposures from
X-ray units with low output.
The ndings of this study show that the patient receives
an approximately 40 - 50% higher absorbed dose (mean
value) when using a tube voltage of 70 kV.
Our results indicate that lowering the tube voltage from
70 to 60 kV will result in a lower dose to the patient
without compromising image quality for evaluation of
carious lesions. Further studies are needed to investigate
if this also applies to other diagnostic tasks in bitewing
imaging, i.e. periodontal bone levels.
CONCLUSIONS
The results of this study clearly indicate that 60 kV
should be used for digital bitewing examinations for
approximal caries detection.
ACKNOWLEDGEMENTS AND DISCLOSURE
STATEMENTS
The authors declare that they have no conict of
interests.
REFERENCES
1. Kullendorff B, Nilsson M, Rohlin M. Diagnostic accuracy of direct digital dental radiography for the detection of
periapical bone lesions: overall comparison between conventional and direct digital radiography. Oral Surg Oral Med
Oral Pathol Oral Radiol Endod. 1996 Sep;82(3):344-50. [Medline: 8884837] [doi: 10.1016/S1079-2104(96)80364-7]
2. Kitagawa H, Scheetz JP, Farman AG. Comparison of complementary metal oxide semiconductor and charge-coupled
device intraoral X-ray detectors using subjective image quality. Dentomaxillofac Radiol. 2003 Nov;32(6):408-11.
[Medline: 15070845] [doi: 10.1259/dmfr/19990417]
3. Paurazas SB, Geist JR, Pink FE, Hoen MM, Steiman HR. Comparison of diagnostic accuracy of digital imaging by using
CCD and CMOS-APS sensors with E-speed lm in the detection of periapical bony lesions. Oral Surg Oral Med Oral
Pathol Oral Radiol Endod. 2000 Mar;89(3):356-62. [Medline: 10710463] [doi: 10.1016/S1079-2104(00)70102-8]
4. Wenzel A. Digital radiography and caries diagnosis. Dentomaxillofac Radiol. 1998 Jan;27(1):3-11. Review.
[Medline: 9482015] [doi: 10.1038/sj.dmfr.4600321]
5. Vandenberghe B, Bosmans H, Yang J, Jacobs R. A comprehensive in vitro study of image accuracy and quality for
periodontal diagnosis. Part 2: the inuence of intra-oral image receptor on periodontal measurements. Clin Oral Investig.
2011 Aug;15(4):551-62. Epub 2010 May 12. [Medline: 20461423] [doi: 10.1007/s00784-010-0417-7]
6. Hellén-Halme K, Nilsson M, Petersson A. Effect of monitors on approximal caries detection in digital radiographs--
standard versus precalibrated DICOM part 14 displays: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol
Endod 2009; 107: 716-20 [Medline: 19201219] [doi: 10.1016/j.tripleo.2008.12.011]
7. Hellén-Halme K, Petersson A, Warfvinge G, Nilsson M. Effect of ambient light and monitor brightness and contrast
settings on the detection of approximal caries in digital radiographs: an in vitro study. Dentomaxillofac Radiol. 2008
Oct;37(7):380-4. [Medline: 18812599] [doi: 10.1259/dmfr/26038913]
8. Cederberg RA, Frederiksen NL, Benson BW, Shulman JD. Inuence of the digital image display monitor on observer
performance. Dentomaxillofac Radiol. 1999 Jul;28(4):203-7. [Medline: 10455382] [doi: 10.1038/sj.dmfr.4600441]
9. Cederberg RA, Frederiksen NL, Benson BW, Shulman JD. Effect of different background lighting conditions on
diagnostic performance of digital and lm images. Dentomaxillofac Radiol. 1998 Sep;27(5):293-7. [Medline: 9879219]
[doi: 10.1038/sj.dmfr.4600377
10. Wenzel A. A review of dentists’ use of digital radiography and caries diagnosis with digital systems. Dentomaxillofac
Radiol. 2006 Sep;35(5):307-14. Review. [Medline: 16940477] [doi: 10.1259/dmfr/64693712]
11. Hellén-Halme K. Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs.
An in vitro study. Clin Oral Investig. 2011 Apr;15(2):209-13. Epub 2010 Feb 9. [Medline: 20143243]
[doi: 10.1007/s00784-010-0383-0]
12. Svenson B, Gröndahl HG, Petersson A, Olving A. Accuracy of radiographic caries diagnosis at different kilovoltages and
two lm speeds. Swed Dent J. 1985;9(1):37-43. [Medline: 3859029]
13. Svenson B, Petersson A. Inuence of tube voltage on radiographic diagnosis ofcaries in premolars and molars. Swed Dent
J. 1991;15(5):245-50. [Medline: 1767370]
14. Vandenberghe B, Corpas L, Bosmans H, Yang J, Jacobs R. A comprehensive in vitro study of image accuracy
and quality for periodontal diagnosis. Part 1: the inuence of X-ray generator on periodontal measurements using
conventional and digital receptors. Clin Oral Investig. 2011 Aug;15(4):537-49. Epub 2010 May 5. [Medline: 20443035]
[doi: 10.1007/s00784-010-0416-8]
15. Vandenberghe B, Jacobs R. The Inuence of Tube Potential on Periodontal Bone Level Measurements and Subjective
Image Quality Using a Digital Photostimulable Storage Phosphor Sensor. J Oral Maxillofac Res 2010; 1: e5.
URL: http://www.ejomr.org/JOMR/archives/2010/1/e5/e5ht.pdf
16. Oren U, Rääf CL, Mattsson S. Gafchromic lm as a fast visual indicator of radiation exposure of rst responders. Radiat
Prot Dosimetry. 2012 Jun;150(1):119-23. Epub 2011 Sep 17. [Medline: 21926417] [doi: 10.1093/rpd/ncr374]
17. Abadi S, Mehrez H, Ursani A, Parker M, Paul N. Direct quantication of breast dose during coronary CT angiography
and evaluation of dose reduction strategies. AJR Am J Roentgenol. 2011 Feb;196(2):W152-8. [Medline: 21257856]
[doi: 10.2214/AJR.10.4626]
18. Eaton DJ, Duck S. Dosimetry measurements with an intra-operative x-ray device. Phys Med Biol. 2010 Jun 21;55(12):N359-
69. Epub 2010 May 26. [Medline: 20505225] [doi: 10.1088/0031-9155/55/12/N02]
19. Memon A, Godward S, Williams D, Siddique I, Al-Saleh K. Dental x-rays and the risk of thyroid cancer: a case-control
study. Acta Oncol. 2010 May;49(4):447-53. [Medline: 20397774] [doi: 10.3109/02841861003705778]
20. Watanabe Y, Patel GN, Patel P. Evaluation of a new self-developing instant lm for imaging and dosimetry. Radiat Prot
Dosimetry. 2006;120(1-4):121-4. Epub 2006 Apr 27. [Medline: 16644995] [doi: 10.1093/rpd/nci551]
21. Radiation protection in dentistry. NCRP report no 145. National Council on Radiation Protection and Measurements.
Bethesda, MD, USA, 2003: European Commission on radiation protection (RP 91) URL: http://ec.europa.eu/energy/
nuclear/radiation_protection/doc/publication/091_en.pdf. Accessed 2011-11-15.
22. European Commission on radiation protection RP (162) URL: http://ec.europa.eu/energy/nuclear/radiation_protection/
consultations/doc/rp_91_update_2009_draft_for_consultation.pdf. Accessed 2012-07-20.
23. Swedish Radiation Safety Authority. URL: http://www.stralsakerhetsmyndigheten.se/In-English/About-the-Swedish-
Radiation-Safety-Authority1. Accessed 2012-07-20.
24. Kaeppler G, Dietz K, Herz K, Reinert S. Factors inuencing the absorbed dose in intraoral radiography. Dentomaxillofac
Radiol. 2007 Dec;36(8):506-13. [Medline: 18033949] [doi: 10.1259/dmfr/32903218]
25. Kitagawa H, Farman AG. Effect of beam energy and ltration on the signal-to-noise ratio of the Dexis intraoral X-ray
detector. Dentomaxillofac Radiol. 2004 Jan;33(1):21-4. [Medline: 15140818] [doi: 10.1259/dmfr/26493631]
26. Hayakawa Y, Farman AG, Scarfe WC, Kuroyanagi K, Rumack PM, Schick DB. Optimum exposure ranges for computed
dental radiography. Dentomaxillofac Radiol. 1996 Apr;25(2):71-5. [Medline: 9446976]
http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.htm J Oral Maxillofac Res 2013 (Jul-Sep) | vol. 4 | No 3 | e2 | p.7
(page number not for citation purposes)
JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH Hellén-Halme and Nilsson
To cite this article:
Hellén-Halme K, Nilsson M. The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube
Voltages of 60 and 70 kV for Bitewing Imaging.
J Oral Maxillofac Res 2013;4(3):e2
URL: http://www.ejomr.org/JOMR/archives/2013/3/e2/v4n3e2ht.pdf
doi: 10.5037/jomr.2013.4302
Copyright © Hellén-Halme K, Nilsson M. Published in the JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH
(http://www.ejomr.org), 1 October 2013.
This is an open-access article, rst published in the JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH, distributed
under the terms of the Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 Unported License, which
permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work and is
properly cited. The copyright, license information and link to the original publication on (http://www.ejomr.org) must be
included.
... Dosimetric film has been suggested as a more accurate alternative for these modalities. 19,20 As far as we know, no study has determined the dose to child patients from all X-ray modalities of interest when specifically diagnosing an impacted canine. The current study aims to determine and compare the dose from both periapical and panoramic radiographs as well as CBCT examinations for children having impacted canines with possible root resorption in neighbouring teeth, using TLD and film measurements. ...
... 14-16 An alternative method, using self-developing radiographic film, has also been used more recently. 19,20,30 Compared to point measurements, such as TLD, film dosimetry has the advantage of allowing for high resolution continuous measurements over a large area, thus making it suitable for dosimetry in panoramic and intraoral radiographs. The use of discreet measurement points is unsuitable for examinations with pronounced dose gradients in the transverse plane, such as panoramic radiographs, due to the large uncertainties in determining the mean organ dose. ...
... Film dosimetry also provides advantages in determining the organ dose in the case of intraoral radiographs, where dose measurements are complicated by the presence of the detector inside the oral cavity as well as the small X-ray field size. 20 The sharp dose gradients for panoramic and intraoral radiographs, compared to CBCT, are illustrated in Figure 1. Therefore, we used film dosimetry for these two modalities instead of TLD measurements. ...
Article
Full-text available
Objectives: To compare the radiation dose to children examined for impacted canines, using two-dimensional examinations (panoramic and periapical radiographs) and cone-beam computed tomography (CBCT). Methods: Organ doses were determined using an anthropomorphic 10-year child phantom. Two CBCT devices, a ProMax3D and a NewTom5G, were examined using thermoluminescent dosimeters. For the panoramic radiograph, a Promax device was used and for periapical radiographs, a Prostyle device with a ProSensor digital sensor was used. Both the panoramic and the intraoral devices were examined using Gafchromic-QR2 dosimetric film placed between the phantom slices. Results: ProMax3D and NewTom5G resulted in an effective dose of 88 µSv and 170 µSv respectively. A panoramic radiograph resulted in an effective dose of 4.1 µSv, while a periapical radiograph resulted in an effective dose of 0.6 µSv and 0.7 µSv using a maxillary lateral projection and central maxillary incisor projection respectively. Conclusions: The effective dose from CBCT ranged from 140 times higher dose (NewTom5G compared to two periapical radiographs) to 15 times higher dose (ProMax3D compared to three periapical and one panoramic radiograph) than a two-dimensional examination.
... Gafchromic film dosimetry, in particular, is one of the more common methods for providing dose verification and for measuring dose maps with high spatial resolution, low energy dependence, and adequate accuracy. [9][10][11] Gafchromic film is a self-developing film and has a radiosensitive layer that contains a crystalline diacetylene monomer, which, when irradiated, polymerizes and darkens. 12 The amount of darkening is related to the absorbed dose. ...
... 12 The film was exposed to 8 doses of radiation between 2000 and 110,000 mGy; 1 nonirradiated film was used for calculating the background value. In a previous study, 11 the film was calibrated for kilovolt (kV) settings from 60 to 120 kV by using the same ionization chamber (which has a flat energy response in this interval). It was found that the calibration curves coincided. ...
Article
Full-text available
Objectives. The aim of this study was to map and compare the distributions of absorbed doses with Gafchromic film for panoramic radiography and cone beam computed tomography (CBCT) examinations of the temporomandibular joint (TMJ) by using adult and child phantoms. Study Design. Gafchromic films were placed at 5 selected levels of anthropomorphic head phantoms of an adult and a child. Clinical protocols for panoramic and CBCT imaging of the TMJ were used for three 2-dimensional or 3-dimensional dental x-ray units. Mean absorbed doses in a set of radiosensitive tissues within the oral and maxillofacial regions were estimated. Results. The absorbed doses varied considerably among and within radiosensitive tissues. The bone surface and the salivary glands received the highest absorbed doses compared with other tissues, in both panoramic and CBCT examinations of the TMJ. The radiation burden to the adult phantom was generally higher than that to the child phantom. Small right and left fields of view were associated with lower amounts of radiation, in contrast to a single larger field of view. Conclusions. The absorbed dose within all radiosensitive tissues varied considerably in relation to examination type, x-ray unit, clinical settings, and patient age. The mean doses were smaller when using 2 (bilateral) 4 £ 4 cm volumes than with use of one
... These diagnostic examinations must perform under standard protocol to achieve sufficient image quality with the lowest possible absorbed dose to the patient. The important factors that affect the absorbed dose to the patient is the exposure time and the tube voltage [13][14]. The main target is to get an optimal condition between the dose delivered to the patient and the diagnostic physical exposure parameters accuracy and reproducibility with an analogue film technique at certain tube potential and time. ...
... The absorbed dose to the patient in dental intraoral units can control by adjusting the exposure time and the tube potential [13]. Significant decreases in radiation dose of dental radiography occur with the use of faster image receptors, intra-oral film holders, rectangular collimation for bitewing radiography, and also use of long, rectangular position indicating devices. ...
Article
Full-text available
In this study, an intra oral dental unit (Siemens-70) at King Abdul Aziz University (KAU) Dental Hospital was selected and investigated for visual image quality assessment and radiation protection purposes. Radiation dosimetry for determining the optimum image quality with the lowest radiation exposure to the patient was carried out. A DXTTR dental radiography trainer phantom head and neck, portable survey meter Model RAD EYE-B20, and radiation dosimetry system RADCAL Acuu-pro were used in this study. RADCAL Accu-pro is a non-invasive kV system, reliable instruments to measure and diagnose all X-ray machines including dental units. The radiation exposure to patients in (mGy) was measured using RADCAL ionization chamber Model 10×6-6. The best image quality with the lowest exposure dose was assessed for conventional intraoral X-ray film (Kodak type E) and the digital processing sensor (RVG 5200). Radiation survey level was done during this study for safety and protection purposes.
... Gafchromic film dosimetry, in particular, is one of the more common methods for providing dose verification and for measuring dose maps with high spatial resolution, low energy dependence, and adequate accuracy. [9][10][11] Gafchromic film is a self-developing film and has a radiosensitive layer that contains a crystalline diacetylene monomer, which, when irradiated, polymerizes and darkens. 12 The amount of darkening is related to the absorbed dose. ...
... 12 The film was exposed to 8 doses of radiation between 2000 and 110,000 mGy; 1 nonirradiated film was used for calculating the background value. In a previous study, 11 the film was calibrated for kilovolt (kV) settings from 60 to 120 kV by using the same ionization chamber (which has a flat energy response in this interval). It was found that the calibration curves coincided. ...
Article
Full-text available
Objectives: The aim of this study was to map and compare the distributions of absorbed doses with Gafchromic film for panoramic radiography and cone beam computed tomography (CBCT) examinations of the temporomandibular joint (TMJ) by using adult and child phantoms. Study design: Gafchromic films were placed at 5 selected levels of anthropomorphic head phantoms of an adult and a child. Clinical protocols for panoramic and CBCT imaging of the TMJ were used for three 2-dimensional or 3-dimensional dental x-ray units. Mean absorbed doses in a set of radiosensitive tissues within the oral and maxillofacial regions were estimated. Results: The absorbed doses varied considerably among and within radiosensitive tissues. The bone surface and the salivary glands received the highest absorbed doses compared with other tissues, in both panoramic and CBCT examinations of the TMJ. The radiation burden to the adult phantom was generally higher than that to the child phantom. Small right and left fields of view were associated with lower amounts of radiation, in contrast to a single larger field of view. Conclusions: The absorbed dose within all radiosensitive tissues varied considerably in relation to examination type, x-ray unit, clinical settings, and patient age. The mean doses were smaller when using 2 (bilateral) 4 × 4 cm volumes than with use of one 14 × 5 cm volume.
... Another limitation owes to the fact that not all potential voltage and current combinations could be evaluated. Around a decade ago, a study by Hellén-Halme and Nilsson has shown that a voltage of 70 kV leads to 40-50% higher absorbed dose in patients than 60 kV [38], and another study found that a voltage of 70 kV does not result in a significantly higher sensitivity than 60 kV [39]. This is in line with the sensor recordings of the present study, even though a slight reduction in specificity was found for 60 kV. ...
Article
Full-text available
Objectives Bitewing radiography is considered to be of high diagnostic value in caries detection, but owing to projections, lesions may remain undetected. The novel bitewing plus (BW +) technology enables scrolling through radiographs in different directions and angles. The present study aimed at comparing BW + with other 2D and 3D imaging methods in terms of sensitivity, specificity, and user reliability. Materials and methods Five human cadavers were used in this study. In three cadavers, natural teeth were transplanted post-mortem. BW + , two-dimensional (digital sensors, imaging plates, 2D and 3D bitewing radiographs) and 3D methods (high and low dose CBCT) were taken. Carious lesions were evaluated on 96 teeth at three positions (mesial, distal, and occlusal) and scored according to their level of demineralization by ten observers, resulting in 35,799 possible lesions across all observers and settings. For reference, µCT scans of all teeth were performed. Results Overall, radiographic evaluations showed a high rate of false-negative diagnoses, with around 70% of lesions remaining undetected, especially enamel lesions. BW + showed the highest sensitivity for dentinal caries and had comparatively high specificity overall. Conclusions Within the limits of the study, BW + showed great potential for added diagnostic value, especially for dentinal caries. However, the tradeoff of diagnostic benefit and radiation exposure must be considered according to each patient’s age and risk.
... One study showed that reducing the tube voltage from 70 kVp to 60 kVp did not compromise image quality for the evaluation of carious lesions. The same study reported that the patient was exposed to an absorbed dose that was approximately 40-50% higher when a tube voltage of 70 kVp was used [21]. Dehghani et al. suggested the use of PSP with 60 kVp to comply with the As Low As Reasonably Achievable (ALARA) rule, especially when a tooth has clinical signs or discoloration indicative of caries, and they attributed this to the higher sensitivity and lower exposure time of PSP compared to intraoral film [17]. ...
Article
Full-text available
Objective: The aim of this study was to evaluate periapical radiographs of enamel caries, dentin caries, and deep caries with exposed pulp and intact teeth obtained in vitro using photo-stimulated phosphor plates (PSP) under different exposure parameters. Methods: 3 non-carious extracted molars were selected. The obtained molars were embedded in the wax created from pink wax by ensuring approximal contact and a base was created. 14 different imaging protocols were used with 60 kVp, 4 mA 0.02-0.1 second and 70 kVp 7 mA, 0.25-1.25 second exposure parameters. Intact teeth were imaged with these various imaging protocols. Artificial cavities were then created for enamel caries, dentin caries and deep caries with exposed pulp and imaged according to the same protocols. The images were evaluated by 3 clinicians who were blind to the exposure protocol and caries status. Inter-observer agreement with actual situations was examined with Kappa statistics. Results: In the low-dose group, the kappa values of observer 1, observer 2, and observer 3 were 0.905, 0.952, 0.952, respectively. The kappa values of observer 1, observer 2, and observer 3 in the ultralow-dose group were 0.833, 1, 1, and the kappa values of observer 1, observer 2, and observer 3 in the high-dose group were 1, 1, 0.833, respectively. The results obtained in all groups showed a statistically significant-excellent agreement (p<0.001). Conclusion: Approximal caries can be diagnosed with intraoral radiography obtained with low radiation doses with PSP in dentistry. Thus, patients could be exposed to less ionizing radiation.
... Sinar-X merupakan sumber radiasi pengion yang paling banyak digunakan untuk pemeriksaan diagnostik dalam aplikasi klinik (Osei and Darko 2013;Bushong 2001;Ali et al. 2013;Abdemola et al. 2013). Pemeriksaan radiodiagnostik adalah salah satu pemanfaatan dari radiasi pengion untuk penegasan hasil diagnosis yang dibutuhkan oleh pasien dalam rangka mengindentifikasikan abnormalitas dari seorang pasien, dengan paparan radiasi seminim mungkin namun memberikan kualitas pencitraan medis yang baik (Abrahams et al. 2015;Kristina and Nilsson 2013.). Salah satu modalitas yang tersedia secara luas di berbagai rumah sakit adalah radiografi planar yang dapat menghasilkan gambaran dua dimensi menggunakan radiasi sinar-X (Ozbayrak et al. 2015). ...
Article
Full-text available
Sinar-X merupakan sumber radiasi pengion yang paling banyak digunakan untuk pemeriksaan diagnostik dalam aplikasi klinik. Radiasi pengion memang memberikan manfaat yang besar dalam pemeriksaan radiodiagnostik, namun paparan radiasi memiliki resiko yang berbahaya sehingga proteksi radiasi terhadap pasien atau pekerja di sekitar daerah smber radiasi perlu diperhatikan, terutama wanita hamil karena berdampak langsung kepada janin. Dalam penelitian digunakan pesawat radiologi planar dengan filter 2 mmAl, solid water phantom dengan ketebalan sebesar 20 cm yang merepresentasikan ketebalan Pelvis, dan Multimeter X-ray untuk pengukuran dosis x-ray pada Focus to Film Distance (FFD) berturut – turut adalah 100 cm, 110 cm, 120 cm, 130 cm, 140 cm dan 150 cm. Pada setiap FFD, digunakan tegangan mulai dari 70 kVp, 73 kVp, 77 kVp, 81 kVp, 85 kVp, 90 kVp hingga 96 kVp pada beban tabung 10 mAs. Penentuan estimasi dosis janin dilakukan dengan menggunakan program Fet.Dose V5. Hasil penelitian menunjukkan bahwa nilai faktor resiko radiasi terhadap janin sebanding dengan tegangan tabung (kVp) dan berbanding terbalik dengan jarak obyek ke sumber radiasi. Untuk satu kali iradiasi dengan pesawat sinar-x dengan pengaturan yang memungkinkan nilai faktor resiko radiasi tertinggi untuk janin diperoleh pada FFD 100 cm dan tegangan tabung 96 kVp sedangkan nilai faktor resiko radiasi terendah diperoleh saat FFD 150 cm dan tegangan tabung 70 kVp.
... Sinar-X telah menjadi sumber radiasi pengion yang dibuat oleh manusia untuk kepentingan diagnostik dan terapi [1][2] [3]. Sebagai sarana penunjang, sinar-X sangat membantu dalam pemeriksaan diagnostik [4], salah satunya pemeriksaan radio diagnostik yang memberikan manfaat berupa informasi diagnosis yang dibutuhkan oleh pasien [3][5] [6] [7]. Meskipun memberikan manfaat, paparan radiasi pengion yang dihasilkan oleh sinar-X memiliki dampak buruk terhadap pasien yang sedang berada dalam masa kehamilan, mengingat radiosensitivitas janin dan efek biologis yang buruk terhadap perkembangan janin [8] [9][10] [11], seperti malformasi organ pada usia kehamilan 3 -8 minggu [8] [12][13] [20], retardasi mental pada usia kehamilan 9 -25 minggu hingga sterilitas ataupun kanker pada usia kehamilan 26 minggu sampai kelahiran [20]. ...
Conference Paper
Full-text available
Abstrak-Janin memiliki radiosensitivitas terhadap radiasi pengion, sehingga perhitungan dosis yang diserap oleh pasien perlu dilakukan agar janin memperoleh dosis dibawah 100 mGy seperti yang direkomendasikan oleh International Comission on Radiological Protection (ICRP). Penelitian ini bertujuan untuk membandingkan hasil estimasi dosis janin pada pemeriksaan radiografi pelvis dengan metode perhitungan manual dan simulasi berbantukan program FetDose V4 pada faktor eksposi sebesar 70 kV, 73 kV, 77 kV, 81 kV, 85 kV, 90 kV dan 96 kV, beban tabung 10 mAs pada Focus to Film Distance (FFD) yaitu 100 cm, 110 cm, 120 cm, 130 cm, 140 cm dan 150 cm. Sampel penelitian adalah solid water phantom dengan tebal 18 cm yang mewakili ketebalan Pelvis dan X-Ray Multimeter sebagai alat ukur Entrance Surface Dose (ESD). Hasil penelitian menunjukkan perbedaan yang bermakna (p < 0,05) pada pemeriksaan radiografi dengan FFD berturut-turut 100 cm, 110 cm, 120 cm, dan 130 cm dan tidak terdapat perbedaan yang bermakna (p > 0,05) pada pemeriksaan radiografi dengan FFD 140 cm dan 150 cm. Hasil uji korelasi Pearson antara kVp dan ESD menghasilkan hubungan yang kuat dengan nilai r berada pada rentang 0,70 – 1,00. Rerata hasil estimasi dosis janin dengan perhitungan manual dan simulasi pada setiap FFD berada dibawah nilai 100 mGy. Dengan demikian, FFD yang digunakan dalam penelitian ini dapat digunakan sebagai refrensi pemeriksaan radiodiagnostik.
Article
The need for quality assurance (QA) for digital dental radiography has existed since the introduction of digital imaging; however, the methods and phantoms required to achieve it were not available. This resulted in a chaotic approach to address QA based largely upon subjective analysis of image quality. The American National Standards Institute (ANSI)/American Dental Association (ADA) Quality Assurance Standard 1094 for Digital Intraoral Radiographic Systems (DIRS) presents a paradigm shift to a scientific and objective method of QA rather than one based on subjective assessments. This standard takes into account the contributions of all components of the digital imaging chain that affect the final image quality rather than assessing the various components in isolation. The optimal image is determined for each DIRS through objective analysis of the image quality properties of dynamic range, spatial resolution, and contrast perceptibility. Image optimization, a critical component of a quality assurance program, is the proper balance between diagnostic image quality and radiation dose to the patient. This publication counters disseminated myths and misconceptions with scientific evidence and helps dental practitioners appreciate and understand the benefits of the new ANSI/ADA Standard on QA for DIRS. Statement of Clinical Relevance: Clinical examination combined with high quality radiographs are essential in diagnosis and treatment planning. Optimized radiographs with maximal diagnostic information and minimal exposure require an effective quality assurance program free from the myths and misconceptions that negatively influence digital radiography.
Article
Objectives: To estimate the radiation dose reduction to the thyroid for an anterior oblique occlusal view from the use of a thyroid shield, compare this to the variation in thyroid dose resulting from differences in examination positioning and discuss the additional considerations associated with the use of a thyroid shield before making a recommendation on their routine use for this examination. Methods: Doses to the oral mucosa, the salivary glands, the thyroid, the extra-thoracic airways, the oesophagus and the lungs were directly measured for anterior oblique occlusal x-rays of a Rando phantom with and without a thyroid shield using strips of calibrated XRQA Gafchromic film. The examination was also simulated using Monte Carlo software for the without thyroid shield case for a comparison of the dose and to evaluate the dosimetric effect of sub-optimal examination positioning. Results: A 36% reduction in thyroid dose was measured as a result of thyroid shield use; the effective dose reduction is of the order of 22%. Sub-optimal positioning was found to increase thyroid dose by a far more significant amount. Conclusions: Despite the reduction in thyroid dose, cost-benefit considerations mean that the purchase of a thyroid shield is only recommended where a very high number of anterior oblique occlusal views are undertaken. Optimisation efforts for this examination are better focussed on training in examination positioning.
Article
Full-text available
The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor. A digital photostimulable storage phosphor (PSP) sensor (Vistascan) and a multipulse X-ray generator (Prostyle Intra) with two tube voltages were used in this study. The front, premolar and molar region of two adult human cadaver skulls jaws were imaged using the X-ray tube at 63 kV and 70 kV, both at 8 mA and decreasing exposure times (160 ms, 120 ms and 80 ms). A standardized exposure protocol containing waxed occlusal keys and an aiming device ensured proper and reproducible beam alignment. Three observers assessed the digital radiographs for 31 selected periodontal bone loss sites. Radiographic measurements were compared to physical measurements (Standard). Subjective ratings of lamina dura, crater defect and furcation involvement visibility, contrast perception and bone quality were also performed. Multiple regression equation of the variables kV and exposure time demonstrated no significant difference for the periodontal bone level measurements (P > 0.05). In 90.3% and 96.7% of the measurements for 70 kV and 63 kV respectively, deviation was within 1 mm. The subjective ratings produced similar findings in terms of image quality for both tube voltages and the three exposure times. The results of the present study revealed that tube voltages of 63 kV and 70 kV provided similar accuracy and image quality for periodontal disease diagnosis.
Article
Full-text available
The thyroid gland is highly susceptible to radiation carcinogenesis and exposure to high-dose ionising radiation is the only established cause of thyroid cancer. Dental radiography, a common source of low-dose diagnostic radiation exposure in the general population, is often overlooked as a radiation hazard to the gland and may be associated with the risk of thyroid cancer. An increased risk of thyroid cancer has been reported in dentists, dental assistants, and x-ray workers; and exposure to dental x-rays has been associated with an increased risk of meningiomas and salivary tumours. To examine whether exposure to dental x-rays was associated with the risk of thyroid cancer, we conducted a population-based case-control interview study among 313 patients with thyroid cancer and a similar number of individually matched (year of birth +/- three years, gender, nationality, district of residence) control subjects in Kuwait. Conditional logistic regression analysis, adjusted for other upper-body x-rays, showed that exposure to dental x-rays was significantly associated with an increased risk of thyroid cancer (odds ratio = 2.1, 95% confidence interval: 1.4, 3.1) (p=0.001) with a dose-response pattern (p for trend <0.0001). The association did not vary appreciably by age, gender, nationality, level of education, or parity. These findings, based on self-report by cases/controls, provide some support to the hypothesis that exposure to dental x-rays, particularly multiple exposures, may be associated with an increased risk of thyroid cancer; and warrant further study in settings where historical dental x-ray records may be available.
Article
Full-text available
This study evaluated the effect of two different tube voltages on clinicians' ability to diagnose approximal carious lesions in digital radiographs. One hundred extracted teeth were radiographed twice at two voltage settings, 60 and 70 kV, using a standardized procedure. Seven observers evaluated the radiographs on a standard color monitor pre-calibrated according to DICOM part 14. Evaluations were made at ambient light levels below 50 lx. All observations were analyzed with receiver operating characteristic curves. A histological examination of the teeth served as the criterion standard. A paired t test compared the effects of the two voltages. The significance level was set to p < 0.05. Weighted kappa statistics estimated intra-observer agreement. No significant difference in accuracy of approximal carious lesion diagnosis was found between the two voltage settings. But five observers rated dentin lesions on radiographs exposed at 70 kV better than on radiographs exposed at 60 kV. Intra-observer agreement differed from fair to moderate. There was no significant difference in accuracy of approximal carious lesion diagnosis between digital radiographs exposed with 60 or 70 kV.
Article
Full-text available
The aim of this study was to investigate how brightness and contrast settings of the display monitor and ambient light level (illuminance) in the viewing room affect the clinician's ability to diagnose carious lesions in digital radiographs. Standardized radiographs were taken of 100 extracted teeth. Seven observers evaluated the images for approximal carious lesions twice, once under 50 lux and once under 1000 lux room illumination. Monitor brightness and contrast were varied +/-50% and +/-6%, respectively, to mimic the normal limits of monitor adjustment by an inexperienced user and one optimal setting. This was done by adjusting radiograph brightness and contrast by +/-25%. Thus, five radiographs of each tooth were evaluated. Receiver operating characteristic (ROC) analyses were performed. Histological examinations of the teeth served as the criterion standard. A paired t-test was used to evaluate whether differences in the areas under the ROC curves were significant and kappa was used to evaluate intraobserver agreement. When a monitor with optimal brightness and contrast settings was used to detect approximal carious lesions, ambient light levels less than 50 lux were significantly better than levels above 1000 lux (dentin and enamel lesions, P < 0.01; dentin lesions, P < 0.02). Increasing the contrast setting of the monitor by 6% did not change these results; 50 lux was still significantly better than 1000 lux (enamel lesions, P < 0.01; dentin and enamel lesions, P < 0.02) for evaluating radiographs. Intraobserver agreement differed from fair to good. Reducing ambient light to less than 50 lux significantly increased the accuracy of diagnosing approximal carious lesions on a monitor with an optimal brightness setting and an optimal or slightly higher than optimal contrast setting.
Article
Three types of Gafchromic films have been studied to investigate their potential for use as a visually readable dosemeter for persons acting as first responders in connection with radiological or nuclear emergencies. The two most sensitive film types show a pronounced variation in sensitivity by photon energy and are therefore not suitable for use in cases of unknown exposures. The third film type tested (RTQA2), which is intended for quality control in radiation therapy has a sensitivity that is independent of the radiation quality, and is therefore considered as the most optimal for visual reading in situ. Tests carried out on a group of 10 human observers showed that absorbed doses down to 40 mGy can be detected by the eye. Read by a portable densitometer, qualitative absorbed dose estimates down to 9 mGy can be achieved. The colour change is obtained instantaneously, giving first responders immediate information about the presence of beta-, gamma- and X-ray radiation.
Article
The purpose of this study was to quantify the absorbed radiation dose received by the adult female breast during coronary CT angiography (CTA) and to evaluate the effectiveness of various dose reduction strategies. An adult female thoracic anthropomorphic phantom was scanned using eight different clinical coronary CTA protocols that varied in detector configuration (320 × 0.5 mm or 64 × 0.5 mm), x-ray tube activation (full R-R, 65% R-R, or 70-80% R-R), use of tube current modulation, and use of breast shields. Direct dosimetry measurements were performed using Gafchromic film to determine the absorbed breast dose. Retrospective helical data acquisition using a 64-detector array and a full cardiac cycle without dose modulation or breast shielding is associated with an average absorbed breast dose of 82.9 mGy. Optimization of coronary CTA technique using a 320-detector array and a 70-80% cardiac phase reduces the absorbed breast dose by 78.9% to 17.5 mGy, whereas breast shields used in isolation reduces breast dose by up to 46.8%. The implementation of clinically validated coronary CTA protocols using large-area detector acquisition and prospective ECG gating with limited x-ray tube activation results in substantial breast dose savings of up to 78.9% and should be used whenever possible in combination with bismuth breast shields to achieve further dose reduction.
Article
The INTRABEAM 50 kV x-ray device can be used for intra-operative partial breast irradiation. Spherical applicators are added to the source probe to deliver a radially symmetric radiation dose. Dosimetric data for calculation of absorbed dose were measured for this unit and a superficial unit with a similar beam quality, as defined by half value layer (HVL). Chamber calibration factors, N(K), and chamber correction factors, k(ch), were determined based on HVL, according to the IPEMB code of practice and addendum. Depth doses were also measured using an ionization chamber and GafChromic EBT film. HVL was measured as 0.85-1.30 mm Al across the range of applicator sizes. Values for N(K) and k(ch) were found to be similar for the two units and all INTRABEAM applicator sizes. Therefore, calibration of ionization chambers, radiochromic film and other relative dosimeters could be performed on the superficial unit. This has the advantage of higher dose rates and lower dependence on small variations in detector positioning. Depth dose measurements performed using film also agreed with chamber values, published and manufacturer data, giving a simple and robust method for commissioning and regular quality assurance.
Article
The aim of the study was to determine the image receptors' influence on exposure levels, image accuracy, and quality for periodontal diagnosis. Periodontal defects from cadaver specimens were imaged using two intra-oral conventional films (E-, F/E-speed), four indirect receptors (Digora 8 bit, Vistascan 12 bit with/without filter, Vistascan 16 bit), and two solid-state sensors (Sigma 12 bit, VistaRay 14 bit) at rising exposure (20-160 ms). Three observers assessed the standardized radiographs for alveolar bone measurements (1,732, 31 sites) and for subjective rating of lamina dura, contrast, trabecularization, crater, and furcation involvements. The measurements were compared to the gold standard. For the imaging plates, highest measurement accuracy was found with Vistascan 16 bit (100% within 0.5 mm) and for solid-state sensors with VistaRay 14 bit (100%, 0.5 mm), although the latter are mostly not significantly different. Higher contrast resolution imaging plates require up to 50% less exposure time, but for solid-state sensors, the dose remains unchanged. For the latter, a higher bit depth does seem to provide more accurate depiction of the alveolar crest, counteracting blooming artifacts. The use of a dedicated periodontal filter contributes to higher accuracy at all exposure times (p < 0.05-0.0001). Accuracy of periodontal diagnosis increases with higher contrast resolution. Digital exposure levels are thus dependent of image receptor as well as X-ray generator.
Article
The aim of this study was the determination of image accuracy and quality for periodontal diagnosis using various X-ray generators with conventional and digital radiographs. Thirty-one in vitro periodontal defects were evaluated on intraoral conventional (E-, F/E-speed) and digital images (three indirect, two direct sensors). Standardised radiographs were made with an alternating current (AC), a high-frequency (HF) and a direct current (DC) X-ray unit at rising exposure times (20-160 ms with 20-ms interval) with a constant kV of 70. Three observers assessed bone levels for comparison to the gold standard. Lamina dura, contrast, trabecularisation, crater and furcation involvements were evaluated. Irrespective X-ray generator-type, measurement deviations increased at higher exposure times for solid-state, but decreased for photostimulable storage phosphor (PSP) systems. Accuracy for HF or DC was significantly higher than AC (p < 0.0001), especially at low exposure times. At 0.5- to 1-mm clinical deviation, 27-53% and 32-55% dose savings were demonstrated when using HF or DC generators compared to AC, but only for PSP. No savings were found for solid-state sensors, indicating their higher sensitivity. The use of digital sensors compared to film allowed 15-90% dose savings using the AC tube, whilst solid-state sensors allowed approximately 50% savings compared to PSP, depending on tube type and threshold level.. Accuracy of periodontal diagnosis increases when using HF or DC generators and/or digital receptors with adequate diagnostic information at lower exposure times.
Article
The aim of this study was to evaluate if different ways of adjusting brightness and contrast of monitors with different technical standards influence the diagnosis of carious lesions in digital radiographs. One hundred extracted teeth (premolars and molars) were radiographed. Seven observers evaluated images for approximal carious lesions on 3 monitors: a standard color monitor with brightness and contrast manually adjusted for viewing radiographs, a Digital Imaging and Communication in Medicine (DICOM) part 14 precalibrated color monitor, and a DICOM part 14 precalibrated monochromatic monitor. All evaluations were made in ambient light <50 lux. Receiver operating characteristic curves were plotted to evaluate results. The standard criterion was a histologic examination of sliced teeth. Kappa statistic evaluated intraobserver agreement. No significant difference in accuracy of approximal carious lesion diagnosis was found between the monitors. Intraobserver agreement varied between fair and good. No differences that could affect clinicians' abilities to detect carious lesions in digital radiographs existed between the standard monitor and the color and monochrome DICOM part 14 precalibrated monitors.